View history entry 5656 for Licor LI-7500A sn: 0065

Date: 2022-01-11
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-09-08 AGC=101.6 SDM= DAC1= DAC2= Ambient P=101.7kPa, Tair=28.1C, Tdew=7.9C Zero Air at 1000psi, AMP CA04119 524.77 at 1675 Tdewspan=20.00 CO2 zero 0.8911 -> .8912 Offset: .48ppm CO2 span 1.0211 -> 1.0205 Offset: .53ppm H2O zero 0.7313 -> .7314 Offset: .8ppm H2O span 1.0143 -> 1.0259 Offset: -.27ppm Notes:
Entered by: