View history entry 6408 for Licor LI-7500A sn: 0065

Date: 2023-09-18
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-04-10 AGC=101.1 SDM=0 DAC1=- DAC2=- AIU sn AIU-1569, Presure A0=58.004, A1=15.479 Ambient P=100.6kPa, Tair=22.4C, Tdew=14.9C Zero Air at 1750psi, AMP CA04119 524.77 at 1800 Tdewspan=16.0 CO2 zero 0.8918 -> 0.8919 Offset: 0.2ppm CO2 span 1.0217 -> 1.0240 Offset: -1.47ppm H2O zero 0.7295 -> 0.7322 Offset: 21.1ppm H2O span 1.0128 -> 1.0226 Offset: -0.2ppm Notes: After calibration I painted some dings with white nail polish.
Entered by: