Ser No.DateLoc FromTitleLoc ToNotes
425A 0839 2011-12-23 Lab none Tested Lab none Tested with other Handar sonics by placing each sonic at 17cm, 48cm, 78cm and 108cm from a blower. All sonics gave similar resonable readings.
425A 0843 2011-12-23 Lab none Tested Lab none Tested with other Handar sonics by placing each sonic at 17cm, 48cm, 78cm and 108cm from a blower. All sonics gave similar resonable readings.
425A 0854 2011-12-23 Lab none Tested Lab none Found in its original box on the shelf in H105. Tested with other Handar sonics by placing each sonic at 17cm, 48cm, 78cm and 108cm from a blower. All sonics gave similar resonable readings.
LI-7500A 0073 2011-10-17 Lab none Calibration Lab none Last calibration: 2011-10-17
AGC=56 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100kPa, Tair=21.8C, Tdew=5.6C
dry air at 600psi, CO2 436.53 at 1900 Tdewspan=13.6
CO2 zero 0.9584 -> 0.9578 Offset: -2.00ppm
CO2 span 1.0022 -> 1.0000 Offset: 1.09ppm
H2O zero 0.7390 -> 0.7381 Offset: -0.12ppm
H2O span 0.9873 -> 0.9940 Offset: 0.09ppm
Notes: operated by Siyan Ma
425A 0839 2011-01-03 Lab none Loaned Loaned Loaned to Marcia DeLonge and Andrew McDowell from Wendee Silver's lab for a couple of months.
425A 0843 2011-01-03 Lab none Loaned Loaned Loaned to Marcia DeLonge and Andrew McDowell from Wendee Silver's lab for a couple of months.
LI-7500A 75H-2150 2011-11-18 Factory Calibration Lab none Initial Factory Calibration: CO2 zero = 0.9022, CO2 span = 1.0056 (at 622ppm), CO2 span2 = 0, H2O zero = 0.8631, H2O span = 1.0054 (at 12 C), H2O span2 = 0.
425A 0852 2011-05-02 Loaned Returned Lab none Returned by Sara Winsemius.
425A 0852 2011-04-10 Lab none Loaned Loaned Loaned to Sara Winsemius for a class project.
425A 0860 2011-04-10 Loaned Returned Lab none Returned by Sara Winsemius. Sara reports that it was banged twice and no longer reports resonable wind speed.
425A 0860 2011-04-08 Lab none Loaned Loaned Loaned with a CR10X to Sara Winsemius for a class project.
425A 0852 2008-12-11 Tonzi-Tower Removed Lab none Removed with the Tonzi Tower wind profile.
425A 0860 2008-12-11 Tonzi-Tower Removed Lab none Removed with the Tonzi Tower wind profile.
425A 0846 2008-12-11 Tonzi-Tower Removed Lab none Removed with the Tonzi Tower wind profile.
425A 0839 2008-12-11 Tonzi-Tower Removed Lab none Removed with the Tonzi Tower wind profile.
425A 0843 2008-12-11 Tonzi-Tower Removed Lab none Removed with the Tonzi Tower wind profile.
425A 0852 2003-08-01 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower as part of a wind profiles with sonics at 10.61m, 11.76m, 13.21m, 15.04m and 16.86m.
425A 0860 2003-08-01 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower as part of a wind profiles with sonics at 10.61m, 11.76m, 13.21m, 15.04m and 16.86m.
425A 0846 2003-08-01 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower as part of a wind profiles with sonics at 10.61m, 11.76m, 13.21m, 15.04m and 16.86m.
425A 0839 2003-08-01 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower as part of a wind profiles with sonics at 10.61m, 11.76m, 13.21m, 15.04m and 16.86m.
425A 0843 2003-08-01 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower as part of a wind profiles with sonics at 10.61m, 11.76m, 13.21m, 15.04m and 16.86m.
LI-7500A 0418 2011-12-18 Lab none Installed Tonzi-Tower Installed on Tonzi Tower so that sn 0065 could be swapped out for calibration.
LI-7500A 0065 2011-12-18 Tonzi-Tower Removed Lab none Removed from Tonzi Tower for calibration. Replaced by sn 0418.
LI-7500 0976 2011-12-15 Lab none Calibration Lab none Last calibration: 2011-08-24
AGC=58 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.04kPa, Tair=22.44C, Tdew=5.78C
zero air at 700psi, CO2 236.53 at 1900 Tdewspan=13.5
CO2 zero 0.8596 -> 0.8586 Offset: -2.21ppm
CO2 span 1.01 -> 1.0106 Offset: -3ppm
H2O zero 0.6276 -> 0.6298 Offset: -1.82ppm
H2O span 0.9824 -> 0.9855 Offset: 0.2ppm
Notes: operated by Siyan Ma
LI-7500A 0418 2011-12-15 Lab none Calibration Lab none Last calibration: unknow
AGC=48 SDM=- DAC1=0-1000 DAC2=10-25
Ambient P=101.4kPa, Tair=22.6C, Tdew=6.04C
zero air at 700psi, CO2 436.53 at 1900 Tdewspan=13.5
CO2 zero 0.9015 -> 0.9017 Offset: 0.88ppm
CO2 span 1.0037 -> 1.0066 Offset: 1.67ppm
H2O zero 0.8436 -> 0.8493 Offset: 0.09ppm
H2O span 0.9604 -> 0.9719 Offset: 0.1ppm
Notes: operated by Siyan Ma
LI-7500A 0042 2011-12-13 Lab none Calibration Lab none Last calibration: 2011-09-27
AGC=73 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.67kPa, Tair=23.42C, Tdew=4.10C
Zero Air at 800psi, CO2 in Air 436.53 at 1900 Tdewspan=10.08
CO2 zero 0.9592 -> 0.9597 Offset: 2.0ppm
CO2 span 0.9758 -> 0.9840 Offset: 5.23ppm
H2O zero 0.7852 -> 0.7828 Offset: -6.0ppm
H2O span 0.9532 -> 0.9341 Offset: 0.32ppm
Notes: The signal from this IRGA seems noisey and the AGC is pretty high.
LI-6262 IRG3-1079 2011-12-12 Lab none Calibration Lab none Calibrated in differential mode for check of new secondary CO2 standards. Ambient temperature was 18.987C, pressure was 99.746kPa. Using zero air (800psi)and god bottle 370.91ppm CO2 (1650psi).
Flowing zero air in both sample and reference
CO2 = -0.20umol/mol adjusted to zero.
H2O = -0.0.004mmol/mol not adjusted
Flowing zero air in reference and god bottel in sample
CO2 = 370.25ppm adjusted to 370.91ppm.
Flowing zero air in reference and new secondary standard in sample, I wrote down about 60 readings and made an average. One new CO2 secondary standard has 421.63ppm CO2, the other was 436.52ppm CO2. Both tanks were label.
HMP45A U2120013 2011-11-21 Siyan Maintenance Lab none I lengthened the cable on this sensor. New wire colors are: ??
HMP45A U2120013 2011-11-22 Lab none Installed Mayberry RH & Temp Installed on the Mayberry tower to replace sn U3030041. This HMP45 has a lengthened cable to allow mounting with the recently raised eddy sensors.
HMP45AC U3030041 2011-11-22 Mayberry Tower Removed Lab none Removed from the Mayberry tower because the cable was not long enough for the new higher mounting of the eddy sensors. Replaced by HMP45 sn U2120013.
HFP01 000144 2011-12-01 Lab none Installed Tonzi Heat Flux Plate E Reinstalled in the field in its orginial location. The ground was dry and crumbly.
HFP01 000144 2011-11-30 Lab none Repair Lab none I removed the cable that had cracking and rotting inslation and replaced it with new shielded, two conductor, PVC jacketed cable. New red lead to old green and new black lead to old white.
HFP01 000144 2011-11-16 Tonzi Heat Flux Plate E Removed Lab none I removed this heat flux plate from the Tonzi understory location due to bad readings. I found that a gopher had cut the cable. In addition the the insulation on the cable was cracked and rotting.
LI-6262 IRG3-1079 2011-11-18 Lab none Calibration Lab none Calibrated in differential mode for check of Frank's new secondary CO2 standard. Ambient temperature was 21.5C, pressure was 100.45kPa. Using N2 and god bottle 402.69ppm CO2 (1850psi).
Flowing N2 in both sample and reference
CO2 = -0.30umol/mol adjusted to zero.
H2O = -0.15
Flowing N2 in reference and god bottel in sample
CO2 = 401.8ppm adjusted to 402.69ppm.
Flowing N2 in reference and new secondary standard in sample, I wrote down 75 readings and made an average. The new CO2 secondary standard has 398.22ppm CO2.
FMA RMT-200 08-0114 2011-11-18 Lab none Calibration Lab none Calibrated in the lab with god bottle 1.8978ppm CH4 at 180psi. Last calibration Aug 12, 2011. Before running the cal gas the LGR read:
10:33PST, 21.58C, 142.54Torr, 15.466usec, 1Hz, 1.866ppm
I missed writing down the ppm with the cal gas before calibration but it was about 1.80ppm. Afterwards it was right on.
FMA RMT-200 08-0114 2011-11-16 Lab none Maintenance Lab none In the field I pulled the internal filter in the LGR to use as the external 2um filter on the system with the new LGR that I swapped in. Runing in the lab with a loose, dirty filter internally dirtied the mirrors, but the ring time was already low to begin with. I swapped the internal 2um filter with a cleaned one. I checked a fitting in front of the pressure controller looking for another filter but found nothing. I pulled the pressure controller and discovered if you disassemble it you can clean some internal filters. This seems to have solved the flow problem. Then I cleaned the mirrors. The first cleaning seemed to work but when I used a high flow rate than began to get dirty. I may have to clean them again.
FMA RMT-200 08-0114 2011-10-27 Twitchell Rice Removed Lab none I removed the LGR from Twitchell due to low pressure and almost no flow in low flow mode. I swapped in LGR sn 05-0010.
AM416 13961 2011-11-07 Lab none Loaned Loaned Loaned to Gavin in the Silver lab.
32707-30 CP6295 2011-10-20 Factory Calibration Factory Initial calibration. Flow rates at STP, max allowable error +-1.5%, Cal for Air, at 23C, 740mmHg, 12.8VDC

ml/min = 202.56 * VDC - 8.8158, R2 = 0.9998

ml/min = 1.0115 * screen - 7.7203, R2 = 0.9998

Cal sheet on file.
CR1000 38211 2011-11-04 Lab none Upgrade Lab none I updated the OS from version 21 to version 22.
LI-7500 1653 2011-11-03 Lab none Calibration Lab none Last calibration: 2011-03-16
AGC=52 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.12kPa, Tair=18.45C, Tdew=7.99C
Zero Air at 700psi, CO2 in Air 410.49 at 1900 Tdewspan=13.0
CO2 zero 0.9237 -> 0.9232 Offset: -2.05ppm
CO2 span 1.0118 -> 1.0103 Offset: 0.86ppm
H2O zero 0.8762 -> 0.8756 Offset: -3.7ppm
H2O span 0.9831 -> 0.9407 Offset: 0.98ppm
Notes: Frank's briefly borrowed for use on Vaira.
TE525MM 49187-109 2011-11-02 Lab none Installed Twitchell Rice Precipitaion Reinstalled at Twitchell after the cable was spliced in the lab to repair a cut suffered in the field.
TE525MM 49187-109 2011-10-27 Twitchell Rice Precipitaion Removed Lab none Removed from Twitchell because the cable was cut some time during the summer.
HFP01SC 00244 2011-11-02 Lab none Installed Twitchell Rice Heat Flux Plate A Installed in the field after harvest and chopping. The cable was replaced due to cracking and the self calibration heater cable was shortened to a pig tail.
HFP01SC 00244 2011-10-27 Twitchell Rice Heat Flux Plate A Removed Lab none Removed from the field to repair cracking cables in the lab.
HFP01SC 00246 2011-11-02 Lab none Installed Twitchell Rice Heat Flux Plate C Installed in the field after harvest and chopping. The cable was replaced due to cracking and the self calibration heater cable was shortened to a pig tail.
HFP01SC 00246 2011-10-27 Twitchell Rice Heat Flux Plate C Removed Lab none Removed from the field for repair of cracking cables in the lab.
LI-7500A 0073 2011-10-18 Lab none Installed Tonzi-Floor Installed on the floor tower to replace sn 0418 for calibration.
LI-7500A 0418 2011-10-18 Tonzi-Floor Removed Lab none Removed from the Floor Tower for calibration. Replaced by sn 0073.
LI-7500A 1594 2011-10-18 Lab none Installed Vaira Gas Analyzer CO2/H2O This IRGA was installed on the Vaira tower to replace Frank's IRGA that was installed temporarily.
LI-7500A 1594 2011-10-17 Lab none Calibration Lab none Last calibration: 2011-07-05
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.62kPa, Tair=29.05C, Tdew=14.17C
dry air at 700psi, CO2 410.49 at 2000 Tdewspan=18.3
CO2 zero 0.8654 -> 0.8644 Offset: -4.21ppm
CO2 span 1.0098 -> 1.0082 Offset: 0.82ppm
H2O zero 0.8901 -> 0.8898 Offset: -0.04ppm
H2O span 0.9558 -> 0.9721 Offset: 0.3ppm
Notes: removed from Tonzi Tower; calibrated by Siyan
LI-7500A 0418 2011-10-17 Lab none Calibration Lab none Last calibration: 2011-07-05
AGC=61 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.48kPa, Tair=30.7C, Tdew=16.56C
dry air at 700psi, CO2 2000 at 2000 Tdewspan=18.3
CO2 zero 0.9588 -> 0.9584 Offset: -1.13ppm
CO2 span 1 -> 1.0022 Offset: 1.08ppm
H2O zero 0.74 -> 0.7390 Offset: -0.07ppm
H2O span 1.0033 -> 0.9873 Offset: 0.4ppm
Notes: calibrated by Siyan
LI-7500A 0035 2011-09-14 Lab none Calibration Lab none Last calibration: 2011-01-04
AGC=58 SDM=7 DAC1=1-1000 DAC2=10-25
Ambient P=100.31kPa Tair=19.4C, Tdew=14.2C
CO2 free air at 60psi, CO2 in air 405.85ppm 1700psi Tdewspan=14
CO2 zero 0.9334 -> 0.9338 Offset: 0.06ppm
CO2 span 1.0093 -> 1.0108 Offset: 0.9ppm
H2O zero 0.7366 -> 0.7353 Offset: -9.2mmol/m3
H2O span 0.9749 -> 0.9843 Offset: 0.14C
Notes:
LI-7500A 0035 2011-05-24 Tonzi-Floor Calibration Lab none Last calibration: 2011-01-04
AGC=58 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.77kPa Tair=23.5C, Tdew=9.06C
zero air at 1450psi, CO2 in air 597.4ppm 0psi Tdewspan=14.1
CO2 zero 0.9376 -> 0.9334 Offset: -29.4ppm
CO2 span 1.038 -> 1.0093 Offset: 24.2ppm
H2O zero 0.725 -> 0.7366 Offset: 77.1mmol/m3
H2O span 0 -> 0.9749 Offset: -3.32C
Notes:
LI-7500A 0042 2011-06-15 Lab none Calibration Lab none Last calibration: 2011-02-02
AGC=68 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.3kPa Tair=21.98C, Tdew=11.9C
zero air at 1450psi, co2 in air 597.4ppm 100psi Tdewspan=16.9
CO2 zero 0.9605 -> 0.9587 Offset: -8.14ppm
CO2 span 0.9855 -> 0.9796 Offset: 11.6ppm
H2O zero 0.7877 -> 0.7886 Offset: -2.9mmol/m3
H2O span 1.1467 -> 0.9667 Offset: 4.62C
Notes: I initiall tried to calibrate on 6/14 but was rushed and got some odd behavior so did a better calibration on 6/15. The offsets are estimates based on the two calibrations.
LI-7500A 0042 2011-09-27 Lab none Calibration Lab none Last calibration: 2011-06-15
AGC=75 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.8kPa Tair=25.14C, Tdew=11.46C
zero air at 1000psi, CO2 in air 405.85ppm 1500psi Tdewspan=13.4
CO2 zero 0.9587 -> 0.9592 Offset: 1.72ppm
CO2 span 0.9796 -> 0.9758 Offset: 2.1ppm
H2O zero 0.7886 -> 0.7858 Offset: -20mmol/m3
H2O span 0.9667 -> 0.9532 Offset: 0.37C
Notes: The AGC values seemed high at between 70 and 75%. The internal chemicals were changed and allowed to run overnight.
LI-800 AKC-0479 2011-10-17 Lab none Maintenance Lab none I cleaned the bench. There was one black spec inside the bench and another on the source end. After cleaning, I was able to calibrate the IRGA.

Press = 1003mBar, Temp = 52.6C, Heater = ON, Filter = 0sec, span = 1000ppm.

Several cycles of zero air and span CO2 410.49 seemed to bring it back into range.
LI-800 AKC-0479 2011-10-13 Tonzi Removed Lab none Removed from the CO2 profiles system because of strange values and would not calibrate.
LI-7500A 0042 2011-10-13 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower due to no signal and reporting 100% AGC even with clean lenses.
LI-6262 IRG3-1079 2011-10-10 Lab none Calibration Lab none Calibrated in differential mode for check of new secondary CO2 standard. Ambient temperature was 24.554C, pressure was 100.43kPa. Using zero air (800psi) and god bottle 370.91ppm CO2 (1700psi).
Flowing zero air in both sample and reference
CO2 = 0.55umol/mol adjusted to zero.
H2O = 0.0003
Flowing zero air in reference and god bottel in sample
CO2 = 371.25ppm adjusted to 370.91ppm.
Flowing zero air in reference and new secondary standard in sample, I wrote down 90 reading and made an average. The new CO2 secondary standard has 410.49ppm CO2.
FMA RMT-200 05-0010 2011-10-10 Lab none Calibration Lab none I calibrated the LRG with the god bottle and then used it to check two secondary standards. God bottel is 1.8978ppm CH4 and 350psi. Last calibration was 2011-08-23 using 3.276ppm CH4.
Precal: 9:32PST, 23.79C, 156.56Torr, 14.715usec, 1Hz, 1.855ppm
Postcal the CH4 read 1.897ppm.

The older secondary standard with 405.85ppm CO2 read 4.675ppm CH4
The newer secondary standard with 410.49ppm CO2 read 3.228ppm CH4
FMA RMT-200 05-0010 2011-10-10 Lab none Maintenance Lab none No flow in low flow mode. The internal 2um after the pressure controller seemed okay. I disassembled the inlet tubing and fittings to get a 10um filter out from infront of the pressure controller. This was completely plugged. I rinsed it with degreaser and DI water and dried it by blowing air through it. Flow was better but still seemed low. I reinstalled the filter and reassembled the inlet tubing. Now is slow flow the internal pressure is 156Torr.
LAI-2000 PCA-0846 2011-10-11 Factory Repair Lab none Returned from repair at Licor on RMA 166323. They replaced the cable on the wand, replaced the casing and battery holder on the console, replaced the battery connector on the analog board, rinsed and dried the boards, assembled, and checked system operation.
FGGA 09-0011 2011-10-06 Lab none Calibration Lab none This analyzer was calibrated withe the CH4 god bottle then used to measure the CH4 in a 400ppm CO2 secondary tank. In the process the LGR CO2 was calibrated with the scondary CO2 tank.

Last calibration of CH4 and CO2 was on 2011-08-29. The last calibration of H2O was on 2009-03-13.

God bottel: 1.8978ppm CH4 350psi
Secondary CO2: 405.85ppm CO2 -> CH4 determined to be 4.666ppm

CH4 Pre-cal:
8:23PST, 21.67C, 138.20Torr, 13.403usec, 1Hz, 1.9140ppm CH4
After CH4 was 1.8980ppm.

CO2 pre-cal:
8:29PST, 21.65C, 138.18Torr, 13.395usec, 1Hz, 4.666ppm CH4, 413.8ppm CO2
After CO2 was 405.5ppm
LAI-2050 PCA-1546 2011-09-27 Factory Repair Lab none Returned from Licor.
LAI-2050 PCA-1546 2011-09-08 Lab none Repair Factory Sent to Licor on RMA 166208 for a check along with console sn PCA-1478.
LAI-2000 PCA-1478 2011-09-27 Factory Repair Lab none Returned from repair at Licor on RMA 166208 for damage caused by battery acid.
LAI-2000 PCA-1478 2011-09-08 Lab none Repair Factory Sent to Licor on RMA 166208 for repair of damaged caued by battery acid. The case was replaced and the main boards were rinsed and tested. The head (LAI-2050) sn PCH1546 was calibrated.
LI-7500A 1594 2011-09-29 Tonzi-Tower Removed Lab none Removed from Tonzi Tower for calibration - replaced by sn 0065.
LI-7500A 0065 2011-09-29 Lab none Installed Tonzi-Tower Installed on Tonzi tower to replace sn 1594 for calibration.
LI-7500A 0073 2011-09-29 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration - replaced by sn 0042
LI-7500A 0042 2011-09-29 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on Vaira tower to replave sn 0073 for calibration.
LI-7500A 0065 2011-09-15 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry tower for calibration, replaced by sn 0035.
LI-7500A 0035 2011-09-15 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry tower to replace 0065 for calibration.
LI-7500A 0035 2011-08-25 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell tower for calibration, replaced by sn 0041.
LI-7500A 0041 2011-08-25 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell tower to replace sn 0035 for calibration.
LI-7500A 0042 2011-08-25 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman Tower for calibration. Replaced with sn 0976.
LI-7500 0976 2011-08-25 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on Sherman Tower to replace sn 0042 for calibration.
CNR1 990181 2011-09-27 Lab none Tested Lab none Data was recorded on two consecutive days with similar radition (September 27 and 28, 2011) from the window on the south side of the lab. At 11:20am on day 271 the CNR was rotated 180 degrees so that the upward and downward sensors switched places. All channels seem to have similar responses. I conclude this sensor is in working order.
WMP 1352 000022 2011-08-29 Factory none Returned Lab none Returned from Gill factory by way of PP Systems. Gill fully tested the unit for problems in communication and tolerance of low input voltage and found no problems. They fixed a a seal where the shaft goes into the main body.
LAI-2000 PCA-0846 2011-09-20 Lab none Repair Factory Sent to Licor to repair damage from leaked battery acid.
FMA RMT-200 05-0010 2011-09-02 Loaned Returned Lab none This analyzer was returned by Frank after completion of their chamber campaign.
FGGA 09-0011 2011-09-02 Borrowed Borrowed Lab none We re-borrowed this analyzer from Frank of continued soil respiration in the lab.
FGGA 09-0011 2011-08-22 Lab none Returned Returned Returned to Frank for chamber campaign on Twitwell ponds during the week of August 22nd, 2011
FMA RMT-200 05-0010 2011-08-22 Lab none Loaned Loaned Loaned to Frank for chamber campaign in Twitchell ponds during the week of August 22nd, 2011.
CS511 DO-2 2011-01-28 Lab none Installed Twitchell Rice DO Sensor First installation of this sensor in the field.
CS511 DO-2 2011-06-02 Twitchell Rice DO Sensor Maintenance Twitchell Rice DO Sensor Jaclyn replaced the electrolyte and cleaned the sensor before putting it back in the field now that the field is flooded.
ML2x 139/052 2011-07-07 Twitchell Rice Soil Moisture Probe B Removed Lab none Removed from the field after it was discovered that otters had chewed its cable.
CS450 20010026 2011-07-07 Twitchell Rice Water Level Removed Lab none This senor was removed from the Twitchell tower after it was discovered that otters had chewed the cable.
CS450 20010026 2011-07-14 Twitchell Rice Water Level Installed Lab none Reinstalled in the field after repairs to the cable where otters had chewed on it. I spliced the wires and the tube that provides the atmospheric pressure reference.
CS511 DO-2 2011-07-28 Twitchell Rice DO Sensor Removed Lab none Removed from the field after the cable was cut during the clearing of weeds.
CS511 DO-2 2011-08-04 Lab none Installed Twitchell Rice DO Sensor Reinstalled after splicing the cable to repair cut from weed clearing.
CNR1 020476 2011-03-01 Sherman Island Net Radiometer 4-way Removed Lab none Removed from the tower due to bad PRT readings. sn 990181 installed in place.
CNR1 990181 2011-03-01 Lab none Installed Sherman Island Net Radiometer 4-way Installed on the tower to replace the original one (sn 020476) that had a bad PRT reading.
CNR1 020476 2011-08-12 Lab none Installed Sherman Island Net Radiometer 4-way This radiometer was reinstalled on the tower after it was determined that the problem with the PRT was due either to the resistor bridge or possible a programming bug that used the wrong excitation channel. It was installed with a new resistor bridge.
CNR1 990181 2011-08-12 Sherman Island Net Radiometer 4-way Removed Lab none Removed from the Sherman tower.
FMA RMT-200 05-0010 2011-08-16 Lab none Calibration Lab none See calibration log. I also set the time to local standard time and changed the serial settings to 115200. I left it in high flow and shutdown so it should boot correctly next time.
FMA RMT-200 05-0010 2011-08-12 Lab none Repair Lab none This analyzer was removed from the field because it would not boot. Connecting the hard drive to a computer running Knopix Linux I could see that the drive was full and deleted some of the data files. Then it would boot but the analyzer software would crash and kick out into the service shell. There I could see that a bunch of the *.ini files in the icos folder were zero length. LGR was able to email me a copy of these files and replacing them seemed to fix the problems.
LI-6262 IRG3-1079 2011-07-08 Lab none Calibration Lab none I calibrated the Li6262 in differential mode. I flowed zero air through a T into both the reference and sample cells at a flow just detectable on wetted lips. CO2 was low by about a 0.1 ppm CO2. I adjusted both CO2 zero and H2O zero. I then flowed 370.91ppm CO2 in air (god bottle) through the sample while leaving the zero air in the reference. The 6262 read 369.20+-0.05ppm. I adjusted the CO2 span to match the tank value. I tested the nearly empty 597.4ppm tank and it was true. I then test the new 400ppm CO2 in air tank and came up with a value of 405.85ppm and labeled the tank with this value.
LI-7500A 1594 2011-07-06 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower after laboratory calibration. Replaced sn 0041.