View history entry 5773 for Licor LI-7500A sn: 0065

Date: 2022-04-08
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-01-11 AGC=100.9 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=29.96C, Tdew=12.0C Zero Air at 750psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.8912 -> .8908 Offset: -1.86ppm CO2 span 1.0205 -> 1.0198 Offset: .57ppm H2O zero 0.7314 -> .7308 Offset: -4.2ppm H2O span 1.0259 -> 1.0160 Offset: .22ppm Notes: In the field, this sensor had a SS ~110. In the lab the SS was much more normal ~100. I'm not sure why there is a difference, as the config file should be the same.