View history entry 751 for Licor LI-7500A sn: 0065

Date: 2012-08-09
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2012-08-07 AGC=43.8 SDM=10 DAC1= DAC2= Ambient P=100.25kPa, Tair=22.21C, Tdew=5.88C Zero air at 1800psi, CO2 in air 436.53 at 450 Tdewspan=15.08 CO2 zero 0.9473 -> 0.8715 Offset: -576ppm CO2 span 1.3749 -> 1.0183 Offset: 225.47ppm H2O zero 0.6596 -> 0.6361 Offset: -151ppm H2O span 1.2399 -> 1.0392 Offset: 4.02ppm Notes: This was a second calibration after upgrading to a LI75000A head interface and letting new internal chemicals settle for two days. There are large offsets but the new zero/span numbers match the calibration of 8-6 before the upgrade. A second check of zeros CO2 0.8714 0.5 ppm offset, H2O 0.6364 2.3 mmol/m3 offset.