Ser No.DateLoc FromTitleLoc ToNotes
HMP155 S3420870 2021-03-16 Lab none Installed Bouldin Alfalfa RH & Temp Installed at BA to replace sn G4540021, which was brought back to the lab for testing.
HMP60 G4540021 2021-03-16 Bouldin Alfalfa RH & Temp Removed Lab none Removed from BA and brought back to the lab b/c AirT has been drifting low. it was replaced by HMP155 sn S3420870.
LI-7500A 0065 2021-03-16 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End to replace sn 0041 for calibration.
LI-7500A 0041 2021-03-16 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration and brought back to the lab. Replaced by sn 0065.
LI-7500A 0073 2021-03-16 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond to replace sn 0042 for calibration.
LI-7500A 0042 2021-03-16 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration and brought back to the lab. Replaced by sn 0073.
GP-001 GP20150111 2021-03-16 Lab none Installed Mayberry CO2 Probe A Installed at MB to replace sn GP20160070. It was installed at the same height as the miniDOT on the floating boom.
GP-001 GP20160070 2021-03-16 Mayberry CO2 Probe A Removed Lab none Removed from MB b/c it was sporadically reading noise. it was replaced by sn GP20150111.
CR1000 23865 2021-03-16 Sherman Wetland Data Logger Cal Removed Lab none Removed from SW at the conclusion of the 4-way intercomparison test and brought back to the lab
NR01 2420 2021-03-16 Sherman Wetland Net Radiometer 4-way Cal A Removed Lab none Removed from SW at the conclusion of the 4-way intercomparison test and brought back to the lab
CNR4 121038 2021-03-16 Sherman Wetland Net Radiometer 4-way Cal B Removed Lab none Removed from SW at the conclusion of the 4-way intercomparison test and brought back to the lab
LI-7500A 0073 2021-03-15 Lab Calibration Lab Last calibration: 2020-12-08
AGC=99 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=23.7C, Tdew=1.2C
Zero Air at 400psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.9564 -> .957 Offset: 1.71ppm
CO2 span 1.0079 -> 1.0077 Offset: .14ppm
H2O zero 0.7323 -> .7324 Offset: -1.7ppm
H2O span 1.0037 -> 1.0024 Offset: .04ppm
Notes: Calibrated on software v8.8.36 with 7550 firmware v8.8.32
LI-7500A 0065 2021-03-15 Lab Calibration Lab Last calibration: 2020-12-08
AGC=102 SDM= DAC1= DAC2=
Ambient P=99.8kPa, Tair=24.0C, Tdew=1.55C
Zero Air at 400psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.8916 -> .8914 Offset: -1.05ppm
CO2 span 1.0213 -> 1.0153 Offset: .74ppm
H2O zero 0.7303 -> .7311 Offset: 6.2ppm
H2O span 1.0165 -> 1.017 Offset: -.02ppm
Notes:
GP-001 GP20200089 2021-03-09 Factory none Calibration Factory none Factory calibration: y = mV/5000*1.2*2%
GP-001 GP20210002 2021-03-09 Factory none Calibration Factory none Factory calibration y=mV/5000*1.2*2%
eosFD FD20180068 2021-03-15 Factory none Returned Lab none Returned from Eosense. It was sent in for zero CO2 readings but Eosense reported that it seemed to be reading correctly. It was calibrated from CO2 concentration and CO2 flux.
eosFD FD20180068 2021-01-12 Lab none Shipped Factory none Sent to Eosense for repair due to zero CO2 readings
eosFD FD20180047 2021-03-15 Factory none Returned Lab none Returned from Eosense for high temp start problem and noisy CO2 readings. It had a failed clock crystal causing start up issues at high temperature, it was given custom firmware to bypass the component. It was calibrated for CO2 concentration and CO2 flux
eosFD FD20180047 2021-01-12 Lab none Shipped Factory none Sent to Eosense on RMA #210105 to check on its high temperature start up problem and some noise in the CO2 readings.
ML2x 139/052 2021-03-15 Lab none Tested Lab none Tested in the lab. In air it read about 19mv in tap water 1136mV. Seems okay.
KD140SX-UFBS 11ZV0M0836 2021-03-11 Hill Slough Solar Panel C Removed Lab none This panel was removed from the lowest position at Hill Slough because its glass face is broken. It was replaced by sn 11ZV0M0828
RPI0W RPI0W-004 2021-03-11 Lab none Installed Hill Slough Camera Installed at Hill Slough with site setup. Mounted onto the scaffolding and pointing ~WNW.
CS450 20010770 2021-03-11 Lab none Installed Hill Slough Water Level Installed at Hill Slough with site setup. In a well pipe about 1 m from the SW corner of the scaffolding.
CS655 17130 2021-03-11 Lab none Installed Hill Slough none Installed at Hill Slough with site setup. Not sure of sensor depth yet (either -8cm or -16cm).
CS655 17131 2021-03-11 Lab none Installed Hill Slough none Installed at Hill Slough with site setup. Not sure of sensor depth yet (either -8cm or -16cm).
HFP01 10809 2021-03-11 Lab none Installed Hill Slough Heat Flux Plate B Installed at Hill Slough with site setup.
HFP01SC 00245 2021-03-11 Lab none Installed Hill Slough Heat Flux Plate A Installed at Hill Slough with site setup.
HMP155 S3410851 2021-03-11 Lab none Installed Hill Slough RH & Temp Installed at Hill Slough with site setup.
SRS-Ni NDVI 2046503402 2021-03-11 Lab none Installed Hill Slough Narrow Band NDVI Reflected Installed at Hill Slough with site setup.
SRS-Ni NDVI 2046503368 2021-03-11 Lab none Installed Hill Slough Narrow Band NDVI Incoming Installed at Hill Slough with site setup.
PAR-LITE 050596 2021-03-11 Lab none Installed Hill Slough PAR Sensor Reflected Installed at Hill Slough with site setup in reflected position
PAR-LITE 050597 2021-03-11 Lab none Installed Hill Slough PAR Sensor Incoming Installed at Hill Slough with site setup in incoming position
NR-LITE 990350 2021-03-11 Lab none Installed Hill Slough Net Radiometer Installed at Hill Slough with site setup. Temporary Rnet measurement until we have finished testing the 4-way radiometer.
AM25T 2856 2021-03-11 Lab none Installed Hill Slough Multiplexer Thermocouple Installed at Hill Slough with site setup.
CR1000 50889 2021-03-11 Lab none Installed Hill Slough Data Logger Met Installed at Hill Slough with site setup.
LI-7550 AIU-0650 2021-03-11 Lab none Installed Hill Slough Interface LI-7550 Installed at Hill Slough with site setup. Renamed from SWTT-0650 to HS-0650.
WM 1590 120907 2021-03-11 Lab none Installed Hill Slough Anemometer Installed at Hill Slough with site setup.
LI-7500A 75H-2180 2021-03-11 Lab none Installed Hill Slough Gas Analyzer CO2/H2O Installed at Hill Slough with site setup. Initially installed with an older calibration file, but this will be fixed at the next site visit.
PS-30M 11400080 2021-03-11 Lab none Installed Hill Slough Solar Controller Installed at Hill Slough with site setup.
KD140SX-UFBS 11ZV0M0828 2021-03-11 Lab none Installed Hill Slough Solar Panel C Installed at Hill Slough with site setup. This panel replaces panel sn 11ZV0M0836 which was removed because of damage when it fell over.
LI-7700 TG1-0211 2021-03-11 Bouldin Alfalfa Gas Analyzer CH4 Moved Hill Slough Gas Analyzer CH4 Installed at Hill Slough with site setup. This sensor was removed from BA because our spare 7700 is currently at the factory.
BB-Fab BB-SW 2021-03-11 Sherman Wetland Battery Bank Maintenance Sherman Wetland Battery Bank I removed the six group 31 SLA batteries from the metal battery box and reconnected them in three pairs on the boardwalk. The battery box will be used at Hill Slough.
WM 1590 120907 2021-03-10 Lab none Adjusted Lab none Changed baud rate from 19,200 to the max 57,600.
PTB110 H0160093 2021-03-10 Lab none Damaged Lab none Accidentally fried this sensor when I plugged the power strip in reverse polarity. When Joe tried later to plug the sensor in, it browned out the whole sensor.
LI-7500A 75H-2180 2021-03-08 Lab Calibration Lab Last calibration: 2020-11-24
AGC=99.7 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=25.7C, Tdew=3.7C
Zero Air at 400psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.9138 -> .9138 Offset: .18ppm
CO2 span 1.0295 -> 1.0231 Offset: 3.95ppm
H2O zero 0.8911 -> .8923 Offset: 7.9ppm
H2O span 1.0245 -> 1.0243 Offset: 0ppm
Notes: Larger H2O zero offset than I expected
PS-30M 11400080 2021-02-26 Sherman Wetland-Temp1 Solar Controller Removed Lab none Removed from the decommissioned SW temporary location and brought back to the lab to be cleaned up.
LI-7500A 75H-2669 2021-03-04 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at Bouldin Alfalfa to replace sn 75H-2182 for calibration.
LI-7500A 75H-2182 2021-03-04 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa and brought back to the lab for calibration. Replaced by sn 75H-2669.
LI-7500A 75H-2514 2021-03-04 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 1594 for calibration.
LI-7500A 1594 2021-03-04 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn and brought back to the lab for calibration. Replaced by sn 75H-2514.
S2-411 1054 2021-03-04 East End Narrow Band NDVI Reflected B Removed Lab none Removed from East End and brought back to the lab for testing. It had some NANs, but might've been influenced by the misbehaving NDVI_in sensor that was wired to the same control port.
S2-411 1053 2021-03-04 East End Narrow Band NDVI Incoming B Removed Lab none Removed from East End because it had a lot of NANs. Brought back to the lab for testing.
LI-7500A 0418 2021-03-04 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 0073 for calibration.
LI-7500A 0073 2021-03-04 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry and brought back to the lab for calibration. Replaced by sn 0418.
LI-7500A 0035 2021-03-04 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at Sherman Wetland to replace sn 0065 for calibration.
LI-7500A 0065 2021-03-04 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Wetland and brought back to the lab for calibration. Replaced by sn 0035.
PAR-LITE 010159 2004-04-30 Lab none Installed Tonzi PAR Sensor Incoming Cal Installed on the radiation boom in the incoming position for inter-comparison with the existing Licor Quantum sensor.
LI-7500A 0418 2021-03-03 Lab Calibration Lab Last calibration: 2020-11-24
AGC=97.7 SDM= DAC1= DAC2=
Ambient P=99.3kPa, Tair=25.7C, Tdew=8.2C
Zero Air at 450psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.8995 -> .9003 Offset: 3.01ppm
CO2 span 1.0034 -> .9978 Offset: 3.49ppm
H2O zero 0.8465 -> .8461 Offset: -2.0ppm
H2O span 0.9847 -> .9906 Offset: -.14ppm
Notes: I had to add more water the dewpt generator to get it to stay at the set dewpoint, otherwise it would just slowly decline forever. Calibrated earlier in the day, but I had forgotten to update the pressure coefficients in the config file for LI-7550 sn AIU-614. With the incorrect coefficients, offsets were (CO2 zero = 0.78, CO2 span = 0.23, H2O zero = -4.6, H2O span = 0.04).
LI-7500A 0035 2021-03-03 Lab Calibration Lab Last calibration: 2020-11-23
AGC=101.3 SDM= DAC1= DAC2=
Ambient P=99.4kPa, Tair=25.1C, Tdew=7.6C
Zero Air at 450psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.9375 -> .9376 Offset: .7ppm
CO2 span 1.019 -> 1.0151 Offset: 2.44ppm
H2O zero .8983 -> .8983 Offset: 1.9ppm
H2O span 1.035 -> 1.0319 Offset: .07ppm
Notes:
miniDOT-usb 931528 2021-02-26 Mayberry Removed Lab none Removed from MB wetland, where it was buried in the mud, and brought back to the lab to download its data
miniDOT-usb 931528 2020-05-29 Lab none Installed Mayberry none Buried in the mud at 30 cm at the 0.5 footprint site at Mayberry
LI-7500A 75H-2667 2021-03-02 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0418 for calibration.
LI-7500A 0418 2021-03-02 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2667.
WM 1590 153805 2021-03-02 Cal Kit Anemometer Removed Lab none Removed from TZ Tower top with the calibration kit
CR1000 4190 2021-03-02 Cal Kit Data Logger Removed Lab none Removed from TZ Tower top with the calibration kit
PAR-LITE 060831 2021-03-02 Cal Kit PAR Sensor Reflected Removed Lab none Removed from TZ Tower top with the calibration kit
HMP155 S3410851 2021-03-02 Cal Kit RH & Temp Removed Lab none Removed from TS Tower top with the calibration kit
LI-7500A 75H-2176 2021-03-02 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0035 for calibration.
LI-7500A 0035 2021-03-02 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 75H-2176.
LI-7500A 75H-2150 2021-03-02 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2021-03-02 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 75H-2150.
LI-7500A 75H-2669 2021-03-01 Lab Calibration Lab Last calibration: 2020-08-27
AGC=101.8 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=29.0C, Tdew=1.3C
Zero Air at 500psi, AMP CA04396 408.97 at 300 Tdewspan=20.0
CO2 zero 0.9184 -> 0.9190 Offset: 2.3ppm
CO2 span 1.0118 -> 1.0059 Offset: 3.03ppm
H2O zero 0.9012 -> 0.8984 Offset: -17.2ppm
H2O span 1.0121 -> 1.0033 Offset: 0.19ppm
Notes: Flow rated larger than usual - 2lpm instead of 2cfpm. Dew point generator problems before cal - dew point stuck at 34 instead of 20C. The dewpt gen was fixed an ran at 20C before starting H2O span but dewpt initially jumped to 25C but steadily declined to 20C. I double checked CO2 zero/span and H2O zero afterwards and they seemed fine.
LI-7500A 75H-2667 2021-03-01 Lab Calibration Lab Last calibration: 2020-08-25
AGC=102.6 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=27.6C, Tdew=2.4C
Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0
CO2 zero 0.942 -> 0.9423 Offset: 1.2ppm
CO2 span 1.0172 -> 1.0092 Offset: 4.13ppm
H2O zero 0.8403 -> 0.8385 Offset: -10.8ppm
H2O span 1.0244 -> 1.0147 Offset: 0.2ppm
Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm
LI-7500A 75H-2514 2021-03-01 Lab Calibration Lab Last calibration: 2020-12-11
AGC=101.1 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=27.2C, Tdew=0.92C
Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0
CO2 zero 0.9049 -> 0.9048 Offset: -0.73ppm
CO2 span 1.0082 -> 0.9984 Offset: 5.03ppm
H2O zero 0.8398 -> 0.8389 Offset: -6.7ppm
H2O span 1.0093 -> 1.0061 Offset: 0.07ppm
Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm
LI-7500A 75H-2176 2021-03-01 Lab Calibration Lab Last calibration: 2020-12-02
AGC=102.5 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=22.9C, Tdew=4.6C
Zero Air at 450psi, AMP CA04396 408.97 at 250 Tdewspan=20.0
CO2 zero 0.8673 -> 0.8671 Offset: -0.25ppm
CO2 span 1.009 -> 1.000 Offset: 2.53ppm
H2O zero 0.8531 -> 0.8524 Offset: -1.3ppm
H2O span 0.9238 -> 0.9092 Offset: 0.37ppm
Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm.
LI-7500A 75H-2150 2021-03-01 Lab Calibration Lab Last calibration: 2020-08-20
AGC=100.8 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=27.4C, Tdew=2.08C
Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0
CO2 zero 0.9019 -> 0.9019 Offset: -0.3ppm
CO2 span 1.0014 -> 0.9975 Offset: 1.93ppm
H2O zero 0.8665 -> 0.8675 Offset: 6.8ppm
H2O span 1.011 -> 1.0040 Offset: 0.15ppm
Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm.
NR01 2420 2021-02-26 Lab none Installed Sherman Wetland Net Radiometer 4-way Cal A Installed on the Sherman Wetland tower for inter-comparison after repair in the lab. It is clamped near the end of the square tube of the radiation boom pointing south.
CNR4 121038 2021-02-26 Factory none Installed Sherman Wetland Net Radiometer 4-way Cal B Installed on the Sherman Wetland tower for inter-comparison. This is the AMP standard. It is clamped near the end of the square tube of the radiation boom pointing south.
CR1000 23865 2021-02-26 Lab none Installed Sherman Wetland Data Logger Cal Installed on the Sherman Wetland tower to collect data from two 4-way net radiometers for inter-comparison
BB-Fab BB-SW_Temp1 2021-02-26 Sherman Wetland-Temp1 Battery Bank Disposed Disposed none Removed from the decommissioned temporary location. The batteries were brought back to the lab and will be used at Hill Slough but this battery bank has ceased to exist as an equipment entry
KD140SX-UFBS 11ZV0M0829 2021-02-26 Sherman Wetland-Temp1 Solar Panel A Moved Hill Slough Solar Panel A Removed from the decommissioned temporary location for use at Hill Slough
KD140SX-UFBS 11ZV0M0836 2021-02-26 Sherman Wetland-Temp1 Solar Panel B Moved Hill Slough Solar Panel B Removed from the decommissioned temporary location for use at Hill Slough
KD140SX-UFBS 11ZV0M0855 2021-02-26 Sherman Wetland-Temp1 Solar Panel C Moved Hill Slough Solar Panel C Removed from the decommissioned temporary location for use at Hill Slough
PAR-LITE 050597 2021-02-26 Sherman Wetland-Temp1 PAR Sensor Incoming Removed Lab none Removed from the decommissioned temporary location and brought back to the lab to be cleaned up.
LI-610 DPG478B 2021-03-01 Lab none Maintenance Lab none The dew point was stuck at about 34C - not dropping to the set point of 20C. I found a crack in the "from cooler" tubing. I replaced this piece of tubing but this didn't solve the problem. There was an air bubble in this piece of tubing and the machine was making an unusual bubbling sound. Refilling the reservoir with DI water immediately change the sound, got rid of the air bubble in the tubing and the dew point began to drop towards the set point.
Mark II C003800 2021-02-25 Lab none Disposed Disposed none Returned to Arable
Mark II C003789 2021-02-25 Lab none Disposed Disposed none Returned to Arable
Mark II C002469 2020-04-23 Lab none Disposed Disposed none Returned to Arable
Mark II C002445 2020-04-23 Lab none Disposed Disposed none Returned to Arable
Mark II C002714 2020-04-23 Lab none Disposed Disposed none Returned to Arable
Mark II C006322 2021-02-04 Lab none Installed Bouldin Alfalfa Weather Station PL_N-B Installed on the northern Planet Labs post to replace one of the two units eaten by the sheep
Mark II C006282 2021-02-04 Lab none Installed Bouldin Alfalfa Weather Station PL_N-A Installed on the northern Planet Labs post to replace one of the two units eaten by the sheep
Mark I A001311 2021-02-25 Lab none Disposed Disposed none Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II
Mark I A001359 2021-02-25 Lab none Disposed Disposed none Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II
Mark I A001353 2021-02-25 Lab none Disposed Disposed none Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II
CR1000 23865 2021-02-25 Lab none Maintenance Lab none The internal lithium battery was nearly dead 1.6V. I couldn't find a new one but swapped it for one reading 2.6V instead. Be sure the check the clock when using this logger.
CNR4 121038 2019-11-17 Factory none Calibration Factory none Initial calibration by Kipp and Zonen. LW on 2019-11-17: LW_in = 6.83 uV/(W/m2), LW_out = 5.48 uV/(W/m2). SW on 2019-11-15: SW_in = 14.56 uV/(W/m2), SW_out = 14.17 uV/(W/m2)
NR01 2420 2021-02-18 Lab none Repair Lab none Over the last week or so I disassembled and cleaned this sensor. I don't see any water damage but I can't get the LW junction board loose as it seems glued to the case. I totally replaced the four wires running between the LW and SW sides. These wires were broken after having been previously broken and spliced. I soldered new wires directly to the junction boards. I collected about two days of data on the window sill with the sensor partially disassembled so that all four sensors could point up together. The relationship between the paired sensors was very tight (r-squared >0.99) - excluding a portion of the LW data that might have been affected by solar heating. LW had a sloped of 1.01 but SW had a slope of 1.20 - new calibration coefficients are probably needed. Today I completely reassembled the sensor and put it out the window on a boom to collect some data to check the assembly.
PAR-LITE 990017 1999-02-09 Factory none Calibration Factory none Initial factory calibration: 5.16uV/(umol/m2/s)
PAR-LITE 990019 1999-02-09 Factory none Calibration Factory none Initial factory calibration; 5.26uV/(umol/m2/s)