View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7500A
Serial No:0418
Tag No:20021003260
Acquisition Date:2002-08-19
Project:NULL
Location on 2024-02-12:Gilbert Tract, NULL
Notes: NULL

Documents

Calibration: cal_LI-7500_sn75H-0418.pdf
Manual: LI-7500A_Manual_V1_200912.pdf
Manual: LI-7500A_Manual_v2_201104.pdf
Manual: LI-7500A_Manual_V4_201303.pdf
Manual: LI-7500A_Manual_V5_201310.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6566 2024-02-12 Lab none Installed Gilbert Tract Gas Analyzer CO2/H2O Installed at Gilbert Tract to replace sn 75H-2176 for calibration.
6564 2024-02-09 Lab Calibration Lab Last calibration: 2023-10-23
AGC=101.3 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.68kPa, Tair=23.2C, Tdew=5.6C
Zero Air at 475psi, AMP CA04119 524.77 at 700 Tdewspan=20.00
CO2 zero 0.9072 -> .9075 Offset: 1.41ppm
CO2 span 1.0066 -> 1.0063 Offset: .14ppm
H2O zero 0.9331 -> .9317 Offset: -7.5ppm
H2O span 1.0052 -> 1.002 Offset: .08ppm
Notes: Vertical orientation. H2O zero only took 10min to stabilize
6538 2024-01-18 Hill Slough Gas Analyzer CO2/H2O Removed Lab none Removed from Hill Slough for calibration. Replaced by sn 1594.
6455 2023-10-26 Lab none Installed Hill Slough Gas Analyzer CO2/H2O Installed at Hill Slough to replace sn 75H-2176 for calibration.
6444 2023-10-23 Lab Calibration Lab Last calibration: 2023-07-19
AGC=97.3 SDM= DAC1= DAC2=
AIU sn AIU-1569, Presure A0=58.004, A1=15.479
Ambient P=100.1kPa, Tair=23.1C, Tdew=29.0C
Zero Air at 700psi, AMP CA04119 524.77 at 775 Tdewspan=20.00
CO2 zero 0.907 -> .9072 Offset: .71ppm
CO2 span 1.0069 -> 1.0066 Offset: .14ppm
H2O zero 0.933 -> .9331 Offset: .6ppm
H2O span 1.0049 -> 1.0052 Offset: -.01ppm
Notes: Vertical orientation. Changed SS from 97.3 to 100 before calibration.
6435 2023-10-11 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 0042.
6349 2023-07-24 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 75H-2669 for calibration.
6337 2023-07-19 Lab none Calibration Lab none Last calibration: 2023-03-29
AGC=97.4 SDM= DAC1= DAC2=
AIU sn AIU-1569, Presure A0=58.004, A1=15.479
Ambient P=100.5kPa, Tair=22.6C, Tdew=12.4C
Zero Air at 1000psi, AMP CA04119 524.77 at 950 Tdewspan=16.00
CO2 zero 0.9075 -> .907 Offset: -1.98ppm
CO2 span 1.0076 -> 1.0069 Offset: .51ppm
H2O zero 0.9326 -> .9330 Offset: 2.4ppm
H2O span 1.0052 -> 1.0049 Offset: 0.0ppm
Notes: Vertical orientation, H2O span right on
6321 2023-07-11 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for calibration. Replaced by sn 75H-2182.
6239 2023-04-03 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 0035 for calibration.
6227 2023-03-29 Lab Calibration Lab Last calibration: 2022-12-19
AGC=97.8 SDM= DAC1= DAC2=
Ambient P=99.8kPa, Tair=18.7C, Tdew=6.56C
Zero Air at 1300psi, AMP CA04119 524.77 at 1150 Tdewspan=16.00
CO2 zero 0.907 -> .9075 Offset: 2.10ppm
CO2 span 1.0059 -> 1.0076 Offset: -1.13ppm
H2O zero 0.933 -> .9326 Offset: -2.1ppm
H2O span 1.0067 -> 1.0052 Offset: .03ppm
Notes:
6217 2023-03-16 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 1594.
6138 2022-12-21 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi to replace sn 75H-2669 for calibration.
6135 2022-12-19 Lab Calibration Lab Last calibration: 2022-09-12
AGC=97.6 SDM=0 DAC1=- DAC2=-
Ambient P=101.4kPa, Tair=19.05C, Tdew=3.36C
Zero Air at 1650psi, AMP CA04119 524.77 at 1300 Tdewspan=16.00
CO2 zero 0.9086 -> 0.9070 Offset: -6.3ppm
CO2 span 1.005 -> 1.0059 Offset: -0.57ppm
H2O zero 0.9317 -> 0.9330 Offset: 6.5ppm
H2O span 1.0065 -> 1.0067 Offset: -0.01ppm
Notes: We have been watching this one as its SS seemed to drift after overhaul at Licor in the spring. The zero offsets are a little high than is common but otherwise seems okay.
6132 2022-12-06 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi Floor for calibration. Replaced by sn 75H-2180.
5986 2022-09-12 Lab Calibration Lab Last calibration: 2022-08-30
AGC=94.7 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=21.8C, Tdew=15.1C
Zero Air at 1900psi, AMP CA04119 524.77 at 1400 Tdewspan=20.0
CO2 zero 0.9082 -> 0.9086 Offset: 1.4ppm
CO2 span 1.0000 -> 1.0050 Offset: -3.47ppm
H2O zero 0.9313 -> 0.9317 Offset: 2.4ppm
H2O span 1.0008 -> 1.0065 Offset: -0.12ppm
Notes: This head had a drift of SS (and CO2, H2O) after refit by Licor in March. It was calibrated/tested in the lab several times in August. Full calibration on 8/30 but not all the information written down. 8/30 coefficients entered as 'existing' and it seems stable since then. We will try it in the field again.
5959 2022-08-25 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 75H-2669.
5746 2022-03-24 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 75H-2180 for calibration.
5742 2022-03-22 Lab none Calibration Lab none Last calibration: 2022-03-08
AGC=100.0 SDM=0 DAC1=- DAC2=-
Ambient P=100.9kPa, Tair=28.8C, Tdew=10.43C
Zero Air at 900psi, AMP CA04119 524.77 at 1650 Tdewspan=20.0
CO2 zero 0.9108 -> 0.9104 Offset: -31.5ppm
CO2 span 1.0024 -> 1.0063 Offset: -2.77ppm
H2O zero 0.9116 -> 0.9144 Offset: 14.9ppm
H2O span 1.0069 -> 1.0127 Offset: -0.13ppm
Notes: First lab calibration after Licor replaced chopper motor, light source and detector and factory calibration. All the new configuration coefficients were entered by hand. We always see large offset on first cal - not sure why.
5731 2022-03-11 Factory none Received Lab none Received back from Licor. The source lamp was defective and we requested that the lamp, chopper motor, and detector all to be replaced. The sensor was recalibrated at the factory.
5732 2022-03-08 Lab Calibration Lab Last calibration: 2021-09-29
AGC= SDM= DAC1= DAC2=
Ambient P=kPa, Tair=C, Tdew=C
at psi, 750 at Tdewspan=12.00
CO2 zero 0.8995 -> .9108 Offset: ppm
CO2 span 1 -> 1.0024 Offset: ppm
H2O zero 0.849 -> .9116 Offset: ppm
H2O span 0.9783 -> 1.0069 Offset: ppm
Notes: Licor factory calibration
5626 2021-12-08 Lab none Shipped Factory none Shipped back to Licor for repair
5619 2021-12-01 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry and brought back to the lab for testing. It was not giving any data for CO2, H2O or SS. Replaced by sn 1594.
5553 2021-10-21 Mayberry Gas Analyzer CO2/H2O Adjusted Mayberry Gas Analyzer CO2/H2O Rest SS from 104 to 100 after cleaning
5485 2021-10-07 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 75H-2180 for calibration.
5474 2021-09-29 Lab Calibration Lab Last calibration: 2021-09-21
AGC=100 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=26.1C, Tdew=11.5C
Zero Air at 1500psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00
CO2 zero 0.8992 -> .8995 Offset: .96ppm
CO2 span 0.9972 -> 1 Offset: -2.11ppm
H2O zero 0.849 -> .849 Offset: .1ppm
H2O span 0.9774 -> .9783 Offset: -.01ppm
Notes: 2nd calibration since chemical swap. Not much change since the 1st calibration, which was the day after the chemical swap. CO2 span was the only one that changed significantly (although still a pretty small change).
5465 2021-09-21 Lab Calibration Lab Last calibration: 2021-09-20
AGC=99.3 SDM= DAC1= DAC2=
Ambient P=100.7kPa, Tair=29C, Tdew=11.1C
Zero Air at 1600psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00
CO2 zero .9004 -> .8992 Offset: -4.8ppm
CO2 span 1.0016 -> .9972 Offset: 3.22ppm
H2O zero .8444 -> .8490 Offset: 27.3ppm
H2O span .9808 -> .9774 Offset: .08ppm
Notes: I swapped the chemicals immediately before this calibration run. The chemicals were a tad sticky at the top but after a bit of poking with a screwdriver it was easy to dump the chemicals out into the waste jar. I don't think the chemicals needed to be changed necessarily. Before inserting the new chemical bottles, I ran some zero air into the headspace. This run was just to see what the numbers looked like. I will let it run on a bench for a week and calibrate again next week.
5464 2021-09-20 Lab Calibration Lab Last calibration: 2021-09-13
AGC=98 SDM= DAC1= DAC2=
Ambient P=100.3kPa, Tair=26.6C, Tdew=11.3C
Zero Air at 1700psi, at Tdewspan=
CO2 zero 0.8992 -> Offset: ppm
CO2 span 0.9956 -> Offset: ppm
H2O zero 0.8484 -> Offset: 26ppm
H2O span 0.9749 -> Offset: ppm
Notes: Checked the zero and the H2O offset is still around 26 mmol/m3. I'm going to try changing the chemicals and calibrating again.
5463 2021-09-13 Lab Calibration Lab Last calibration: 2021-06-01
AGC=96.8 SDM= DAC1= DAC2=
Ambient P=99.9kPa, Tair=26.4C, Tdew=14.6C
Zero Air at 1725psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00
CO2 zero 0.9004 -> .8992 Offset: -4.69ppm
CO2 span 1.0016 -> .9956 Offset: -4.2ppm
H2O zero 0.8444 -> .8484 Offset: 23.0ppm
H2O span 0.9808 -> .9749 Offset: .15ppm
Notes: Large H2O zero offset. Check again next week.
5442 2021-09-01 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 75H-2669.
5300 2021-06-02 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 75H-2176 for calibration.
5273 2021-06-01 Lab Calibration Lab Last calibration: 2021-03-03
AGC=100 SDM=0 DAC1=- DAC2=-
Ambient P=100.5kPa, Tair=21.8C, Tdew=10.7C
Zero Air at 300psi, AMP CA04119 524.77 at 1800 Tdewspan=16.0
CO2 zero 0.9003 -> 0.9004 Offset: 0.6ppm
CO2 span 0.9978 -> 1.0016 Offset: -2.67ppm
H2O zero 0.8461 -> 0.8444 Offset: -9.7ppm
H2O span 0.9906 -> 0.9808 Offset: 0.2ppm
Notes: I reset the SS from 95 to 100 before calibration.
5287 2021-05-26 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for calibration. Replaced by sn 0042.
5163 2021-03-04 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 0073 for calibration.
5158 2021-03-03 Lab Calibration Lab Last calibration: 2020-11-24
AGC=97.7 SDM= DAC1= DAC2=
Ambient P=99.3kPa, Tair=25.7C, Tdew=8.2C
Zero Air at 450psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.8995 -> .9003 Offset: 3.01ppm
CO2 span 1.0034 -> .9978 Offset: 3.49ppm
H2O zero 0.8465 -> .8461 Offset: -2.0ppm
H2O span 0.9847 -> .9906 Offset: -.14ppm
Notes: I had to add more water the dewpt generator to get it to stay at the set dewpoint, otherwise it would just slowly decline forever. Calibrated earlier in the day, but I had forgotten to update the pressure coefficients in the config file for LI-7550 sn AIU-614. With the incorrect coefficients, offsets were (CO2 zero = 0.78, CO2 span = 0.23, H2O zero = -4.6, H2O span = 0.04).
5153 2021-03-02 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2667.
4998 2020-12-01 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 75H-2176 for calibration.
4987 2020-11-24 Lab Calibration Lab Last calibration: 2020-05-22
AGC=96.5 SDM= DAC1= DAC2=
Ambient P=100.9kPa, Tair=24.3C, Tdew=8.4C
Zero Air at 550psi, AMP CA4396 408.97 at 375 Tdewspan=2
CO2 zero 0.8992 -> .8995 Offset: 1.24ppm
CO2 span 0.9991 -> 1.0034 Offset: -2.35ppm
H2O zero 0.8462 -> .8465 Offset: 1.5ppm
H2O span 0.9764 -> .9847 Offset: -.19ppm
Notes:
4721 2020-08-28 Sherman Wetland-Temp1 Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Wetland-Temp1 for calibration. Replaced by sn 75H-2180.
4490 2020-05-21 Lab none Installed Sherman Wetland-Temp1 Gas Analyzer CO2/H2O Installed at Sherman Wetland Temp Tower to replace sn 75H-2514 for calibration.
4479 2020-05-20 Lab Calibration Lab none Last calibration: 2020-02-19
AGC=94.4 SDM=0 DAC1=- DAC2=-
Ambient P=100.9kPa, Tair=21.0C, Tdew=10.0C
Zero Air at 800psi, AMP CA04396 408.97 at 600 Tdewspan=16.0
CO2 zero 0.8994 -> 0.8992 Offset: -0.8ppm
CO2 span 0.9987 -> 0.9991 Offset: -0.27ppm
H2O zero 0.8456 -> 0.8462 Offset: 3.3ppm
H2O span 0.9845 -> 0.9764 Offset: 0.2ppm
Notes:
4460 2020-05-19 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond and brought back to the lab for calibration. Replaced by sn 0041.
4385 2020-02-20 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond to replace sn 75H-2667 for calibration.
4393 2020-02-19 Lab Calibration Lab Last calibration: 2019-11-18
AGC=93.2 SDM= DAC1= DAC2=
Ambient P=100.8kPa, Tair=28.1C, Tdew=8.2C
Zero Air at 800psi, AMP CA4396 408.97 at 550 Tdewspan=16.00
CO2 zero 0.8991 -> .8994 Offset: 1.12ppm
CO2 span 0.9988 -> .9987 Offset: -.74ppm
H2O zero 0.8482 -> .8456 Offset: -5.4ppm
H2O span 0.9812 -> .9845 Offset: -.06ppm
Notes:
4374 2020-02-12 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from Tonzi Floor for calibration. Replaced by sn 1594.
4169 2019-11-19 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed on Tonzi Floor after testing and calibration in the lab.
4155 2019-11-18 Lab none Calibration Lab none Last calibration: 2019-08-21
AGC=94.4 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=25.2C, Tdew=8.5C
Zero Air at 950psi, AMP CA04396 408.97 at 600 Tdewspan=20.0
CO2 zero 0.899 -> 0.8991 Offset: 0.35ppm
CO2 span 0.9976 -> 0.9988 Offset: -0.72ppm
H2O zero 0.8466 -> 0.8482 Offset: 9.4ppm
H2O span 0.9818 -> 0.9812 Offset: 0.03ppm
Notes: Removed from Vaira due to no readings but has test fine in the lab - no missing data in about a week.
4152 2019-11-13 Lab none Tested Lab none Ran overnight in the lab on the bench and the 30-minute data looked fine for CO2 and H2O. I'll let it run for a few more days in a sunny window.
4146 2019-11-07 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira due to failure - probably bad source or detector. Replaced by sn 75H-2180 from Tonzi Floor.
4067 2019-08-22 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on Vaira to replace sn 75H-2668 for calibration.
4059 2019-08-21 Lab none Calibration Lab none Last calibration: 2019-04-22
AGC=93.7 SDM=0 DAC1=- DAC2=-
Ambient P=100.3kPa, Tair=26.98C, Tdew=14.83C
Zero Air at 1000psi, AMP CO2 408.97 at 800 Tdewspan=20.0
CO2 zero 0.8992 -> 0.8990 Offset: -0.55ppm
CO2 span 1.0011 -> 0.9976 Offset: 1.71ppm
H2O zero 0.8464 -> 0.8466 Offset: -0.8ppm
H2O span 0.9866 -> 0.9818 Offset: 0.11ppm
Notes:
4041 2019-08-07 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from WP and brought back to the lab for calibration. It was replaced by sn 75H-2182.
3948 2019-04-25 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on West Pond to replace sn 75H-2668 for calibartion
3929 2019-04-22 Lab none Calibration Lab none Last calibration: 2019-01-07
AGC=95 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=24.2C, Tdew=11.4C
Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00
CO2 zero 0.8992 -> .8992 Offset: -.04ppm
CO2 span 1.0033 -> 1.0011 Offset: 1.21ppm
H2O zero 0.8465 -> .8464 Offset: -.4ppm
H2O span 0.9832 -> .9866 Offset: -.07ppm
Notes:
3921 2019-04-18 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from SW and brought back to the lab for calibration. It was replaced by sn 75H-2150.
3841 2019-01-18 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at SW to replace 75H-2180, which was brought back to the lab for calibration.
3836 2019-01-07 Lab none Calibration Lab none Last calibration: 2018-10-10
AGC=97.2 SDM=0 DAC1=0 DAC2=0
Ambient P=100.9kPa, Tair=17.9C, Tdew=9.7C
Zero Air at 1200psi, AMP Air 408.97 at 900 Tdewspan=16.00
CO2 zero 0.8982 -> .8992 Offset: 3.69ppm
CO2 span 0.9965 -> 1.0033 Offset: -3.58ppm
H2O zero 0.8454 -> .8465 Offset: 6.1ppm
H2O span 0.9813 -> .9832 Offset: -.03ppm
Notes:
3826 2019-01-04 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 75H-2176.
3741 2018-10-17 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on Vaira to replace sn 0065 that had no CO2, H2O or signal strength.
3731 2018-10-10 Lab none Calibration Lab none Last calibration: 2018-04-13
AGC=93.2 SDM=0 DAC1=0 DAC2=0
Ambient P=99.98kPa, Tair=29.11C, Tdew=15.00C
Zero Air at 1500psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9001 -> .8982 Offset: -7.43ppm
CO2 span 1.0057 -> .9965 Offset: 4.86ppm
H2O zero 0.8426 -> .8454 Offset: 16.8ppm
H2O span 0.9892 -> .9813 Offset: 0.16ppm
Notes: calibrated by Katrina
3726 2018-10-04 East Pond Gas Analyzer CO2/H2O Removed Lab none Removed from the East Pond tower for calibration. Replaced by sn 0035.
3558 2018-06-28 East Pond Gas Analyzer CO2/H2O Moved East Pond Gas Analyzer CO2/H2O Sensor moved up 26cm to try to address rotation angle issues. The water level was 68cm and the sensor was 540cm above water level.
3401 2018-04-17 Lab none Installed East Pond Gas Analyzer CO2/H2O Installed at EP tower as part of initial site set up.
3395 2018-04-13 Lab none Calibration Lab none Last calibration: 2017-11-27
AGC=96.6 SDM=0 DAC1=0 DAC2=0
Ambient P=101.9kPa, Tair=25.1C, Tdew=2.5C
Zero Air at 100psi, LBL Air 408.97 at 1500 Tdewspan=16.00
CO2 zero 0.8998 -> .9001 Offset: 1.28ppm
CO2 span 1.006 -> 1.0057 Offset: 1.14ppm
H2O zero 0.8437 -> .8426 Offset: -6.1ppm
H2O span 0.9821 -> .9892 Offset: -.15ppm
Notes:
3363 2018-03-19 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from MB for calibration, replaced with SN 75H-2150.
3223 2017-11-30 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at MB to replace sn 75H-2669.
3202 2017-11-27 Lab none Calibration Lab none Last calibration: 2017-07-02
AGC=97.1 SDM=0 DAC1=0 DAC2=0
Ambient P=100.9kPa, Tair=22.8C, Tdew=6.8C
Zero Air at 350psi, LBL Air 408.97 at 1750 Tdewspan=16.00
CO2 zero 0.8995 -> .8998 Offset: .87ppm
CO2 span 1.0066 -> 1.006 Offset: .4ppm
H2O zero 0.8436 -> .8437 Offset: .2ppm
H2O span 0.9847 -> .9821 Offset: .05ppm
Notes: Calibrated on 7550 software version 8.5.0
3147 2017-10-13 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration. Replaced by sn 75H-2150
2918 2017-07-03 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond to replace 75H-2668.
2890 2017-07-02 Lab none Calibration Lab none Last calibration: 2017-06-21
AGC=97.9 SDM=1 DAC1=- DAC2=-
Ambient P=100.37kPa, Tair=19.11C, Tdew=17.16C
Zero Air at 1300psi, AMP CO2 513.22 at 350 Tdewspan=16.0
CO2 zero 0.8992 -> 0.8995 Offset: -2.76ppm
CO2 span 0.9985 -> 1.0066 Offset: -0.12ppm
H2O zero 0.8432 -> 0.8436 Offset: 19.4ppm
H2O span 0.9806 -> 0.9847 Offset: 4.79ppm
Notes: Large changes were seen on 6/13 after upgrading to 7500A. These were checked on 6/20 and 6/21 and the large changes were rolled back but apparently no configuration files were saved, so the last saved coefficients were the bad ones from 6/13 where CO2zero=0.9903, CO2span=1.0066, H2Ozero=0.8410 and H2Ospan=1.2209. The current cal looks good in comparison to the coefficient written down from 6/21.
2847 2017-06-23 Lab none Calibration Lab none Slated to replace 75H-2668 at West Pond; measured unreasonable H20 levels and returned to lab for further calibration.
2842 2017-06-21 Lab none Calibration Lab none Last calibration: 2017-06-20
AGC=96.8 SDM= DAC1= DAC2=
Ambient P=99.4kPa, Tair=24.7C, Tdew=12.2C
Zero Air at 1450psi, NOAA Air 402.69 at 1600 Tdewspan=16.00
CO2 zero 0.8994 -> .8992 Offset: -.84ppm
CO2 span 0.9961 -> .9985 Offset: -1.31ppm
H2O zero 0.8423 -> .8432 Offset: 4.9ppm
H2O span 0.9658 -> .9806 Offset: -0.33ppm
Notes: Re-testing zero and spans after large change in H2O zero/span and a smaller change in CO2 zero after upgrade from 7500 to 7500A. The coefficients seem stable now. The changes in H2O and CO2 zeros after upgrade might be from changing out the old chemicals for fresh chemicals during the upgrade.
2841 2017-06-20 Lab none Calibration Lab none Last calibration: 2017-06-13
AGC=95.4 SDM= DAC1= DAC2=
Ambient P=99.5kPa, Tair=26.9C, Tdew=13.2C
Zero Air at 1500psi, NOAA Air 402.69 at 1650 Tdewspan=16.00
CO2 zero 0.9003 -> .8994 Offset: -3.25ppm
CO2 span 1.0066 -> .9961 Offset: 5.59ppm
H2O zero 0.841 -> .8423 Offset: 7.6ppm
H2O span 1.2209 -> .9658 Offset: 5.25ppm
Notes: Checking the zero and spans again because there was a large H2O zero offset last time. H2O zero is pretty stable, but there was a relatively large shift in H2O span. I'll re-test tomorrow.
2833 2017-06-13 Lab none Calibration Lab none Last calibration: 2017-01-12
AGC=100 SDM= DAC1= DAC2=
Ambient P=100.2kPa, Tair=23.2C, Tdew=7.69C
Zero Air at 1500psi, NOAA Air 402.69 at 1600 Tdewspan=16.00
CO2 zero 0.9009 -> .9003 Offset: -2.33ppm
CO2 span 1.0062 -> 1.0066 Offset: -0.23ppm
H2O zero 0.857 -> .8410 Offset: -96.9ppm
H2O span 0.9788 -> 1.2209 Offset: -4.6ppm
Notes: Large H2O Zero offset. First calibration since upgrade from 7500 to 7500A. Copied coefficients for CO2 (A, B, C, D, E, XS, Z) and H2O (A, B, C, XS, Z) from sn00418_20160815post.l75 before the calibration.
2823 2017-06-12 Lab none Upgrade Lab none Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). I confirmed the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength to 100% after cleaning the windows (Calibration icon > Signal Strength tab > Reset Signal Strength).
2783 2017-05-24 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor tower. Replaced by the new Licor GHG system from Ameriflux.
2588 2017-01-17 Lab none Installed Tonzi-Floor Installed on the Floor tower to replace sn 0073 that had bad CO2 calibration coefficients.
2577 2017-01-12 Lab none Calibration Lab none Last calibration: 2016-08-15
AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.44kPa, Tair=22.75C, Tdew=6.33C
Zero Air at 1800psi, God CO2 421.22 at 400 Tdewspan=16.00
CO2 zero 0.9012 -> 0.9009 Offset: ppm
CO2 span 1.007 -> 1.0062 Offset: ppm
H2O zero 0.8558 -> 0.8570 Offset: 0.17ppm
H2O span 0.9911 -> 0.9788 Offset: 0.02ppm
Notes: Using a new tank of Zero Air. Forgot to write offsets for CO2. The serial port on bottom of Licor box has missing pin. Connect to laptop using a different cable that has one serial end and one green-guillotine-port end. Connect the green-guillotine-port end inside the box for the "Serial" connection.
2542 2016-12-07 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower for calibration. Replaced by sn 0035.
2428 2016-08-23 Lab none Installed Tonzi-Tower Installed on Tonzi Tower to replace sn 0041 for calibration.
2397 2016-08-15 Lab none Calibration Lab none Last calibration: 2016-02-22
AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.59kPa, Tair=22.61C, Tdew=12.50C
Zero Air at 500psi, God CO2 421.22 at 500 Tdewspan=16.00
CO2 zero 0.9013 -> 0.9012 Offset: -0.60ppm
CO2 span 1.0076 -> 1.0070 Offset: 0.38ppm
H2O zero 0.8564 -> 0.8558 Offset: -3.6ppm
H2O span 1.007 -> 0.9911 Offset: 0.34ppm
Notes: At first the diagnostic value was 24 with flags set on PLL, detector and chopper. Reconnecting the head fixed the problem.
2265 2016-05-10 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower for calibration. Replaced by sn 0041.
2185 2016-02-23 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower to replace sn 0035 for calibration.
2202 2016-02-22 Lab none Calibration Lab none Last calibration: 2015-07-15
AGC=48 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.62kPa, Tair=18.94C, Tdew=8.87C
Zero Air at 650psi, God CO2 421.22 at 650 Tdewspan=11.09
CO2 zero 0.9016 -> 0.9013 Offset: -0.92ppm
CO2 span 1.0064 -> 1.0076 Offset: -0.62ppm
H2O zero 0.855 -> 0.8564 Offset: 0.21ppm
H2O span 0.9897 -> 1.0070 Offset: -0.34ppm
Notes:
2033 2015-11-03 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower for calibration. Replaced by sn 0035
1917 2015-07-20 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower to replace sn 0035 for calibration.
1904 2015-07-15 Lab none Calibration Lab none Last calibration: 2015-01-13
AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.4kPa, Tair=21.72C, Tdew=16.62C
Zero Air at 800psi, God CO2 421.22 at 850 Tdewspan=11.08
CO2 zero 0.9011 -> 0.9016 Offset: 1.75ppm
CO2 span 1.0015 -> 1.0064 Offset: -2.52ppm
H2O zero 0.8534 -> 0.8550 Offset: 8.9ppm
H2O span 0.9867 -> 0.9897 Offset: 0.04ppm
Notes: External serial port did not work
1797 2015-04-23 East End Gas Analyzer CO2/H2O Removed Lab none Removed from the East End tower for calibration. Replaced by sn 1594.
1722 2015-01-20 Lab none Installed East End Gas Analyzer CO2/H2O Installed on the repositioned scaffolding at East End with the rest of the sensors after the scaffolding fell over in the wind and was rebuilt.
1661 2015-01-13 Lab none Calibration Lab none Last calibration: 2014-09-23
AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.42kPa, Tair=25.54C, Tdew=10.08C
Zero Air at 900psi, God CO2 421.22 at 1300 Tdewspan=15.95
CO2 zero 0.9013 -> 0.9011 Offset: -.63ppm
CO2 span 0.9992 -> 1.0015 Offset: -1.32ppm
H2O zero 0.851 -> 0.8534 Offset: 14.5ppm
H2O span 0.9823 -> 0.9867 Offset: -0.1ppm
Notes:
1657 2015-01-07 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman Tower for calibration. Replaced by sn 0042.
1574 2014-09-25 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman tower to replace sn 0042 for calibration.
1567 2014-09-23 Lab none Calibration Lab none Last calibration: 2014-06-10
AGC=51 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.65kPa, Tair=27.35C, Tdew=15.91C
Zero Air at 1050psi, God CO2 421.22 at 1350 Tdewspan=15.8
CO2 zero 0.9015 -> 0.9013 Offset: -0.84ppm
CO2 span 1.0058 -> 0.9992 Offset: 3.53ppm
H2O zero 0.8533 -> 0.8510 Offset: -13.2ppm
H2O span 0.9973 -> 0.9823 Offset: 0.32ppm
Notes:
1551 2014-09-11 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower tower for calibration. Replaced by sn 0976.
1348 2014-06-11 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower to replace sn 0073 for calibration.
1345 2014-06-10 Lab none Calibration Lab none Last calibration: 2013-12-05
AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.10kPa, Tair=22.66C, Tdew=12.27C
Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26
CO2 zero 0.9014 -> 0.9015 Offset: 0.6ppm
CO2 span 1.0016 -> 1.0058 Offset: -2.42ppm
H2O zero 0.8521 -> 0.8533 Offset: 6.8ppm
H2O span 0.9806 -> Offset: -0.37ppm
Notes:
1237 2014-03-13 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman tower for calibration. Replaced by sn 0976.
1214 2013-12-06 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman tower to replace sn 1594 for calibration.
1169 2013-12-05 Lab none Calibration Lab none Last calibration: 2013-08-12
AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.09kPa, Tair=22.90C, Tdew=-7.14C
Zero air at 1500psi, God CO2 421.22 at 1500 Tdewspan=15.17
CO2 zero 0.9014 -> 0.9014 Offset: -0.3ppm
CO2 span 1.0027 -> 1.0016 Offset: 0.48ppm
H2O zero 0.8522 -> 0.8521 Offset: -0.3ppm
H2O span 0.9825 -> 0.9806 Offset: 0.02ppm
Notes: The serial connector on the bottom of the box had a broken pin. I replaced it with one from the old sn 0065 box.
1146 2013-11-27 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower for calibration. Replaced by sn 0041
1043 2013-08-13 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn0042 for calibration
1035 2013-08-12 Lab none Calibration Lab none Last calibration: 2013-04-18
AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.58kPa, Tair=20.28C, Tdew=14.85C
Zero air at 1600psi, God co2 421.22 at 1600 Tdewspan=15.20
CO2 zero 0.9017 -> 0.9014 Offset: -1.0ppm
CO2 span 1.0104 -> 1.0027 Offset: 4.28ppm
H2O zero 0.8497 -> 0.8522 Offset: 0.26ppm
H2O span 0.9871 -> 0.9825 Offset: 0.09ppm
Notes: This 7500 just had its cable (box to head) replaced with the one from the old sn0065 box. The head connector was a little tight.
1041 2013-08-05 Lab none Maintenance Lab none The cable from the box to the head got caught in the elevator door and was sheared off. I replace the whole cable assembly with the one from the old sn0065 box that is no longer in use.
1040 2013-07-28 Tonzi-Tower Removed Lab none Removed from the Floor tower for calibration. Replaced by sn0073
962 2013-04-19 Lab none Installed Tonzi-Tower Swapped on to the Tower to replace sn 0035 for calibration
959 2013-04-18 Lab none Calibration Lab none Last calibration: 2012-10-23
AGC=43 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.71kPa, Tair=22.02C, Tdew=4.42C
Zero air at 1700psi, CO2 in air 392.30 at 1700 Tdewspan=15.04
CO2 zero 0.9011 -> 0.9017 Offset: 2.26ppm
CO2 span 1.0085 -> 1.0104 Offset: -0.9ppm
H2O zero 0.8516 -> 0.8497 Offset: -10.7ppm
H2O span 1 -> 0.9871 Offset: ppm
Notes:
954 2013-02-27 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower by Siyan for calibration. Replaced by sn 0976.
813 2012-10-24 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira Tower to swap out 0041 for calibration.
840 2012-10-23 Lab none Calibration Lab none Last calibration: 2012-06-06
AGC=48 SDM=0 DAC1=0-1000 DAC2=10-25
Ambient P=100.79kPa, Tair=22.2C, Tdew=0C
dry air at 1800psi, CO2 398.79 at 1600 Tdewspan=15.07
CO2 zero 0.9011 -> 0.9011 Offset: -0.16ppm
CO2 span 1.0032 -> 1.0085 Offset: -2.3ppm
H2O zero 0.8456 -> 0.8516 Offset: 0.03ppm
H2O span 0.9835 -> 1 Offset: 0.27ppm
Notes: operated by Siyan Ma
729 2012-08-09 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman tower for calibration. Replaced by sn 0035.
648 2012-06-07 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman tower to replace sn 0041 for calibration.
641 2012-06-06 Lab none Calibration Lab none Last calibration: 2011-12-15
AGC=51 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100kPa, Tair=24-57C, Tdew=6.06C
zero air at 2000psi, 436.53 at 1600 Tdewspan=
CO2 zero 0.9017 -> 0.9011 Offset: -2.40ppm
CO2 span 1.0066 -> 1.0032 Offset: 1.89ppm
H2O zero 0.8493 -> 0.8456 Offset: -20.4ppm
H2O span 0.9719 -> 0.9835 Offset: -0.26ppm
Notes:
604 2012-05-18 Tonzi-Tower Removed Lab none Swapped off of the Tonzi tower for calibration, replaced by sn 1594
467 2011-12-18 Lab none Installed Tonzi-Tower Installed on Tonzi Tower so that sn 0065 could be swapped out for calibration.
460 2011-12-15 Lab none Calibration Lab none Last calibration: unknow
AGC=48 SDM=- DAC1=0-1000 DAC2=10-25
Ambient P=101.4kPa, Tair=22.6C, Tdew=6.04C
zero air at 700psi, CO2 436.53 at 1900 Tdewspan=13.5
CO2 zero 0.9015 -> 0.9017 Offset: 0.88ppm
CO2 span 1.0037 -> 1.0066 Offset: 1.67ppm
H2O zero 0.8436 -> 0.8493 Offset: 0.09ppm
H2O span 0.9604 -> 0.9719 Offset: 0.1ppm
Notes: operated by Siyan Ma
434 2011-10-18 Tonzi-Floor Removed Lab none Removed from the Floor Tower for calibration. Replaced by sn 0073.
431 2011-10-17 Lab none Calibration Lab none Last calibration: 2011-07-05
AGC=61 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.48kPa, Tair=30.7C, Tdew=16.56C
dry air at 700psi, CO2 2000 at 2000 Tdewspan=18.3
CO2 zero 0.9588 -> 0.9584 Offset: -1.13ppm
CO2 span 1 -> 1.0022 Offset: 1.08ppm
H2O zero 0.74 -> 0.7390 Offset: -0.07ppm
H2O span 1.0033 -> 0.9873 Offset: 0.4ppm
Notes: calibrated by Siyan
380 2011-07-06 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 0073.
636 2011-03-30 Lab none Installed Vaira Gas Analyzer CO2/H2O Swapped on to the Vaira tower to replace sn 0065 for calibration.
322 2011-02-08 Sherman Island Gas Analyzer CO2/H2O Removed Lab none This LI7500 was removed from the Sherman Island tower for calibration. sn 0073 was installed.
253 2010-11-18 Lab none Installed Sherman Island Gas Analyzer CO2/H2O This IRGA was installed at Sherman Island to replace sn 0976 for calibration.
252 2010-11-16 Lab none Calibration Lab none Last cal 2010-08-13, SMD 7 DAC1 H2O 0-1000 DAC2 CO2 10-25
AGC = 49%, T = 24.35C, P = 100.92kPa, Td = 12C
Zero Air 1700psi, CO2 597.4ppm 200psi, Td = 9.6C
CO2 zero = 0.9038-> 0.9017, offset -8.3ppm
CO2 span = 1.0047-> 0.9983, offset +5.6ppm
H2O zero = 0.8467-> 0.8479, offset +7.2mmol/m3
H2O span = 0.9841-> 0.9644, offset +0.43C
229 2010-10-07 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira Tower for calibration. sn 1594 swapped in.
186 2010-08-14 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on Vaira tower by Siyan after lab calibration.
151 2010-07-01 Sherman Island-Levee Gas Analyzer CO2/H2O Removed Lab none Removed from leveen site due to theft of everything else.
124 2010-06-03 Lab none Installed Sherman Island-Levee Gas Analyzer CO2/H2O Installed on the Sherman Levee tower after fresh lab calibration.
123 2010-06-02 Lab none Calibration Lab none Last cal 2010-05-18, SMD 7 DAC1 H2O 0-1000 DAC2 CO2 10-25
AGC = 49%, T = 23.11C, P = 100.88kPa, Td = 12.44C
N2 = 0.0 400psi CO2 = 597.4ppm <300psi, Td = 9.7C
CO2 zero = 0.9011-> 0.9015, offset +1.44ppm
CO2 span = 1.0045-> 1.0022, offset +1.5ppm
H2O zero = 0.8524-> 0.8499, offset -0.262mmol/m3
H2O span = 1.0097-> 0.9921, offset +0.31C
119 2010-05-28 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none This analyzer was removed from the Twitchell tower for calibration.
90 2010-03-18 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O This LI7500 was installed at Twitchell Island. Its SDM address was changed to 7 to match CR1000 programs.
89 2010-03-18 Lab none Calibration Lab none Calibrated in the lab 100.70kPa, 18.36C, 49% AGC 13.9C Td
N2 tank 0.0 @ 700psi, CO2 tank 597.4ppm @ 200psi
Last cal 11-4-2009
CO2 zero 0.9020 -> 0.9011 offset -3.18ppm
CO2 span 1.0000 -> 1.0042 offset +3.5ppm
H2O zero 0.8486 -> 0.8524 offset +0.410g/m3
H2O span 0.9807 -> 1.0097 offset 0.45C

SDM address 3
86 2010-03-17 Lab none Maintenance Lab none Embedded software upgraded from v3.0.0 to v3.0.1
Internal chemicals changed
4009 2009-01-08 Lab Gas Analyzer CO2/H2O Installed Tonzi-Floor Installed on the Tonzi Tower to replace sn 0065 for calibration
4926 2003-10-02 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 0041 for calibration.
4915 2003-06-20 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0073 for calibration.
4907 2003-05-09 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0065 for calibration.
4902 2003-04-06 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0042.
4878 2002-12-23 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0065.
4807 2002-11-05 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0042 for calibration.
4876 2002-09-20 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0042.
4871 2002-08-15 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0042 for calibration.
 
 