Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 [17] 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65
Ser No.DateLoc FromTitleLoc ToNotes
PRB-K Type #8 000000402FF3 2021-02-10 Tonzi Soil Moisture Probe F -0.50m Removed Lab none Removed from Tonzi floor due to fixed readings for a long while. This sensor was out of the ground at the edge of the trees near the path to the tower. It was buried vertically with the top of the sensor 31cm below the surface - the time were 44-50cm below the surface.
PAR-LITE 990031 2021-02-10 Tonzi PAR Sensor Incoming Cal Removed Lab none Removed from Tonzi Tower after three weeks of inter comparison with the AMP standard
PAR-LITE 990025 2021-02-10 Cal Kit PAR Sensor Incoming Removed Lab none Removed from TZ Tower after three weeks of inter comparison with the AMP standard
WM 1590 153805 2021-02-10 Lab none Installed Cal Kit Anemometer Installed on the TZ Tower for inter-comparison. This sensor is the one that got wet when the BC tower fell over. It's mounted on the post just north of the main eddy sensors, at the same height, and oriented to true north.
eosFD FD20200001 2021-02-10 Lab none Installed Vaira Forced Diffusion North Installed near the old north soil co2 profile. This unit has a homemade solar shield to try to keep it cooler. Could use a soil collar.
eosFD FD20200001 2021-02-04 Factory none Returned Lab none Returned from Eosense where its pump relay board was replaced under warranty repair.
GP-001 GP20140026 2021-02-04 West Pond CO2 Probe B Removed Lab none Removed from WP and brought back to the lab b/c this sensor has been reading noise since late Nov
LI-7500A 1594 2021-02-01 Lab Calibration Lab Last calibration: 2020-08-20
AGC=99.3 SDM=0 DAC1=- DAC2=-
Ambient P=100.3kPa, Tair=28.4C, Tdew=11.0C
Zero Air at 0psi, AMP CA04396 408.97 at 0 Tdewspan=20.0
CO2 zero 0.8651 -> 0.8646 Offset: -2.1ppm
CO2 span 1.02124 -> 1.0131 Offset: -0.32ppm
H2O zero 0.8838 -> 0.8890 Offset: 29.85ppm
H2O span 1.015 -> 1.000 Offset: 0.35ppm
Notes: The usual 7550 box in the lab is running the newest firmware v8.8 and is complaining that the v8.5 config file has bad number formats for GPS locations. I used 7550 sn AIU-1866 for calibration instead. This box and head will go to replace equipment on the BC tower that fell in the water.
NR01 2420 2021-01-30 Lab none Damaged Lab none This sensor was removed from a submerged position on the fallen BC tower. The shortwave side had been torn loose breaking the wires similar to last time it was damaged. After returning to the lab on the 29th I pulled it apart and tried to air it out. On the 30th it was still damp inside and I ran dry air through the holes under the circuit boards for two days.
WM 1590 153805 2021-02-01 Lab none Tested Lab none After being removed from a partially submerged position on the fallen BC tower, I tested this sensor in the lab. There was no evidence that water had leaked inside. I connected it to a Licor system and used fans to blow air through in on the floor. The response sign and magnitude looked about right. I plan to run it on the Tonzi tower for a couple weeks to check its performance.
LI-7500A 75H-2176 2021-02-01 Lab none Tested Lab none On the 29th after being retrieved from a partially submerged position on the fallen BC tower, I removed the desiccant bottles and saw no evidence that water had gotten inside. I flowed dry air into the desiccant bottle openings over night. On the 1st, after consultation with Licor, I powered up this head on the calibration rig in the lab. It seemed to operate fine. I flooded the base with zero air before sealing it back up but it needs to run for a while to equilibrate and be calibrated.
LI-7700 TG1-0155 2021-02-04 Lab none Shipped Factory none Shipped to Licor and RMA 00225226 for evaluation after being submerged when the BC tower fell in the storm last week.
LI-7700 TG1-0155 2021-02-01 Lab none Tested Lab none On the 29th I removed this sensor from being submerged after the BC tower fell just after midnight on the morning of the 27th. On the evening of the 29th, about 1/2 cup of water came out of the bottom when I removed the base. I removed the desiccant bottle in the head and saw no evidence that water had leaked in there. I left dry air flowing inside the base over night. On the 30th I removed the bottom case and only saw minor water damage on the pressure sensor. On the 1st, after consultation with Licor, I powered up the analyzer and got green power and status lights but no communications on the Ethernet. It will need to be returned to Licor for further evaluation.
SRS-Pi PRI 897103002 2021-02-03 Lab none Installed Bouldin Corn Narrow Band PRI Reflected Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
SRS-Ni NDVI 2046503357 2021-02-03 Lab none Installed Bouldin Corn Narrow Band NDVI Reflected Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
SRS-Ni NDVI 882103020 2021-02-03 Lab none Installed Bouldin Corn Narrow Band NDVI Incoming Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
CNR1 990181 2021-02-03 Lab none Installed Bouldin Corn Net Radiometer 4-way Installed on the BC tower to replace the Hukseflux NR01 that broke when the tower fell and still needs repair and cross comparison.
Mark II C006123 2021-02-03 Lab none Installed Bouldin Corn Weather Station B Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
Mark II C003251 2021-02-03 Lab none Installed Bouldin Corn Weather Station C Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
PA-II-SD 68c63acc07a9 2021-02-03 Lab none Installed Bouldin Corn Air Quality Sensor Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
SD130BN 0030F4D21491 2021-02-03 Lab none Installed Bouldin Corn Camera Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
PQS 1 140441 2021-02-03 Lab none Installed Bouldin Corn PAR Sensor Reflected Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
PQS 1 140453 2021-02-03 Lab none Installed Bouldin Corn PAR Sensor Incoming Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
SRS-Pi PRI 913003004 2021-02-03 Lab none Installed Bouldin Corn Narrow Band PRI Incoming Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
LI-7500A 1594 2021-02-03 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed on the BC tower to replace sensors damaged when the tower fell last week
LI-7550 AIU-1866 2021-02-03 Lab none Installed Bouldin Corn Interface LI-7550 Installed on the BC tower to replace sensors damaged when the tower fell last week
LI-7700 TG1-0224 2021-02-03 Lab none Installed Bouldin Corn Gas Analyzer CH4 Installed on the BC tower to replace sensors damaged when the tower fell last week
WM 1590 153803 2021-02-03 Lab none Installed Bouldin Corn Anemometer Installed on the BC tower to replace sensors damaged when the tower fell last week
WL16U 57278 2018-05-08 Lab none Repair Lab none I stripped off the old rotting insulation and then slid on a long continuous piece of heat shrink. I shrunk the whole length of the cable. I disassembled the data logging enclosure and kept only the data logger board and USB connector in the end cap. I cut off the twin 9V battery plugs and will power the sensor from solar connected rechargeable battery.
WL16U 57278 2018-03-27 Tonzi Water Level Floor Removed Lab none Removed from the Floor Well. The cable insulation is rotting and needs to be replaced. Replaced by the new METER water level sensor a couple of weeks ago.
CMTD-As4s0 3317052234210 2018-05-11 Lab none Installed Tonzi none Installed at the Road Well to power Global Water sensor.
WL16U 57278 2018-05-11 Lab none Installed Tonzi Water Level Road Installed at the Road Well to measure water level at 5min intervals storing 30min averages
Mark II C006123 2021-01-29 Bouldin Corn Weather Station B Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Maybe some rain in the Tair bulge
Mark II C003251 2021-01-29 Bouldin Corn Weather Station C Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Maybe some rain in the Tair bulge
PA-II-SD 84f3ebd5ffb1 2021-01-29 Bouldin Corn Air Quality Sensor Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Got rained on and the power connector is corroded
SD130BN 0030F4D21491 2021-01-29 Bouldin Corn Camera Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
SRS-Pi PRI 897103002 2021-01-29 Bouldin Corn Narrow Band PRI Reflected Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
SRS-Pi PRI 913003004 2021-01-29 Bouldin Corn Narrow Band PRI Incoming Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27.
SRS-Ni NDVI 2046503357 2021-01-29 Bouldin Corn Narrow Band NDVI Reflected Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
SRS-Ni NDVI 882103020 2021-01-29 Bouldin Corn Narrow Band NDVI Incoming Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
PQS 1 140441 2021-01-29 Bouldin Corn PAR Sensor Reflected Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
PQS 1 140453 2021-01-29 Bouldin Corn PAR Sensor Incoming Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
NR01 2420 2021-01-29 Bouldin Corn Net Radiometer 4-way Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged and broken in two pieces
LI-7550 AIU-0614 2021-01-29 Bouldin Corn Interface LI-7550 Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27.
WM 1590 153805 2021-01-29 Bouldin Corn Anemometer Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Partially submerged
LI-7700 TG1-0155 2021-01-29 Bouldin Corn Gas Analyzer CH4 Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
LI-7500A 75H-2176 2021-01-29 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Partially submerged
Mark II C003800 2021-01-21 Bouldin Alfalfa Weather Station PL_N-A Removed Lab none Removed from the northeast Planet Labs location due to broken antenna caused by sheep
Mark II C003789 2021-01-21 Bouldin Alfalfa Weather Station PL_N-B Removed Lab none Removed from the northwest Planet Labs location due to broken antenna caused by sheep
Mark II C006164 2021-01-21 Lab none Installed Bouldin Alfalfa Weather Station B Installed at Bouldin Alfalfa to replace Mark I sn A001311 which are being phased out.
Mark I A001311 2021-01-21 Bouldin Alfalfa Weather Station B Removed Lab none Removed from Bouldin Alfalfa as Mark I's are being phased out. Replaced by Mark II sn C006164.
Mark II C006123 2021-01-21 Lab none Installed Bouldin Corn Weather Station B Installed at Bouldin Corn to replace Mark I sn A001359 which is being phased out.
Mark I A001359 2021-01-21 Bouldin Corn Weather Station B Removed Lab none Removed from Bouldin Corn as Mark I's are phased out. Replaced by Mark II sn C006123.
Mark II C006160 2021-01-21 Lab none Installed Vaira Weather Station A Installed at Vaira to replace Mark I sn A001353 which is being phased out.
Mark I A001353 2021-01-21 Vaira Weather Station A Removed Lab none Removed from Vaira as Mark I's are being phased out. Replaced by Mark II sn C006160.
Mark I A001353 2019-04-30 Lab none Installed Vaira Weather Station A Arable Mark I installed for CalVal inter-comparison
PAR-LITE 060831 2021-01-21 Lab none Installed Cal Kit PAR Sensor Incoming Installed at TZ in the incoming position as part of the inter-comparison kit to test PAR, AirT and RH on the top of the Tower.
PAR-LITE 990031 2021-01-21 Lab none Installed Tonzi none Installed as part of the inter-comparison kit to test PAR, AirT and RH on the top of the Tower.
PAR-LITE 990025 2021-01-21 Lab none Installed Cal Kit PAR Sensor Incoming Installed at TZ as part of the inter-comparison kit to test PAR, T/RH on the top of the Tower.
HMP155 S3410851 2021-01-21 Lab none Installed Cal Kit RH & Temp Installed as part of the inter-comparison kit to test PAR, AirT and RH on the top of the Tower.
CR1000 4190 2021-01-21 Lab none Installed Cal Kit Data Logger Installed at TZ as part of the inter-comparison kit to test PAR, T/RH on the top of the Tower.
GP-001 GP20140084 2021-01-21 Lab none Tested Lab none This sensor is dead. Tested in the lab and no response from CO2 or T. The connector at the top of the sensor twists loose from the body of the sensor. There was some corrosion in the connector head and the cable.
2300-000 2300565 2021-01-17 Mayberry Fluorometer Removed Lab none Removed from MB and brought back to the lab for further testing b/c it has noisy data
GP-001 GP20140084 2021-01-17 Sherman Wetland CO2 Probe Removed Lab none Removed from SW and brought back to the lab b/c the sensor is bad and only outputting noise
LI-7550 AIU-1570 2021-01-17 West Pond Interface LI-7550 Upgrade West Pond Interface LI-7550 Upgraded embedded software from v8.5.0 to v8.8.32 in hopes this will solve the persistent 7700 sync issue
HMP155 R0360206 2021-01-17 Cal Kit RH & Temp Moved West Pond RH & Temp Moved from Comparison Set to tower to replace sn G4540025 which had high RH readings.
HMP60 G4540025 2021-01-17 West Pond RH & Temp Removed Lab none Removed from WP and brought back to the lab b/c RH was railing high. It was replaced by HMP155 from the Comparison Set, sn R0360206.
NL115 9774 2016-08-12 Lab none Installed Bouldin Alfalfa none Installed at BA with initial site set up
NL115 13056 2017-04-26 Lab none Installed Bouldin Corn none Installed at BC with initial set up of site
eosFD FD20160001 2020-12-18 Lab none Installed Bouldin Alfalfa Forced Diffusion A Installed to replace the FD Chamber with the high temperature startup problems.
eosFD FD20180047 2020-12-18 Bouldin Alfalfa Forced Diffusion A Removed Lab none Removed from Bouldin Alfalfa. This one has high temperature start up issues and was running on a relay to restart it every night. I removed the relay too but did not update the CR1000 program.
CR23X 2299 2020-11-11 Lab none Installed Vaira Data Logger Soil CO2 Installed at Vaira in the Soil CO2 system. The 0-5V range on the 23x avoids the problem of needing voltage dividers.
eosFD FD20190028 2020-12-18 Lab none Installed Vaira Forced Diffusion West Installed at Vaira West to replace sn FD20200001 for test of high temp readings in the lab.
eosFD FD20200001 2020-12-18 Vaira Forced Diffusion West Removed Lab none Removed from Vaira West for testing to high temp reading in the lab. Replaced by sn FD20190028
CR23X 2299 2020-12-18 Vaira Data Logger Soil CO2 Location Vaira Data Logger Soil CO2 Confirmed serial number and location
TE525 48130-611 2020-12-18 Tonzi none Removed Lab none Removed from OSU SP1 for experiments in the lab. The cable was disconnected inside the tipping bucket and left in place in the field.
CS547A 5688 2020-12-16 Lab none Installed Sherman Wetland Conductivity Sensor Installed at SW. Temperature sensor is broken, so 2 temperature wires were left loose.
eosFD FD20160001 2020-12-17 Factory none Returned Lab none Returned from repair at Eosense. They replaced the CO2 board and sensor, the internal battery and the conical membranes. They did and FD calibration and a FD FGen on 2020-12-14.
eosFD FD20190028 2020-12-17 Factory none Returned Lab none Returned from Eosense after repair. They replaced relay board 1 and 2 and did a Flux calibration on 2020-12-01
LI-7500A 75H-2514 2020-12-11 Lab Calibration Lab Last calibration: 2020-11-30
AGC=99.9 SDM= DAC1= DAC2=
Ambient P=100.9kPa, Tair=21.5C, Tdew=7.8C
Zero Air at 600psi, AMP CA4396 408.97 at 300 Tdewspan=20.00
CO2 zero 0.905 -> .9049 Offset: -.14ppm
CO2 span 1.0152 -> 1.0082 Offset: 3.4ppm
H2O zero 0.8398 -> .8398 Offset: .1ppm
H2O span 1.0238 -> 1.0093 Offset: .32ppm
Notes: 2nd calibration after chemical swap on 11/24. CO2/H2O zeros calibrated on 12/11. CO2/H2O spans calibrated on 12/14 (ambient dewpt=6.08C). CO2 span offset large.
CS547A 5688 2020-12-11 Lab none Tested Lab none Tested in the lab with both Campbell interface and homemade interface sn 547-01. In either case, the temperature read high, around 40 degC when it should've been around 20-25 degC. I unwrapped an old splice in the cable, but it looked perfectly fine, so I think the temperature sensor is just bad. We can still use this sensor in the future, but we'll need to be careful to do the temperature correction with another temperature sensor in the water (e.g., from water level sensor or a water thermocouple).
A547 547-01 2020-12-09 Mayberry Interface A547 Removed Lab none Removed from MB along with conductivity sensor sn 5688. Temperature reading is high so brought back to the lab for testing.
A547 547-02 2020-12-09 Lab none Installed Mayberry Interface A547 Installed with conductivity sensor sn 7157 at MB.
CS547A 7157 2020-12-09 Lab none Installed Mayberry Conductivity Sensor A Installed at MB with interface sn 547-02 to replace sn 5688, which was brought back to the lab for testing.
CS547A 5688 2020-12-09 Mayberry Conductivity Sensor A Removed Lab none Removed from MB and brought back to the lab for testing. Temperature reading is too high.
LI-7500A 0073 2020-12-09 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 75H-2669 for calibration.
LI-7500A 75H-2669 2020-12-09 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for calibration. Replaced by sn 0073.
LI-7500A 0065 2020-12-09 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at Sherman Wetland to replace sn 75H-2667 for calibration.
LI-7500A 75H-2667 2020-12-09 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Wetland for calibration. Replaced by sn 0065.
RPI0W RPI0W-004 2020-11-12 Sherman Wetland-Temp1 Camera Removed Lab none Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup.
LI-7500A 0073 2020-12-08 Lab Calibration Lab Last calibration: 2020-08-20
AGC=99.8 SDM= DAC1= DAC2=
Ambient P=100.7kPa, Tair=24.4C, Tdew=2.3C
Zero Air at 550psi, AMP CA4396 408.97 at 300 Tdewspan=20.00
CO2 zero 0.9575 -> .9564 Offset: -3.46ppm
CO2 span 1.0092 -> 1.0079 Offset: .42ppm
H2O zero 0.7312 -> .7323 Offset: 6.7ppm
H2O span 1.0118 -> 1.0037 Offset: .18ppm
Notes:
LI-7500A 0065 2020-12-08 Lab Calibration Lab Last calibration: 2020-08-20
AGC=101 SDM= DAC1= DAC2=
Ambient P=100.8kPa, Tair=23.1C, Tdew=2.5C
Zero Air at 550psi, AMP CA4396 408.97 at 300 Tdewspan=20.00
CO2 zero 0.8917 -> .8916 Offset: -.35ppm
CO2 span 1.0217 -> 1.0213 Offset: .27ppm
H2O zero 0.7297 -> .7303 Offset: 4.5ppm
H2O span 1.0186 -> 1.0165 Offset: .05ppm
Notes:
LI-7500A 0042 2020-12-03 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond to replace sn 75H-2150 for calibration.
LI-7500A 75H-2150 2020-12-03 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration. Replaced by sn 0042.
TR-525M 50229-112 2020-12-03 East End Precipitation Tested East End Precipitation Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket (1/2 of the bottle at a time) which should result in 100 tips. Got 110 tips. Did not adjust - will retest with better water metering system.
LI-7500A 0041 2020-12-03 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End to replace sn 1594 for calibration.
LI-7500A 1594 2020-12-03 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration. Replaced by sn 0041.
TE525MM 65548-615 2020-12-03 Bouldin Corn Precipitation Tested Bouldin Corn Precipitation Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket (1/2 of the bottle at a time) which should result in 100 tips. Got 117 tips. Did not adjust - will retest with better water metering system.
LI-7500A 75H-2176 2020-12-03 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 0073 for calibration.
LI-7500A 0073 2020-12-03 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 75H-2176.