Equipment and Calibration History

View Equipment |

View history entry 5465 for Licor LI-7500A sn: 0418

Date: 2021-09-21
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-09-20 AGC=99.3 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=29C, Tdew=11.1C Zero Air at 1600psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero .9004 -> .8992 Offset: -4.8ppm CO2 span 1.0016 -> .9972 Offset: 3.22ppm H2O zero .8444 -> .8490 Offset: 27.3ppm H2O span .9808 -> .9774 Offset: .08ppm Notes: I swapped the chemicals immediately before this calibration run. The chemicals were a tad sticky at the top but after a bit of poking with a screwdriver it was easy to dump the chemicals out into the waste jar. I don't think the chemicals needed to be changed necessarily. Before inserting the new chemical bottles, I ran some zero air into the headspace. This run was just to see what the numbers looked like. I will let it run on a bench for a week and calibrate again next week.
Entered by: