View history entry 723 for Licor LI-7500A sn: 0065

Date: 2012-08-06
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2012-05-18 AGC=65 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=99.63kPa, Tair=22.07C, Tdew=8.90C Zero air at 1700psi, CO2 in air 436.53 at 1450 Tdewspan=15.08 CO2 zero 0.8741 -> 0.8740 Offset: -0.36ppm CO2 span 0.9958 -> 0.9973 Offset: -0.93ppm H2O zero 0.6415 -> 0.6387 Offset: -19.3ppm H2O span 1.0009 -> 1.0898 Offset: -1.76ppm Notes: Calibration in preparation to be upgraded to a LI-7500A head.