Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 [42] 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65
Ser No.DateLoc FromTitleLoc ToNotes
LI-7500A 75H-2668 2016-12-06 Lab none Calibration Lab none Last calibration: 2016-05-20
AGC=103.3 SDM=6 DAC1= DAC2=
Ambient P=100.29kPa, Tair=15.36C, Tdew=3.00C
Zero Air at 350psi, God CO2 421.22 at 400 Tdewspan=16.02
CO2 zero 0.9385 -> 0.9385 Offset: 0.0ppm
CO2 span 1.0136 -> 1.0171 Offset: -1.92ppm
H2O zero 0.8969 -> 0.9017 Offset: 30.2ppm
H2O span 1.0146 -> 1.0841 Offset: -1.41ppm
Notes:
PRB-K Type #8 00000FEDDAA1 2016-11-17 Lab none Installed Tonzi Installed at Tonzi near the main tower to replace sn 00000128B8072 that has two broken pins
PRB-K Type #8 00000128B8072 2016-11-17 Tonzi Removed Lab none Removed from Tonzi near the main tower. This probe has two broken pins.
FMA RMT-200 05-0010 2014-05-20 Lab none Damaged Lab none After being moved, but not treated roughly, the analyzer stopped reporting CH4. The ring time is variable from 2 to 20usec and the spectrum looks odd. Cleaning the mirrors made no difference. LGR thinks there is something wrong with the laser. At this time there is an 8 week wait for repairs and a minimum cost of 00 for cleaning and calibration - any repairs would be an additional cost. We don't need this analyzer right now and will wait to repair it.
U24-002-C 11020258 2016-11-16 Lab none Installed Mayberry Conductivity Sensor B The HOBO conductivity sensor was put in the water near the existing conductivity sensor at the southeast corner of the scaffolding for an intercomparison with the Campbell CS547A sensor b/c we're technically out of the CS547A range. It is tied to the tower with some rope.
CR10X 31307 2016-11-15 Lab none Maintenance Lab none The internal lithium battery was dead. I replaced the battery.
PQS 1 140457 2016-11-02 Vaira PAR Sensor Incoming Cal Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
NR01 2420 2016-11-02 Vaira Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
NR01 1720 2016-11-02 Vaira Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
CR6 2113 2016-11-02 Vaira Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
AM16-32A 1238 2016-11-02 Vaira Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
LI-7700 TG1-0224 2016-10-27 Lab none Loaned Loaned Loaned to Frank Anderson. He will use it temporarily at his Nisqually Marsh site while his 7700 is tested for no signal problems.
LI-7700 TG1-0224 2016-10-27 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2014-03-07
RSSI=79.3%, Optical RH=17%
Ambient P=100.4kPa, Tair=22.8C
Zero Air at 400psi, CH4 2nd 4.995 at 500
CH4 Zero Offset: 0.042ppm
CH4 Span Offset: 0.057ppm
Notes: Last calibration at Licor after replacement of laser due to unknown failure.
LI-7500A 75H-2176 2016-10-25 Lab none Calibration Lab none Last calibration: 2016-06-03
AGC=91.2 SDM=6 DAC1=- DAC2=-
Ambient P=100.28kPa, Tair=20.11C, Tdew=12.3C
Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.01
CO2 zero 0.8688 -> 0.8689 Offset: 0.22ppm
CO2 span 1.0175 -> 1.0176 Offset: -0.12ppm
H2O zero 0.8571 -> 0.8563 Offset: -4.9ppm
H2O span 1.0281 -> 1.0134 Offset: 0.31ppm
Notes:
DO6400/T DO-06 2016-10-25 Lab none Maintenance Lab none This sensor was serviced in the lab. The probe was taken apart and cleaned with vinegar and DI water. The membrane was replace and the probe was reassembled with fresh solution. In ambient air it was reading 80.0mV
DO6400/T DO-06 2016-10-26 Lab none Installed East End DO Sensor Re-installed at East End tower after testing in the lab.
LI-7500A 75H-2176 2016-10-26 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry tower after lab calibration to replace 75H-2669
LI-7500A 75H-2669 2016-10-26 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry tower for calibration. Replaced by 75H-2176
GP-001 GP20150111 2016-10-26 Lab none Installed Mayberry CO2 Probe B Installed probe at original 'mid' level with other water sensors. This level is at water surface right now due to low water table. This probe has its cable connector painted and covered with silicon sealant.
PQS 1 140457 2016-10-20 Twitchell Rice Moved Vaira PAR Sensor Incoming Cal Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
NR01 2420 2016-10-20 Twitchell Rice Moved Vaira Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
NR01 1720 2016-10-20 Twitchell Rice Moved Vaira Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
CR6 2113 2016-10-20 Twitchell Rice Moved Vaira Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
AM16-32A 1238 2016-10-20 Twitchell Rice Moved Vaira Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
AM16/32A 1236 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station north of the road
CR10X 47772 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station north of the road
AM16/32A 1430 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station within the Tower cow fence
AM16/32A 1429 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station within the Tower cow fence
CR10X 48042 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station within the Tower cow fence
LI-820 CGA-556 2016-10-13 Tonzi-Tower Calibration Tonzi-Tower none Last calibration was 2016-09-13 Cell T = 52.2, Pressure = 96.2, N2 at 1150psi, CO2 406ppm at 1800
zero offset = -1.5ppm
Span offset = -7.5ppm
Zero 0.9429 -> 0.9424
Span 0.8368 -> 0.8492
DO6400/T DO-06 2016-10-11 East End DO Sensor Removed Lab none Removed from East End for service. It was out od the water, reading high and had bite marks on the cable.
miniDOT 1220 2016-10-11 Mayberry DO Sensor D Moved Mayberry DO Sensor F Due to the lowering of the water to flush the salt, this sensor was no longer in the water, but could not be removed because it was rusted to the chain. It was moved to the bottom/lower location by inverting the chain.
miniDOT 1221 2016-10-11 Mayberry DO Sensor F Removed Lab none Removed from Mayberry due to lowering water for flushing salt. This sensor was at the bottom location. The top sensor (sn 1220) was no longer in water, but could not be removed due to rust and was moved to the bottom location.
GP-001 GP20140024 2016-10-11 Mayberry CO2 Probe C Removed Lab none Removed from Mayberry due to odd readings and lowering water for flushing salt
GP-001 GP20140025 2016-10-11 Mayberry CO2 Probe B Removed Lab none Removed from Mayberry due to odd readings and lowering water for flushing salt
LI-7700 TG1-0357 2016-10-11 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: Mayberry
Last calibration: 2016-04-13
RSSI=73.6%, Optical RH=13%
Ambient P=101.6kPa, Tair=18.3C
zero air at 850psi, AMP CH4 4.030 at 1900
CH4 Zero Offset: 0.020ppm
CH4 Span Offset: 0.025ppm
Notes: Windy.
LI-7700 TG1-0423 2016-10-11 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2016-05-20
RSSI=78.7%, Optical RH=23%
Ambient P=101.6kPa, Tair=16.4C
zero air at 850psi, AMP CH4 4.030 at 1900
CH4 Zero Offset: 0.026ppm
CH4 Span Offset: 0.023ppm
Notes: Windy.
LI-7700 TG1-0223 2016-10-11 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: west pond
Last calibration: 2016-04-13
RSSI=71%, Optical RH=27%
Ambient P=101.3kPa, Tair=22C
zero air at 700psi, AMP CH4 4.030 at 1200
CH4 Zero Offset: 0.001ppm
CH4 Span Offset: -0.006ppm
Notes: Breezy.
LI-7700 TG1-0155 2016-10-11 East End Gas Analyzer CH4 Calibration East End Gas Analyzer CH4 Calibration Location: East End
Last calibration: 2016-04-13
RSSI=82%, Optical RH=21%
Ambient P=101.6kPa, Tair=21.1C
zero air at 800psi, AMP CH4 4.030 at 1400
CH4 Zero Offset: 0.020ppm
CH4 Span Offset: 0.010ppm
Notes: Breezy
LI-7700 TG1-0211 2016-10-11 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell Rice
Last calibration: 2016-04-13
RSSI=68.0%, Optical RH=26%
Ambient P=101.6kPa, Tair=20.8C
zero air at 850psi, AMP CH4 4.030 at 1600
CH4 Zero Offset: -0.01ppm
CH4 Span Offset: 0.021ppm
Notes: windy, rice just harvested, tower out of field.
LI-7500A 75H-2514 2016-09-12 Lab none Calibration Lab none Last calibration: 2016-06-03
AGC=103.2 SDM=6 DAC1=- DAC2=-
Ambient P=99.87kPa, Tair=18.64C, Tdew=9.66C
Zero Air at 400psi, God CO2 421.22 at 400 Tdewspan=16.0
CO2 zero 0.9047 -> 0.9054 Offset: 2.80ppm
CO2 span 1.0055 -> 1.0092 Offset: -2.07ppm
H2O zero 0.8368 -> 0.8371 Offset: -10.5ppm
H2O span 1.015 -> 1.0239 Offset: -0.18ppm
Notes:
LI-7500A 75H-2150 2016-09-12 Lab none Calibration Lab none Last calibration: 2016-05-13
AGC=99.0 SDM=6 DAC1=- DAC2=-
Ambient P=99.83kPa, Tair=19.51C, Tdew=10.68C
Zero Air at 400psi, God CO2 421.22 at 400 Tdewspan=10.6
CO2 zero 0.902 -> 0.9019 Offset: -0.58ppm
CO2 span 1.008 -> 1.0084 Offset: -0.29ppm
H2O zero 0.8668 -> 0.8662 Offset: -4.1ppm
H2O span 1.0109 -> 1.0155 Offset: -0.1ppm
Notes:
CS655 17134 2016-10-07 Lab none Disposed Disposed This sensors was hit by s mower. One tine was ripped off and the other twisted. The cable is cut near the body. Put in the dead sensor box in the lab.
WM 1590 153806 2016-10-06 Lab none Installed Bouldin Alfalfa Anemometer Reinstalled on the Bouldin Alfalfa tower replacing sn 124215. This sonic is working correctly. Spikes were caused by the soil moisture probes powering up and browning out the system.
WM 1590 124215 2016-10-06 Bouldin Alfalfa Anemometer Removed Lab none Removed from the Bouldin Alfalfa tower. This sensor was installed to test the system. Replaced by sn 153806.
CS655 17134 2016-10-06 Bouldin Alfalfa Soil Moisture Probe B Removed Lab none Removed with soil sensor Set B from Bouldin Alfalfa. This set was hit by a mower.
HFP01 11334 2016-10-06 Bouldin Alfalfa Heat Flux Plate B Removed Lab none Removed with soil sensor Set B from Bouldin Alfalfa. This set was hit by a mower.
LI-190/R Q100421 2016-10-06 Twitchell Rice Removed Lab none Removed from the Golden Radiometer experiment at Twtichell Rice.
PQS 1 140456 2016-10-06 Twitchell Rice Removed Lab none Removed from the Golden Radiometer experiment at Twtichell Rice.
NR01 1720 2016-10-06 Lab none Installed Twitchell Rice Installed in the Golden Radiometer experiment at Twtichell Rice.
CR6 2113 2016-10-06 Lab none Installed Twitchell Rice Installed in the Golden Radiometer experiment at Twtichell Rice.
AM16-32A 1238 2016-10-06 Lab none Installed Twitchell Rice Installed in the Golden Radiometer experiment at Twtichell Rice.
CR6 2115 2016-10-06 Twitchell Rice Removed Lab none Removed from the Golden Radiometer experiment at Twtichell Rice.
PTB110 H0160069 2016-09-29 Lab none Installed East End Barometer Tested okay in the lab. Reinstalled to try again.
LI-7500A 1594 2016-09-29 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed on the Sherman Wetland tower to replace sn 75H-2668 for calibration.
LI-7500A 75H-2668 2016-09-29 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman Wetland tower for calibration. Replaced by sn 1594.
SD130BN 0030F4D214A3 2016-09-29 Lab none Installed East End Camera Installed on the East End tower replacing a Canon A series camera. A compute stick was also installed to collect the images.
NR01 2420 2016-09-22 Lab none Installed Twitchell Rice Installed on an independent tripod at Twitchell Rice. This is part of a "golden" radiometers experiment to compare incoming PAR and SW across the delta
CMP 11 060039 2016-09-22 Lab none Installed Twitchell Rice Installed on an independent tripod at Twitchell Rice. This is part of a "golden" radiometers experiment to compare incoming PAR and SW across the delta
LI-190/R Q100421 2016-09-22 Lab none Installed Twitchell Rice Installed on an independent tripod at Twitchell Rice. This is part of a "golden" radiometers experiment to compare incoming PAR and SW across the delta
PQS 1 140457 2016-09-22 Lab none Installed Twitchell Rice Installed on an independent tripod at Twitchell Rice. This is part of a "golden" radiometers experiment to compare incoming PAR and SW across the delta
PQS 1 140456 2016-09-22 Lab none Installed Twitchell Rice Installed on an independent tripod at Twitchell Rice. This is part of a "golden" radiometers experiment to compare incoming PAR and SW across the delta
CR6 2115 2016-09-22 Lab none Installed Twitchell Rice Installed on an independent tripod at Twitchell Rice. This is part of a "golden" radiometers experiment to compare incoming PAR and SW across the delta
LI-6400 PSC-0785 2007-02-20 Factory Calibration Factory Sent to Licor to check for flow restriction in chemical tubes, test PAR IN stability and calibrate light source. Findings: No CO2 source cover sent. In checking for blockages in the chemical tubes, initially there were none, so I had them rebuilt. In that process we discovered that the small tubes in the bypass valve had been pinched flat. Most likely, they had been left in either the scrub or bypass position for an extended period of time, causing them to be flattened-shut. We replaced them. It is a good idea to out the valve in the center, most relaxed, position for storage. The light source is stable and I could not get it to do anything unusual, even with lots of cable stressing/wiggling.
CR23X 2299 2007-07-24 Factory Returned Lab none Returned from repair at Campbell on RMA 4887: "Verified that it would not communicate when received. Unit had several low 12V errors and corrupt intermediate location which caused the communication problems. Most likely caused by lightning strike or repeated low supply problems. Updated Operating System to latest revision 19. calibrated with documentation, temperature tested and given a bench test."
CR10X 35474 2007-07-24 Factory Returned Lab none Returned from factory repair on RMA 4887: "Verified that it would not communicate when received. Unit had several low 12V errors and corrupt intermediate location which caused the communication problems. Most likely caused by lightning strike or repeated low supply problems. Updated Operating System to latest revision 22. calibrated with documentation, temperature tested and given a bench test."
USB2000-VIS-NIR USB2G4461 2007-07-16 Factory Returned Lab none Returned from factory repair
AM416 13961 2016-09-19 Lab none Loaned Loaned Loaned to Ron Amundson's lab for their ecosystem in a bottle experiment.
LI-7500A 75H-2514 2016-09-13 Lab none Installed East End Gas Analyzer CO2/H2O Installed at EE tower after calibration to replace sn 1594. Accidentally swapped the LI7500 at EE at this date, had not reached it's official swap period.
LI-7500A 1594 2016-09-13 East End Gas Analyzer CO2/H2O Removed Lab none Removed from EE tower for calibration. Replaced with 75H-2514. Accidentally swapped the LI7500 at EE at this date, had not reached it's official swap period.
PTB110 H0160069 2016-09-13 East End Barometer Removed Lab none Was reading low in the field again (around 98kPa) even though it was just returned from Campbell. Removed from EE tower for testing in the lab.
LI-7500A 75H-2150 2016-09-13 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond after calibration. Replacing 75H-2176.
LI-7500A 75H-2176 2016-09-13 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed Li7500 from West Pond tower for calibration. Replaced by 75H-2150.
GP-001 GP20160003 2016-09-13 Tonzi-Tower Removed Lab none Removed from Tonzi tower CO2 profile second lowest 1.8m location and channels 4-6 on the data logger
GP-001 GP20150084 2016-09-13 Tonzi-Tower Removed Lab none Removed from Tonzi tower CO2 profile lowest 0.35m location and channels 1-3 on the data logger
CR6 2115 2016-09-13 Tonzi-Tower Removed Lab none Removed from Tonzi tower were it collected data from Eosence CO2 profile experiment
LI-6262 IRG3-1079 2016-09-09 Lab none Calibration Lab none Calibrated in differential mode for double check of secondary standard. Ambient temperature was 24.37C, pressure was 100.54kPa. Using Zero air (400psi) and god bottle 421.22ppm CO2 (400psi).
Flowing zero air in both sample and reference
CO2 0.22ppm - adjusted to 0.06
H2O 0.25 - could not adjust - railed
Flowing zero air in reference and god bottle in sample
CO2 = 420.54ppm adjusted to 421.26ppm.
Flowing zero air in reference and secondary standard in sample, I recorded serial data with Hyperterm and took an average of 100 points. The new tank is 405.83ppm
I accidentally changed the H2O span. I reset to approximately the same value but did not check it with the dew point generator.
LI-7500A 75H-2180 2016-08-30 Lab none Calibration Lab none Last calibration: 2016-02-09
AGC=100.3 SDM=6 DAC1=- DAC2=-
Ambient P=100.13kPa, Tair=23.80C, Tdew=13.90C
Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.0
CO2 zero 0.9145 -> 0.9142 Offset: -1.25ppm
CO2 span 1.0203 -> 1.0227 Offset: -1.22ppm
H2O zero 0.8847 -> 0.8858 Offset: 7.0ppm
H2O span 1.0241 -> 1.0211 Offset: 0.06ppm
Notes:
LI-7500A 75H-2182 2016-08-30 Lab none Calibration Lab none Last calibration: 2016-04-11
AGC=87.48 SDM=6 DAC1=- DAC2=-
Ambient P=100.13kPa, Tair=22.36C, Tdew=13.81C
Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.0
CO2 zero 0.8832 -> 0.8831 Offset: -0.53ppm
CO2 span 1.0198 -> 1.0142 Offset: 2.98ppm
H2O zero 0.8466 -> 0.8464 Offset: -0.9ppm
H2O span 1.0138 -> 1.0230 Offset: -0.19ppm
Notes: Bandwidth setting changed from 5Hz to 10Hz
LI-7500A 1594 2016-08-30 Lab none Calibration Lab none Last calibration: 2016-02-22
AGC=101.4 SDM=6 DAC1=- DAC2=-
Ambient P=100.12kPa, Tair=24.79C, Tdew=13.72C
Zero Air at 400psi, God CO2 421.22 at 400 Tdewspan=16.0
CO2 zero 0.865 -> 0.8648 Offset: -0.70ppm
CO2 span 1.0196 -> 1.0164 Offset: 1.68ppm
H2O zero 0.8863 -> 0.8859 Offset: -2.6ppm
H2O span 1.0139 -> 1.0113 Offset: 0.05ppm
Notes:
CS547A 6097 2016-07-18 Factory Tested Factory Campbell says, "Tested and inspected sensor. Sensor show 'NAN' readings for cond and temp. Sensor is damaged internally. Sensor is beyond repair."
PTB110 H0160069 2016-08-15 Factory Calibration Factory Sent to Vaisala by Campbell for calibration. Campbell says, "Tested against reference. Inspected inside of sensor, no problems found. Successfully calibrated by Vaisala. Tested and confirmed proper function when received back."
PTB110 H0160069 2016-08-31 Lab none Installed East End Barometer Reinstalled on the East End tower after calibration at Vaisala.
CS451 20010046 2016-07-28 Factory Calibration Factory Campbell Sci "Tested unit, it has calcium on the element and needs [to be] cleaned. Cleaned and ran through the environmental calibration chamber. Unit meets specs"
CS450 20010770 2016-07-28 Factory Calibration Factory Campbell Sci "Tested unit and ran through he environmental calibration chamber. Unit meets specs"
LI-7500A 75H-2182 2016-08-31 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed on the Twitchell Alfalfa tower to replace sn 75H-2514 for calibration.
LI-7500A 75H-2514 2016-08-31 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell Alfalfa tower for calibration. Replaced by sn 75H-2182
LI-7500A 1594 2016-08-31 Lab none Installed East End Gas Analyzer CO2/H2O Installed on the East End tower to replace sn 75H-2150 for calibration.
LI-7500A 75H-2150 2016-08-31 East End Gas Analyzer CO2/H2O Removed Lab none Removed from the East End tower for calibration. Replaced by sn 1594.
LI-7500A 75H-2180 2016-08-31 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell Rice tower to replace sn 0065 for calibration.
LI-7500A 0065 2016-08-31 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell Rice tower for calibration. Replaced by sn 75H-2180.
WM 1590 153806 2016-08-31 Bouldin Alfalfa Anemometer Removed Lab none Removed from the Bouldin Alfalfa tower due to daytime spiking problem likely due to new firmware. Replaced by sn 124215
WM 1590 124215 2016-08-31 Lab none Installed Bouldin Alfalfa Anemometer Installed on the Bouldin Alfalfa tower to replace sn 153806 that has a spiking problem.
CS547A 6097 2016-08-30 Lab none Disposed Disposed Put in dead sensor box in lab.
CS547A 6097 2016-08-24 Factory Returned Lab none Returned from Campbell on RMA 24692 - unrepariable
PTB110 H0160069 2016-08-24 Factory Returned Returned Returned from Campbell on RMA 24692 - tested ok and calibrated by Vaisala.
CS451 20010046 2016-08-24 Factory Returned Lab none Returned from Campbell on RMA 24692 - cleaned, tested ok and calibrated.
CS450 20010770 2016-08-24 Factory Returned Lab none Returned from Campbell on RMA 24692 - tested ok and calibrated.
LI-7500A 0035 2016-08-23 Tonzi-Floor Removed Lab none Removed from the Floor Tower for calibration, replaced by sn 0976
LI-7500 0976 2016-08-23 Lab none Installed Tonzi-Floor Installed on the Floor Tower to replace sn 0035 for calibration.