Equipment and Calibration History

View Equipment |

View history entry 1495 for Los Gatos Research FMA RMT-200 sn: 05-0010

Date: 2012-01-25
From: Twitchell Rice -
To: Twitchell Rice -
Title: Calibration
Notes: With the new external filter I calibrated the LGR. The cal tank has <200psi of 3.276ppm CH4. Last calibration was Dec 20, 2011. Pre Cal: 8:49PST, 29.61C, 156.07Torr, 13.871usec, 3.282ppm, 1Hz The first try at calibration failed somehow ending in a value (3.317ppm) farther from the span gas. A second calibration seemed to work fine and ended up right on. When I left the LGR read 9:56PST, 23.93C, 153.81Torr, 13.060usec, 10Hz, 2.311ppm
Entered by: Sara Knox
Field notes