Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 [51] 52 53 54 55 56 57 58 59 60 61 62 63 64 65
Ser No.DateLoc FromTitleLoc ToNotes
GP-001ex GP2013ex02 2014-10-30 Lab none Installed Twitchell Alfalfa Installed at -16cm at Alfalfa
LI-7550 AIU-0614 2014-10-21 Factory Received Lab none Received back from Licor where they replaced the analog board (connector board) and updated the front panel. They also updated the embedded software to v7.3.1
LI-7550 AIU-0614 2014-10-02 Lab none Shipped Factory Shipped to Licor for upgrade (sealing) of bottom connector circuit board.
CR10X 12213 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
CS547A 5688 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
KD140SX-UFBS 11ZV0M1127 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
LI-7550 AIU-0397 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
WM 1590 124215 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
PS-15M 12501584 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
LI-7700 TG1-0357 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
LI-7500A 0065 2014-10-09 Mayberry-Portable Removed Lab none Removed with the rest of the portable tower equipment. End of the portable tower campaign.
GP-001 GP20140025 2014-10-16 Lab none Installed Mayberry CO2 Probe C Installed at 108cm below the surface on the fence post in the open water channel.
GP-001 GP20140024 2014-10-16 Lab none Installed Mayberry CO2 Probe B Installed at 65cm below the surface on the fence post in the open water channel.
GP-001 GP20140010 2014-06-26 Lab none Installed Mayberry CO2 Probe A Installed about 5cm below the surface on the fence post in the open water channel.
CR10X 12213 2013-01-06 Tonzi Removed Lab none Removed from the non-functional east sap flow station at Tonzi.
CR10X 31308 2013-11-21 Factory Returned Lab none Returned from factory repair
CR10X 31308 2013-11-06 Factory Repair Factory Could not upload new OS. Lost its brains while running on the tram. Campbell: Replaced PLD U11E, which was causing intermittent communication errors/watchdogs. Replaced water damaged CS I/O connector P23B. Loaded lastest OS 23. Bench calibrated and tested unit, then tested unit in environmental chamber. Retested unit at bench. All problems resolved.
CR10X 20139 2011-07-21 Sherman Island Removed Lab none This silver box was removed from the Sherman Pasture met wiring panel. It was replaced with sn 47253
CR10X 47253 2011-07-21 Lab none Installed Sherman Island Data Logger Met The silver box of this data logger was reinstalled in the met wiring panel at the Sherman Pasture. The silver box sn 20139 was removed.
miniDOT 1221 2014-10-09 Lab none Installed Mayberry DO Sensor F Installed at 109cm below the water surface on the fence post in the channel.
miniDOT 1220 2014-10-16 Lab none Installed Mayberry DO Sensor D Installed at 26cm below the water surface on the fence post in the channel.
miniDOT 1219 2014-10-16 Lab none Installed Mayberry DO Sensor E Installed at 66cm below the water surface on the fence post in the channel.
miniDOT 1221 2014-10-09 West Pond DO Sensor C Removed Lab none Removed from West Pond for installation at Mayberry where a longer profile is possible.
miniDOT 1220 2014-10-09 West Pond DO Sensor A Removed Lab none Removed from West Pond for installation at Mayberry where a longer profile is possible.
miniDOT 1219 2014-10-09 West Pond DO Sensor B Removed Lab none Removed from West Pond for installation at Mayberry where a longer profile is possible.
GMP222 Z2720011 2014-10-09 Lab none Installed Twitchell Alfalfa Installed in the FD ambient chamber to replace sn Z2720014 that had high readings.
GMP222 Z2720014 2014-10-09 Twitchell Alfalfa Removed Lab none Removed from the FD ambient chamber due to high readings. Replaced by sn Z2720011.
LI-7550 AIU-1285 2014-10-02 Mayberry Interface LI-7550 Upgrade Mayberry Interface LI-7550 Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
LI-7550 AIU-0397 2014-10-02 Mayberry-Portable Upgrade Mayberry-Portable Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
LI-7550 AIU-0397 2014-03-07 Mayberry Gas Analyzer CO2/H2O Upgrade Mayberry Gas Analyzer CO2/H2O Updated the 7550 firmware from v6.5.5 to v7.0.1. Improved interface, support for SmartFlux, bug fixes.
LI-7550 AIU-0674 2014-10-02 Twitchell Alfalfa Interface LI-7550 Upgrade Twitchell Alfalfa Interface LI-7550 Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
LI-7550 AIU-0674 2014-03-07 Twitchell Alfalfa Interface LI-7550 Upgrade Twitchell Alfalfa Interface LI-7550 Updated the 7550 firmware from v6.5.5 to v7.0.1. Improved interface, support for SmartFlux, bug fixes.
LI-7550 AIU-0666 2014-10-02 West Pond Interface LI-7550 Upgrade West Pond Interface LI-7550 Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
LI-7550 AIU-0666 2014-02-28 Lab none Upgrade Lab none Upgraded firmware from v6.5.5 to v7.0.1 - FPGA v6.0.3 not needing update. Improved interface, support for SmartFlux, bug fixes.
LI-7550 AIU-0650 2014-10-02 Twitchell Rice Interface LI-7500 Upgrade Twitchell Rice Interface LI-7500 Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
LI-7550 AIU-0650 2014-03-07 Twitchell Rice Interface LI-7500 Upgrade Twitchell Rice Interface LI-7500 Updated the 7550 firmware from v6.5.5 to v7.0.1. Improved interface, support for SmartFlux, bug fixes.
miniDOT 1220 2014-09-25 Lab none Installed West Pond DO Sensor A Installed at West Pond in a profile at 13cm above soil in water column.
miniDOT 1219 2014-09-25 Lab none Installed West Pond DO Sensor B Installed at West Pond in a profile at 7.5cm above soil in water column.
miniDOT 1221 2014-09-25 Lab none Installed West Pond DO Sensor C Installed at West Pond in a profile at 2cm above soil in water column.
GP-001 GP20140009 2014-10-02 West Pond CO2 Probe B Removed Lab none Removed from West Pond due to over range readings. It was replaced by a 0 to 4% probe sn GP20140021.
GP-001 GP20140009 2014-06-26 Lab none Installed West Pond CO2 Probe B This 0 to 2% probe was installed at West Pond to hang off scaffold leg 5 cm below surface. No program change, using existing SE channels from last Forerunner probe.
GP-001ex GP2013ex01 2014-02-11 Lab none Tested Disposed I took apart this probe in the lab. It is definitely dead. I sent photos to Forerunner.
GP-001ex GP2013ex01 2014-02-10 West Pond CO2 Probe B Removed Lab none Removed for West Pond due to bad readings - probably water damage.
GP-001ex GP2013ex01 2014-01-23 Lab none Installed West Pond CO2 Probe B This 0 to 3% probe was installed at West Pond. The CR1000 program was updated for three SE measurements: CO2, temp and RH.
GP-001 GP20140021 2014-10-02 Lab none Installed West Pond CO2 Probe B Installed at West Pond to replace the current 2% probe sn GP20140009.
LI-7500A 0042 2014-09-30 Lab none Installed Tonzi-Floor Installed on the Tonzi Floor tower to replace sn 1594 for calibration.
LI-7500A 1594 2014-09-30 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor tower for calibration. Replaced by sn 0042.
LI-7500A 0042 2014-10-02 Lab none Calibration Lab none Last calibration: 2014-06-24
AGC=54 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.76kPa, Tair=23.23C, Tdew=14.05C
Zero air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.81
CO2 zero 0.9506 -> 0.9500 Offset: -2.15ppm
CO2 span 1.0074 -> 0.9989 Offset: 4.78ppm
H2O zero 0.8506 -> 0.8505 Offset: -0.3ppm
H2O span 0.996 -> 0.9819 Offset: 0.3ppm
Notes: Exterior serial connector doesn't work. This calibration was done using the thermistor in the box and not on the calibration tube.
LI-7500A 75H-2176 2014-09-25 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell tower to replace sn 75H-2150 for calibration.
LI-7500A 75H-2150 2014-09-25 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell tower for calibration. Replaced by sn 75H-2176.
LI-7500A 0418 2014-09-25 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman tower to replace sn 0042 for calibration.
LI-7500A 0042 2014-09-25 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman tower for calibration. Replaced by sn 0418.
LI-7500A 0073 2014-09-25 Lab none Installed East End Gas Analyzer CO2/H2O Installed on the East End tower to replace sn 0041 for calibration.
LI-7500A 0041 2014-09-25 East End Gas Analyzer CO2/H2O Removed Lab none Removed from the East End tower for calibration. Replaced by sn 0073.
miniDOT 1221 2014-09-23 Lab none Maintenance Lab none Using the miniDOTControl program I set the time and set the sample interval to one minute.
miniDOT 1220 2014-09-24 Lab none Maintenance Lab none Using the miniDOTControl program I set the time and set the sample interval to one minute.
miniDOT 1219 2014-09-24 Lab none Maintenance Lab none Using the miniDOTControl program I set the time and set the sample interval to one minute.
LI-7500A 0418 2014-09-23 Lab none Calibration Lab none Last calibration: 2014-06-10
AGC=51 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.65kPa, Tair=27.35C, Tdew=15.91C
Zero Air at 1050psi, God CO2 421.22 at 1350 Tdewspan=15.8
CO2 zero 0.9015 -> 0.9013 Offset: -0.84ppm
CO2 span 1.0058 -> 0.9992 Offset: 3.53ppm
H2O zero 0.8533 -> 0.8510 Offset: -13.2ppm
H2O span 0.9973 -> 0.9823 Offset: 0.32ppm
Notes:
LI-7500A 75H-2176 2014-09-23 Lab none Calibration Lab none Last calibration: 2014-03-17
AGC=88.9 SDM=1 DAC1=- DAC2=-
Ambient P=100.22kPa, Tair=26.52C, Tdew=16.19C
Zero Air at 1050psi, God CO2 421.22 at 1350 Tdewspan=15.80
CO2 zero 0.8687 -> 0.8685 Offset: -0.73ppm
CO2 span 1.0113 -> 1.0145 Offset: -1.74ppm
H2O zero 0.8531 -> 0.8547 Offset: 8.8ppm
H2O span 0.9904 -> 1.0075 Offset: -0.37ppm
Notes:
LI-7500A 0073 2014-09-23 Lab none Calibration Lab none Last calibration: 2014-06-24
AGC=59 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.46kPa, Tair=25.61C, Tdew=17.80C
Zero Air at 1050psi, God CO2 421.22 at 1350 Tdewspan=15.80
CO2 zero 0.958 -> 0.9586 Offset: 2.05ppm
CO2 span 1.0018 -> 1.0023 Offset: -0.37ppm
H2O zero 0.7389 -> 0.7382 Offset: -3.9ppm
H2O span 1.0138 -> 1.0109 Offset: 0.06ppm
Notes:
LI-820 CGA-556 2014-08-21 Tonzi-Tower Removed Lab none Removed from the tower due to dirty cell that could not be cleaned well enough in the field.
LI-820 CGA-556 2014-09-11 Lab none Installed Tonzi-Tower Reinstalled in the CO2 profile system on the Tonzi Tower. The IRGA seemed to be working but something wrong with the plumbing made it impossible to calibrate on the tower.
LI-820 CGA-556 2014-09-09 Lab none Maintenance Lab none I cleaned the cell and replaced the internal inlet tubing. The tubing had some dirty looking deposit inside it. I calibrated the IRGA: P = 100.3kPa, CellT = 52.2C, Zero = 0.9656, Span = 0.8022
LI-7500 0976 2014-09-09 Lab none Calibration Lab none Last calibration: 2014-03-12
AGC=59 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.32kPa, Tair=23.50C, Tdew=13.94C
Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25
CO2 zero 0.86 -> 0.8597 Offset: -1.06ppm
CO2 span 1.0068 -> 1.0070 Offset: -0.12ppm
H2O zero 0.6254 -> 0.6281 Offset: 19.4ppm
H2O span 1.0088 -> 1.0157 Offset: -0.15ppm
Notes:
LI-7500A 0035 2014-09-09 Lab none Calibration Lab none Last calibration: 2014-03-17
AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.49kPa, Tair=23.33C, Tdew=13.88C
Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25
CO2 zero 0.9389 -> 0.9386 Offset: -1.05ppm
CO2 span 1.0139 -> 1.0181 Offset: -3.02ppm
H2O zero 0.8932 -> 0.8957 Offset: 20ppm
H2O span 0.9462 -> 0.9838 Offset: -0.79ppm
Notes: The signal seems a little noisy: +-1ppm CO2 and +-20mmol/m3 H2O
LI-7500A 0065 2014-09-09 Lab none Calibration Lab none Last calibration: 2014-04-02
AGC=73.73 SDM=1 DAC1=- DAC2=-
Ambient P=100.35kPa, Tair=22.06C, Tdew=12.61C
Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25
CO2 zero 0.8899 -> 0.8899 Offset: 0.0ppm
CO2 span 1.0062 -> 1.0070 Offset: -0.52ppm
H2O zero 0.704 -> 0.7053 Offset: 8.6ppm
H2O span 0.9932 -> 1.0101 Offset: -0.34ppm
Notes:
CS511 DO-8 2014-09-10 Twitchell Rice DO Sensor Removed Lab none Removed from the Twitchell rice because the field in drained.
SRS-Pi PRI 759503109 2014-09-11 Lab none Installed Tonzi-Tower I installed this sensor on the top of the Tonzi Tower on a post near the pivot of the radiation boom. This sensor is measuring incoming radiation.
SRS-Pi PRI 759503109 2014-09-09 Lab none Maintenance Lab none I updated the SDI-12 address from the default 0 (zero) to 1 (one) using a CR1000 in SDI-12 transparent mode. I was not able to do this with a 23x.
SRS-Pi PRI 759503109 2014-08-21 Lab none Tested Lab none I tried to install this sensor on the Tonzi Tower. This is a SDI-12 sensor and the 23x was having trouble reading it. This might be due to a default SDI-12 address of 0 (zero). I could not change the address with the 23x terminal in SDI-12 transparent mode, so I brought it back to the lab.
SRS-Pi PRI 816903083 2014-09-11 Tonzi-Tower Maintenance Tonzi-Tower I updated the SDI-12 of this sensor from 0 (zero) to 2 (two). This solved the problem of the 23x reading the sensor. The update was done using a CR1000 in SDI-12 transparent mode.
SRS-Pi PRI 816903083 2014-08-21 Lab none Installed Tonzi-Tower I installed this sensor on the radiation boom at the tower top. It is facing down collecting reflected radiation. This is an SDI-12 sensor, but the 23x was having trouble reading it. I brought in its twin to troubleshoot in the lab.
LI-7500 0976 2014-09-11 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower tower to replace sn 0418 for calibration.
LI-7500A 0418 2014-09-11 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower tower for calibration. Replaced by sn 0976.
LI-7500A 0035 2014-09-11 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 0073 for calibration.
LI-7500A 0073 2014-09-11 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 0035.
LI-7500A 0065 2014-09-10 Lab none Installed Mayberry-Portable Installed on the Mayberry Portable tower to replace sn 75H-2176 for calibration.
LI-7500A 75H-2176 2014-09-10 Mayberry-Portable Removed Lab none Removed from the Mayberry Portable tower for calibration. Replaced by sn 0065.
LI-7700 TG1-0155 2014-09-10 East End Gas Analyzer CH4 Calibration East End Gas Analyzer CH4 Calibration Location: East End
Last calibration: 2014-03-07
RSSI=80%, Optical RH=23%
Ambient P=101.23kPa, Tair=29.9C
Zero Air at 1400psi, 2nd CH4 4.995 at 700
CH4 Zero Offset: 0.085ppm
CH4 Span Offset: -0.195ppm
Notes: Very hot day with no wind. Offsets larger than usual.
LI-7700 TG1-0223 2014-09-10 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2014-03-07
RSSI=82%, Optical RH=12%
Ambient P=100.88kPa, Tair=31.5C
Zero Air at 1400psi, 2nd CH4 4.995 at 750
CH4 Zero Offset: 0.025ppm
CH4 Span Offset: -0.085ppm
Notes: Very hot day with no wind. Span offset a little large.
LI-7700 TG1-0211 2014-09-10 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell
Last calibration: 2014-03-07
RSSI=69%, Optical RH=22%
Ambient P=kPa, Tair=25.9C
Zero Air at 1400psi, 2nd CH4 4.995 at 800
CH4 Zero Offset: 0.07ppm
CH4 Span Offset: -0.075ppm
Notes: Very warm day with no wind. Offsets seem a little large.
LI-7500A 75H-2180 2014-09-08 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on the West Pond tower with 7550 sn AIU-0666. There was no 7550-7500A pair on the tower due to issues when we tried to swap the head on 9/4.
LI-7550 AIU-0666 2014-09-08 Lab none Installed West Pond Interface LI-7550 Installed on the West Pond tower with head sn 75H-2180. There was now 7550-7500A pair on the tower because of issues with swapping the head on 9/4. This box has a new door gasket.
LI-7700 TG1-0155 2014-02-25 Twitchell Alfalfa Gas Analyzer CH4 Installed East End Gas Analyzer CH4 Installed on the East End Tower with its wash box to replace sn TG1-0203 that was returned to Frank. This instrument came directly from Alfalfa.
LI-7700 TG1-0155 2014-02-25 Twitchell Alfalfa Gas Analyzer CH4 Removed East End Gas Analyzer CH4 Removed from the Alfalfa tower along with its wash box to be installed on the East End Tower to replace sn TG1-0203 that was returned to Frank
LI-7700 TG1-0223 2012-05-02 Sherman Island Gas Analyzer CH4 Installed West Pond Gas Analyzer CH4 Installed at West Pond on the folding boom at the top of the tower.
LI-7700 TG1-0223 2012-05-02 Sherman Island Gas Analyzer CH4 Removed West Pond Gas Analyzer CH4 Removed from the Sherman tower for use at West pond. Replaced by LGR (sn 08-114).
LI-7700 TG1-0357 2014-06-13 Lab none Installed Mayberry-Portable Installed on the Mayberry Portable tower to replace sn TG1-224 that was having line lock problems.
LI-7550 AIU-0614 2014-09-04 West Pond Interface LI-7550 Removed Lab none Removed from the West Pond tower. In trying to swap the head, this box went into a reboot loop and we could not get it to complete a boot. This box also needs its connector circuit board upgraded to prevent condensation problems.
LI-7500A 0065 2014-09-04 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed for calibration. Problems with the 7550 box sn AIU-0614 prevented installation of a new 7500.
LI-7500A 75H-2180 2014-09-02 Lab none Calibration Lab none Last calibration: 2014-03-26
AGC=94.8 SDM=1 DAC1=- DAC2=-
Ambient P=99.73kPa, Tair=23.3C, Tdew=14.27C
Zero Air at 1000psi, God CO2 421.22 at 1350 Tdewspan=15.25
CO2 zero 0.9147 -> 0.9143 Offset: -1.66ppm
CO2 span 1.0173 -> 1.0134 Offset: 1.98ppm
H2O zero 0.8803 -> 0.8812 Offset: 5.3ppm
H2O span 0.9946 -> 1.0193 Offset: -0.5ppm
Notes:
LI-7500A 75H-2182 2014-08-25 Lab none Calibration Lab none Last calibration: 2014-02-28
AGC=83.4 SDM=1 DAC1=- DAC2=-
Ambient P=99.77kPa, Tair=23.79C, Tdew=12.26C
Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25
CO2 zero 0.8815 -> 0.8816 Offset: 0.90ppm
CO2 span 1.0064 -> 1.0074 Offset: -0.54ppm
H2O zero 0.8476 -> 0.8463 Offset: -7.7ppm
H2O span 0.9875 -> 1.0039 Offset: -0.33ppm
Notes:
CR23X 2071 2011-01-24 Borrowed Borrowed Lab none Borrowed from Robin Weber in Allen Goldstein's lab.
CR23X 2071 2014-08-29 Lab none Upgrade Lab none The operating system was updated from Rev9 to Rev19.
GMP221 K1340006 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in profile B at 8cm to replace sn F4960018 due to zero reading.
GMT221 F4960018 2014-08-27 Twitchell Alfalfa Removed Lab none Removed from profile B at 8cm due to reading zero. Replaced by sn K1340006.
GMP221 Y2009100 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in profile B at 2cm to replace sn F4960019 due to zero reading.
GMT221 F4960019 2014-08-27 Twitchell Alfalfa Removed Lab none Removed from profile B at 2cm due to reading zero. Replaced by sn Y2009100.
GMP222 Z2720014 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in Forced Diffusion probe 1 to replace sn Z2720011 due to zero reading
GMP222 Z2720011 2014-08-27 Twitchell Alfalfa Removed Lab none Removed from Forced Diffusion probe 1 because it was reading zero. Replaced by sn Z2720014