Field Notes
<--2003-04-05 00:00:00 | 2003-04-08 09:20:00-->Other sites visited today: Tonzi
Phenocam link: Searching...
Vaira: 2003-04-06 17:00:00 (DOY 96)
Author: Ted Hehn
Others:
Summary: Replace Li7500s
17:00 Arrive—all appears well Gillsonic functioning and all data present with 05:00 hr resets. Swapped LI7500s sn 0418 came off and sn 0042 went on. I removed a good power connector from #418 and installed it on #42. Note that the intermittent failure problem with #42 was caused by a bad bulkhead connector on the white translator box. 18:49 Leave Vaira site |
Field Data
No tank data
No VR Well data found
No TZ Moisture Point data found
No VR Moisture Point data found
No TZ grass heigth found
No VR grass heigth found
No VR Tree DBH found
No water potential data found
No VR Grass Biomass data found
No VR OakLeaves Biomass data found
No VR PineNeedles Biomass data found
0 photos found
9 sets found
16 VR_met graphs found
No data found for VR_flux.
No data found for VR_soilco2.
No data found for VR_pond.
Can't check TZVR_mixed
No data found for VR_fd.
No data found for VR_processed.
No data found for VR_arable.
No data found for VR_soilvue.