Equipment and Calibration History

View Equipment |

View history entry 3377 for Licor LI-7500A sn: 0065

Date: 2018-04-04
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-11-28 AGC=100.4 SDM=0 DAC1=0 DAC2=0 Ambient P=100.8kPa, Tair=22.5C, Tdew=9.3C Zero Air at 100psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.8906 -> .8905 Offset: -.52ppm CO2 span 1.0163 -> 1.0231 Offset: -3.35ppm H2O zero 0.7083 -> .7082 Offset: -.7ppm H2O span 1.0126 -> 1.0185 Offset: -.13ppm Notes: Calibrated on LI-7550 sn 674 with 8.5.0 embedded software version
Entered by: