Equipment and Calibration History

View Equipment |

View history entry 665 for Licor LI-7500A sn: 0065

Date: 2011-01-18
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 0000-00-00 AGC=66 SDM=0 DAC1= DAC2= Ambient P=99.48kPa, Tair=28.8C, Tdew=0C zero air at 0psi, CO2 597.4 at 0 Tdewspan=15.9 CO2 zero 0.6397 -> 0.6399 Offset: 0ppm CO2 span 0.8741 -> 0.8744 Offset: 0ppm H2O zero 0.9917 -> 0.9919 Offset: 0ppm H2O span 1 -> 1.0049 Offset: 0ppm Notes: performed by Siyan