Equipment and Calibration History

View Equipment |

View history entry 3707 for Licor LI-7500A sn: 0065

Date: 2018-09-26
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-04-04 AGC=99.7 SDM=0 DAC1=0 DAC2=0 Ambient P=100.4kPa, Tair=26.6C, Tdew=13.8C Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00 CO2 zero 0.8905 -> .8903 Offset: -1.09ppm CO2 span 1.0231 -> 1.0184 Offset: 2.48ppm H2O zero 0.7082 -> .7086 Offset: 2.7ppm H2O span 1.0185 -> 1.0155 Offset: .07ppm Notes:
Entered by: