Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 [39] 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65
Ser No.DateLoc FromTitleLoc ToNotes
LI-7500A 75H-2176 2017-06-05 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Twitchell Alfalfa and brought back to the lab for calibration. It was replaced by sn 1594.
BB-Fab BB-TT 2017-01-17 Tonzi Maintenance Tonzi The Tower has been losing power over night recently. All the batteries were disconnected and tested them with the load tester. Four batteries tested bad. We removed them and reconnected the others.
GP-001 GP20140017 2017-06-05 Lab none Installed West Pond CO2 Probe B Installed at West Pond to replace sn GP2015019.
GP-001 GP20150109 2017-06-05 West Pond CO2 Probe B Removed Lab none Removed from West Pond because it was giving funky numbers after the water level went down.
SRS-Ni NDVI 2046503376 2017-06-05 Lab none Installed Mayberry Narrow Band NDVI Incoming Installed in incoming position at Mayberry.
SRS-Ni NDVI 1066703780 2017-06-05 Lab none Installed West Pond Narrow Band NDVI Incoming Installed in incoming position at West Pond.
LI-7500A 0073 2017-06-12 Lab none Upgrade Lab none Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). I confirmed the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength to 100% after cleaning the windows (Calibration icon > Signal Strength tab > Reset Signal Strength).
LI-7500A 0418 2017-06-12 Lab none Upgrade Lab none Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). I confirmed the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength to 100% after cleaning the windows (Calibration icon > Signal Strength tab > Reset Signal Strength).
LI-7500A 75H-2176 2017-06-12 Lab none Calibration Lab none Last calibration: 2017-03-07
AGC=90.4 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=19.8C, Tdew=8.83C
Zero Air at 1500psi, NOAA Air 402.69 at 1600 Tdewspan=16.00
CO2 zero 0.8691 -> .8690 Offset: -.41ppm
CO2 span 1.0195 -> 1.0160 Offset: 1.77ppm
H2O zero 0.8561 -> .8572 Offset: 5.9ppm
H2O span 1.0102 -> 1.0022 Offset: 0.18ppm
Notes:
LI-7500 0976 2017-06-12 Lab none Calibration Lab none Last calibration: 2017-04-07
AGC=57 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.7kPa, Tair=20.3C, Tdew=7.04C
Zero Air at 1500psi, NOAA Air 402.69 at 1600 Tdewspan=16.00
CO2 zero 0.8594 -> .8597 Offset: 1.39ppm
CO2 span 1.0097 -> 1.0077 Offset: 1.01ppm
H2O zero 0.6348 -> .6335 Offset: -.23ppm
H2O span 0.9689 -> 1.0310 Offset: -1.36ppm
Notes: Set span temperature in software to 15.92 in order for dew point value to be 16.00 after pressing "Span"
SRS-Ni NDVI 2046503357 2017-06-05 Lab none Installed Bouldin Corn Narrow Band NDVI Reflected Installed at BC tower to replace downward-facing NDVI sensor with field-stop lens.
SRS-Nr NDVI 943103015 2017-06-05 Bouldin Corn Narrow Band NDVI Reflected Removed Lab none Removed from BC tower and replaced by a downward-facing NDVI sensor with hemispherical view.
SRS-Nr NDVI 943203017 2017-06-05 Bouldin Alfalfa Narrow Band NDVI Reflected Removed Lab none Removed because we have already installed a downward-facing NDVI sensor with hemispherical view.
BB-Fab BB-TW 2017-06-01 Twitchell Rice Removed Twitchell Rice The Twitchell Rice solar system was disassembled and removed. Two of the batteries were dead and will be recycled. The four other batteries load tested okay and were taken to Tonzi for the Floor solar power system. The solar panels and charge controller were removed and will be used at the temporary tower at East End this summer.
HMP45AC T2130005 2017-06-01 Tonzi-Tower Removed Lab none Removed from the higher middle position of CO2 profile system on the Tonzi Tower. At least one has not been working and all need to be compared against each other for drift.
HMP45AC T1250012 2017-06-01 Tonzi-Tower Removed Lab none Removed from the lower middle position of CO2 profile system on the Tonzi Tower. At least one has not been working and all need to be compared against each other for drift.
HMP45AC U3030041 2017-06-01 Tonzi-Tower Removed Lab none Removed from the lowest position of CO2 profile system on the Tonzi Tower. At least one has not been working and all need to be compared against each other for drift.
BB-Fab BB-TF 2017-06-01 Tonzi-Floor Maintenance Tonzi-Floor I disconnected all seven batteries and load tested them. Three were very weak, two dipped a little under load but still in the green and two were rock solid. I replaced the three weak ones and the two that dipped under load with four batteries we brought from the old Twitchell Rice solar system.
LI-7550 AIU-0650 2017-06-01 Lab none Installed East End-Temp1 Interface LI-7550 Installed at open water temporary tower in East End wetland
LI-7700 TG1-0423 2017-06-01 Lab none Installed East End-Temp1 Gas Analyzer CH4 Installed at open water temporary tower in East End wetland
LI-7500A 0065 2017-06-01 Lab none Installed East End Gas Analyzer T1 CO2/H2O Installed at open water temporary tower in East End wetland
WM 1590 153803 2017-06-01 Lab none Installed East End-Temp1 Anemometer Installed at open water temporary tower in East End wetland
GP-001 GP20140009 2017-05-31 Factory Disposed Disposed Disposed of at Eosense factory because it was not repairable
GP-001 GP20140010 2017-05-31 Factory Disposed Disposed Disposed of at Eosense factory because it was not repairable
GP-001 GP20140021 2017-05-31 Factory Disposed Disposed Disposed of at Eosense factory because it was not repairable
GP-001 GP20140022 2017-05-31 Factory Disposed Disposed Disposed of at Eosense factory because it was not repairable
GP-001 GP20140023 2017-05-31 Factory Disposed Disposed Disposed of at Eosense factory because it was not repairable
GP-001 GP20140024 2017-05-31 Factory Disposed Disposed Disposed of at Eosense factory because it was not repairable
GP-001 GP20140025 2017-05-31 Factory Disposed Disposed Disposed of at Eosense factory because it was not repairable
GP-001 GP20150100 2017-05-31 Factory Disposed Disposed Disposed of at Eosense factory because it was not repairable
BB-Fab BB-TW 2015-12-28 Twitchell Rice Upgrade Twitchell Rice I added two batteries to the battery bank. I added two batteries from the lab. One form 5/14 and one from 5/11. Both tested okay in the lab. The existing batteries were two from 2/12 and two from 6/14.
BB-Fab BB-TW 2016-01-21 Twitchell Rice Repair Twitchell Rice The solar charge controller was replaced.
BB-Fab BB-TW 2016-01-12 Twitchell Rice Lost Twitchell Rice The old 8D batteries, the solar charge controller were stolen along with the old solar panels that were on top on the big box. Power is down until the charge controller can be replaced.
LI-7500A 0065 2017-05-30 Lab none Calibration Lab none Last calibration: 2016-12-06
AGC=100 SDM= DAC1= DAC2=
Ambient P=99.7kPa, Tair=19.8C, Tdew=10.1C
Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00
CO2 zero 0.8899 -> .8901 Offset: 1.08ppm
CO2 span 1.0142 -> 1.0130 Offset: -.84ppm
H2O zero 0.7086 -> .7087 Offset: .6ppm
H2O span 1.0127 -> 1.0166 Offset: -.08ppm
Notes: Signal strength was 73.9 after thorough cleaning of lenses, so I reset the Signal Strength to 100%.
LI-7500A 75H-2150 2017-05-30 Lab none Calibration Lab none Last calibration: 2017-03-14
AGC=97.2 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=19.9C, Tdew=12.2C
Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00
CO2 zero 0.9014 -> .9024 Offset: 4.23ppm
CO2 span 1 -> 1.0194 Offset: -9.8ppm
H2O zero 0.8651 -> .8664 Offset: 9.3ppm
H2O span 1.0094 -> .9975 Offset: .26ppm
Notes:
LI-7500A 1594 2017-05-30 Lab none Calibration Lab none Last calibration: 2017-01-27
AGC=101.4 SDM= DAC1= DAC2=
Ambient P=99.7kPa, Tair=19.3C, Tdew=9.5C
Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00
CO2 zero 0.8654 -> .8653 Offset: .79ppm
CO2 span 1.0194 -> 1.0187 Offset: 0.42ppm
H2O zero 0.8868 -> .8861 Offset: -4.1ppm
H2O span 1.0067 -> 1.0256 Offset: -.39ppm
Notes:
LI-7700 TG1-0423 2017-05-31 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2016-10-11
RSSI=80.7%, Optical RH=10%
Ambient P=100.3kPa, Tair=20.4C
Zero Air at 1500psi, NOAA Air 1.89919 at 1600
CH4 Zero Offset: 0.017ppm
CH4 Span Offset: -0.021ppm
Notes: Replaced desiccant and flushed cavity with zero air. Optics RH was 26% before replacement and 10% after replacement.
LI-7700 TG1-0211 2017-05-31 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2016-10-11
RSSI=71.2%, Optical RH=14%
Ambient P=100.3kPa, Tair=C
Zero Air at 1500psi, NOAA Air 1.89919 at 1600
CH4 Zero Offset: -0.1ppm
CH4 Span Offset: -0.002ppm
Notes: Replaced temperature component. Replaced desiccant and flushed cavity with zero air. Optics RH was 26% before replacement and 14% afterwards.
CFM100 4994 2017-05-24 Vaira Removed Lab none Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux.
CR1000 23865 2017-05-24 Vaira Removed Lab none Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux.
LI-7500A 0073 2017-05-24 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux.
1189PC 000192 2017-05-24 Vaira Removed Lab none Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux.
WMP 1352 000028 2017-05-24 Vaira Anemometer Removed Lab none Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux.
LI-7500 0976 2017-05-24 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower. Replaced by the new Licor GHG system from Ameriflux.
WMP 1352 000019 2017-05-24 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower. Replaced by the new Licor GHG system from Ameriflux.
NL115 6761 2017-05-24 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor tower. Replaced by the new Licor GHG system from Ameriflux.
CR1000 50889 2017-05-24 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor tower. Replaced by the new Licor GHG system from Ameriflux.
LI-7500A 0418 2017-05-24 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor tower. Replaced by the new Licor GHG system from Ameriflux.
1189PC 000193 2017-05-24 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor tower. Replaced by the new Licor GHG system from Ameriflux.
WMP 1352 000023 2017-05-24 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor tower. Replaced by the new Licor GHG system from Ameriflux.
HFP01SC 00246 2017-05-26 Lab none Tested Disposed Sensor is not working. Resistance of sensor is 0 and there is no connection. The wires look pulled out at the plate.
HFP01SC 00245 2017-05-26 Lab none Repair Lab none Added silicone around the connection between wire and plate. Soldered new cable onto sensor and used clear heat shrink to attach sn and calibration information onto new cable.
CR1000 66250 2017-05-22 Lab none Upgrade Lab none OS upgraded from v27 to v31. CPU drive is reformated from FAT16 to FAT32 so complete reset of all settings. Lots of changes see CR1000_OS_31_readme.pdf
BB-Fab BB-WP 2012-05-02 Lab none Installed West Pond Battery Bank This is the initial installation of this battery bank for the West Pond Tower. Six deep cycle marine batteries from O'Reilly Auto Parts part #27DCM 175min reserve capacity were installed in parallel for a 12V system.
KU265-6MCA 158APW4M0000188 2017-05-18 Lab none Installed West Pond Solar Panel A Installed on West Pond to replace old 12V system.
KU265-6MCA 158APW4M0000189 2017-05-18 Lab none Installed West Pond Solar Panel B Installed on West Pond to replace old 12V system.
KD140SX-UFBS 11ZV0M0836 2017-05-18 West Pond Solar Panel A Removed Lab none Removed from West Pond to make way for new 24V system.
KD140SX-UFBS 11ZV0M0856 2017-05-18 West Pond Solar Panel B Removed Lab none Removed from West Pond to make way for new 24V system.
KD140SX-UFBS 11ZV0M0855 2017-05-18 West Pond Solar Panel C Removed Lab none Removed from West Pond to make way for new 24V system.
BB-Fab BB-WP 2017-05-18 West Pond Battery Bank Upgrade West Pond Battery Bank All the old batteries but one load tested week and the tower has been losing power overnight. The old batteries were replaced with new batteries from Napa Auto Parts, part #BAT 8270, 180min reserve capacity. The three old 12V 140W solar panels were replaced with two new 24V 265W panels and the battery bank was rewired as three pairs for a 24V system.
LI-7700 TG1-0211 2017-04-04 Twitchell Rice Gas Analyzer CH4 Removed Lab none Removed from Twitchell Rice site with Twitchell Rice decommissioning.
LI-7550 AIU-0650 2017-05-15 Lab none Upgrade Lab none Updated the embedded software in the LI-7550 from 8.0.0 to 8.5.0.
LI-7550 AIU-0614 2017-05-15 Lab none Upgrade Lab none Updated the embedded software in the LI-7550 from 8.0.0 to 8.5.0.
WM 1590 170523 2017-05-11 Lab none Installed Tonzi-Tower Installed on the Tower top replacing original sonic which was moved to temporary mount for inter-comparison.
LI-7550 AIU-1866 2017-05-11 Lab none Installed Tonzi-Tower Installed on the Tower one level below the top on the southwest corner post.
LI-7500A 0041 2017-05-11 Lab none Installed Tonzi-Tower Installed on the Tower top replacing original 7500 which was moved to temporary mount for inter-comparison.
LI-7550 AIU-1869 2017-05-11 Lab none Installed Tonzi-Floor Installed at Tonzi Floor tower. Most of the interface settings seemed to be correct (update rate, data stored, CO2/H2O separation), but we could not figure out how to save the sonic analog settings. The Li-7500 v 8.5 software was cut off on our computer monitor.
LI-7500A 0042 2017-05-11 Lab none Installed Tonzi-Floor Installed at the Tonzi Floor tower. I uploaded a new config file and updated the pressure coefficients. The old IRGA head, sn 0418, was moved to a temporary mount on the same tower so we can compare the old and new heads.
WM 1590 170525 2017-05-11 Lab none Installed Tonzi-Floor Installed at Tonzi Floor tower. sn 000023 was moved to a temporary mount on the same tower so that we can compare the old and new sonics.
CNR1 990181 2017-05-10 Lab none Installed Bouldin Corn Net Radiometer 4-way Installed at Bouldin Corn to calibrate against NR01 sn 2420. Joe had replaced the sensor cables recently.
LI-7550 AIU-1870 2017-05-11 Lab none Installed Vaira Interface LI-7550 Installed at Vaira to replace interface interface AIU-650.
LI-7500A 0035 2017-05-11 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 75H-2150.
WM 1590 170524 2017-05-11 Lab none Installed Vaira Installed at Vaira to replace sn 153805.
LI-7550 AIU-0650 2017-05-11 Vaira Removed Lab none Removed from Vaira because the Ameriflux intercomparison finished and because we received new 7500A and sonics from Ameriflux. This interface was replaced by interface AIU-1870.
WM 1590 153805 2017-05-11 Vaira Removed Lab none Removed from Vaira because the Ameriflux intercomparison finished and because we received new 7500A and sonics from Ameriflux. This sonic was replaced with brand new sonic from Ameriflux, sn 170524.
LI-7500A 75H-2150 2017-05-11 Vaira Removed Lab none Removed from Vaira because the Ameriflux intercomparison finished and because we received new 7500A and sonics from Ameriflux. This head was replaced with sn 0035 that has been freshly upgraded from 7500 to 7500A.
CM11 996985 2017-04-04 Twitchell Rice Pyranometer Removed Lab none Removed from Twitchell Rice site with Twitchell Rice decommissioning.
LI-7500A 0035 2017-05-08 Lab none Calibration Lab none Last calibration: 2016-12-05
AGC=100 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=21.9C, Tdew=7.8C
Zero Air at 1550psi, LBL Air 513.22 at 350 Tdewspan=16.00
CO2 zero 0.939 -> .9381 Offset: -3.83ppm
CO2 span 1.0265 -> 1.0230 Offset: 2.41ppm
H2O zero 0.9025 -> .8974 Offset: -28.2ppm
H2O span 1.0029 -> 1.0261 Offset: -.48ppm
Notes: First calibration since upgrade from 7500 to 7500A.
HMP60 G4540021 2017-05-05 Factory Received Lab none Sensor received at lab after servicing at Campbell Sci.
HMP60 G4540021 2017-04-03 Lab none Serviced Factory Shipped to Campbell Sci for calibration. CSI Technician notes:
Tested & inspected. Refurbished RH chip. Confirmed proper function.
HMP45AC U3030032 2017-05-05 Factory Disposed Disposed Not repairable, so disposed of at CSI.
HMP45AC U3030032 2017-04-03 Lab none Serviced Factory Shipped to Campbell Sci for calibration. CSI Technician notes:
Tested & inspected. 2x4 connectors have corrosion, need replaced. Needs new o-rings. While testing sensor head, also found temp chip out of spec, drifting high in high humidity. Temperature chips are no longer available for this sensor. Sensor can not be repaired. Dispose of at CSI.
HMP45AC T0720012 2017-05-05 Factory Disposed Disposed Not repairable, so disposed of at CSI.
HMP45AC T0720012 2017-04-03 Lab none Serviced Factory Shipped to Campbell Sci for calibration. CSI Technician notes:
Inspected. Housing severely damaged, handle piece PCB non-functional, 2x4 connectors have corrosion damage. Sensor is beyond repair. Dispose of at CSI.
HMP45AC W0730036 2017-05-05 Factory Calibration Lab none Sensor received from Campbell Sci with Certificate of Calibration. Next calibration due 15-Apr-2018.
HMP45AC W0730036 2017-04-03 Lab none Serviced Factory Shipped to Campbell Sci for calibration. CSI Technician notes:
Tested & inspected. Replaced o-rings. RH chip has corrosion damage. Took before readings with original RH chip. Replaced RH chip & calibrated with documentation.
LI-7500A 0041 2017-05-05 Lab none Calibration Lab none Last calibration: 2016-12-05
AGC=99.9 SDM= DAC1= DAC2=
Ambient P=100.2kPa, Tair=22.7C, Tdew=9.22C
Zero Air at 1600psi, LBL Air 513.22 at 400 Tdewspan=16.00
CO2 zero 1.0004 -> .9969 Offset: -11.7ppm
CO2 span 1.0124 -> 1.0177 Offset: -3.72ppm
H2O zero 0.8255 -> .8210 Offset: -25.0ppm
H2O span 1 -> 1.0320 Offset: -.69ppm
Notes: First calibration since upgrade from LI-7500 to LI-7500A
LI-7500A 0042 2017-05-05 Lab none Calibration Lab none Last calibration: 2016-08-15
AGC=99.9 SDM= DAC1= DAC2=
Ambient P=100.2kPa, Tair=25.5C, Tdew=7.8C
Zero Air at 1600psi, LBL Air 513.22 at 350 Tdewspan=16.00
CO2 zero 0.9504 -> .9847 Offset: -6.7ppm
CO2 span 1.001 -> 1.0244 Offset: -15.52ppm
H2O zero 0.8519 -> .8434 Offset: -54.8ppm
H2O span 1.003 -> 1.1404 Offset: -2.72ppm
Notes: First calibration since upgrade from LI-7500 to LI-7500A
LI-7500A 0042 2017-05-03 Lab none Upgrade Lab none Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). After the upgrade, I attached this sensor to a LI-7550 Analyzer Interface Unit and manually input the CO2 coefficients (A, B, C, D, E, Z, XS) and H2O coefficients (A, B, C, ZX, S). I also updated the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength (Calibration icon > Signal Strength tab > Reset Signal Strength).
LI-7500A 0041 2017-05-03 Lab none Upgrade Lab none Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). After the upgrade, I attached this sensor to a LI-7550 Analyzer Interface Unit and manually input the CO2 coefficients (A, B, C, D, E, Z, XS) and H2O coefficients (A, B, C, ZX, S). I also updated the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength (Calibration icon > Signal Strength tab > Reset Signal Strength).
LI-7500A 0035 2017-05-03 Lab none Upgrade Lab none Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). After the upgrade, I attached this sensor to a LI-7550 Analyzer Interface Unit and manually input the CO2 coefficients (A, B, C, D, E, Z, XS) and H2O coefficients (A, B, C, ZX, S). I also updated the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength (Calibration icon > Signal Strength tab > Reset Signal Strength).
WM 1590 170523 2017-05-02 Lab none Initial settings Lab none All Checksum fields report PASS. Anemometer Firmware Version: 2329-701

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Resolution settings : Normal resolution
Resolution settings : High resolution
Output rate : P1 1 Hz
Output rate : P8 20 Hz
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER ========================

Serial Number : W170523

---- SW VERSION --------------------------------------------
Anemometer Firmware Version: 2329-701

---- HW DETAILS OF ANEMOMETER ------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER ----------------------
Baud Rate : 19200
Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : Normal resolution
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound
Synchronised Polling : Off
WM 1590 170524 2017-05-02 Lab none Initial settings Lab none All Checksum fields report PASS. Anemometer Firmware Version: 2329-701

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Resolution settings : Normal resolution
Resolution settings : High resolution
Output rate : P1 1 Hz
Output rate : P8 20 Hz
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : 0v to 5V
Analogue output mode : -5V to 5V
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER ========================

Serial Number : W170524

---- SW VERSION --------------------------------------------
Anemometer Firmware Version: 2329-701

---- HW DETAILS OF ANEMOMETER ------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER ----------------------
Baud Rate : 19200
Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : Normal resolution
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : 0v to 5V
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound
Synchronised Polling : Off
WM 1590 170525 2017-05-02 Lab none Initial settings Lab none All Checksum fields report PASS. Anemometer Firmware Version: 2329-701

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Resolution settings : Normal resolution
Resolution settings : High resolution
Output rate : P1 1 Hz
Output rate : P8 20 Hz
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : -2.5V to +2.5V
Analogue output mode : -5V to 5V
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER ========================

Serial Number : W170525

---- SW VERSION --------------------------------------------
Anemometer Firmware Version: 2329-701

---- HW DETAILS OF ANEMOMETER ------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER ----------------------
Baud Rate : 19200
Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : Normal resolution
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : -2.5V to +2.5V
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound
Synchronised Polling : Off

LI-7500A 75H-2180 2017-04-27 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on Mayberry to replace sn 1594 for calibration.
LI-7500A 1594 2017-04-27 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for calibration. Replaced by sn 75H-2180.
AM25T 7908 2017-04-26 Lab none Installed Bouldin Corn Multiplexer Thermocouple Installed at the Bouldin Corn site with the initial set up of the tower.
SRS-Nr NDVI 943103015 2017-04-26 Lab none Installed Bouldin Corn Narrow Band NDVI Reflected Installed at the Bouldin Corn site with the initial set up of the tower.
SRS-Ni NDVI 882103020 2017-04-26 Lab none Installed Bouldin Corn Narrow Band NDVI Incoming Installed at the Bouldin Corn site with the initial set up of the tower.
SRS-Pi PRI 913003004 2017-04-26 Lab none Installed Bouldin Corn Narrow Band PRI Incoming Installed at the Bouldin Corn site with the initial set up of the tower.