Equipment and Calibration History
Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 [57] 58 59 60 61 62 63 64 65 66 67 68 69 70Ser No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
FGGA 09-0012 | 2010-01-14 | Twitchell Rice | Calibration | Twitchell Rice | Last calibration: 2009-03-03 00:00:00
Cell Pressure=148.63kPa, Cell Temperature=28.27C, Ring Time=17.277usec 2nd Standard at 0psi: CH4: 3.276ppm, CO2: 433ppm, H2O: 0ppm CH4 = 3.368ppm -> 3.275ppm Offset: -0.092ppb CO2 = 432.1ppm -> 433.4ppm Offset: 0.9ppm Notes: LGR pressure very low ~16Torr, but very easily adjusted using bypass valve. |
LI-7500A 0041 | 2009-07-31 | Lab none | Installed | Twitchell Rice Gas Analyzer CO2/H2O | Installed at Rice. Can't figure out which sn it replaced though.... |
FMA RMT-200 08-0114 | 2009-08-06 | Twitchell Rice | Moved | Sherman Island Gas Analyzer CH4 Closed-path | Removed from Rice and replaced with Frank's FGGA analyzer (sn 09-0012). Installed to replace LGR sn 05-0010 for maintenance. |
FMA RMT-200 08-0114 | 2009-06-04 | Lab none | Installed | Twitchell Rice | Installed at the Rice. |
WMP 1352 000024 | 2009-04-03 | Lab none | Installed | Twitchell Rice Anemometer | Installed at the Rice. Z= 3.05 m at middle of path & North spar pointing to 318 magnetic.
Need to apply Schotanus correction since model 1352. |
USB2000-VIS-NIR USB2G4461 | 2014-07-08 | Factory | Returned | Lab none | Returned from Ocean Optics. They fixed the coax connector where the fiber optics connects to the box. They did a "level two repair" which seems to include a calibration (document on file). |
USB2000-VIS-NIR USB2G4461 | 2014-06-13 | Lab none | Shipped | Factory | Sent to Ocean Optics for repair on RMA 1415292. The coax connector where the fiber optics connects to the box broke off. |
LI-6400 PSC-0785 | 2014-06-25 | Lab none | Calibration | Lab none | Madeleine and I calibrated the 6400 in the lab
Zero: CO2 1.45 to 0.0 H2O -0.21 to -0.001 Span CO2 with 421.22 CO2 gas --did not need to span--left at 421.0 H2O span (dew point set at 15.26C) 14.63 to 15.26C |
LI-7500A 0073 | 2014-06-24 | Lab none | Calibration | Lab none | Last calibration: 2014-02-28 AGC=60 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.46kPa, Tair=22.08C, Tdew=12.44C Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26 CO2 zero 0.9578 -> 0.9580 Offset: 0.65ppm CO2 span 1.0009 -> 1.0018 Offset: -0.72ppm H2O zero 0.7355 -> 0.7389 Offset: 21.0ppm H2O span 0.9972 -> 1.0138 Offset: -0.28ppm Notes: |
LI-7500A 0042 | 2014-06-24 | Lab none | Calibration | Lab none | Last calibration: 2014-02-28 AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.62kPa, Tair=21.30C, Tdew=11.17C Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26 CO2 zero 0.9506 -> 0.9506 Offset: -0.12ppm CO2 span 1.0078 -> 1.0074 Offset: 0.18ppm H2O zero 0.8506 -> 0.8506 Offset: 0.3ppm H2O span 0.9853 -> 0.9960 Offset: -0.19ppm Notes: |
LI-7500A 75H-2150 | 2014-06-24 | Lab none | Calibration | Lab none | Last calibration: 2013-12-18 AGC=95.08 SDM=1 DAC1=- DAC2=- Ambient P=100.15kPa, Tair=21.78C, Tdew=11.81C Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26 CO2 zero 0.9012 -> 0.9012 Offset: 0.07ppm CO2 span 1.0016 -> 1.0011 Offset: 0.28ppm H2O zero 0.8643 -> 0.8652 Offset: 5.8ppm H2O span 0.9987 -> 1.0087 Offset: -0.21ppm Notes: |
GMP221 Y2009100 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.233 + -902.3; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
GMP221 Y0850004 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.743 + -1252; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
GMP221 Y1720006 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.6676 + -1028; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
GMP222 Z2720011 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.176 + -287.8; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
GMP221 Y0710013 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.028 + -18.33; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
GMT221 G0150037 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.017 + -200.7; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
GMP222 H1630066 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.746 + -27.77; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
GMP222 Z2720014 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.248 + 0.9176; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
GMP222 Z2720014 | 2014-04-08 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.027 + 0.2028; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP221 Y2009100 | 2014-04-08 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.047 + -347.2; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP221 K1340006 | 2014-04-08 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.002 + -37.28; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP221 K1340007 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.01 + -11.96; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP221 K1340005 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9908 + -5.527; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP222 K1350103 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.022 + 63.48; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP222 K1350102 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.02 + 41.22; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP222 K1310104 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.008 + 43.09; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP222 K1350101 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.014 + 51.82; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP222 K1310103 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.094 + 0.7501; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP222 K1310102 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.014 + 42.61; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |
GMP222 Y1950005 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.073 + -174.1; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 Z2720015 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9937 + 44.53; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMT222 G0230062 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.004 + 19.27; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP221 Y1720006 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9255 + -857.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP221 Y0850004 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.056 + -1146; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMT222 G0230061 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.01 + 67.88; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 Y2030049 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.095 + 39.49; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 H1630066 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.753 + -29.94; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 Z2720014 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.076 + -9.057; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 X5050014 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.287 + 388; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 Z2720011 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.11 + 190.2; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP221 Y0710013 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.023 + -15.82; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP221 Y2009100 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.031 + -299.9; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 Z2720012 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*2.256 + 934.1; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 Z2720013 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.327 + 209.8; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMT221 F4960019 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.067 + -469.7; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMT221 F4960018 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9972 + -319.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 Y2009109 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9809 + 65.37; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP222 X4840007 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9548 + -78.1; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMT221 F4960020 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.042 + -361.7; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMT221 G0150036 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9924 + -154.6; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMT221 G0150037 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.007 + -79.91; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMT221 G0150035 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.015 + -365.7; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
GMP221 H1310008 | 2013-01-08 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9955 + -27.3; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 421ppm, 3056ppm, and 10000ppm. Ta=22.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMP222 H0630070 | 2013-01-08 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9745 + 35.22; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 421ppm, 3056ppm, and 10000ppm. Ta=22.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMP222 H1630066 | 2013-01-08 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9828 + 15.78; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 421ppm, 3056ppm, and 10000ppm. Ta=22.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMP221 H0960006 | 2012-06-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.003 + 133; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 3056ppm, and 10000ppm. Ta=25.9C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMP221 H1310020 | 2012-06-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9873 + 114.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 3056ppm, and 10000ppm. Ta=25.9C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMP222 X1920001 | 2012-02-15 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.098 + -121.7; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.7C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 Y2009109 | 2012-02-15 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.08 + -56.27; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.7C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMT221 F4960019 | 2012-02-15 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.08 + -386.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.7C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP221 Y1720006 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.08 + 78.35; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP221 Y2009100 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.036 + -68.6; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP221 Y0710013 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9552 + -2922; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP221 Y0850004 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.016 + -32.48; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 Y2009109 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9987 + 7.368; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 Z2720012 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.029 + 14.22; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 Z2720011 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9639 + -59.19; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 X5050014 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.063 + -34.27; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 Z2720013 | 2012-02-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.043 + -24.84; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP221 Y1720006 | 2012-02-03 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9432 + -643; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.6C, P=101.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP221 Y2009100 | 2012-02-03 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.003 + -355.6; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.6C, P=101.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP221 Y0710013 | 2012-02-03 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.07 + -227; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.6C, P=101.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP221 Y0850004 | 2012-02-03 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.001 + -1162; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.6C, P=101.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 Z2720015 | 2012-01-30 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.005 + 20.82; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=27.4C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 Y1950005 | 2012-01-30 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.08 + -104.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=27.4C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMT222 G0230062 | 2012-01-30 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.017 + 4.211; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=27.4C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMP222 Y2030049 | 2012-01-30 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.054 + 39.65; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=27.4C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMT222 G0230061 | 2012-01-30 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.02 + 31.82; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=27.4C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
GMT221 F4960018 | 2011-12-13 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.039 + 10.06; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.6C, P=101.7kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 G0150036 | 2011-12-13 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.026 + -13.15; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.6C, P=101.7kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 G0150037 | 2011-12-13 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.035 + -6.175; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.6C, P=101.7kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 F4960019 | 2011-10-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.02 + -56.65; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.2C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 G0150035 | 2011-10-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.04 + 90.53; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.2C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 G0150034 | 2011-10-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.014 + 23.21; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.2C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 F4960020 | 2011-10-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.007 + -3.598; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.2C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 F4960018 | 2011-10-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.028 + 1.298; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.2C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 G0150036 | 2011-10-06 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.011 + -52.12; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.2C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
GMT221 F4960019 | 2011-03-17 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.034 + 102.5; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.6C, P=102.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT221 F4960020 | 2011-03-17 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.023 + -14.32; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.6C, P=102.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT222 G0230061 | 2011-03-14 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.046 + 56.59; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT221 F4960018 | 2011-03-14 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.062 + 47.47; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT221 G0150037 | 2011-02-10 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.03 + 52.52; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.5C, P=102.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT221 G0150034 | 2011-02-10 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.055 + 27.37; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.5C, P=102.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT221 G0150035 | 2011-02-10 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.047 + -6.387; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.5C, P=102.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT221 F4960019 | 2011-02-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.048 + 150; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.8C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT221 F4960020 | 2011-02-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.034 + 46.31; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.8C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT222 G0230062 | 2011-02-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.075 + 6.9; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.8C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMT221 G0150036 | 2011-02-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.023 + 34.36; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.8C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
GMP222 Y2009109 | 2011-02-01 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.041 + 3.308; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, and 597ppm. Ta=25.5C, P=102.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |