TE525 48130-611 2012-01-09 Lab none Installed Tonzi Installed on the ground at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
AM16/32B 10194 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
CR3000 38970 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
CE4 1329 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
OS136-1-MVC NO1107536 2012-01-09 Lab none Installed Tonzi Installed at SP2 near the Tower. The sensor is pointing at the ground to the south at about a 45deg angle from vertical. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX350J T21009229B43579 2012-01-09 Lab none Installed Tonzi Installed at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
RF401 35787 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15836 2012-01-09 Lab none Installed Tonzi Installed at -70cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15818 2012-01-09 Lab none Installed Tonzi Installed at -20cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15846 2012-01-09 Lab none Installed Tonzi Installed at -5cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15822 2012-01-09 Lab none Installed Tonzi Installed at -10cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15840 2012-01-09 Lab none Installed Tonzi Installed at -30cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15856 2012-01-09 Lab none Installed Tonzi Installed at -50cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1405 2012-01-09 Lab none Installed Tonzi Installed at -70cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1308 2012-01-09 Lab none Installed Tonzi Installed at -50cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1407 2012-01-09 Lab none Installed Tonzi Installed at -30cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1396 2012-01-09 Lab none Installed Tonzi Installed at -20cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1282 2012-01-09 Lab none Installed Tonzi Installed at -10cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1285 2012-01-09 Lab none Installed Tonzi Installed at -5cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1306 2012-01-09 Lab none Installed Tonzi Installed at -2cm at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
TE525 48132-611 2012-01-09 Lab none Installed Tonzi Installed on the ground at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
AM16/32B 10195 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
CR1000 41909 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
CE4 1328 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
OS136-1-MVC NO1107535 2012-01-09 Lab none Installed Tonzi Installed at SP3 southeast of the Tower. The sensor is pointing at the ground to the south at about a 45deg angle from vertical. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX350J T21009229B43638 2012-01-09 Lab none Installed Tonzi Installed at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
RF401 35754 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15839 2012-01-09 Lab none Installed Tonzi Installed at -45cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15833 2012-01-09 Lab none Installed Tonzi Installed at -20cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15825 2012-01-09 Lab none Installed Tonzi Installed at -30cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15824 2012-01-09 Lab none Installed Tonzi Installed at -5cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15942 2012-01-09 Lab none Installed Tonzi Installed at -65cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15843 2012-01-09 Lab none Installed Tonzi Installed at -10cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1296 2012-01-09 Lab none Installed Tonzi Installed at -65cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1395 2012-01-09 Lab none Installed Tonzi Installed at -30cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1297 2012-01-09 Lab none Installed Tonzi Installed at -45cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1388 2012-01-09 Lab none Installed Tonzi Installed at -20cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1315 2012-01-09 Lab none Installed Tonzi Installed at -10cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1286 2012-01-09 Lab none Installed Tonzi Installed at -5cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TM 5TM-1342 2012-01-09 Lab none Installed Tonzi Installed at -2cm at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
TE525 48102-611 2012-01-09 Lab none Installed Tonzi Installed on the ground at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
AM16/32B 10208 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
CR1000 41907 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
CE4 1327 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
eosFD FD20160001 2018-07-26 Bouldin Alfalfa Forced Diffusion A Removed Lab none Removed from BA because it was not turning on after power cycling and brought back to the lab for testing. I think the chamber is fine, just needs to be off for 1 minute before being turned back on.
SRS-Pr PRI 930103961 2018-07-26 Lab none Installed Bouldin Alfalfa Narrow Band PRI Reflected Installed again at BA after spider's nest was cleaned out of one of its lenses.
miniDOT 1219 2018-07-17 Lab none Installed East End DO Sensor Turned on and reinstalled with fresh batteries. The water level is very low so the sensor end is just sitting on the muck.
BB-Fab BB-EP 2018-07-17 East Pond Battery Bank Upgrade East Pond Battery Bank I turned off all the solar power and removed the four remaining size 27 batteries. These batteries were from Mayberry and were about seven years old. I replaced them with five brand new (July 2018 tags) size 31 Duracell deep cycle batteries. These are part number SLI31MDC and have a capacity of 105Ah. Size 27 are rated at 90Ah, so about the same capacity as six size 27s. After the swap the solar panels were feeding the batteries 20A.
ML2x 377/010 2018-07-11 Lab none Installed Sherman Barn Soil Moisture Probe C Installed at Sherman Barn to the SW of the scaffolding at -5cm. This is the initial set up of the site.
ML2x 54/016 2018-07-11 Lab none Installed Sherman Barn Soil Moisture Probe B Installed at Sherman Barn to the W of the scaffolding at -5cm. This is the initial set up of the site.
ML2x 377/008 2018-07-11 Lab none Installed Sherman Barn Soil Moisture Probe A Installed at Sherman Barn to the NW of the scaffolding at -5cm. This is the initial set up of the site.
HFP01 6498 2018-07-11 Lab none Installed Sherman Barn Heat Flux Plate C Installed at Sherman Barn to the SW of the scaffolding at -2cm. This is the initial set up of the site.
HFP01 6491 2018-07-11 Lab none Installed Sherman Barn Heat Flux Plate B Installed at Sherman Barn to the W of the scaffolding at -2cm. This is the initial set up of the site.
HFP01 6342 2018-07-11 Lab none Installed Sherman Barn Heat Flux Plate A Installed at Sherman Barn to the NW of the scaffolding at -2cm. This is the initial set up of the site.
miniDOT 1219 2018-07-12 East End DO Sensor Removed Lab none Brought back to the lab to replace batteries.
LI-7500A 75H-2176 2018-07-12 Lab none Installed East End Gas Analyzer CO2/H2O Installed at EE to replace sn 0073, which was brought back to the lab for calibration.
LI-7500A 0073 2018-07-12 East End Gas Analyzer CO2/H2O Removed Lab none This sensor was removed from EE and brought back to the lab for calibration. it was replaced by sn 75H-2176.
LI-6400 PSC-2474 2018-07-16 Lab none Calibration Lab none Elke Eichelmann & Katrina Cone

We borrowed this LI-6400 from the Silver lab and they had not zeroed and spanned it for about 10 years. We checked the zero with the chemicals and they were a little off, so we set the zero and the span using cal gas and a dewpoint generator. Considering how long this analyzer had not been calibrated the settings were actually pretty good. But it also hadn't been used for most of that time.

Then we zeroed and spanned:
Zero:
Checked using chemicals: CO2_S: 15.2 CO2_R: 7.8 H2O_S: 5.7 H2O_R: 5.7

Set zero using cal gas (zero air):
Before: CO2_S:16.8 CO2_R: 6.82 H2O_S: 0.32 H2O_R: 0.282
After: *All fluctuating around zero.

Span with 408.97ppm cal gas:
Before: CO2_R: 404.4 CO2_S: 404.8
After: *Fluctuating around 409

Span with 16.0C dew point:
Before: H2O_R: 15.06C H2O_S:15.02C
After: *Fluctuating between 15.99 and 16.01
LI-6400 PSC-0785 2018-07-16 Lab none Calibration Lab none Elke Eichelmann & Katrina Cone

The 6400 had not been zeroed and spanned for a while. We checked the zero with the chemicals and they were very off, so we set the zero and the span using cal gas and dewpoint generator.

Then I zeroed and spanned:
Zero:
Chemicals: CO2_S: 34.0 CO2_R: 25.4 H2O_S: 13.7 H2O_R: 14.1
Cal Gas (zero air):
Before: CO2_S:41.0 CO2_R: 24.3 H2O_S: 3.5 H2O_R: 2.3
After: *All fluctuating around zero.

Span with 408.97ppm cal gas:
Before: CO2_R: 399.9 CO2_S: 397.1
After: *Fluctuating around 409

Span with 16.0C dew point:
Before: H2O_R: 14.69C H2O_S:13.23C
After: H2O_R: 16.01C H2O_S: 15.99C
**Note: we had to override the maximum slope parameter for the H2O_S to be able to set the slope above 1.100 and to set the correct dewpoint. 'Shift' + 'J' will allow you to override the maximum slope.
CS451 70011229 2018-07-12 Mayberry Water Level Removed Lab none Removed from MB because it had diurnal fluctuations, which seem suspect. It was replaced by sn 70011236.
CS451 70011236 2018-07-12 Lab none Installed Mayberry Water Level Installed at MB to replace sn 70011229, which was removed b/c of suspect diurnal fluctuations.
PTB110 L1920571 2018-07-12 Bouldin Alfalfa Barometer Removed Lab none Removed from BA because pressure has been slowly decreasing since Feb and does not match BC pressure data.
SRS-Pr PRI 930103961 2018-07-12 Bouldin Alfalfa Narrow Band PRI Reflected Removed Lab none Removed from BA because 570nm reading has been close to 0 since 2018-06-20.
LI-7500A 0042 2018-07-12 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at WP to replace sn 75H-2514, which was brought back to the lab for calibration.
LI-7500A 75H-2514 2018-07-12 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from WP and brought back to the lab for calibration. It was replaced by sn 0042.
eosFD FD20160001 2018-07-12 Lab none Installed Bouldin Alfalfa Forced Diffusion A Installed at BA under the scaffolding. Sampling frequency is 5 min.
LI-7500A 0065 2018-07-12 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from BA and brought back to the lab for calibration. It was replaced by sn 75H-2150.
LI-7500A 75H-2150 2018-07-12 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at BA to replace sn 0065, which was brought back to the lab for calibration.
LI-7500A 75H-2182 2018-07-12 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at BC to replace sn 75H-2668, which was brought back to the lab for calibration.
LI-7500A 75H-2668 2018-07-12 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from BC and brought back to the lab for calibration. It was replaced by sn 75H-2182.
HMP45AC U3030041 2018-07-11 Lab none Installed Sherman Barn RH & Temp Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
TE525MM 49187-109 2018-07-11 Lab none Installed Sherman Barn Precipitation Installed at Sherman Barn on a T-post in the northwest corner of the cow fence. This is the initial set up of the site.
NR01 1570 2018-07-11 Lab none Installed Sherman Barn Net Radiometer 4-way Installed at Sherman Barn on the radiation boom just outside the southwest corner of the cow fence. This is the initial set up of the site.
SRS-Ni NDVI 882103061 2018-07-11 Lab none Installed Sherman Barn Narrow Band NDVI Reflected Installed at Sherman Barn on the radiation boom just outside the southwest corner of the cow fence. This is the initial set up of the site.
SRS-Ni NDVI 882103036 2018-07-11 Lab none Installed Sherman Barn Narrow Band NDVI Incoming Installed at Sherman Barn on the radiation boom just outside the southwest corner of the cow fence. This is the initial set up of the site.
SRS-Pi PRI 913003007 2018-07-11 Lab none Installed Sherman Barn Narrow Band PRI Reflected Installed at Sherman Barn on the radiation boom just outside the southwest corner of the cow fence. This is the initial set up of the site.
SRS-Pi PRI 913003006 2018-07-11 Lab none Installed Sherman Barn Narrow Band PRI Incoming Installed at Sherman Barn on the radiation boom just outside the southwest corner of the cow fence. This is the initial set up of the site.
PQS 1 110338 2018-07-11 Lab none Installed Sherman Barn PAR Sensor Reflected Installed at Sherman Barn on the radiation boom just outside the southwest corner of the cow fence. This is the initial set up of the site.
PQS 1 110339 2018-07-11 Lab none Installed Sherman Barn PAR Sensor Incoming Installed at Sherman Barn on the radiation boom just outside the southwest corner of the cow fence. This is the initial set up of the site.
CS451 20010046 2018-07-11 Lab none Installed Sherman Barn Water Level Installed at Sherman Barn just to the north of the scaffolding. This is the initial set up of the site.
WM 1590 153802 2018-07-11 Lab none Installed Sherman Barn Anemometer Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
LI-7700 TG1-0357 2018-07-11 Lab none Installed Sherman Barn Gas Analyzer CH4 Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
LI-7500A 0041 2018-07-11 Lab none Installed Sherman Barn Gas Analyzer CO2/H2O Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
LI-7550 AIU-1557 2018-07-11 Lab none Installed Sherman Barn Interface LI-7550 Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
AM25T 6767 2018-07-11 Lab none Installed Sherman Barn Multiplexer Thermocouples Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
CR1000 47108 2018-07-11 Lab none Installed Sherman Barn Data Logger Met Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
AM16/32B 11662 2018-07-11 Lab none Installed Sherman Barn Multiplexer Met Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
NL115 6763 2018-07-11 Lab none Installed Sherman Barn Interface NL115 Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
SD130BN 0030F4D2148D 2018-07-11 Lab none Installed Sherman Barn Camera Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
LI-7500A 0042 2018-07-10 Lab none Calibration Lab none Last calibration: 2018-07-03
AGC=98 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=26.8C, Tdew=C
Zero Air at 200psi, LBL Air 408.97 at 1400 Tdewspan=16.00
CO2 zero 0.9485 -> .9485 Offset: .03ppm
CO2 span 1.0054 -> 1.0039 Offset: .77ppm
H2O zero 0.8407 -> .8406 Offset: -.9ppm
H2O span 1.0418 -> 1.0158 Offset: .55ppm
Notes: Calibrated again b/c previous calibration on 2018-07-03 had a large H2O zero offset, but the offsets today are fine, so the calibration seemed fine.
LI-7500A 75H-2182 2018-07-10 Lab none Calibration Lab none Last calibration: 2018-03-05
AGC=87 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=26.3C, Tdew=13.00C
Zero Air at 200psi, LBL Air 408.97 at 1400 Tdewspan=16.00
CO2 zero 0.8844 -> .8841 Offset: -.91ppm
CO2 span 1.0217 -> 1.0168 Offset: 2.33ppm
H2O zero 0.8453 -> .845 Offset: -1.7ppm
H2O span 1.0169 -> 1.0202 Offset: -.08ppm
Notes:
LI-7500A 75H-2176 2018-07-10 Lab none Calibration Lab none Last calibration: 2018-03-05
AGC=101.3 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=27C, Tdew=13.3C
Zero Air at 175psi, LBL Air 408.97 at 1400 Tdewspan=16.00
CO2 zero 0.8679 -> .8671 Offset: -3.65ppm
CO2 span 1.0057 -> 1.0029 Offset: 1.23ppm
H2O zero 0.8486 -> .848 Offset: -3.4ppm
H2O span 0.9112 -> .9088 Offset: .05ppm
Notes:
LI-7500A 75H-2150 2018-07-10 Lab none Calibration Lab none Last calibration: 2018-03-05
AGC=97 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=27.2C, Tdew=12.7C
Zero Air at 150psi, LBL Air 408.97 at 1300 Tdewspan=16.00
CO2 zero 0.9023 -> .9017 Offset: -2.64ppm
CO2 span 1.0093 -> 1.002 Offset: 3.79ppm
H2O zero 0.9655 -> .8662 Offset: 5.1ppm
H2O span 1.0096 -> 1.0069 Offset: .06ppm
Notes:
LI-7500A 0042 2018-07-03 Lab none Calibration Lab none Last calibration: 2018-03-05
AGC=97.8 SDM= DAC1=- DAC2=-
Ambient P=100.1kPa, Tair=23.25C, Tdew=11.75C
Zero Air at 245psi, AMP 408.97 at 1300 Tdewspan=16.0
CO2 zero 0.9485 -> 0.9485 Offset: 0.04ppm
CO2 span 1.0124 -> 1.0054 Offset: 3.53ppm
H2O zero 0.843 -> 0.8407 Offset: -14ppm
H2O span 1.0225 -> 1.0418 Offset: ppm
Notes: Robert in Training: Forgot to write down H2O span offset in C but it was -19mmol/m3. Large change in H2O zero offset. Daphne redid the calibration on 7/10 and the H2O zero offset was stable <2mmol/m3 change.
LI-7500A 0041 2018-07-03 Lab none Calibration Lab none Last calibration: 2018-03-05
AGC=00 SDM=0 DAC1=0 DAC2=0
Ambient P=100.1kPa, Tair=22.2C, Tdew=12.5C
Zero Air at 300psi, LBL Air 408.97 at 1300 Tdewspan=16.00
CO2 zero 0.9972 -> .9966 Offset: -1.94ppm
CO2 span 1.0196 -> 1.0173 Offset: 2.0ppm
H2O zero 0.8194 -> .8192 Offset: -.9ppm
H2O span 1.0267 -> 1.0285 Offset: -.03ppm
Notes: Training Robert
LI-7700 TG1-0357 2018-07-10 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-11-09
RSSI=74.6%, Optical RH=2%
Ambient P=100.5kPa, Tair=25.8C
Zero Air at 220psi, LBL Air 1.9497 at 1400
CH4 Zero Offset: -0.001ppm
CH4 Span Offset: 0.0127ppm
Notes:
PAR-LITE 060868 2018-07-06 Lab none Returned Returned Returned to Ameriflux for calibration. Hopefully to be replaced by another freshly calibrated sensor.
WM 1590 153803 2018-06-28 East Pond Anemometer Moved East Pond Anemometer Sensor moved up 26cm to try to address rotation angle issues. The water level was 68cm and the sensor was 540cm above water level.
LI-7700 TG1-0211 2018-06-28 East Pond Gas Analyzer CH4 Moved East Pond Gas Analyzer CH4 Sensor moved up 26cm to try to address rotation angle issues. The water level was 68cm and the sensor was 534cm above water level.
LI-7500A 0418 2018-06-28 East Pond Gas Analyzer CO2/H2O Moved East Pond Gas Analyzer CO2/H2O Sensor moved up 26cm to try to address rotation angle issues. The water level was 68cm and the sensor was 540cm above water level.