View history entry 5985 for Licor LI-7500A sn: 0065

Date: 2022-09-12
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-04-08 AGC=101.0 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=22.8C, Tdew=15.2C Zero Air at 1900psi, AMP CA04119 524.77 at 1400 Tdewspan=20.0 CO2 zero 0.8908 -> 0.8918 Offset: 4.5ppm CO2 span 1.0198 -> 1.0227 Offset: -1.97ppm H2O zero 0.7308 -> 0.7316 Offset: 6.7ppm H2O span 1.016 -> 1.0167 Offset: -0.01ppm Notes:
Entered by: