Equipment and Calibration

View History

Equipment Entry

Instrument:Interface
Manufacture:Licor
Model No:LI-7550
Serial No:AIU-0666
Tag No:NULL
Acquisition Date:2012-02-07
Project:DWR-Wetland
Location on 2021-06-02:Tonzi-Floor, NULL
Notes: Part of GHG1 system with LI-7700 sn TG1-0224 and LI-7500A sn 75H-2182

Documents

Calibration: cal_Licor_LI-7550_snAIU-0666_20111130.pdf
PackingSlip: ps_Licor_20120207.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
5296 2021-06-02 Tonzi-Floor Interface LI-7550 Upgrade Tonzi-Floor Interface LI-7550 The firmware was upgraded from v8.5.0 to v8.8.32 for standardization with the Delta Licor systems
3979 2019-05-07 Lab none Installed Tonzi-Floor Interface LI-7500 Installed on the Floor tower to replace the one that was moved to Tonzi Tower last week.
3416 2018-04-19 East Pond Gas Analyzer CO2/H2O Removed Lab none Removed from East Pond because it is supposed to be use in the lab for calibrations.
3400 2018-04-17 Lab none Installed East Pond Gas Analyzer CO2/H2O Installed at EP tower as part of initial site set up.
3356 2018-03-09 Lab none Maintenance Lab none Cleaned out the box of bug detritus and replaced the gasket on the front lid. This box is missing the pressure port (the hole is covered by electrical tape) and the reset button is missing, so this box should stay in the lab for now for calibration purposes.
3315 2018-01-31 West Pond Interface LI-7550 Removed Lab none Brought back to the lab for some TLC. It was replaced by AIU-1570.
3306 2018-01-16 West Pond Interface LI-7550 Upgrade West Pond Interface LI-7550 Downgraded embedded Licor software from v8.7.5 to 8.5.0 because 8.7.5 caused mismatch problems between data columns and headers
2240 2016-04-13 West Pond Interface LI-7550 Upgrade West Pond Interface LI-7550 Updated the 7550 firmware from v7.3.0 to v8.0.0 skiping version 7.5.0
Relevant 7.5.0 changes: Main Window is now resizable, Added serial number entry for sonics, Help update, Fixed timeout when saving config files
Relevant 8.0.0 changes: Separate bandwidth settings for analog inputs and gas channels, Detection of new RS instruments, Additional calibration coefficients, Added sonic software version entry, Added chopper temp logging
1592 2014-10-02 West Pond Interface LI-7550 Upgrade West Pond Interface LI-7550 Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
1542 2014-09-08 Lab none Installed West Pond Interface LI-7550 Installed on the West Pond tower with head sn 75H-2180. There was now 7550-7500A pair on the tower because of issues with swapping the head on 9/4. This box has a new door gasket.
1591 2014-02-28 Lab none Upgrade Lab none Upgraded firmware from v6.5.5 to v7.0.1 - FPGA v6.0.3 not needing update. Improved interface, support for SmartFlux, bug fixes.
1055 2013-08-21 West Pond Interface LI-7550 Removed Lab none Removed from the West Pond tower. This box has been suffer an ant infestation for many weeks. It was removed to be de-ant-ified, cleaned up and tested in the lab.
875 2013-01-03 Lab none Installed West Pond Interface LI-7550 Installed on the West Pond tower with head 0065 to replace AIU-0397 for repair of condensation problem.
868 2012-12-21 Factory Returned Lab none Returned from Licor with 75H-0065 after repair of condensation problem. The external connector board has been completely sealed and new firmware v6.0.0 has been loaded to the box. Pressure Sensor sn PX-2763: A0=57.595, A1=15.231
849 2012-11-02 Lab none Shipped Factory Shipped to Licor for repair of condensation problem, RMA 172095
829 2012-10-30 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from the Mayberry Tower due to corrosion on the connector circuit board. Replaced by sn 0614
547 2012-04-24 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry eddy tower to swap out LI-7500 sn 1594
 
 