Equipment and Calibration History

View Equipment |

View history entry 1084 for Licor LI-7500A sn: 0065

Date: 2013-09-11
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2013-05-08 AGC=71.77 SDM=1 DAC1=- DAC2=- Ambient P=100.08kPa, Tair=23.13C, Tdew=14.04C Zero air at 1650psi, God CO2 421.22 at 1600 Tdewspan=15.8 CO2 zero 0.8904 -> 0.8897 Offset: -2.78ppm CO2 span 1.0198 -> 1.0041 Offset: 8.53ppm H2O zero 0.7022 -> 0.7040 Offset: 12.5ppm H2O span 0.9726 -> 0.9959 Offset: -1.1ppm Notes:
Entered by: