View history entry 867 for Licor LI-7500A sn: 0065

Date: 2012-12-21
From: Lab - none
To: Lab - none
Title: Returned
Notes: Returned from Licor with AIU-0666 after repair of condesation problem. Condesation was removed from the source lens and the detector was noisy so was also replaced. The 7500 was recalibrated. CO2: A=1.57076E2, B=5.35482E3, C=5.64100E7, D=-1.93457E10, E=2.76431E12, XS=0.0025, Z=-9.00000E-4, zero=0.8901, span=1.0010 H2O: A=4.99824E3, B=6.61401E6, C=-1.74913E9, XS=-0.0007, Z=1.06000E-2, zero=0.7016, span=0.9935, CX=32000 SS: B=2.691, C=4.178