Equipment and Calibration History

View Equipment |

View history entry 5461 for Licor LI-7500A sn: 0065

Date: 2021-09-08
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2021-06-01 AGC=102 SDM= DAC1= DAC2= Ambient P=100.3kPa, Tair=27.75C, Tdew=15.5C Zero Air at 1875psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.8913 -> .8911 Offset: -1.15ppm CO2 span 1.0201 -> 1.0211 Offset: .7ppm H2O zero 0.7305 -> .7313 Offset: 5.7ppm H2O span 1.0158 -> 1.0143 Offset: .03ppm Notes: Showing Carlos how to calibrate 7500s
Entered by: