Equipment and Calibration History

View Equipment |

View history entry 3 for Licor LI-7500A sn: 0065

Date: 2009-12-03
From: Lab - none
To: Lab - none
Title: Calibration
Notes: AGC = 62, Pressure = 99.73kPa, Temperature = 21.63C span CO2 = 594.7ppm, span H2O = 16.7C from dewpoint generator Last calibration Aug 26, 2009 Last CO2 zero = 0.8738, span = 0.9939 Last H2O zero = 0.6414, span = 1.0055 New CO2 zero = 0.8739, span = 0.9931 New H2O zero = 0.6378, span = 0.9992 Configuration downloaded and saved
Entered by: