Equipment and Calibration History

History for Licor LI-7500A 0042

Rec No.DateLoc FromTitleLoc ToNotes
7015 2024-11-06 Hill Slough Gas Analyzer CO2/H2O Removed Lab none Removed from Hill Slough for calibration. Replaced by sn 75H-2176.
6763 2024-07-16 Lab none Installed Hill Slough Gas Analyzer CO2/H2O Installed at Hill Slough to replace sn 0035 for calibration.
6752 2024-07-12 Lab Calibration Lab Last calibration: 2024-03-23
AGC=101.5 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=99.9kPa, Tair=23.17C, Tdew=13.9C
Zero Air at 1750psi, AMP CA04119 524.77 at 350 Tdewspan=20.00
CO2 zero 0.9483 -> .9482 Offset: -.53ppm
CO2 span 1.0098 -> 1.0114 Offset: -1.07ppm
H2O zero 0.8451 -> .8455 Offset: 2.1ppm
H2O span 1.02 -> 1.0264 Offset: -.14ppm
Notes:
6747 2024-06-19 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2182.
6628 2024-03-26 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0035 for calibration.
6622 2024-03-23 Lab none Calibration Lab none Last calibration: 2023-10-05
AGC=100.8 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.08kPa, Tair=24.9C, Tdew=9.91C
Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00
CO2 zero 0.9473 -> .9483 Offset: 3.82ppm
CO2 span 1.0033 -> 1.0098 Offset: -4.49ppm
H2O zero 0.8474 -> .8451 Offset: -13.6ppm
H2O span 1.0075 -> 1.02 Offset: -.28ppm
Notes: H2O span was oscillating within 1ppm (but average value was stable), but this went away after adding more DI water to the dewpoint generator reservoir. CO2 span seems larger than usual. Painted some chips in the paint with nail polish
6511 2024-01-04 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 0065.
6436 2023-10-11 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 0418 for calibration.
6431 2023-10-05 Lab Calibration Lab Last calibration: 2023-06-13
AGC=99.7 SDM= DAC1= DAC2=
AIU sn AIU-1569, Presure A0=58.004, A1=15.479
Ambient P=100.48kPa, Tair=27.9C, Tdew=14.5C
Zero Air at 750psi, AMP CA04119 524.77 at 800 Tdewspan=20.00
CO2 zero 0.9486 -> .9473 Offset: -5.2ppm
CO2 span 1.0184 -> 1.0033 Offset: 10.51ppm
H2O zero 0.8461 -> .8474 Offset: 7.7ppm
H2O span 1.039 -> 1.0075 Offset: -.08ppm
Notes: Vertical orientation. CO2 span offset larger than usual.
6417 2023-09-19 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 75H-2180.
6310 2023-06-16 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 75H-2667 for calibration.
6319 2023-06-13 Lab Calibration Lab Last calibration: 2022-11-29
AGC=100.3 SDM= DAC1= DAC2=
AIU sn AIU-1569, Presure A0=58.004, A1=15.479
Ambient P=100.7kPa, Tair=19.8C, Tdew=11.0C
Zero Air at 1150psi, AMP CA04119 524.77 at 1050 Tdewspan=16.01
CO2 zero 0.9478 -> 0.9486 Offset: 3.0ppm
CO2 span 1.0145 -> 1.0184 Offset: -2.67ppm
H2O zero 0.8472 -> 0.8461 Offset: -6.8ppm
H2O span 1.0267 -> 1.039 Offset: -0.26ppm
Notes: Calibrated with the IRGA vertical
6274 2023-04-25 Hill Slough Gas Analyzer CO2/H2O Removed Lab none Removed from Hill Slough for calibration. Replaced by sn 0041.
6104 2022-11-30 Lab none Installed Hill Slough Gas Analyzer CO2/H2O Installed at Hill Slough to replace sn 75H-2176 for calibration.
6126 2022-11-29 Lab Calibration Lab Last calibration: 2022-11-15
AGC=99.3 SDM= DAC1= DAC2=
Ambient P=99.3kPa, Tair=22.8C, Tdew=2.02C
Zero Air at 1750psi, AMP CA 04119 524.77 at 1325 Tdewspan=16.00
CO2 zero 0.9481 -> .9478 Offset: -.94ppm
CO2 span 1.0127 -> 1.0145 Offset: -1.15ppm
H2O zero 0.847 -> .8472 Offset: 1.3ppm
H2O span 1.0292 -> 1.0267 Offset: .05ppm
Notes: Typed in "15.99" for 16.00C span point. Seemed to be more accurate than typing in "16.00" for 16.00C span point.
6064 2022-11-15 Lab Calibration Lab Last calibration: 2022-04-19
AGC=99.1 SDM= DAC1= DAC2=
Ambient P=101.2kPa, Tair=23.0C, Tdew=2.87C
Zero Air at 1850psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9483 -> .9481 Offset: -.73ppm
CO2 span 1.0087 -> 1.0127 Offset: -2.76ppm
H2O zero 0.8444 -> .847 Offset: 16.3ppm
H2O span 1.0191 -> 1.0292 Offset: -.22ppm
Notes: Used AMP LI-7550 sn AIU-1039
6079 2022-11-02 Bouldin Alfalfa-ADEX Gas Analyzer Edge CO2/H2O Removed Lab none Removed from top of the Edge tower at 3.96m alg at the end of the 2022 advection experiment at BA
5853 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Gas Analyzer Edge CO2/H2O Installed on top of the Edge tower at 3.96m alg for the advection experiment at BA
5800 2022-04-19 Lab Calibration Lab Last calibration: 2021-12-09
AGC=100.4 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=25.8C, Tdew=7.5C
Zero Air at 650psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00
CO2 zero 0.9481 -> .9483 Offset: .85ppm
CO2 span 1.0091 -> 1.0087 Offset: .14ppm
H2O zero 0.8452 -> .8444 Offset: -4.9ppm
H2O span 1.0213 -> 1.0191 Offset: .05ppm
Notes:
5793 2022-04-13 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration and brought back to the lab. Replaced by sn 0065.
5637 2021-12-14 Lab none Installed Bouldin Corn none Installed at Bouldin Corn to replace sn 0041 for calibration.
5627 2021-12-09 Lab none Calibration Lab none Last calibration: 2021-08-31
AGC=100.7 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=25.9C, Tdew=5.9C
Zero Air at 100psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00
CO2 zero 0.9479 -> .9481 Offset: .73ppm
CO2 span 1.0063 -> 1.0091 Offset: -1.82ppm
H2O zero 0.8448 -> .8452 Offset: 2.4ppm
H2O span 1.0163 -> 1.0123 Offset: -.11ppm
Notes: <100psi of air left in the zero air tank, but the zero was stable for 2+ min after setting the zeros, so I feel comfortable the calibration was good.
5601 2021-12-02 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2182.
5439 2021-09-01 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0065 for calibration.
5436 2021-08-31 Lab Calibration Lab Last calibration: 2021-05-25
AGC=99.8 SDM=0 DAC1=- DAC2=-
Ambient P=99.7kPa, Tair=26.4C, Tdew=13.15C
Zero Air at 1800psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0
CO2 zero 0.9481 -> 0.9479 Offset: -0.78ppm
CO2 span 1.0041 -> 1.0063 Offset: -1.62ppm
H2O zero 0.8437 -> 0.8448 Offset: 7.1ppm
H2O span 1.0149 -> 1.0163 Offset: -0.04ppm
Notes:
5330 2021-06-16 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for due to bad pressure reading. Replaced by sn 75H-2180.
5288 2021-05-26 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 0418 for calibration.
5269 2021-05-25 Lab Calibration Lab Last calibration: 2020-12-02
AGC=99.5 SDM=0 DAC1=- DAC2=-
Ambient P=100.7kPa, Tair=26.2C, Tdew=10.8C
Zero Air at 350psi, AMP CA03923 474.19 at 500 Tdewspan=20.0
CO2 zero 0.9482 -> 0.9481 Offset: -0.4ppm
CO2 span 1.0111 -> 1.0041 Offset: 4.31ppm
H2O zero 0.8436 -> 0.8437 Offset: 0.7ppm
H2O span 1.0264 -> 1.0149 Offset: 0.3ppm
Notes:
5210 2021-03-16 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration and brought back to the lab. Replaced by sn 0073.
5023 2020-12-03 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond to replace sn 75H-2150 for calibration.
4995 2020-12-02 Lab Calibration Lab Last calibration: 2020-08-25
AGC=101 SDM= DAC1= DAC2=
Ambient P=100.7kPa, Tair=24.3C, Tdew=2.2C
Zero Air at 450psi, AMP CA4396 408.97 at 300 Tdewspan=20.00
CO2 zero 0.9479 -> .9482 Offset: 1.16ppm
CO2 span 1.0073 -> 1.0111 Offset: -2.03ppm
H2O zero 0.8441 -> .8436 Offset: -3.1ppm
H2O span 1.0268 -> 1.0264 Offset: .01ppm
Notes: 2nd time doing calibration today. The first calibration had large offsets, so I decided to redo it (CO2 zero offset 3.93, CO2 span offset -3.77, H2O zero offset -13.5, H2O span offset 0.39). The 2nd time looks much better with expected range of offsets, but I'm not sure why it was different.
5012 2020-12-01 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 0035.
4705 2020-08-26 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 75H-2514 for calibration.
4673 2020-08-25 Lab Calibration Lab Last calibration: 2020-05-13
AGC=99.6 SDM=0 DAC1=- DAC2=-
Ambient P=100.2kPa, Tair=23.3C, Tdew=14.3C
Zero Air at 700psi, AMP CA04396 408.97 at 500 Tdewspan=20.0
CO2 zero 0.9481 -> 0.9479 Offset: -0.8ppm
CO2 span 1.0141 -> 1.0073 Offset: 3.43ppm
H2O zero 0.8441 -> 0.8441 Offset: -0.5ppm
H2O span 1.0271 -> 1.0268 Offset: 0ppm
Notes:
4679 2020-08-21 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration. Replaced by sn 1594.
4459 2020-05-19 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End to replace sn 75H-2182 for calibration.
4447 2020-05-13 Lab Calibration Lab Last calibration: 2019-12-04
AGC=99.9 SDM= DAC1= DAC2=
Ambient P=100.9kPa, Tair=19.9C, Tdew=9.18C
Zero Air at 750psi, AMP CA4396 408.97 at 550 Tdewspan=16.00
CO2 zero 0.949 -> .9481 Offset: -3.44ppm
CO2 span 1.0182 -> 1.0141 Offset: 2.03ppm
H2O zero 0.8441 -> .8441 Offset: 0.1ppm
H2O span 1.0341 -> 1.0271 Offset: 0.15ppm
Notes:
4379 2020-02-20 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 75H-2668.
4304 2019-12-13 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Re-installed at Bouldin Corn after re-assembling tower.
4308 2019-12-11 Lab none Tested Lab none Tested for 2+ days in the lab connected to the lab 7550, and numbers looked ok.
4232 2019-12-04 Lab none Calibration Lab none Last calibration: 2019-11-12
AGC=100.5 SDM= DAC1= DAC2=
Ambient P=100.1kPa, Tair=20.3C, Tdew=10.11C
Zero Air at 850psi, AMP CA4396 408.97 at 550 Tdewspan=20.00
CO2 zero 0.9481 -> .9490 Offset: 3.19ppm
CO2 span 1.0091 -> 1.0182 Offset: -4.69ppm
H2O zero 0.8449 -> .8441 Offset: -4.7ppm
H2O span 1.0139 -> 1.0341 Offset: -.45ppm
Notes: This sensor was on the BC tower when it fell down because of a wind/rain storm on 2019-11-26. First calibration since it was brought back to the lab. The CO2 span is larger than I expected. More testing needed before putting it back up in the field.
4271 2019-12-02 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26.
4162 2019-11-14 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 75H-2669 for calibration.
4149 2019-11-12 Lab none Calibration Lab none Last calibration: 2019-07-24
AGC=99.5 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=26.7C, Tdew=11.3C
Zero Air at 950psi, AMP CA4396 408.97 at 700 Tdewspan=16.00
CO2 zero 0.9481 -> .9481 Offset: .26ppm
CO2 span 1.0049 -> 1.0091 Offset: -2.31ppm
H2O zero 0.8433 -> .8449 Offset: 10.0ppm
H2O span 1.0179 -> 1.0139 Offset: .08ppm
Notes:
4138 2019-10-31 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for calibration. Replaced by sn 1594.
4027 2019-07-25 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at MB to replace 75H-2182, which was brought back to the lab for calibration.
4016 2019-07-24 Lab none Calibration Lab none Last calibration: 2019-01-02
AGC=97.7 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=27.3C, Tdew=14.4C
Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00
CO2 zero 0.9489 -> .9481 Offset: -3.43ppm
CO2 span 1.0187 -> 1.0049 Offset: 7.19ppm
H2O zero 0.8434 -> .8433 Offset: -1.0ppm
H2O span 1.0334 -> 1.0179 Offset: .35ppm
Notes: When SS was 97.7, I reset it to 100.0
3976 2019-04-30 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from the Tower for calibration. Replaced by sn 1594. The Tower 7550 box died and was replaced by the Floor 7550 box. No reason to believe the Head was damaged by the box crash but might be worth checking.
3831 2019-01-04 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed on the Tonzi Tower to replace sn 0073 for calibration.
3822 2019-01-02 Lab none Calibration Lab none Last calibration: 2018-07-10
AGC=98.7 SDM=0 DAC1=0 DAC2=0
Ambient P=101.3kPa, Tair=19.2C, Tdew=.7C
Zero Air at 1400psi, AMP Air 408.97 at 875 Tdewspan=16.00
CO2 zero 0.9485 -> .9489 Offset: 1.45ppm
CO2 span 1.0039 -> 1.0187 Offset: -7.39ppm
H2O zero 0.8406 -> .8434 Offset: 18.2ppm
H2O span 1.0158 -> 1.0334 Offset: -.37ppm
Notes: Dewpoint generator was set at 16.03C and I adjusted it to 16.00 before beginning calibration. H2O zero offset larger than I expected, but H2O span seems fine.
3769 2018-10-25 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from WP and brought back to the lab for calibration. it was replaced by sn 75H-2182.
3594 2018-07-12 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at WP to replace sn 75H-2514, which was brought back to the lab for calibration.
3568 2018-07-10 Lab none Calibration Lab none Last calibration: 2018-07-03
AGC=98 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=26.8C, Tdew=C
Zero Air at 200psi, LBL Air 408.97 at 1400 Tdewspan=16.00
CO2 zero 0.9485 -> .9485 Offset: .03ppm
CO2 span 1.0054 -> 1.0039 Offset: .77ppm
H2O zero 0.8407 -> .8406 Offset: -.9ppm
H2O span 1.0418 -> 1.0158 Offset: .55ppm
Notes: Calibrated again b/c previous calibration on 2018-07-03 had a large H2O zero offset, but the offsets today are fine, so the calibration seemed fine.
3564 2018-07-03 Lab none Calibration Lab none Last calibration: 2018-03-05
AGC=97.8 SDM= DAC1=- DAC2=-
Ambient P=100.1kPa, Tair=23.25C, Tdew=11.75C
Zero Air at 245psi, AMP 408.97 at 1300 Tdewspan=16.0
CO2 zero 0.9485 -> 0.9485 Offset: 0.04ppm
CO2 span 1.0124 -> 1.0054 Offset: 3.53ppm
H2O zero 0.843 -> 0.8407 Offset: -14ppm
H2O span 1.0225 -> 1.0418 Offset: ppm
Notes: Robert in Training: Forgot to write down H2O span offset in C but it was -19mmol/m3. Large change in H2O zero offset. Daphne redid the calibration on 7/10 and the H2O zero offset was stable <2mmol/m3 change.
3547 2018-06-26 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from Tonzi Floor for calibration. Replaced by sn 0035.
3374 2018-03-06 Lab none Installed Tonzi-Floor Installed on Tonzi Floor to replace sn 1594, which was brought back to the lab for calibration.
3347 2018-03-05 Lab none Calibration Lab none Last calibration: 2017-05-05
AGC=98 SDM=0 DAC1=0 DAC2=0
Ambient P=101.4kPa, Tair=25.5C, Tdew=2.0C
Zero Air at 150psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.9847 -> .9485 Offset: 1.08ppm
CO2 span 1.0244 -> 1.0124 Offset: -1.35ppm
H2O zero 0.8434 -> .843 Offset: .7ppm
H2O span 1.1404 -> 1.0225 Offset: -.11ppm
Notes:
3276 2017-12-19 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from Bouldin Alfalfa for calibration. Replaced by sn 75H-2667.
3078 2017-09-19 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at BA tower to replace sn 0073 for calibration.
2969 2017-08-17 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 75H-2180
2763 2017-05-11 Lab none Installed Tonzi-Floor Installed at the Tonzi Floor tower. I uploaded a new config file and updated the pressure coefficients. The old IRGA head, sn 0418, was moved to a temporary mount on the same tower so we can compare the old and new heads.
2743 2017-05-05 Lab none Calibration Lab none Last calibration: 2016-08-15
AGC=99.9 SDM= DAC1= DAC2=
Ambient P=100.2kPa, Tair=25.5C, Tdew=7.8C
Zero Air at 1600psi, LBL Air 513.22 at 350 Tdewspan=16.00
CO2 zero 0.9504 -> .9847 Offset: -6.7ppm
CO2 span 1.001 -> 1.0244 Offset: -15.52ppm
H2O zero 0.8519 -> .8434 Offset: -54.8ppm
H2O span 1.003 -> 1.1404 Offset: -2.72ppm
Notes: First calibration since upgrade from LI-7500 to LI-7500A
2742 2017-05-03 Lab none Upgrade Lab none Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). After the upgrade, I attached this sensor to a LI-7550 Analyzer Interface Unit and manually input the CO2 coefficients (A, B, C, D, E, Z, XS) and H2O coefficients (A, B, C, ZX, S). I also updated the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength (Calibration icon > Signal Strength tab > Reset Signal Strength).
2540 2016-12-07 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower for calibration. Replaced by sn 0041.
2426 2016-08-23 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 0073 for calibration.
2396 2016-08-15 Lab none Calibration Lab none Last calibration: 2016-02-22
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.52kPa, Tair=22.05C, Tdew=12.36C
Zero Air at 500psi, God CO2 421.22 at 500 Tdewspan=16.00
CO2 zero 0.9505 -> 0.9504 Offset: -0.42ppm
CO2 span 1.0107 -> 1.0010 Offset: 0.63ppm
H2O zero 0.8522 -> 0.8519 Offset: -1.8ppm
H2O span 0.999 -> 1.0030 Offset: -0.08ppm
Notes: Exterior serial connector did not work.
2267 2016-05-10 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 0035.
2187 2016-02-23 Lab none Installed Tonzi-Floor Installed on the Tonzi Floor to replace sn 0073 for calibration.
2201 2016-02-22 Lab none Calibration Lab none Last calibration: 2015-07-15
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.33kPa, Tair=21.07C, Tdew=9.49C
Zero Air at 650psi, God CO2 421.22 at 650 Tdewspan=11.09
CO2 zero 0.9503 -> 0.9505 Offset: 0.98ppm
CO2 span 1.0097 -> 1.0107 Offset: -0.62ppm
H2O zero 0.8526 -> 0.8522 Offset: -0.06ppm
H2O span 1.001 -> 0.9990 Offset: 0.02ppm
Notes:
2035 2015-11-03 Tonzi-Floor Removed Lab none Removed from Tonzi Floor for calibration. Replaced by sn 0073.
1927 2015-08-12 Lab none Installed Tonzi-Floor Installed on the Floor Tower to replace sn 0073 for calibration.
1906 2015-07-15 Lab none Calibration Lab none Last calibration: 2015-01-06
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.34kPa, Tair=21.50C, Tdew=14.82C
Zero Air at 800psi, God CO2 421.22 at 850 Tdewspan=11.08
CO2 zero 0.9504 -> 0.9503 Offset: -0.50ppm
CO2 span 1.0086 -> 1.0097 Offset: -0.62ppm
H2O zero 0.851 -> 0.8526 Offset: 9.9ppm
H2O span 1 -> 1.0010 Offset: -0.02ppm
Notes: External serial port flakey
1855 2015-05-12 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed with the decommissioning of the Sherman Cow Pasture site.
1658 2015-01-07 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman Tower to replace sn 0418 for calibration.
1653 2015-01-06 Lab none Calibration Lab none Last calibration: 2014-10-02
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.62kPa, Tair=21.21C, Tdew=5.11C
Zero Air at 950psi, God CO2 421.22 at 1300 Tdewspan=15.95
CO2 zero 0.95 -> 0.9504 Offset: 1.58ppm
CO2 span 0.9989 -> 1.0086 Offset: -5.12ppm
H2O zero 0.8505 -> 0.8510 Offset: 3.0ppm
H2O span 0.9819 -> 1.000 Offset: -0.4ppm
Notes: External serial connector on the box did not work.
1643 2014-12-23 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 0041
1577 2014-10-02 Lab none Calibration Lab none Last calibration: 2014-06-24
AGC=54 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.76kPa, Tair=23.23C, Tdew=14.05C
Zero air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.81
CO2 zero 0.9506 -> 0.9500 Offset: -2.15ppm
CO2 span 1.0074 -> 0.9989 Offset: 4.78ppm
H2O zero 0.8506 -> 0.8505 Offset: -0.3ppm
H2O span 0.996 -> 0.9819 Offset: 0.3ppm
Notes: Exterior serial connector doesn't work. This calibration was done using the thermistor in the box and not on the calibration tube.
1579 2014-09-30 Lab none Installed Tonzi-Floor Installed on the Tonzi Floor tower to replace sn 1594 for calibration.
1573 2014-09-25 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman tower for calibration. Replaced by sn 0418.
1518 2014-06-26 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman tower to replace sn 0976 for calibration.
1465 2014-06-24 Lab none Calibration Lab none Last calibration: 2014-02-28
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.62kPa, Tair=21.30C, Tdew=11.17C
Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26
CO2 zero 0.9506 -> 0.9506 Offset: -0.12ppm
CO2 span 1.0078 -> 1.0074 Offset: 0.18ppm
H2O zero 0.8506 -> 0.8506 Offset: 0.3ppm
H2O span 0.9853 -> 0.9960 Offset: -0.19ppm
Notes:
1349 2014-06-11 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 1594
1227 2014-03-04 Lab none Installed Tonzi-Floor Installed on the Floor Tower to replace sn 0976 for calibration.
1223 2014-02-28 Lab none Calibration Lab none Last calibration: 2013-08-20
AGC=52 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=98.58kPa, Tair=19.15C, Tdew=8.52C
Zero air at 1400psi, God CO2 421.22 at 1500 Tdewspan=15.17
CO2 zero 0.9503 -> 0.9506 Offset: 0.88ppm
CO2 span 1 -> 1.0078 Offset: -4.12ppm
H2O zero 0.8499 -> 0.8506 Offset: 3.8ppm
H2O span 0.9929 -> 0.9853 Offset: 0.18ppm
Notes: H2O not zeroed last time - April 23, 2013 last zero
1153 2013-11-25 DWR Corn Removed Lab none This sensor was removed for calibration. It was replaced by sn 0073
1060 2013-08-21 Lab none Installed DWR Corn Installed on the Corn / Construction tower to replace sn 0976 for calibration
1062 2013-08-20 Lab none Calibration Lab none Last calibration: 2013-04-23
AGC=54 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.28kPa, Tair=24.51C, Tdew=13.40C
Zero Air at 1700psi, God CO2 421.22 at 1600 Tdewspan=15.19
CO2 zero 0.9504 -> 0.9503 Offset: -0.16ppm
CO2 span 1.0033 -> 1.0 Offset: 1.23ppm
H2O zero 0.8499 -> 0.8499 Offset: -0.06ppm
H2O span 0.9779 -> 0.9929 Offset: -0.33ppm
Notes:
1044 2013-08-13 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn0418
973 2013-04-24 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to swap out sn 0976 for calibration
985 2013-04-23 Lab none Calibration Lab none Last calibration: 2013-02-04
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.36kPa, Tair=26.27C, Tdew=-1.5C
Zero air at 1750psi, CO2 in air 392.30 at 1900 Tdewspan=15.03
CO2 zero 0.9508 -> 0.9504 Offset: -1.81ppm
CO2 span 1.0084 -> 1.0033 Offset: 2.55ppm
H2O zero 0.8575 -> 0.8499 Offset: -3.4ppm
H2O span 0.9872 -> 0.9779 Offset: 0.23ppm
Notes: Last H2O zero and span were slightly different than recorded. H2O zero was 0.8505 and span was 0.9892
963 2013-04-19 Tonzi-Floor Removed Lab none Removed from the Floor tower for calibration. Replaced by sn 1594
902 2013-02-06 Lab none Installed Tonzi-Floor Installed on the Floor Tower to replce 0976 for calibration
891 2013-02-04 Lab none Calibration Lab none Last calibration: 2012-08-02
AGC=52 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.2kPa, Tair=19.9C, Tdew=6.6C
dry air at 1500psi, CO2 421.22 at 1700 Tdewspan=15.04
CO2 zero 0.9513 -> 0.9508 Offset: -1.84ppm
CO2 span 1.0044 -> 1.0084 Offset: 1.57ppm
H2O zero 0.8472 -> 0.8575 Offset: 0.54ppm
H2O span 0.9696 -> 0.9872 Offset: 0.96ppm
Notes: by Siyan
886 2013-01-30 DWR Corn Removed Lab none Removed from the DRW-Corn tower for calibration. Replaced with sn 0073
722 2012-08-02 Lab none Calibration Lab none Last calibration: 2012-05-17
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.49kPa, Tair=21.28C, Tdew=11.87C
Zero Air at 1900psi, CO2 in air 436.53 at 1400 Tdewspan=15.08
CO2 zero 0.9546 -> 0.9513 Offset: -12.72ppm
CO2 span 1.0182 -> 1.0044 Offset: 7.97ppm
H2O zero 0.8346 -> 0.8472 Offset: 72.5ppm
H2O span 0.8284 -> 0.9696 Offset: -3.2ppm
Notes: Relatively large offsets seem to counter similar large offsets seen at last calibration. H2O zero offset may be a typo.
720 2012-08-02 Lab none Installed DWR Corn Installed on the DWR Corn Tower to replace sn 0065 for calibration and upgrade.
721 2012-07-25 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 0041.
602 2012-05-18 Lab none Installed Vaira Gas Analyzer CO2/H2O Swapped onto the to tower to replace sn 0073 for calibration
569 2012-05-17 Lab none Calibration Lab none Last calibration: 2012-05-16
AGC=54 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.19kPa, Tair=23.04C, Tdew=8.4C
dry air at 16psi, CO2 436.53 at 1500 Tdewspan=13.7
CO2 zero 0.9512 -> 0.9546 Offset: 13.62ppm
CO2 span 0.8486 -> 1.0182 Offset: 17.44ppm
H2O zero 1.0024 -> 0.8346 Offset: -1.89ppm
H2O span 0.999 -> 0.8284 Offset: 5.11ppm
Notes: calibrated in the lab; >10ppm offset although the IRGA was just fixed by Li-cor. Joe called Li-cor, but no clue.
566 2012-05-16 Lab none Calibration Lab none Last calibration: 2011-12-13
AGC=54 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.19kPa, Tair=23.04C, Tdew=8.4C
dry air at 16psi, CO2 436.53 at 1500 Tdewspan=13.7
CO2 zero 0.9597 -> 0.9512 Offset: ppm
CO2 span 0.984 -> 0.8486 Offset: ppm
H2O zero 0.7828 -> 1.0024 Offset: ppm
H2O span 0.9341 -> 0.9990 Offset: ppm
Notes: recently fixed by Li-cor
double check in the lab conditions
no calibration applied yet
562 2012-05-10 Factory Returned Lab none Returned from repair at Licor on RMA 168677. This head went in for a bad source. Because of its age and time and cost of recalibration we had the detector and chopper motor replaced as well. New calibrations coefficients (2012-04-25): CO2 Cal: A = 1.48568E2, B = 4.72425E3, C = 4.31507E7, D = -1.33645E10, E = 1.69762E12, XS = 0.0002, Z = 9.00000E-4. H2O Cal: A = 5.38386E3, B = 3.60338E6, C = 9.95283E7, XS = -0.0002, Z = 1.53000E-2. Pressure Cal: A0 = 10.485, A1 = 26.036. Zero/Span: CO2 zero = 0.9512, CO2 span = 1.0024, H2O zero = 0.8486, H2O span = 0.9990
561 2012-03-27 Lab none Shipped Factory Shipped to Licor for repair on RMA 168677. The source was intermittently bad.
1212 2012-02-28 Sherman Island Gas Analyzer CO2/H2O Removed Lab none The LI-7500 sn 0042 was malfunctioning - CO2 values were 5.85, and water was above 40,000. We pulled it from the field, and it was not working in the lab either
1211 2011-12-14 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman tower to replace sn 0976 for calibration.
459 2011-12-13 Lab none Calibration Lab none Last calibration: 2011-09-27
AGC=73 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.67kPa, Tair=23.42C, Tdew=4.10C
Zero Air at 800psi, CO2 in Air 436.53 at 1900 Tdewspan=10.08
CO2 zero 0.9592 -> 0.9597 Offset: 2.0ppm
CO2 span 0.9758 -> 0.9840 Offset: 5.23ppm
H2O zero 0.7852 -> 0.7828 Offset: -6.0ppm
H2O span 0.9532 -> 0.9341 Offset: 0.32ppm
Notes: The signal from this IRGA seems noisey and the AGC is pretty high.
424 2011-10-13 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower due to no signal and reporting 100% AGC even with clean lenses.
411 2011-09-29 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on Vaira tower to replave sn 0073 for calibration.
427 2011-09-27 Lab none Calibration Lab none Last calibration: 2011-06-15
AGC=75 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.8kPa Tair=25.14C, Tdew=11.46C
zero air at 1000psi, CO2 in air 405.85ppm 1500psi Tdewspan=13.4
CO2 zero 0.9587 -> 0.9592 Offset: 1.72ppm
CO2 span 0.9796 -> 0.9758 Offset: 2.1ppm
H2O zero 0.7886 -> 0.7858 Offset: -20mmol/m3
H2O span 0.9667 -> 0.9532 Offset: 0.37C
Notes: The AGC values seemed high at between 70 and 75%. The internal chemicals were changed and allowed to run overnight.
406 2011-08-25 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman Tower for calibration. Replaced with sn 0976.
571 2011-06-21 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Swapped onto the to tower to replace sn 0073 for calibration
428 2011-06-15 Lab none Calibration Lab none Last calibration: 2011-02-02
AGC=68 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.3kPa Tair=21.98C, Tdew=11.9C
zero air at 1450psi, co2 in air 597.4ppm 100psi Tdewspan=16.9
CO2 zero 0.9605 -> 0.9587 Offset: -8.14ppm
CO2 span 0.9855 -> 0.9796 Offset: 11.6ppm
H2O zero 0.7877 -> 0.7886 Offset: -2.9mmol/m3
H2O span 1.1467 -> 0.9667 Offset: 4.62C
Notes: I initiall tried to calibrate on 6/14 but was rushed and got some odd behavior so did a better calibration on 6/15. The offsets are estimates based on the two calibrations.
365 2011-05-25 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from the Mayberry tower for calibration. Replaced by sn 0065.
295 2011-02-03 Lab none Installed Mayberry Gas Analyzer CO2/H2O This sensor was installed on the tower. sn 0073 was removed for calibration.
294 2011-02-02 Lab none Calibration Lab none Last calibration Nov 16, 2010
AGC=67, SDM=7, DAC1 0-1000 H2O, DAC2 10-25 CO2
P=101.84 kPa Tair=21.06C, Tdew=~-1C
Zero air no scrub 1600psi, CO2 597.4ppm 150psi, Tdew=-7.1C
Precal CO2 zero=0.9622, span=0.9827
Postcal CO2 zero=0.9605, span=0.9855
Precal H2O zero=0.7829, span=0.9693
Postcal H2O zero=0.7877, span=1.1467
CO2 zero off by -6.13ppm, span off by -2.3ppm
H2O zero off by +30.7mmol/m3, span off by -2.4C

The weather is warm and dry making the dew point very low and difficult to calibrate.
286 2011-01-28 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none This sensor was removed from the Twitchell tower for calibration in the lab. sn 1594 was swapped in.
255 2010-11-18 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O This IRGA was installed on the Twitchell tower after the move to replace sn 0035 for calibration.
251 2010-11-16 Lab none Calibration Lab none Last calibration Aug 13, 2010
AGC=66, SDM=7, DAC1 0-1000 H2O, DAC2 10-25 CO2
P=100.96kPa Tair=25.29C, Tdew=15C
Zero air no scrubs 1700psi, CO2 597.4ppm 200psi, Tdew=9.7C
Precal CO2 zero=0.9676, span=0.9814
Postcal CO2 zero=0.9622, span=0.9827
Precal H2O zero=0.7680, span=0.1.0441
Postcal H2O zero=0.7829, span=0.9693
CO2 zero off by -20.10ppm, span off by -1ppm
H2O zero off by +113.7mmol/m3, span off by +1.62C
228 2010-10-07 Tonzi-Floor Removed Lab none Removed from the Floor Tower for calibration. sn 0041 installed in place.
188 2010-08-14 Lab none Installed Tonzi-Floor Installed on the Tonzi floor tower by Siyan after lab calibration.
125 2010-06-03 Sherman Island-Levee Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman Levee tower for lab calibration.
1198 2010-04-06 Lab none Installed Sherman Island-Levee Gas Analyzer CO2/H2O Installed on the new Sherman Levee tower.
98 2010-04-05 Lab none Calibration Lab none Last calibration Dec 8, 2009
AGC=68, Pressure=100.491kPa Temperature=20.42C
CO2 span=597.4ppm, H2O span=17.6C, N2 650psi, CO2 300psi
Precal CO2 zero=0.9615, span=0.9897
Postcal CO2 zero=0.9620, span=0.9862
Precal H2O zero=0.7844, span=0.9773
Postcal H2O zero=0.7840, span=1.0151
CO2 zero off by +1.81ppm, span off by +2.6+-0.2ppm
H2O zero off by -0.05g/m3, span off by -0.77C
94 2010-03-18 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none This LI7500 was removed from Twitchell tower for calibration.
1197 2009-12-09 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell Rice tower to replace 0041 for calibration.

Not sure what the note on 2009-09-09 indicates since this sensor was only installed at the rice on 20091209 and prior to that sn 0041 was installed at the rice on 20090731 (there are no indications in the field notes that would suggest this sensor was installed at the rice on 2009-09-09).
16 2009-12-08 Lab none Calibration Lab none Embedded software updated to v3.0.1
Last calibration Sept 15 2009
AGC=67, Pressure=101.01kPa Temperature=17.3C
CO2 span=597.4ppm, H2O span=14.5C
Precal CO2 zero=0.9609, span=0.9818
Postcal CO2 zero=0.9615, span=0.9897
Precal H2O zero=0.7855, span=1.0086
Postcal H2O zero=0.7844, span=0.9773
CO2 zero off by 2.2ppm, span off by ~7ppm
H2O span off by ~0.7C
15 2009-12-07 Lab none Upgrade Lab none Internal chemicals changed.
AGC = 69, Embedded software v3.0.0
Last calibration Sep 15 2009
77 2009-12-04 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from permanent tower for calibration.
76 2009-09-09 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Freshly calibrated, installed.
195 2009-01-29 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman tower after lab calibration.
4010 2009-01-08 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 0976.
5259 2007-12-18 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Island for calibration. Replaced by sn 0035.
5248 2007-04-05 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed with the initial setup of the site.
3952 2004-04-07 Lab none Calibration Lab none Calibrated in the lab after being recently returned from Licor. There is a change in the power wiring. Red/white is positive, green is negative. They also changed the delay to 240 ms.
4924 2003-10-02 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0065 for calibration.
4919 2003-06-20 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0065 for calibration.
4908 2003-05-09 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0073.
4903 2003-04-06 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0418 for calibration.
4883 2002-12-23 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 0035 for calibration.
4806 2002-11-05 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0418.
4877 2002-09-20 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0418 for calibration.
4870 2002-08-15 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0418.
4849 2002-07-19 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0073 for calibration.
4844 2002-07-11 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 0065.
4813 2002-06-06 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0073 for calibration.
4838 2002-05-09 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0065.
4837 2002-04-18 Vaira Gas Analyzer CO2/H2O Location Vaira Gas Analyzer CO2/H2O Tried to swap out for calibration but replacement (sn 0073) was reading out of range. Left in place
4820 2002-02-28 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0041 for calibration.
3805 2002-02-08 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from the Tonzi Tower for calibration. Replaced by sn 0073.
4774 2001-11-10 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Tonzi to replace sn 0041 for calibration.
4771 2001-08-17 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 0065.
4766 2001-07-19 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed on Tonzi Tower to replace sn 0035 for calibration
4763 2001-06-28 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi Floor for calibration. Replaced with sn 0073.
3163 2001-04-18 Lab none Installed Tonzi-Floor Installed on the Floor tower to replace existing IRGA for calibration.
4539 2000-12-14 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 0035 for calibration: H2O = 0-1000mmol/m3, CO2 = 13-20mmol/m3.
3446 2000-11-19 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace existing 7500 sn 0035 for calibration: H2O 0-1000 mmol/m3, CO2 13-20 mmol/m3
2348 2000-11-05 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed for calibration. Replaced by sn 0035
4890 2000-10-12 Lab none Calibration Lab none Zero with N2 was 2.0, after running several weeks in the lab.Cal gas 402.25 read 411.1 ppm at P=101.31 kPa and T 23.17 C. Licor Dewpoint sensor read 14.57 Tdew. Vaisala standard read 13.9 C. We are using the Vaisala as the standard for the experiment. Rezeroed and spanned

D/A #1, H2O mmol m-3
0v=0
5 v=1000

D/A #2, CO2 mmol m-3
0v=13 mmol m-3
5v=20 mmol m-3
2347 2000-10-07 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed with the initial set up of the Vaira site. D/A settings: H2O 0-1000 mmol/m3, CO2 13-20 mmol/m3