Equipment and Calibration History

View Equipment |

View history entry 4059 for Licor LI-7500A sn: 0418

Date: 2019-08-21
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-04-22 AGC=93.7 SDM=0 DAC1=- DAC2=- Ambient P=100.3kPa, Tair=26.98C, Tdew=14.83C Zero Air at 1000psi, AMP CO2 408.97 at 800 Tdewspan=20.0 CO2 zero 0.8992 -> 0.8990 Offset: -0.55ppm CO2 span 1.0011 -> 0.9976 Offset: 1.71ppm H2O zero 0.8464 -> 0.8466 Offset: -0.8ppm H2O span 0.9866 -> 0.9818 Offset: 0.11ppm Notes:
Entered by: