Equipment and Calibration History

View Equipment |

View history entry 3836 for Licor LI-7500A sn: 0418

Date: 2019-01-07
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-10-10 AGC=97.2 SDM=0 DAC1=0 DAC2=0 Ambient P=100.9kPa, Tair=17.9C, Tdew=9.7C Zero Air at 1200psi, AMP Air 408.97 at 900 Tdewspan=16.00 CO2 zero 0.8982 -> .8992 Offset: 3.69ppm CO2 span 0.9965 -> 1.0033 Offset: -3.58ppm H2O zero 0.8454 -> .8465 Offset: 6.1ppm H2O span 0.9813 -> .9832 Offset: -.03ppm Notes:
Entered by: