View history entry 431 for Licor LI-7500A sn: 0418

Date: 2011-10-17
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2011-07-05 AGC=61 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.48kPa, Tair=30.7C, Tdew=16.56C dry air at 700psi, CO2 2000 at 2000 Tdewspan=18.3 CO2 zero 0.9588 -> 0.9584 Offset: -1.13ppm CO2 span 1 -> 1.0022 Offset: 1.08ppm H2O zero 0.74 -> 0.7390 Offset: -0.07ppm H2O span 1.0033 -> 0.9873 Offset: 0.4ppm Notes: calibrated by Siyan