Equipment and Calibration History

View Equipment |

View history entry 1486 for Los Gatos Research FMA RMT-200 sn: 08-0114

Date: 2010-09-08
From: Twitchell Rice -
To: Twitchell Rice -
Title: Calibration
Notes: I adjusted the bypass valve. When Frank arrived we did a calibration. Last cal was done on Aug 6, 2010. The field tank is 3.276ppm CH4 Precal: 9:52PST, 25.54C, 111.2Torr, 16.112usec, 1Hz, 3.377ppm Low flow pressure seems low here too but I did not check the filters. Being 100ppb off seems a lot, last cal uses a standard closer to 1.8ppm though. Post cal CH4 read 3.276ppm. When we left the LGR read: 10:02PST, 25.49C, 144.34Torr, 16.080usec, 10Hz, 1.840ppm
Entered by: Sara Knox
Field notes