Equipment and Calibration History

View Equipment |

View history entry 1559 for Licor LI-7500A sn: 0065

Date: 2014-09-09
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-04-02 AGC=73.73 SDM=1 DAC1=- DAC2=- Ambient P=100.35kPa, Tair=22.06C, Tdew=12.61C Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25 CO2 zero 0.8899 -> 0.8899 Offset: 0.0ppm CO2 span 1.0062 -> 1.0070 Offset: -0.52ppm H2O zero 0.704 -> 0.7053 Offset: 8.6ppm H2O span 0.9932 -> 1.0101 Offset: -0.34ppm Notes:
Entered by: