View history entry 724 for Licor LI-7500A sn: 0065

Date: 2012-08-07
From: Lab - none
To: Lab - none
Title: Upgrade
Notes: This IRAG has been selected to be upgraded to a 7500A head interface. This is the oldest IRGA that has not been overhauled. The upgrade went smoothly, o-rings were greased, the internal chemicals were changed, coefficients were manually entered in the LI-7550 box (sn AIU-0674). The gain potentiometer was adjusted, per instructions, about a half turn before maxing out. The AGC was 50.0 before adjusting and 43.75 afterwards. See initial calibration this date. A subsequent calibration will be done in two days to let the internal chemicals settle.