Equipment and Calibration History

Page: 1 2 3 4 5 [6] 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65
Ser No.DateLoc FromTitleLoc ToNotes
LI-7500A 75H-2180 2023-03-29 Lab Calibration Lab Last calibration: 2022-12-05
AGC=100.7 SDM= DAC1= DAC2=
Ambient P=99.8kPa, Tair=18.7C, Tdew=6.3C
Zero Air at 1300psi, AMP CA04119 524.77 at 1150 Tdewspan=16.00
CO2 zero 0.9145 -> .9149 Offset: 2.13ppm
CO2 span 1.0433 -> 1.044 Offset: -.33ppm
H2O zero 0.8933 -> .8913 Offset: -12.8ppm
H2O span 1.0381 -> 1.0275 Offset: .22ppm
Notes: Had to use 524.45 as "span point" for sensor to read 524.77 after spanning. I didn't really notice this for other sensors I calibrated today but I'll need to pay more attention next time. Previously, there was a large offset in CO2 span. Need to keep an eye on this sensor.
GP-001 GP20210002 2023-03-23 Lab none Installed Gilbert Tract CO2 Probe Installed at Gilbert Tract on a post (just off the far end of the boardwalk) with the other water quality sensors. Should be at the same height as before (about 5cm above the soil)
GP-001 GP20200089 2023-03-23 Gilbert Tract CO2 Probe Removed Lab none Removed from Gilbert Tract (was installed by the staff gauge) and brought back to the lab for testing. I think this sensor is dead.
LI-7700 TG1-0421 2023-02-02 Lab none Installed Loaned none Re-installed on the BWW tower after calibration in the lab
LI-7500A 75H-2514 2023-03-16 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 75H-2176 for calibration.
LI-7500A 75H-2176 2023-03-16 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2514.
LI-7500A 75H-2182 2023-03-16 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2023-03-16 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 75H-2182.
LI-7500A 1594 2023-03-16 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 0418 for calibration.
LI-7500A 0418 2023-03-16 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 1594.
LI-7500A 75H-2182 2023-03-15 Lab Calibration Lab Last calibration: 2023-02-23
AGC=100.4 SDM=0 DAC1=- DAC2=-
Ambient P=100.6kPa, Tair=22.15C, Tdew=6.27C
Zero Air at 1400psi, AMP CA04119 524.77 at 1250 Tdewspan=16.0
CO2 zero 0.8849 -> 0.8852 Offset: 1.3ppm
CO2 span 1.003 -> 1.0166 Offset: -9.47ppm
H2O zero 0.8262 -> 0.8292 Offset: 0.09ppm
H2O span 1.0144 -> 1.0332 Offset: -0.39ppm
Notes: First calibration since return from Licor. I entered all the coefficients by hand. Afterwards it was pretty right on except for CO2 span. Not sure why but decided to do the calibration.
LI-7550 AIU-1569 2023-03-14 Lab none Tested Lab none I tested this box in the lab. It had been removed from the field because the door was left open and it got wet inside. I had cleaned it up earlier but I'm not sure if I ran it before. Today, it was running but very slowly/laggy. At first I was concerned it was the new laptop but it was the same on the old laptop. I pulled the shroud off and found a little corrosion on one of the ribbon cable connectors. I cleaned this off and it seems to be working ok.
LI-7500A 75H-2514 2023-03-14 Lab Calibration Lab Last calibration: 2022-03-17
AGC=101.4 SDM=0 DAC1=- DAC2=-
Ambient P=99.2kPa, Tair=19.7C, Tdew=12.3C
Zero Air at 1400psi, AMP CA04119 524.77 at 1250 Tdewspan=16.0
CO2 zero 0.905 -> 0.9053 Offset: 1.4ppm
CO2 span 1.0125 -> 1.0169 Offset: -3.02ppm
H2O zero 0.8403 -> 0.8408 Offset: 3.4ppm
H2O span 1.0196 -> 1.0243 Offset: -0.9ppm
Notes: Using AMP loaner AIU-2312
LI-7500A 1594 2023-03-14 Lab Calibration Lab Last calibration: 2023-01-25
AGC=100.7 SDM=0 DAC1=- DAC2=-
Ambient P=99.1kPa, Tair=19.3C, Tdew=12.77C
Zero Air at 1400psi, AMP CA04119 524.77 at 1250 Tdewspan=16.0
CO2 zero 0.8656 -> 0.8656 Offset: 0.04ppm
CO2 span 1.0205 -> 1.0225 Offset: -1.37ppm
H2O zero 0.8869 -> 0.8885 Offset: 9.0ppm
H2O span 1.0127 -> 1.0196 Offset: -0.14ppm
Notes: Using AMP loaner AIU-2312
eosFD FD20190027 2023-03-01 Lab none Installed Tonzi-SoilCO2 Forced Diffusion Open Installed in the open location fresh from factory repair.
SS-6L-12V 21501223 2023-03-01 Lab none Installed Tonzi Solar Controller Soil CO2 Installed at the soil CO2 battery bank to replace sn 12501588 that doesn't seem to be charging batteries.
PS-15M 12501588 2023-03-01 Tonzi Solar Controller Soil CO2 Removed Lab none This solar controller has not been showing solar amps and doesn't seem to charge batteries. I brought it in for testing
BB-Fab BB-TZ_soilco2 2023-03-01 Tonzi Battery Bank soil co2 Replaced Tonzi Battery Bank soil co2 I brought the three batteries last remove from HS and put them under the solar panels. I rewired all the panels into a terminal strip. I used one of the old batteries for the fence charger and brought the rest in.
PAR-LITE 030382 2023-03-07 Loaned none Returned Lab none Returned by Sophie Ruehr from lab measurements of grow lights
PAR-LITE 030382 2023-03-01 Lab none Loaned Loaned none Loaned to Sophie Ruehr with 23x sn 2303 for lab measurements of grow lights
CR23X 2303 2023-03-07 Loaned none Returned Lab none Returned from Sophie Ruehr from lab measurements of grow lights
CR23X 2303 2023-03-01 Lab none Loaned Loaned none Loaned to Sophie Ruehr with PAR sn 030382 for lab measurements of grow lights
LAI-2270C PCA-3914 2023-03-03 Loaned none Returned Lab none Sensor and all accessories were returned from Manuela.
LI-7550 AIU-1866 2023-03-02 Bouldin Corn Interface LI-7550 Removed Lab none Unit is receiving 24V but no LED lights and couldn't connect. Removed from BC tower and brought back to the lab for testing.
PS-15M 12501588 2023-03-02 Lab none Tested Disposed none This charge controller hasn't been showing solar amps. I took it apart and cleaned out a bunch or large dead ants. I found two diodes and one capacitor that looked a little fouled but not obviously broken. I'm not sure I have the tools/skills to do any more. I cleaned it up, put it back together and tested it. Now the screen shows nonsense - different on each power cycle. I think it's a goner. Put it in the e-waste box.
eosFD FD20190027 2023-02-28 Factory none Returned Lab none Returned after repair at Eosense: CO2 sensor replaced, concentration calibration and flux calibration.
eosFD FD20190027 2022-08-23 Lab none Shipped Factory none Shipped to Eosense on RMA220809 for repair
eosFD FD20180026 2022-08-23 Lab none Shipped Factory none Shipped to Eosense on RMA220809 for repair
eosFD FD20180026 2023-02-28 Factory none Returned Lab none Received after repair at Eosense: CO2 sensor replaced, relay board replaced, pump replaced, concentration calibration, flux calibration
LI-7500A 75H-2182 2023-02-23 Factory none Received Lab none Received from Licor. They inspected the unit and found condensation on the source lens, the chopper motor, source and heat bottom defective. They replaced defective parts, subjected the unit to environmental tests, and calibrated.
EXO Turbidity 22C100485 2023-01-25 Gilbert Tract none Removed Lab none USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab.
EXO fDOM 21K102591 2023-01-25 Gilbert Tract none Removed Lab none USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab.
EXO fDOM 21K102592 2023-01-25 Gilbert Tract none Removed Lab none USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab.
EXO2 ADINAEXO1 2023-01-25 Gilbert Tract none Removed Lab none USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab.
EXO Turbidity 22C100485 2023-02-15 Lab none Returned Returned none Returned to Adina during our Elkhorn Slough visit
EXO fDOM 21K102591 2023-02-15 Lab none Returned Returned none Returned to Adina during our Elkhorn Slough visit
EXO fDOM 21K102592 2023-02-15 Lab none Returned Returned none Returned to Adina during our Elkhorn Slough visit
EXO2 ADINAEXO1 2023-02-15 Lab none Returned Returned none Returned to Adina during our Elkhorn Slough visit
LI-7500A 75H-2182 2023-02-09 Lab none Calibration Lab none Last calibration: 2022-03-22
AGC= SDM= DAC1= DAC2=
Ambient P=kPa, Tair=C, Tdew=C
at psi, 750 at Tdewspan=12.00
CO2 zero 0.8849 -> .8849 Offset: ppm
CO2 span 1.0222 -> 1.0030 Offset: ppm
H2O zero 0.8483 -> .8262 Offset: ppm
H2O span 1.0312 -> 1.0144 Offset: ppm
Notes: Calibrated at Licor after repair. Source lens, chopper motor, and source were replaced.
HygroVUE10 E3452 2023-02-14 Lab none Installed Gilbert Tract RH & Temp Installed on GT tower to replace HMP45 W0730036, which was removed for railing RH
HMP45AC W0730036 2023-02-14 Gilbert Tract RH & Temp Removed Lab none Removed from GT tower because RH has been maxing out last winter. Replaced by Hygrovue10 sn E3452
LI-7700 TG1-0421 2023-02-02 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2022-06-01
RSSI=79%, Optical RH=23%
Ambient P=101.3kPa, Tair=19.5C
Zero Air at 1400psi, AMP CA04119 4.2238 at 1275
CH4 Zero Offset: -.011ppm
CH4 Span Offset: .0138ppm
Notes: First calibration since sensor was installed on BWW urban tower 7 months ago. After calibration, swapped desiccant and flushed headspace with zero air until new optics RH=10%. Sensor removed this morning and will go back up on the BWW urban tower this afternoon.
LI-7500A 75H-2150 2023-02-02 Lab none Calibration Lab none Last calibration: 2022-03-02
AGC=101.8 SDM= DAC1= DAC2=
Ambient P=101.3kPa, Tair=17.9C, Tdew=4.5C
Zero Air at 1400psi, AMP CA04119 524.77 at 1250 Tdewspan=16.00
CO2 zero 0.9016 -> 0.9035 Offset: 7.88ppm
CO2 span 1.0075 -> 1.0257 Offset: -12.34ppm
H2O zero 0.8707 -> .8731 Offset: 16.0ppm
H2O span 1.0158 -> 1.0331 Offset: -.36ppm
Notes: First calibration in 11 months. Sensor installed on BWW urban tower for the past 10 months. Sensor removed this morning and will go back up on the BWW urban tower this afternoon. I updated the calibration coefficients directly on the BWW computer instead of uploading the new config file. I also swapped the HMP fan its screw terminal that was all corroded.
LAI-2270C PCA-3914 2023-02-02 Lab none Loaned Loaned none Loaned to Manuela Girotto for snow field canopy cover survey. Items loaned include console, 2 wands, 8 AA Ni-MH batteries, Ni-MH battery charger, tripod, and black plastic case with manual, USB cable, and various view caps.
LI-7500A 1594 2023-01-27 Lab none Tested Lab none Ran in the lab on the bench for 2 days and used eddypro to calculate the 30-min concentrations. CO2 and H2O data seemed reasonable: CO2 ranged from 420 to 510ppm and H2O ranged from 250 to 450 mmol/m3. Not sure why this sensor wasn't working in the field.
SRS-Ni NDVI 2046503354 2023-01-25 Lab none Tested Disposed none Removed the spliced cable and tested in the lab. Still no response on the terminal emulator. This sensor is probably dead. It's weird that both incoming and outgoing sensors went bad at the the same time. Both sensors had spliced cables and were connected to the same C port and the same pluggable power block. One of the spliced wires for data (red) had a few cracks in the plastic sheath; maybe it shorted against the bare G wire and killed both sensors.
SRS-Ni NDVI 2046503376 2023-01-25 Lab none Tested Disposed none Removed the spliced cable and tested in the lab. Still no response on the terminal emulator. This sensor is probably dead. It's weird that both incoming and outgoing sensors went bad at the the same time. Both sensors had spliced cables and were connected to the same C port and the same pluggable power block. One of the spliced wires for data (red) had a few cracks in the plastic sheath; maybe it shorted against the bare G wire and killed both sensors.
LI-7500A 1594 2023-01-25 Lab Calibration Lab Last calibration: 2022-11-28
AGC=99 SDM= DAC1= DAC2=
Ambient P=101.5kPa, Tair=23.4C, Tdew=2.53C
Zero Air at 1450psi, AMP CA04119 524.77 at 1250 Tdewspan=16.00
CO2 zero 0.8643 -> .8656 Offset: 5.31ppm
CO2 span 1.0168 -> 1.0205 Offset: -2.43ppm
H2O zero 0.8889 -> .8869 Offset: -10.6ppm
H2O span 1.0136 -> 1.0127 Offset: .02ppm
Notes: This sensor was not working at BC, but it worked in the lab and the calibration seems within normal range. I'll run it for a few days in the lab to see if it's stable.
RPI0W RPI0W-006 2023-01-18 Lab none Installed West Pond none Installed at West Pond for an intercomparison with the Canon camera. Temporarily ziptied to the bottom of the Canon enclosure arm. Facing west, same direction as the Canon camera.
LI-7500A 75H-2667 2023-01-19 Lab none Installed Gilbert Tract Gas Analyzer CO2/H2O Installed at Gilbert Tract to replace sn 75H-2418 for calibration.
LI-7500A 75H-2418 2023-01-19 Gilbert Tract Gas Analyzer CO2/H2O Removed Lab none Removed from Gilbert Tract for calibration. Replaced by sn 75H-2667.
LI-7500A 0035 2023-01-18 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 75H-2514 for calibration.
LI-7500A 75H-2514 2023-01-18 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for calibration. Replaced by sn 0035.
SRS-Ni NDVI 2046503376 2023-01-18 Mayberry Narrow Band NDVI Incoming Removed Lab none Removed from MB for testing in the lab. It was not responding on the terminal emulator.
SRS-Ni NDVI 2046503354 2023-01-18 Mayberry Narrow Band NDVI Reflected Removed Lab none Removed from MB for testing in the lab. It was not responding on the terminal emulator.
LI-7700 TG1-0223 2023-01-18 Mayberry Calibration Mayberry Calibration Location: Mayberry
Last calibration: 2022-06-07
RSSI=84%, Optical RH=15%
Ambient P=101.9kPa, Tair=9.9C
Zero Air at 1500psi, AMP CA04396 1.9317 at 1350
CH4 Zero Offset: .011ppm
CH4 Span Offset: .0093ppm
Notes: Did not flush headspace or change desiccant afterwards
LI-7700 TG1-0424 2023-01-19 Gilbert Tract Calibration Gilbert Tract Calibration Location: Gilbert Tract
Last calibration: 2022-06-09
RSSI=86%, Optical RH=23%
Ambient P=102.3kPa, Tair=8.0C
Zero Air at 1350psi, AMP CA04396 1.9317 at 1250
CH4 Zero Offset: .001ppm
CH4 Span Offset: .0017ppm
Notes: I wanted to flush the headspace with zero air, but the tubing did not quite reach. I swapped the desiccant.
LI-7700 TG1-0423 2023-01-18 East End Calibration East End Calibration Location: East End
Last calibration: 2022-06-09
RSSI=77%, Optical RH=21%
Ambient P=101.9kPa, Tair=12.6C
Zero Air at 1450psi, AMP CA04396 1.9317 at 1200
CH4 Zero Offset: .025ppm
CH4 Span Offset: .0023ppm
Notes: After calibration, changed desiccant and flushed headspace with zero air until optics RH=14%.
LI-7700 TG1-0155 2023-01-18 West Pond Calibration West Pond Calibration Location: West Pond
Last calibration: 2022-06-09
RSSI=81%, Optical RH=10%
Ambient P=101.9kPa, Tair=12.3C
Zero Air at 1500psi, AMP CA04396 1.9317 at 1125
CH4 Zero Offset: .008ppm
CH4 Span Offset: .0103ppm
Notes: Since optics RH = 10%, I did not change desiccant or flush headspace after calibration.
LI-7500A 75H-2667 2023-01-13 Lab Calibration Lab Last calibration: 2022-09-15
AGC=102 SDM= DAC1= DAC2=
Ambient P=100.3kPa, Tair=20.6C, Tdew=12.3C
Zero Air at 1600psi, AMP CA04119 524.77 at 1275 Tdewspan=16.00
CO2 zero 0.9413 -> .9418 Offset: 1.72ppm
CO2 span 1.0204 -> 1.0233 Offset: -1.91ppm
H2O zero 0.8407 -> .8421 Offset: 8.4ppm
H2O span 1.0338 -> 1.0314 Offset: .05ppm
Notes: Set span dew point to 16.00C because the usual 20.00C > room's dew point temp.
LI-7500A 0035 2023-01-13 Lab Calibration Lab Last calibration: 2022-08-25
AGC=99.4 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=18.8C, Tdew=11.5C
Zero Air at 1625psi, AMP CA04119 524.77 at 1275 Tdewspan=16.00
CO2 zero 0.9374 -> .9374 Offset: .2ppm
CO2 span 1.0234 -> 1.0295 Offset: -3.97ppm
H2O zero 0.8995 -> .9003 Offset: 4.9ppm
H2O span 1.0334 -> 1.0369 Offset: -.06ppm
Notes: Setting the CO2 span to 524.77ppm seemed to force the value to ~524.85 instead of 524.77. I think it's ok for now but something to keep an eye on. H2O value was not stabilizing when dewpt generator was set to 20.00C, probably because the room's dewpt<20.00C. I set the span point to 16.00C instead, which worked well.
LI-7500A 75H-2667 2022-09-15 Lab Calibration Lab Last calibration: 2022-03-02
AGC=100.4 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=20.6C, Tdew=13.4C
Zero Air at 1875psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9412 -> .9413 Offset: .3ppm
CO2 span 1.0171 -> 1.0204 Offset: -2.27ppm
H2O zero 0.841 -> .8407 Offset: -1.7ppm
H2O span 1.018 -> 1.0338 Offset: .01ppm
Notes: Calibrated with AMP 7550 sn AIU-1039. At first, large H2O span offset (-0.32) than usual. The condenser water tube in the back was low on water. I opened the cap to let water level rise again. Afterwards, H2O span was right on (offset 0.01).
RPI0W RPI0W-003 2023-01-04 Lab none Tested Lab none Plugged in in the lab but the picam could not boot all the way. The microSD card is probably busted from flickering power at the end of December. Joe put in a new microSD card and loaded the fwb code (r=1.8, b=1.8). We had some trouble getting the RTC recognized by the picam, but in the end it worked with a brand-new RTC and brand-new coin cell battery. Ran overnight in the lab and it seems fine.
RPI0W RPI0W-003 2023-01-06 Lab none Installed Mayberry Camera Redeployed at Mayberry on the same spot, attached to the wooden post on the southeast corner of the scaffolding.
LI-7500A 75H-2669 2023-01-06 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End to replace sn 75H-2667 for calibration.
LI-7500A 75H-2667 2023-01-06 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration. Replaced by sn 75H-2669.
LI-7500A 0065 2023-01-06 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond to replace sn 0035 for calibration.
LI-7500A 0035 2023-01-06 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration. Replaced by sn 0065.
LI-7500A 0041 2023-01-06 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 1594 because sn 1594 was not giving any data.
LI-7500A 1594 2023-01-06 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn because no CO2/H2O data since mid December. Replaced by sn 0041.
EXO2 22C103416 2023-01-06 Lab none Installed Hill Slough Water Level Installed in the channel at Hill Slough (same height as before) after calibrating in the lab yesterday. Installed without batteries in the sonde.
EXO pH-ORP 22D102641 2023-01-06 Lab none Installed Hill Slough pH-ORP Installed in the channel at Hill Slough (same height as before) after calibrating in the lab yesterday.
EXO fDOM 22E100649 2023-01-06 Lab none Installed Hill Slough fDOM Installed in the channel at Hill Slough (same height as before) after calibrating in the lab yesterday.
EXO Turbidity 22E106072 2023-01-06 Lab none Installed Hill Slough Turbidity Installed in the channel at Hill Slough (same height as before) after calibrating in the lab yesterday.
RPI0W RPI0W-003 2023-01-03 Mayberry Camera Removed Lab none Removed from the Mayberry tower because its web server was not booting. Brought back to the lab for further testing.
EXO2 22C103416 2023-01-03 Hill Slough Removed Lab none Removed from Hill Slough channel and brought back to the lab for calibration.
EXO pH-ORP 22D102641 2023-01-03 Hill Slough pH-ORP Removed Lab none Removed from Hill Slough channel and brought back to the lab for calibration.
EXO fDOM 22E100649 2023-01-03 Hill Slough fDOM Removed Lab none Removed from Hill Slough channel and brought back to the lab for calibration.
EXO Turbidity 22E106072 2023-01-03 Hill Slough Turbidity Removed Lab none Removed from Hill Slough channel and brought back to the lab for calibration.
eosFD FD20180067 2023-01-10 Lab none Tested Lab none Tested in the lab and it turned on once we plugged in the power with its original cable, hallelujah. Data started being weird (spiky and/or 0) in mid-December when the field started flooding. However, during the performance check the CO2 concentrations were consistently all over the place, even after trying to reset it. This chamber probably needs to go back to the factory.
eosFD FD20180067 2023-01-03 Bouldin Corn Forced Diffusion C Removed Lab none Removed from BC and brought the chamber and cable back to the lab for testing. We found it floating in the water underneath the tower. The cable connector was underwater.
LI-7550 AIU-1039 2023-01-03 none Installed Gilbert Tract Interface LI-7550 Installed at Gilbert Tract to replace sn AIU-1569, which had some corrosion inside and was acting funny.
LI-7550 AIU-1569 2023-01-03 Gilbert Tract Interface LI-7550 Removed Lab none Removed from Gilbert Tract because it was acting funny (took a long time to eject/mount USB, consistent 7700 time sync error) and there was some corrosion inside from being left open last month. it was replaced with AMP box AIU-1039.
GMP343 N2220008 2023-01-06 Tonzi CO2 Probe Tower Profile 23 Removed Lab none Removed from the top of the tower. This sensor has not been working for a while. I don't see any issues - brought in for testing.
LI-7500A 75H-2669 2023-01-04 Lab Calibration Lab Last calibration: 2022-08-23
AGC=101.1 SDM=0 DAC1=- DAC2=-
Ambient P=99.5kPa, Tair=17.6C, Tdew=8.8C
Zero Air at 1650psi, AMP CA04119 524.77 at 1250 Tdewspan=15.0
CO2 zero 0.9188 -> 0.9187 Offset: -0.3ppm
CO2 span 1.0083 -> 1.0187 Offset: -7.17ppm
H2O zero 0.904 -> 0.9029 Offset: -7.1ppm
H2O span 1.0174 -> 1.0330 Offset: -0.31ppm
Notes: Largish offset in CO2 span.
LI-7500A 0065 2023-01-04 Lab Calibration Lab Last calibration: 2022-09-12
AGC=101.1 SDM=0 DAC1=- DAC2=-
Ambient P=99.5kPa, Tair=18.1C, Tdew=10.7C
Zero Air at 1650psi, AMP CA04119 524.77 at 1250 Tdewspan=15.0
CO2 zero 0.8918 -> 0.8915 Offset: -1.5ppm
CO2 span 1.0227 -> 1.0227 Offset: 0.03ppm
H2O zero 0.7316 -> 0.7323 Offset: 5.1ppm
H2O span 1.0167 -> 1.0226 Offset: -0.12ppm
Notes:
LI-7500A 0041 2023-01-05 Lab Calibration Lab Last calibration: 2022-11-28
AGC=100.7 SDM=0 DAC1=- DAC2=-
Ambient P=100.1kPa, Tair=17.4C, Tdew=10.1C
Zero Air at 1600psi, AMP CA04119 524.77 at 1250 Tdewspan=15.0
CO2 zero 0.9968 -> 0.9961 Offset: -2.3ppm
CO2 span 1.0199 -> 1.0234 Offset: -2.37ppm
H2O zero 0.8221 -> 0.8232 Offset: 6.5ppm
H2O span 1.0304 -> 1.0382 Offset: -0.15ppm
Notes:
LI-7500A 0418 2022-12-21 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi to replace sn 75H-2669 for calibration.
LI-7500A 75H-2669 2022-12-21 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 0418.
LI-7500RS 75H-0128 2021-09-20 Factory none Calibration Factory none Last calibration:
AGC= SDM= DAC1= DAC2=
Ambient P=kPa, Tair=C, Tdew=C
at psi, 750 at Tdewspan=12
CO2 zero -> 1.1626 Offset: ppm
CO2 span -> 1.0087 Offset: ppm
H2O zero -> 1.0599 Offset: ppm
H2O span -> 1.0374 Offset: ppm
Notes: Full calibration values on yellow slips
LI-7500A 0418 2022-12-19 Lab Calibration Lab Last calibration: 2022-09-12
AGC=97.6 SDM=0 DAC1=- DAC2=-
Ambient P=101.4kPa, Tair=19.05C, Tdew=3.36C
Zero Air at 1650psi, AMP CA04119 524.77 at 1300 Tdewspan=16.00
CO2 zero 0.9086 -> 0.9070 Offset: -6.3ppm
CO2 span 1.005 -> 1.0059 Offset: -0.57ppm
H2O zero 0.9317 -> 0.9330 Offset: 6.5ppm
H2O span 1.0065 -> 1.0067 Offset: -0.01ppm
Notes: We have been watching this one as its SS seemed to drift after overhaul at Licor in the spring. The zero offsets are a little high than is common but otherwise seems okay.
LI-7500RS 75H-0128 2022-11-15 Lab Calibration Lab Last calibration: 2021-09-21
AGC=102 SDM=0 DAC1=- DAC2=-
Ambient P=101.1kPa, Tair=23.3C, Tdew=0.53C
Zero Air at 1800psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 1.1643 -> 1.1626 Offset: -3.68ppm
CO2 span 0.9978 -> 1.0087 Offset: -7.23ppm
H2O zero 1.0603 -> 1.0599 Offset: -2.2ppm
H2O span 1.0172 -> 1.0374 Offset: -0.44ppm
Notes: Calibrated on box sn AIU-2312. This pair is kept together because they are an RS model and belong to AMP.
LI-7500A 75H-2180 2022-12-06 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi Floor to replace sn 0418 for calibration.
LI-7500A 0418 2022-12-06 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi Floor for calibration. Replaced by sn 75H-2180.
LI-7500A 75H-2176 2022-12-06 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0065 for calibration.
LI-7500A 0065 2022-12-06 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2176.
LI-7700 TG1-0223 2022-12-13 Mayberry Gas Analyzer CH4 Maintenance Mayberry Gas Analyzer CH4 The humidity reading was 31%. I swapped the desiccant bottle refilling it with silica gel desiccant. The humidity didn't change and hopefully will go down. Swap with regular dri-rite at calibration next time.
Mark I A001911 2022-12-13 Gilbert Tract Weather Station Removed Lab none Removed from the GT tower. This unit has not be sending reliable data for a long time and Arable no longer supports this model. To be disposed of.
LI-7500A 75H-2180 2022-12-05 Lab Calibration Lab Last calibration: 2022-08-30
AGC=101.0 SDM=0 DAC1=- DAC2=-
Ambient P=100.6kPa, Tair=21.3C, Tdew=9.2C
Zero Air at 1700psi, AMP CA04119 524.77 at 1300 Tdewspan=16.0
CO2 zero 0.9148 -> 0.9145 Offset: -1.53ppm
CO2 span 1.0318 -> 1.0433 Offset: -7.27ppm
H2O zero 0.8916 -> 0.8933 Offset: 10.7ppm
H2O span 1.0122 -> 1.0381 Offset: -0.53ppm
Notes: Large offset in CO2 span but also large offset when removed from tower and similar offset in sn 2176 CO2 span calibrated at the same time. Used our 7550 box 0674 instead on the AMP loaner used previously. I think this IRGA is okay.
LI-7500A 0042 2022-11-29 Lab Calibration Lab Last calibration: 2022-11-15
AGC=99.3 SDM= DAC1= DAC2=
Ambient P=99.3kPa, Tair=22.8C, Tdew=2.02C
Zero Air at 1750psi, AMP CA 04119 524.77 at 1325 Tdewspan=16.00
CO2 zero 0.9481 -> .9478 Offset: -.94ppm
CO2 span 1.0127 -> 1.0145 Offset: -1.15ppm
H2O zero 0.847 -> .8472 Offset: 1.3ppm
H2O span 1.0292 -> 1.0267 Offset: .05ppm
Notes: Typed in "15.99" for 16.00C span point. Seemed to be more accurate than typing in "16.00" for 16.00C span point.