Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7700
Serial No:TG1-0211
Tag No:20111000791
Acquisition Date:2011-12-12
Project:USDA
Location on 2023-07-24:Mayberry, 5.11
Notes: Came with LI7550 sn AIU-614 and LI7540 sn 75H-2150. Settings: lasermoddepth=13265, laserstarttemp=11.4, blockstarttemp=30, blockstarttemplowrange=5, rssidropthresh=0, pzero=57.71, pspan=15.31, samplegain=0, refgain=0, mirrorpos=0, offset1=32768, delta1=77, offset2=0, delta2=0, dither=30000, sampledcoffset=420, sampleacoffset=-1000, sampleopticaloffset=0.

Documents

Manual: Collecting_LI-7700_Ethernet_Data.pdf
Manual: LI-7700_Manual_V2_201103.pdf
Manual: LI-7700_Manual_V3_201312.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6526 2024-01-11 Mayberry Calibration Mayberry Calibration Location: Mayberry
Last calibration: 2023-07-20
RSSI=71%, Optical RH=3%
Ambient P=102.1kPa, Tair=11.5C
Zero Air at 1850psi, AMP CA03923 3.3865 at 1700
CH4 Zero Offset: .032ppm
CH4 Span Offset: .0095ppm
Notes: Breezy day
6353 2023-07-24 Lab none Installed Mayberry Gas Analyzer CH4 Installed at Mayberry to replace sn TG1-0223 for calibration.
6340 2023-07-20 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2023-06-15
RSSI=73.4%, Optical RH=24%
Ambient P=100.4kPa, Tair=23.4C
Zero Air at 975psi, AMP CA04119 4.2238 at 925
CH4 Zero Offset: -.014ppm
CH4 Span Offset: -.0048ppm
Notes: Could only connect after laptop restart. After calibration, flushed headspace with zero air until optics RH=9-10% and swapped headspace desiccant.
6327 2023-07-11 Hill Slough Gas Analyzer CH4 Removed Lab none Removed from Hill Slough for calibration. Replaced by sn TG1-0421.
6306 2023-06-15 Hill Slough Calibration Hill Slough Calibration Location: Hill Slough
Last calibration: 2022-11-30
RSSI=73%, Optical RH=23%
Ambient P=100.9kPa, Tair=17.0C
Zero Air at 1200psi, AMP CA04119 4.2238 at 1000
CH4 Zero Offset: -.003ppm
CH4 Span Offset: n/a
Notes: Windy! Did not set the span because the sensor value would "stabilize" at different number that would vary by >0.25ppm. The "stable" value would change whenever a gust would blow through.
6099 2022-11-30 Hill Slough Calibration Hill Slough Calibration Location: Hill Slough
Last calibration: 2022-06-08
RSSI=71%, Optical RH=26%
Ambient P=101.4kPa, Tair=10.3C
Zero Air at 1650psi, AMP CA04396 1.9317 at 1100
CH4 Zero Offset: .005ppm
CH4 Span Offset: .0003ppm
Notes: After calibration we flushed headspace with zero air until optics RH = 8%. We swapped in fresh drierite. The current thermocouple assembly is missing an o-ring (has putty instead in the o-ring groove), so we swapped in a brand-new thermocouple assembly.
5898 2022-06-08 Hill Slough Calibration Hill Slough Calibration Location: Hill Slough
Last calibration: 2021-12-01
RSSI=73%, Optical RH=21%
Ambient P=100.7kPa, Tair=24.0C
Zero Air at 550psi, AMP CA04119 4.2238 at 1525
CH4 Zero Offset: -.002ppm
CH4 Span Offset: .0008ppm
Notes: Very windy day of course. Did not change drierite or flush headspace with zero air. Changed the top/bottom heater to start at 18:00 instead of at 16:00 for summertime.
5614 2021-12-01 Hill Slough Calibration Hill Slough Calibration Location: Hill Slough
Last calibration: 2021-06-01
RSSI=72%, Optical RH=38%
Ambient P=101.6kPa, Tair=16.6C
Zero Air at 950psi, AMP CA04396 1.9317 at 1500
CH4 Zero Offset: -.066ppm
CH4 Span Offset: .0633ppm
Notes: Relatively still day. After calibration, flushed out headspace until optics RH=10% and swapped out desiccant bottle for one with fresh drierite. Zero air tank pressure uncertain--wonky regulator.
5274 2021-05-26 Hill Slough Calibration Hill Slough Calibration Location: Hill Slough
Last calibration: 2020-11-17
RSSI=70%, Optical RH=7%
Ambient P=101.4kPa, Tair=22.3C
AMP FB04161 at 1150psi, AMP CA04119 4.223 at 2100
CH4 Zero Offset: 0.077ppm
CH4 Span Offset: -0.123ppm
Notes: Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. Zero offset and span are higher than I usually see on 7700s.
5175 2021-03-11 Bouldin Alfalfa Gas Analyzer CH4 Moved Hill Slough Gas Analyzer CH4 Installed at Hill Slough with site setup. This sensor was removed from BA because our spare 7700 is currently at the factory.
4971 2020-11-19 Lab none Installed Bouldin Alfalfa Gas Analyzer CH4 Installed at Bouldin Alfalfa to replace sn TG1-0421 for calibration.
4946 2020-11-17 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2020-05-22
RSSI=72%, Optical RH=22%
Ambient P=100.2kPa, Tair=20.5C
Zero Air at 575psi, AMP CA4396 1.9497 at 400
CH4 Zero Offset: -.064ppm
CH4 Span Offset: .011ppm
Notes: Swapped optics drierite after calibration and flushed headspace with zero air. New optics RH was 11%.
4951 2020-11-12 Sherman Wetland-Temp1 Gas Analyzer CH4 Removed Lab none Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup.
4485 2020-05-21 Sherman Wetland-Temp1 Gas Analyzer CH4 Calibration Sherman Wetland-Temp1 Gas Analyzer CH4 Calibration Location: Sherman Wetland Temp
Last calibration: 2019-10-22
RSSI=73%, Optical RH=23%
Ambient P=101.3kPa, Tair=25.8C
Zero Air at 700psi, AMP CA3923 3.121 at 700
CH4 Zero Offset: 0.023ppm
CH4 Span Offset: -0.005ppm
Notes: Breezy day. Swapped optics drierite after calibration and used cal gas to dry head space. New optics RH was 8%.
4333 2020-01-30 East Pond Gas Analyzer CH4 Moved Sherman Wetland-Temp1 Gas Analyzer CH4 Removed with the decommissioning of the EP tower and installed at the new SW Temporary site 1250m upwind of the Permanent Tower.
4116 2019-10-22 East Pond Gas Analyzer CH4 Calibration East Pond Gas Analyzer CH4 Calibration Location: East Pond
Last calibration: 2019-04-11
RSSI=72%, Optical RH=19%
Ambient P=101.9kPa, Tair=26.1C
Zero Air at 1300psi, AMP CA03923 3.121 at 1200
CH4 Zero Offset: 0.003ppm
CH4 Span Offset: -0.001ppm
Notes:
4096 2019-09-04 East Pond Gas Analyzer CH4 Upgrade East Pond Gas Analyzer CH4 We updated the 7700 firmware from v1.0.23 to v1.0.27. First try failed part way through, second try was successful.
3902 2019-04-11 East Pond Gas Analyzer CH4 Calibration East Pond Gas Analyzer CH4 Calibration Location: East Pond
Last calibration: 2018-10-26
RSSI=70%, Optical RH=14%
Ambient P=101.9kPa, Tair=17.1C
Zero Air M4002 at 1600psi, AMP CA03923 3.121 at 1800
CH4 Zero Offset: 0.014ppm
CH4 Span Offset: 0.003ppm
Notes: Windy day
3762 2018-10-25 East Pond Gas Analyzer CH4 Calibration East Pond Gas Analyzer CH4 Calibration Location: East Pond
Last calibration: 2018-04-13
RSSI=74%, Optical RH=10%
Ambient P=101.6kPa, Tair=26.4C
Zero Air at 1950psi, AMP Air 3.121 at 2100
CH4 Zero Offset: .029ppm
CH4 Span Offset: .006ppm
Notes: Calm day
3559 2018-06-28 East Pond Gas Analyzer CH4 Moved East Pond Gas Analyzer CH4 Sensor moved up 26cm to try to address rotation angle issues. The water level was 68cm and the sensor was 534cm above water level.
3402 2018-04-17 Lab none Installed East Pond Gas Analyzer CH4 Installed at EP tower as part of initial site set up.
3398 2018-04-13 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-10-02
RSSI=72%, Optical RH=4%
Ambient P=101.6kPa, Tair=23.7C
Zero Air at 500psi, LBL Air 1.9497 at 1500
CH4 Zero Offset: 0.009ppm
CH4 Span Offset: -.0057ppm
Notes: CH4 Span calibrated on 2018-04-16 under these conditions: P=100.5, T=26.1, RSSI=72, Optics RH =4.
3384 2018-04-05 Bouldin Corn Gas Analyzer CH4 Removed Lab Gas Analyzer CH4 Removed from BC and brought back to the lab for calibration. it was replaced by TG1-0155.
3104 2017-10-04 Lab none Installed Bouldin Corn Gas Analyzer CH4 Installed on the Bouldin Corn tower to replace sn TG1-0421 for calibration and testing in the lab for drop outs when washing problem.
3102 2017-10-02 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-05-31
RSSI=73.3%, Optical RH=4%
Ambient P=99.9kPa, Tair=28.4C
Zero Air at 550psi, AMP Air 1.9497 at 2050
CH4 Zero Offset: 0.010ppm
CH4 Span Offset: -.0087ppm
Notes:
3084 2017-09-26 East End-Temp2 Gas Analyzer CH4 Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
2838 2017-06-14 Lab none Installed East End-Temp2 Gas Analyzer CH4 Installed with initial setup of East End closed vegetation temporary tower for the summer of 2017
2793 2017-05-31 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2016-10-11
RSSI=71.2%, Optical RH=14%
Ambient P=100.3kPa, Tair=C
Zero Air at 1500psi, NOAA Air 1.89919 at 1600
CH4 Zero Offset: -0.1ppm
CH4 Span Offset: -0.002ppm
Notes: Replaced temperature component. Replaced desiccant and flushed cavity with zero air. Optics RH was 26% before replacement and 14% afterwards.
2770 2017-04-04 Twitchell Rice Gas Analyzer CH4 Removed Lab none Removed from Twitchell Rice site with Twitchell Rice decommissioning.
2490 2016-10-11 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell Rice
Last calibration: 2016-04-13
RSSI=68.0%, Optical RH=26%
Ambient P=101.6kPa, Tair=20.8C
zero air at 850psi, AMP CH4 4.030 at 1600
CH4 Zero Offset: -0.01ppm
CH4 Span Offset: 0.021ppm
Notes: windy, rice just harvested, tower out of field.
2234 2016-04-13 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell Rice
Last calibration: 2015-09-30
RSSI=66.1%, Optical RH=23%
Ambient P=102.1kPa, Tair=18.6C
Zero Air at 1000psi, AMP CH4 1.9966 at 950
CH4 Zero Offset: 0.018ppm
CH4 Span Offset: -0.010ppm
Notes: Gusty breeze, mirrors hard to get clean.
1949 2015-09-30 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell Rice
Last calibration: 2015-03-24
RSSI=70.8%, Optical RH=17%
Ambient P=101.6kPa, Tair=17.1C
Zero Air at 1100psi, AMP CH4 1.9966 at 1550
CH4 Zero Offset: 0.0015ppm
CH4 Span Offset: -0.0006ppm
Notes: light, sporadic rain
1762 2015-03-24 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell
Last calibration: 2014-09-10
RSSI=56%, Optical RH=25%
Ambient P=102.5kPa, Tair=18.8C
Zero air at 1200psi, 2nd CH4 4.995 at 600
CH4 Zero Offset: -0.05ppm
CH4 Span Offset: 0.065ppm
Notes: Very windy
1544 2014-09-10 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell
Last calibration: 2014-03-07
RSSI=69%, Optical RH=22%
Ambient P=kPa, Tair=25.9C
Zero Air at 1400psi, 2nd CH4 4.995 at 800
CH4 Zero Offset: 0.07ppm
CH4 Span Offset: -0.075ppm
Notes: Very warm day with no wind. Offsets seem a little large.
1231 2014-03-07 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell
Last calibration: 2013-08-21
RSSI=63%, Optical RH=17%
Ambient P=102.1kPa, Tair=18.5C
Zero Air at 1600psi, 2nd CH4 4.995 at 1100
CH4 Zero Offset: 0.01ppm
CH4 Span Offset: 0.004ppm
Notes: Calibrated by Cove and Sara. Firmware upgraded to v1.0.21 after calibration. Windy day.
1049 2013-08-21 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell
Last calibration: 2013-02-21
RSSI=64%, Optical RH=13%
Ambient P=101.2kPa, Tair=18.3C
Zero Air at 1750psi, CH4 in air 4.995 at 1300
CH4 Zero Offset: 0.01ppm
CH4 Span Offset: -0.025ppm
Notes: It was windy but seemed to have no effect. The concentration of the secondary ch4 standard has been reevaluated since last cal. Last time 4.982, now 4.995
933 2013-02-21 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell
Last calibration: 2012-09-06
RSSI=65%, Optical RH=6%
Ambient P=102.1kPa, Tair=7.87C
Zero Air at 1800psi, 2nd CH4 in Air 4.982 at 1300
CH4 Zero Offset: 0.005ppm
CH4 Span Offset: 0.018ppm
Notes:
838 2012-09-06 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell
Last calibration: 2011-12-03
RSSI=63%, Optical RH=%
Ambient P=101.45kPa, Tair=22.5C
ultra zero air at 1900psi, ch4 in air 4.982 at 950
CH4 Zero Offset: 0.050ppm
CH4 Span Offset: -0.002ppm
Notes: Calibrated on the Twitchell tower by Joe and Cove. Last cal is the original done by Licor.
764 2012-01-17 Lab none Installed Twitchell Rice Gas Analyzer CH4 Installed on the Twitchell tower with the old 7550 box (sn AIU-0397). Connected to the CR1000 by SDM.
 
 