Equipment and Calibration History

History for Licor LI-7500A 0065

Rec No.DateLoc FromTitleLoc ToNotes
6865 2024-09-27 Lab none Installed Tonzi Gas Analyzer CO2/H2O Floor Installed at Tonzi Floor to replace sn 0041 for calibration.
6855 2024-09-26 Lab Calibration Lab Last calibration: 2024-04-15
AGC=95.1 SDM=0 DAC1=- DAC2=-
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.0kPa, Tair=24.7C, Tdew=14.6C
Zero Air at 1350psi, AMP CA03923 488.75 at 1450 Tdewspan=20.0
CO2 zero 0.8916 -> 0.8913 Offset: -1.5ppm
CO2 span 1.024 -> 1.0229 Offset: 0.7ppm
H2O zero 0.7329 -> 0.7336 Offset: 5.7ppm
H2O span 1.021 -> 1.0269 Offset: -0.12ppm
Notes:
6784 2024-07-22 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa for calibration. Replaced by sn 75H-2667.
6666 2024-04-16 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at Bouldin Alfalfa to replace sn 75H-2514 for calibration.
6657 2024-04-15 Lab Calibration Lab Last calibration: 2024-01-03
AGC=95.57 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.49
Ambient P=100.86kPa, Tair=21.18C, Tdew=C
Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00
CO2 zero 0.8918 -> .8916 Offset: -.84ppm
CO2 span 1.0255 -> 1.024 Offset: 1.01ppm
H2O zero 0.7329 -> .7329 Offset: -.1ppm
H2O span 1.0238 -> 1.021 Offset: .06ppm
Notes: Used 524.67 as CO2 span point in software for post-span CO2 value to be ~524.77.
6637 2024-03-25 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 75H-2180.
6512 2024-01-04 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 0042 for calibration.
6499 2024-01-03 Lab Calibration Lab Last calibration: 2023-09-18
AGC=100 SDM=0 DAC1=- DAC2=-
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.8kPa, Tair=22.7C, Tdew=7.3C
Zero Air at 550psi, AMP CA04119 524.77 at 700 Tdewspan=20.0
CO2 zero 0.8919 -> 0.8918 Offset: -0.27ppm
CO2 span 1.024 -> 1.0255 Offset: -1.3ppm
H2O zero 0.7322 -> 0.7329 Offset: 4.9ppm
H2O span 1.0226 -> 1.0238 Offset: -0.02ppm
Notes: Signal strength rest from 107 to 100 before calibration. Used CO2 span pt of 524.67
6505 2024-01-02 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0035.
6422 2023-09-19 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 1594 for calibration.
6408 2023-09-18 Lab Calibration Lab Last calibration: 2023-04-10
AGC=101.1 SDM=0 DAC1=- DAC2=-
AIU sn AIU-1569, Presure A0=58.004, A1=15.479
Ambient P=100.6kPa, Tair=22.4C, Tdew=14.9C
Zero Air at 1750psi, AMP CA04119 524.77 at 1800 Tdewspan=16.0
CO2 zero 0.8918 -> 0.8919 Offset: 0.2ppm
CO2 span 1.0217 -> 1.0240 Offset: -1.47ppm
H2O zero 0.7295 -> 0.7322 Offset: 21.1ppm
H2O span 1.0128 -> 1.0226 Offset: -0.2ppm
Notes: After calibration I painted some dings with white nail polish.
6360 2023-08-08 Gilbert Tract Gas Analyzer CO2/H2O Removed Lab none Removed from Gilbert Tract for calibration. Replaced by sn 75H-2669.
6254 2023-04-11 Lab none Installed Gilbert Tract Gas Analyzer CO2/H2O Installed at Gilbert Tract to replace sn 75H-2667 for calibration.
6251 2023-04-10 Lab Calibration Lab Last calibration: 2023-01-04
AGC=102 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=23.7C, Tdew=10.5C
Zero Air at 1300psi, AMP CA04119 524.77 at 1150 Tdewspan=16.00
CO2 zero 0.8915 -> .8918 Offset: 1.72ppm
CO2 span 1.0227 -> 1.0217 Offset: .83ppm
H2O zero 0.7323 -> .7295 Offset: -20.4ppm
H2O span 1.0226 -> 1.0128 Offset: .21ppm
Notes:
6240 2023-04-03 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration. Replaced by sn 75H-2180.
6159 2023-01-06 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond to replace sn 0035 for calibration.
6140 2023-01-04 Lab Calibration Lab Last calibration: 2022-09-12
AGC=101.1 SDM=0 DAC1=- DAC2=-
Ambient P=99.5kPa, Tair=18.1C, Tdew=10.7C
Zero Air at 1650psi, AMP CA04119 524.77 at 1250 Tdewspan=15.0
CO2 zero 0.8918 -> 0.8915 Offset: -1.5ppm
CO2 span 1.0227 -> 1.0227 Offset: 0.03ppm
H2O zero 0.7316 -> 0.7323 Offset: 5.1ppm
H2O span 1.0167 -> 1.0226 Offset: -0.12ppm
Notes:
6130 2022-12-06 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2176.
5995 2022-09-14 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 75H-2667 for calibration.
5985 2022-09-12 Lab Calibration Lab Last calibration: 2022-04-08
AGC=101.0 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=22.8C, Tdew=15.2C
Zero Air at 1900psi, AMP CA04119 524.77 at 1400 Tdewspan=20.0
CO2 zero 0.8908 -> 0.8918 Offset: 4.5ppm
CO2 span 1.0198 -> 1.0227 Offset: -1.97ppm
H2O zero 0.7308 -> 0.7316 Offset: 6.7ppm
H2O span 1.016 -> 1.0167 Offset: -0.01ppm
Notes:
5979 2022-08-31 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 75H-2180.
5794 2022-04-13 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 0042 for calibration.
5773 2022-04-08 Lab Calibration Lab Last calibration: 2022-01-11
AGC=100.9 SDM= DAC1= DAC2=
Ambient P=100.7kPa, Tair=29.96C, Tdew=12.0C
Zero Air at 750psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00
CO2 zero 0.8912 -> .8908 Offset: -1.86ppm
CO2 span 1.0205 -> 1.0198 Offset: .57ppm
H2O zero 0.7314 -> .7308 Offset: -4.2ppm
H2O span 1.0259 -> 1.0160 Offset: .22ppm
Notes: In the field, this sensor had a SS ~110. In the lab the SS was much more normal ~100. I'm not sure why there is a difference, as the config file should be the same.
5767 2022-03-31 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration. Replaced by sn 0041.
5666 2022-01-20 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End to replace sn 75H-2667 for calibration.
5656 2022-01-11 Lab Calibration Lab Last calibration: 2021-09-08
AGC=101.6 SDM= DAC1= DAC2=
Ambient P=101.7kPa, Tair=28.1C, Tdew=7.9C
Zero Air at 1000psi, AMP CA04119 524.77 at 1675 Tdewspan=20.00
CO2 zero 0.8911 -> .8912 Offset: .48ppm
CO2 span 1.0211 -> 1.0205 Offset: .53ppm
H2O zero 0.7313 -> .7314 Offset: .8ppm
H2O span 1.0143 -> 1.0259 Offset: -.27ppm
Notes:
5645 2022-01-02 Gilbert Tract Gas Analyzer CO2/H2O Removed Lab none Removed from Gilbert Tract for calibration. Replaced by sn 0041.
5549 2021-10-21 Gilbert Tract Gas Analyzer CO2/H2O Adjusted Gilbert Tract Gas Analyzer CO2/H2O Reset signal strength from 108 to 100
5528 2021-09-15 Lab none Installed Gilbert Tract Gas Analyzer CO2/H2O Installed at Gilbert Tract to replace sn 0041 for calibration.
5461 2021-09-08 Lab none Calibration Lab none Last calibration: 2021-06-01
AGC=102 SDM= DAC1= DAC2=
Ambient P=100.3kPa, Tair=27.75C, Tdew=15.5C
Zero Air at 1875psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00
CO2 zero 0.8913 -> .8911 Offset: -1.15ppm
CO2 span 1.0201 -> 1.0211 Offset: .7ppm
H2O zero 0.7305 -> .7313 Offset: 5.7ppm
H2O span 1.0158 -> 1.0143 Offset: .03ppm
Notes: Showing Carlos how to calibrate 7500s
5438 2021-09-01 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0042.
5304 2021-06-02 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 75H-2667 for calibration.
5271 2021-06-01 Lab Calibration Lab Last calibration: 2021-03-15
AGC=101.2 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=21.2C, Tdew=10.6C
Zero Air at 300psi, AMP CA04119 524.77 at 1800 Tdewspan=16.0
CO2 zero 0.8914 -> 0.8913 Offset: -0.2ppm
CO2 span 1.0153 -> 1.0201 Offset: -3.27ppm
H2O zero 0.7311 -> 0.7305 Offset: -4.1ppm
H2O span 1.017 -> 1.0158 Offset: 0.0ppm
Notes:
5282 2021-05-26 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration. Replaced by sn 1594.
5213 2021-03-16 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End to replace sn 0041 for calibration.
5203 2021-03-15 Lab Calibration Lab Last calibration: 2020-12-08
AGC=102 SDM= DAC1= DAC2=
Ambient P=99.8kPa, Tair=24.0C, Tdew=1.55C
Zero Air at 400psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.8916 -> .8914 Offset: -1.05ppm
CO2 span 1.0213 -> 1.0153 Offset: .74ppm
H2O zero 0.7303 -> .7311 Offset: 6.2ppm
H2O span 1.0165 -> 1.017 Offset: -.02ppm
Notes:
5160 2021-03-04 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Wetland and brought back to the lab for calibration. Replaced by sn 0035.
5028 2020-12-09 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at Sherman Wetland to replace sn 75H-2667 for calibration.
5024 2020-12-08 Lab Calibration Lab Last calibration: 2020-08-20
AGC=101 SDM= DAC1= DAC2=
Ambient P=100.8kPa, Tair=23.1C, Tdew=2.5C
Zero Air at 550psi, AMP CA4396 408.97 at 300 Tdewspan=20.00
CO2 zero 0.8917 -> .8916 Offset: -.35ppm
CO2 span 1.0217 -> 1.0213 Offset: .27ppm
H2O zero 0.7297 -> .7303 Offset: 4.5ppm
H2O span 1.0186 -> 1.0165 Offset: .05ppm
Notes:
5014 2020-12-03 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa for calibration. Replaced by sn 75H-2182.
4686 2020-08-21 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at Bouldin Alfalfa to replace sn 75H-2667 for calibration.
4668 2020-08-20 Lab Calibration Lab Last calibration: 2020-02-04
AGC=99.4 SDM= DAC1= DAC2=
Ambient P=99.9kPa, Tair=28.6C, Tdew=14.7C
Zero Air at 700psi, AMP CA4396 408.97 at 500 Tdewspan=20.00
CO2 zero 0.8916 -> .8917 Offset: 0.11ppm
CO2 span 1.0209 -> 1.0217 Offset: -0.39ppm
H2O zero 0.7291 -> 0.7297 Offset: 4.8ppm
H2O span 1.0182 -> 1.0186 Offset: -0.01ppm
Notes: Air quality bad inside from wildfire smoke leaking in through windows
4493 2020-05-21 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry and brought back to the lab for calibration. Replaced by sn 0035.
4367 2020-02-06 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 1594, which was brought back to the lab for calibration.
4358 2020-02-04 Lab Calibration Lab Last calibration: 2019-08-21
AGC=99.95 SDM=0 DAC1=- DAC2=-
Ambient P=101.4kPa, Tair=25.8C, Tdew=-3.3C
Zero Air at 850psi, AMP CA04396 408.97 at 550 Tdewspan=16.0
CO2 zero 0.8916 -> 0.8916 Offset: 0.15ppm
CO2 span 1.0191 -> 1.0209 Offset: -0.97ppm
H2O zero 0.7278 -> 0.7291 Offset: 9.9ppm
H2O span 1.0165 -> 1.0182 Offset: -0.03ppm
Notes:
4167 2019-11-19 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi Tower for calibration. Replaced by sn 75H-2182.
4063 2019-08-22 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed on Tonzi Tower to replace sn 1594 for calibration.
4061 2019-08-21 Lab none Calibration Lab none Last calibration: 2019-04-15
AGC=100 SDM=0 DAC1=- DAC2=-
Ambient P=100.3kPa, Tair=27.6C, Tdew=15.0C
Zero Air at 1000psi, AMP CO2 408.97 at 800 Tdewspan=20.0
CO2 zero 0.8911 -> 0.8916 Offset: 2.25ppm
CO2 span 1.0171 -> 1.0191 Offset: -0.97ppm
H2O zero 0.726 -> 0.7278 Offset: 14.2ppm
H2O span 1.0118 -> 1.0165 Offset: -0.1ppm
Notes: Water zero offset a little large - maybe still acclimating from factory service and new internal chemicals. Signal Strength reset from 101.1 to 100.0
4044 2019-08-07 East Pond Gas Analyzer CO2/H2O Removed Lab none Removed from EP and brought back to the lab for calibration. It was replaced by sn 0035.
3920 2019-04-18 Lab none Installed East Pond Gas Analyzer CO2/H2O Installed at EP to replace sn 0073, which was brought back to the lab for calibration.
3914 2019-04-15 Lab none Calibration Lab none Last calibration: 2019-02-11
AGC=99.2 SDM=0 DAC1=- DAC2=-
Ambient P=100.16kPa, Tair=27.11C, Tdew=6.82C
Zero Air 369068 at 1100psi, AMP CA04396 408.97 at 750 Tdewspan=15.99
CO2 zero 0.892 -> 0.8911 Offset: -4.4ppm
CO2 span 1.0073 -> 1.0171 Offset: -5.12ppm
H2O zero 0.7263 -> 0.7260 Offset: -2.4ppm
H2O span 1.0104 -> 1.0118 Offset: -0.02ppm
Notes: First lab calibration since factory calibration. I manually typed in the new coefficients and zero/span values. I saved a config file then opened it in a text editor and changed the calibration dates, standard gas values and WX coefficient. I saved and uploaded this config. This IRGA now has the SD* temperature coefficients and its signal strength was reset at the factory. Our zeros were lower than the factory calibration so I calibrated and saved a new lab calibration config file.
3890 2019-02-22 Factory Received Lab none Received from Licor after repair and calibration. Service notes: Inspected the unit and found the source lamp intermittent. Replaced the defective source and the chopper motor per the customer's request. Subjected the unit to environmental tests and calibrated.
3891 2019-02-11 Lab none Calibration Lab none Last calibration: 2018-09-26
AGC= SDM= DAC1= DAC2=
Ambient P=kPa, Tair=C, Tdew=C
at psi, at Tdewspan=12.00
CO2 zero 0.8903 -> .8920 Offset: ppm
CO2 span 1.0184 -> 1.0073 Offset: ppm
H2O zero 0.7086 -> 07263 Offset: ppm
H2O span 1.0155 -> 1.0104 Offset: ppm
Notes: CX=34338.1, WX=34091.7. Calibrated at Licor factory after source/chopper replacement.
3889 2019-01-21 Lab none Shipped Factory Shipped back to factory for repair. Received by Licor on 2019-01-21.
3740 2018-10-17 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira due to no CO2, H2O or Signal Strength - maybe bad source or detector. Replaced by sn 0418
3709 2018-09-27 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira Tower to replace sn 1594 for calibration.
3707 2018-09-26 Lab none Calibration Lab none Last calibration: 2018-04-04
AGC=99.7 SDM=0 DAC1=0 DAC2=0
Ambient P=100.4kPa, Tair=26.6C, Tdew=13.8C
Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.8905 -> .8903 Offset: -1.09ppm
CO2 span 1.0231 -> 1.0184 Offset: 2.48ppm
H2O zero 0.7082 -> .7086 Offset: 2.7ppm
H2O span 1.0185 -> 1.0155 Offset: .07ppm
Notes:
3591 2018-07-12 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from BA and brought back to the lab for calibration. It was replaced by sn 75H-2150.
3387 2018-04-05 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at BA to replace sn 75H-2667, which was brought back to the lab for calibration.
3377 2018-04-04 Lab none Calibration Lab none Last calibration: 2017-11-28
AGC=100.4 SDM=0 DAC1=0 DAC2=0
Ambient P=100.8kPa, Tair=22.5C, Tdew=9.3C
Zero Air at 100psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.8906 -> .8905 Offset: -.52ppm
CO2 span 1.0163 -> 1.0231 Offset: -3.35ppm
H2O zero 0.7083 -> .7082 Offset: -.7ppm
H2O span 1.0126 -> 1.0185 Offset: -.13ppm
Notes: Calibrated on LI-7550 sn 674 with 8.5.0 embedded software version
3367 2018-03-06 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed for calibration and replaced by sn 75H-2182.
3210 2017-11-29 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 75H-2514 for calibration.
3206 2017-11-28 Lab none Calibration Lab none Last calibration: 2017-05-30
AGC=100.0 SDM=0 DAC1=0 DAC2=0
Ambient P=100.5kPa, Tair=23.0C, Tdew=8.7C
Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.8901 -> .8906 Offset: 2.12ppm
CO2 span 1.013 -> 1.0163 Offset: -1.35ppm
H2O zero 0.7087 -> .7083 Offset: -3.0ppm
H2O span 1.0166 -> 1.0126 Offset: 0.08ppm
Notes: Calibrated on 7550 software version 8.7.5
3168 2017-10-19 East End Gas Analyzer T1 CO2/H2O Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
2810 2017-06-01 Lab none Installed East End Gas Analyzer T1 CO2/H2O Installed at open water temporary tower in East End wetland
2797 2017-05-30 Lab none Calibration Lab none Last calibration: 2016-12-06
AGC=100 SDM= DAC1= DAC2=
Ambient P=99.7kPa, Tair=19.8C, Tdew=10.1C
Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00
CO2 zero 0.8899 -> .8901 Offset: 1.08ppm
CO2 span 1.0142 -> 1.0130 Offset: -.84ppm
H2O zero 0.7086 -> .7087 Offset: .6ppm
H2O span 1.0127 -> 1.0166 Offset: -.08ppm
Notes: Signal strength was 73.9 after thorough cleaning of lenses, so I reset the Signal Strength to 100%.
2677 2017-03-23 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa and brought back to the lab for calibration. It was replaced by sn 75H-2667.
2539 2016-12-07 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed on the Bouldin Alfalfa tower to replace sn 75H-2667 for calibration.
2530 2016-12-06 Lab none Calibration Lab none Last calibration: 2016-05-13
AGC=74.7 SDM=6 DAC1= DAC2=
Ambient P=100.29kPa, Tair=16.04C, Tdew=0.60C
Zero Air at 350psi, God CO2 421.22 at 400 Tdewspan=16.02
CO2 zero 0.8898 -> 0.8899 Offset: 0.36ppm
CO2 span 1.0149 -> 1.0142 Offset: 0.43ppm
H2O zero 0.7086 -> 0.7086 Offset: 0.2ppm
H2O span 1.0139 -> 1.0127 Offset: 0.01ppm
Notes: H2O span done later in the day.
2439 2016-08-31 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell Rice tower for calibration. Replaced by sn 75H-2180.
2281 2016-05-24 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed at Twitchell Rice tower to replace sn 75H-2176 for calibration
2262 2016-05-13 Lab none Calibration Lab none Last calibration: 2015-12-07
AGC=74.96 SDM=- DAC1=- DAC2=-
Ambient P=100.14kPa, Tair=20.14C, Tdew=8.2C
Zero Air at 600psi, God CO2 421.22 at 500 Tdewspan=15.09
CO2 zero 0.89 -> 0.8898 Offset: -0.57ppm
CO2 span 1.0164 -> 1.0149 Offset: 0.91ppm
H2O zero 0.7097 -> 0.7086 Offset: -7.4ppm
H2O span 1.0231 -> 1.0139 Offset: 0.19ppm
Notes: Calibration from H2O performed at 15.09C dew point instead of 11.09 which has been used previously. Use 15.09 from now on for a larger calibrated H2O range.
2205 2016-02-25 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from the West Pond tower for calibration. Replaced by sn 1594.
2123 2015-12-08 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on the West Pond tower to replace sn 75H-2180 for calibration.
2127 2015-12-07 Lab none Calibration Lab none Last calibration: 2015-04-20
AGC=76.5 SDM=1 DAC1=- DAC2=-
Ambient P=100.92kPa, Tair=19.81C, Tdew=14.18C
Zero Air at 650psi, God CO2 421.22 at 700 Tdewspan=11.09
CO2 zero 0.8897 -> 0.8900 Offset: 1.20ppm
CO2 span 1.0074 -> 1.0164 Offset: -4.88ppm
H2O zero 0.7055 -> 0.7097 Offset: 31.2ppm
H2O span 1.0071 -> 1.0231 Offset: -0.32ppm
Notes:
2108 2015-11-25 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell Rice tower for calibration. REplaced by sn 75H-2182
1925 2015-08-06 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell Rice tower to replace sn 75H-2514 for calibration.
1918 2015-07-22 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from the Alfalfa Tower for calibration. Replaced by sn 75H-2150
1795 2015-04-23 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed on the Alfalfa tower to replace sn 75H-2176 for calibration.
1783 2015-04-20 Lab none Calibration Lab none Last calibration: 2014-12-22
AGC=72.8 SDM=1 DAC1=- DAC2=-
Ambient P=100.0kPa, Tair=20.13C, Tdew=12.49C
Zero Air at 850psi, God CO2 421.22 at 1100 Tdewspan=15.96
CO2 zero 0.8909 -> 0.8897 Offset: -5.4ppm
CO2 span 0.9835 -> 1.0074 Offset: -13.02ppm
H2O zero 0.705 -> 0.7055 Offset: 4.0ppm
H2O span 1.1245 -> 1.0071 Offset: 2.34ppm
Notes: The internal chemicals were changes a few weeks ago - the old ones still looked good though. Relatively large offsets in span, H2O span offset similar to other 7500s calibrated at this time - dew point generator and Vaisala probe agreed with in 0.5C
1676 2015-01-20 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Rice tower because it was the wrong head from the coefficients loaded to the 7550 box. Replaced by sn 75H-2150
1633 2014-12-23 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry tower to replace sn 75H-2514 for calibration.
1650 2014-12-22 Lab none Calibration Lab none Last calibration: 2014-09-09
AGC=73.11 SDM=6 DAC1=- DAC2=-
Ambient P=101.19kPa, Tair=23.76C, Tdew=14.50C
Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.95
CO2 zero 0.8899 -> 0.8909 Offset: 4.46ppm
CO2 span 1.007 -> 0.9835 Offset: 13.26ppm
H2O zero 0.7053 -> 0.7050 Offset: -2.0ppm
H2O span 1.0101 -> 1.1245 Offset: -2.24ppm
Notes: Started calibration on sn AIU-0397 box. It suddenly restarted and then quit. No power light. Same cables worked on sn AIU-614 and the calibration was completed - however the large offsets are unusual and weren't seen on the 0397 box.
1614 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
1548 2014-09-10 Lab none Installed Mayberry-Portable Installed on the Mayberry Portable tower to replace sn 75H-2176 for calibration.
1559 2014-09-09 Lab none Calibration Lab none Last calibration: 2014-04-02
AGC=73.73 SDM=1 DAC1=- DAC2=-
Ambient P=100.35kPa, Tair=22.06C, Tdew=12.61C
Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25
CO2 zero 0.8899 -> 0.8899 Offset: 0.0ppm
CO2 span 1.0062 -> 1.0070 Offset: -0.52ppm
H2O zero 0.704 -> 0.7053 Offset: 8.6ppm
H2O span 0.9932 -> 1.0101 Offset: -0.34ppm
Notes:
1534 2014-09-04 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed for calibration. Problems with the 7550 box sn AIU-0614 prevented installation of a new 7500.
1315 2014-04-03 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on the West Pond tower to replace sn 74H-2150 from calibration.
1279 2014-04-02 Lab none Calibration Lab none Last calibration: 2013-09-11
AGC=73.1 SDM=1 DAC1=- DAC2=-
Ambient P=100.27kPa, Tair=19.36C, Tdew=6.05C
Zero Air at 1400psi, God CO2 421.22 at 1500 Tdewspan=15.27
CO2 zero 0.8897 -> 0.8899 Offset: .82ppm
CO2 span 1.0041 -> 1.0062 Offset: -0.68ppm
H2O zero 0.704 -> 0.7040 Offset: 0.2ppm
H2O span 0.9959 -> 0.9932 Offset: 0.06ppm
Notes: Calibrated by Joe
1277 2014-03-27 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell Alfalfa tower for calibration. Replaced by sn 75H-2180.
1085 2013-09-11 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O installed on the Alfalfa tower to replace sn 2180 for calibration
1084 2013-09-11 Lab none Calibration Lab none Last calibration: 2013-05-08
AGC=71.77 SDM=1 DAC1=- DAC2=-
Ambient P=100.08kPa, Tair=23.13C, Tdew=14.04C
Zero air at 1650psi, God CO2 421.22 at 1600 Tdewspan=15.8
CO2 zero 0.8904 -> 0.8897 Offset: -2.78ppm
CO2 span 1.0198 -> 1.0041 Offset: 8.53ppm
H2O zero 0.7022 -> 0.7040 Offset: 12.5ppm
H2O span 0.9726 -> 0.9959 Offset: -1.1ppm
Notes:
1065 2013-08-28 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell Rice tower for calibration. Replaced by sn 2176
1045 2013-05-09 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell Rice tower to replace sn2180 for calibration.
989 2013-05-08 Lab none Calibration Lab none Last calibration: 2012-12-13
AGC=15.5 SDM=1 DAC1=none DAC2=none
Ambient P=101.55kPa, Tair=23.16C, Tdew=10.26C
Zero air at 1700psi, God bottle 421.22 at 1700 Tdewspan=15.03
CO2 zero 0.8901 -> 0.8904 Offset: 0.90ppm
CO2 span 1.001 -> 1.0198 Offset: -10.07ppm
H2O zero 0.7016 -> 0.7022 Offset: 4.2ppm
H2O span 0.9935 -> 0.9726 Offset: 0.44ppm
Notes: Secondary CO2 standard seemed low at 344.62. I rechecked it with the LI6262 and got even lower value about 337. I calibrated with the god bottle but the IGRA was still off about 10ppm. I double checked the secondary standard and it read 335 - okay.
1048 2013-04-26 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from the West Pond tower for calibration. Replaced by sn2176
872 2013-01-03 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on the West Pond tower with AIU-0666 to swap out sn 2182 for calibration and AIU-0397 for repair of the condensation problem.
867 2012-12-21 Lab none Returned Lab none Returned from Licor with AIU-0666 after repair of condesation problem. Condesation was removed from the source lens and the detector was noisy so was also replaced. The 7500 was recalibrated.
CO2: A=1.57076E2, B=5.35482E3, C=5.64100E7, D=-1.93457E10, E=2.76431E12, XS=0.0025, Z=-9.00000E-4, zero=0.8901, span=1.0010
H2O: A=4.99824E3, B=6.61401E6, C=-1.74913E9, XS=-0.0007, Z=1.06000E-2, zero=0.7016, span=0.9935, CX=32000
SS: B=2.691, C=4.178
988 2012-12-13 Factory Calibration Lab none Last calibration: 2012-08-09
AGC= SDM= DAC1= DAC2=
Ambient P=kPa, Tair=C, Tdew=C
at psi, at Tdewspan=
CO2 zero 0.8715 -> 0.8901 Offset: ppm
CO2 span 1.0183 -> 1.0010 Offset: ppm
H2O zero 0.6361 -> 0.7016 Offset: ppm
H2O span 1.0392 -> 0.9935 Offset: ppm
Notes: Factory calibration
848 2012-11-02 Lab none Shipped Factory Shipped to Licor for repair of condensation problem caused by problem in the 7550 box, RMA 172095
823 2012-10-25 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from the Mayberry Permanent Tower due to failure. Will be returned to Licor for repair.
753 2012-08-16 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry permanent tower. sn 2182 was removed for calibration. The existing coefficeints and zero/span values were downloaded from the 7550 box. The new coefficients and zero/span values were uploaded to the 7550 box. The existing head sn 2182 was removed and this IRGA was installed. All looked well.
752 2012-08-09 Lab none Installed Lab none This IRGA has recently been upgraded to a 7500A interface. I tried to install it on the 7550 box at Twitchell but it sent the 7550 box into a reboot cycle. So now it is back in the lab to be tried on another tower.
751 2012-08-09 Lab none Calibration Lab none Last calibration: 2012-08-07
AGC=43.8 SDM=10 DAC1= DAC2=
Ambient P=100.25kPa, Tair=22.21C, Tdew=5.88C
Zero air at 1800psi, CO2 in air 436.53 at 450 Tdewspan=15.08
CO2 zero 0.9473 -> 0.8715 Offset: -576ppm
CO2 span 1.3749 -> 1.0183 Offset: 225.47ppm
H2O zero 0.6596 -> 0.6361 Offset: -151ppm
H2O span 1.2399 -> 1.0392 Offset: 4.02ppm
Notes: This was a second calibration after upgrading to a LI75000A head interface and letting new internal chemicals settle for two days. There are large offsets but the new zero/span numbers match the calibration of 8-6 before the upgrade. A second check of zeros CO2 0.8714 0.5 ppm offset, H2O 0.6364 2.3 mmol/m3 offset.
750 2012-08-07 Lab none Calibration Lab none Last calibration: 2012-08-06
AGC=37.5 SDM=10 DAC1= DAC2=
Ambient P=100.68kPa, Tair=22.52C, Tdew=C
Zero air at 1700psi, Co2 in air 436.53 at 1450 Tdewspan=15.08
CO2 zero 0.874 -> 0.9473 Offset: 294.5ppm
CO2 span 0.9973 -> 1.3749 Offset: -146.53ppm
H2O zero 0.6387 -> 0.6596 Offset: 187.9ppm
H2O span 1.0898 -> 1.2399 Offset: -2.67ppm
Notes: Initial calibration after upgrading to a 7500A head. Calibration coefficients and zero/span values were entered by hand into the LI-7550. New internal chemicals will need to settle for at least a day.
724 2012-08-07 Lab none Upgrade Lab none This IRAG has been selected to be upgraded to a 7500A head interface. This is the oldest IRGA that has not been overhauled. The upgrade went smoothly, o-rings were greased, the internal chemicals were changed, coefficients were manually entered in the LI-7550 box (sn AIU-0674). The gain potentiometer was adjusted, per instructions, about a half turn before maxing out. The AGC was 50.0 before adjusting and 43.75 afterwards. See initial calibration this date. A subsequent calibration will be done in two days to let the internal chemicals settle.
723 2012-08-06 Lab none Calibration Lab none Last calibration: 2012-05-18
AGC=65 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=99.63kPa, Tair=22.07C, Tdew=8.90C
Zero air at 1700psi, CO2 in air 436.53 at 1450 Tdewspan=15.08
CO2 zero 0.8741 -> 0.8740 Offset: -0.36ppm
CO2 span 0.9958 -> 0.9973 Offset: -0.93ppm
H2O zero 0.6415 -> 0.6387 Offset: -19.3ppm
H2O span 1.0009 -> 1.0898 Offset: -1.76ppm
Notes: Calibration in preparation to be upgraded to a LI-7500A head.
719 2012-08-02 DWR Corn Removed Lab none Removed from the DWR Corn Tower for calibration and upgrade to LI-7500A. Replaced by sn 0042.
662 2012-05-21 Lab none Installed DWR Corn Installed on the DWR Corn tower.
595 2012-05-07 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the tower.  Replaced by the new DWR Li7500A (sn 2150).
649 2012-01-25 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell tower to replace sn 0041 for calibration.
498 2012-01-16 Lab none Calibration Lab none Last calibration: 2011-09-27
AGC=65 SDM=7 DAC1=0-1000 DAC2=12-25
Ambient P=100.22kPa, Tair=26.28C, Tdew=-0.53C
zero air at 500psi, CO2 in air 436.53 at 1750 Tdewspan=14.15
CO2 zero 0.8741 -> 0.8738 Offset: -1.15ppm
CO2 span 0.9936 -> 0.9941 Offset: -0.53ppm
H2O zero 0.6341 -> 0.6393 Offset: ppm
H2O span 0.9914 -> 1.0174 Offset: 0.49ppm
Notes: 5Hz band width. Internal chemicals changed on 01-12-2012.
496 2012-01-12 Lab none Maintenance Lab none Internal chemicals changed. The old chemicals still looked okay, but changed just the same.
466 2011-12-18 Tonzi-Tower Removed Lab none Removed from Tonzi Tower for calibration. Replaced by sn 0418.
413 2011-09-29 Lab none Installed Tonzi-Tower Installed on Tonzi tower to replace sn 1594 for calibration.
667 2011-09-27 Lab none Calibration Lab none Last calibration: 2011-05-24
AGC=65 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=99.28kPa, Tair=26.94C, Tdew=10.69C
zero air at 1000psi, co2 in air 405.85 at 1500 Tdewspan=13.4
CO2 zero 0.8748 -> 0.8741 Offset: -2.5ppm
CO2 span 0.9939 -> 0.9936 Offset: 0.05ppm
H2O zero 0.6408 -> 0.6341 Offset: -44.6ppm
H2O span 1.014 -> 0.9914 Offset: 0.55ppm
Notes:
410 2011-09-15 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry tower for calibration, replaced by sn 0035.
363 2011-05-25 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry tower to swap out sn 0042 for calibration
666 2011-05-24 Lab none Calibration Lab none Last calibration: 2011-04-21
AGC=62 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=99.67kPa, Tair=23.61C, Tdew=9.06C
zero air at 1450psi, CO2 in air 597.4 at 0 Tdewspan=14.2
CO2 zero 0.8741 -> 0.8748 Offset: 2.56ppm
CO2 span 1.0093 -> 0.9939 Offset: 13.41ppm
H2O zero 0.7366 -> 0.6408 Offset: -7.5ppm
H2O span 0.9749 -> 1.014 Offset: -0.58ppm
Notes:
637 2011-03-30 Vaira Gas Analyzer CO2/H2O Removed Lab none Swapped off the Vaira tower for calibration. Replaced by sn 0418.
275 2011-01-19 Lab none Installed Vaira Gas Analyzer CO2/H2O This sensor was installed on the Vaira tower after laboratory calibration. sn 1597 was removed.
665 2011-01-18 Lab none Calibration Lab none Last calibration: 0000-00-00
AGC=66 SDM=0 DAC1= DAC2=
Ambient P=99.48kPa, Tair=28.8C, Tdew=0C
zero air at 0psi, CO2 597.4 at 0 Tdewspan=15.9
CO2 zero 0.6397 -> 0.6399 Offset: 0ppm
CO2 span 0.8741 -> 0.8744 Offset: 0ppm
H2O zero 0.9917 -> 0.9919 Offset: 0ppm
H2O span 1 -> 1.0049 Offset: 0ppm
Notes: performed by Siyan
272 2011-01-05 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower for calibration. Replaced by sn 0976
226 2010-10-07 Lab none Installed Tonzi-Tower Installed on tower to swap out cn 0073 for calibration.
189 2010-08-14 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower by Siyan for lab calibration.
115 2010-05-13 Lab none Installed Tonzi-Tower Installed after fresh calibration in the lab.
92 2010-03-18 Sherman Island Gas Analyzer CO2/H2O Removed Lab none This was LI7500 was removed from the Sherman tower for calibartion.
78 2009-12-04 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Freshly calibrated. Installed on permanent tower.
3 2009-12-03 Lab none Calibration Lab none AGC = 62, Pressure = 99.73kPa, Temperature = 21.63C
span CO2 = 594.7ppm, span H2O = 16.7C from dewpoint generator
Last calibration Aug 26, 2009
Last CO2 zero = 0.8738, span = 0.9939
Last H2O zero = 0.6414, span = 1.0055

New CO2 zero = 0.8739, span = 0.9931
New H2O zero = 0.6378, span = 0.9992

Configuration downloaded and saved
2 2009-11-30 Lab none Upgrade Lab none Current embedded software v3.0.0 upgraded to v3.0.1. Internal chemicals changed.
4008 2009-01-08 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from the Tonzi Tower for calibration. Replaced by sn 0418.
6095 2005-01-27 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira replacing current IRGA presumably for calibration
6615 2004-04-07 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0041
4923 2003-10-02 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 0042.
4918 2003-06-20 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0042.
4910 2003-05-30 Tonzi-Floor Gas Analyzer CO2/H2O Moved Vaira Gas Analyzer CO2/H2O Moved from Tonzi-Floor to Vaira because the Floor sonic was dead and the Vaira 7500 box had a bad power connector
4906 2003-05-09 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 0418.
4901 2003-03-17 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0073 for calibration.
4879 2002-12-23 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0418 for calibration.
4808 2002-11-05 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 0073.
4875 2002-09-20 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0073 for calibration.
4866 2002-08-15 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 0073.
4845 2002-07-11 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0042 for calibration.
4842 2002-06-20 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0073.
4839 2002-05-09 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0042 for calibration.
4833 2002-04-18 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa for calibration. Replaced by sn 0041.
4821 2002-02-28 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0035.
4772 2001-08-17 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 0042 for calibration.
4767 2001-07-19 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi Floor for calibration. Replaced by sn 0041
6613 2001-03-22 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0041.