View history entry 2127 for Licor LI-7500A sn: 0065

Date: 2015-12-07
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2015-04-20 AGC=76.5 SDM=1 DAC1=- DAC2=- Ambient P=100.92kPa, Tair=19.81C, Tdew=14.18C Zero Air at 650psi, God CO2 421.22 at 700 Tdewspan=11.09 CO2 zero 0.8897 -> 0.8900 Offset: 1.20ppm CO2 span 1.0074 -> 1.0164 Offset: -4.88ppm H2O zero 0.7055 -> 0.7097 Offset: 31.2ppm H2O span 1.0071 -> 1.0231 Offset: -0.32ppm Notes: