Ser No.DateLoc FromTitleLoc ToNotes
PRB-K Type #8 000000573FAA 2010-07-07 Lab none Repair Factory Sent to ESI - Environmental Sensors Inc (www.esica.com)for evaluation and possible repair.
PR K Type 8 00000FED8688 2010-07-07 Lab none Repair Factory Sent to ESI - Environmental Sensors Inc (www.esica.com)for evaluation and possible repair.
PRB-K Type #8 000000A6D8356 2010-07-07 Lab none Repair Factory Sent to ESI - Environmental Sensors Inc (www.esica.com)for evaluation and possible repair.
PR K Type 8 00000FED7C67 2010-07-07 Lab none Repair Factory Sent to ESI - Environmental Sensors Inc (www.esica.com)for evaluation and possible repair.
GVSP30 DG002 2010-08-03 Lab none Repair Factory The pump was pickup for service at http://provac.com/
FMA RMT-200 08-0114 2010-08-06 Sherman Island Calibration Sherman Island Last calibrated June 2, 2010. Calibrated with 1.8978ppm CH4 God bottle. Before calibration: 1.908+-0.002ppm. After calibration: 1.898+-0.001ppm.
LI-6262 IRG3-1079 2010-08-04 Lab none Calibration Lab none I calibrated the LI-6262 with ultra pure N2, zero air and primary standard 370.91ppm CO2 (God bottle). I changed the chemicals for the chopper/reference path and let the 6262 run over night. The N2 and zero air gave very similar results with the zero air possibly being 1ppm CO2 and 14ppm H2O higher than the N2 tank. I filled a new scrub tube with Ascarite II and Drierite and used this to scrub the zero air. I spanned with the 370.91ppm CO2 God bottle. Then I checked a new tank of 400ppm CO2 that will go to Tonzi. This tank checked out at 384.08ppm CO2.
ESDP30A 983071037 2010-07-23 Factory Disposed Disposed I was notified by a technician (Paul) at http://www.provac.com/ that this pump it toast and I asked that they recycle it for us.
ISP-500C SV1864 2010-07-23 Factory Disposed Disposed I was notified by a technician (Paul) at http://www.provac.com/ that this pump it toast and I asked that they recycle it for us.
ESDP30A 973071056 2010-07-23 Factory Disposed Disposed I was notified by a technician (Paul) at http://www.provac.com/ that this pump it toast and I asked that they recycle it for us.
ESDP30A 973071056 2010-07-20 Lab none Repair Factory Sent to http://www.provac.com/ for evaluation and repair.
ESDP30A 973071056 2008-09-15 Lab none Found Lab none Found in parts when I started. Obvious damage to one of the scrolls.
FMA RMT-200 08-0114 2010-07-29 Lab none Installed Sherman Island Installed in the box at Sherman Island to replace the other FMA for mirror cleaning.
FMA RMT-200 05-0010 2009-08-06 Sherman Island Removed Lab none Removed from the box at Sherman Island for mirror cleaning because the ring time was 10.8usec.
FMA RMT-200 08-0114 2009-08-25 Lab none Installed Sherman Island Installed at Sherman Island after the LGR was removed to check problems caused by a blown fues. The scroll pump blew the system main fuse but I didn't know that and removed the LGR to check in the lab.
ISP-500C SV1864 2009-08-25 Sherman Island Removed Lab none This pump was removed from the box at sherman Island after blowing the fuse in the LGR and probably seizing. Lots or soot in both the intake and outlet. It had 9604.5 hours on it when removed and a note saying it had a minor rebuild at 1994.7 hours, May 2008.
ISP-500C SV1864 2010-07-20 Lab none Repair Factory This pump was sent to http://www.provac.com/ for evaluation.
GVSP30 DG002 2010-02-04 Twitchell Rice Removed Lab none This pump was removed from the box at Twitchell Island where it had been sitting idle.
ESDP30A 983071037 2010-03-11 Sherman Island Removed Lab none This pump was removed from the box at Sherman Island because of high pressure in the LGR FMA cell (>155Torr). The pump had 7322.2 hours on it when it was removed.
GVSP30 DG002 2010-03-11 Lab none Installed Sherman Island This pump was installed in the box at Sherman Island to run the LGR FMA there. The pump had 140 hours on it when it was installed.
GVSP30 DG002 2009-10-22 Lab none Installed Twitchell Rice This pump was installed in the box at Twitchell Island to run the LGR FGGA there.
ESDP30A 983071037 2010-07-20 Lab none Repair Factory This pump was sent to http://www.provac.com/ for evaluation.
GVSP30 DG002 2010-07-28 Sherman Island Removed Lab none This pump was removed from the Sherman Island LGR for maintenance. It has 3467 hours on it.
GVSP30 DG002 2010-07-29 Lab none Notes Lab none This pump has four serial numbers on it:
End cap: 043071002
Side Plate: DG002
Side Sticker: 044012498
Motor Plate: 84010040551
LI-7500A 0418 2010-07-01 Sherman Island-Levee Gas Analyzer CO2/H2O Removed Lab none Removed from leveen site due to theft of everything else.
PR K Type 8 00000FED7C67 2010-06-24 Sherman Island Removed Lab none This probe was removed from the Sherman Island cow pasture. It was giving bad readings.
PRB-K Type #8 000000A6D8356 2010-06-22 Sherman Island Removed Lab none This probe was removed from the Sherman Island cow pasture. It was outside the cow fence giving bad readings. Second pin from the top is broken.
PR K Type 8 00000FED8688 2010-06-22 Sherman Island Removed Lab none Removed from the Sherman Island cow pasture. This probe was outside the cow fence and giving bad readings.
PRB-K Type #8 000000573FAA 2010-06-24 Vaira Moisture Point Solar Removed Lab none Removed from Vaira between the solar panels and the diffuse radiation tripod. The top pin in the connector is broken.
AM416 14049 2010-06-24 Lab none Installed Tonzi This multiplexor was installed with two soil heat flux plates in the floor radiation system. The HMP45C lines were moved from the CR10X to the multiplexor to make room for the multiplexor.
HFP01 000141 2010-06-24 Lab none Installed Tonzi-Floor This sensor was installed near the open net radiation sensor in the floor site.
HFP01SC 00067 2010-06-24 Lab none Installed Tonzi-Floor This sensor was installed near the under tree radiation sensor in the floor site.
HFP01SC 00067 2010-06-15 Lab none Repair Lab none One of the wires was cut by animals. Spliced.
HFP01 000141 2010-06-15 Lab none Repair Lab none Wire is exposed at the juction to the plate - covered with liquid electric tape. The wire is spliced in several locations - cut and resplice to new wire.
HFP01SC 00068 1999-09-09 Factory Calibration Factory Initial factory calibration: 61.1 uV/W/m2
Heater resistance: 105.0 Ohm
Heater area: 0.003885 m2
HFP01SC 00067 1999-09-09 Factory Calibration Factory Initial factory calibration: 58.4 uV/W/m2
Heater resistance: 105.8 Ohm
Heater area: 0.003885 m2
HFP01 000144 1999-09-09 Factory Calibration Factory Initial factory calibration: 59.5 uV/W/m2
HFP01 000143 1999-09-09 Factory Calibration Factory Initial factory calibration: 58.6 uV/W/m2
HFP01 000142 1999-09-09 Factory Calibration Factory Initial factory calibration: 60.1 uV/W/m2
HFP01 000141 1999-09-09 Factory Calibration Factory Initial factory calibration: 59.0 uV/W/m2
HFP01 000140 1999-09-09 Factory Calibration Factory Initial factory calibration: 59.3 uV/W/m2
HFP01 000139 1999-09-09 Factory Calibration Factory Initial factory calibration: 59.3 uV/W/m2
HFP01 000138 1999-09-09 Factory Calibration Factory Initial factory calibration: 58.9 uV/W/m2
HFP01 000137 1999-09-09 Factory Calibration Factory Initial Factory calibration: 60.0 uV/W/m2
HFP01SC 00246 2001-05-25 Factory Calibration Factory Initial factory calibration: 61.2 uV/W/m2
Heater resistance: 99.4 Ohm
Heater area: 0.003885 m2
HFP01 002211 2007-05-09 Factory Calibration Factory Initial factory calibration: 63.1 uV/W/m2
HFP01SC 00244 2001-05-25 Factory Calibration Factory Initial factory calibration: 61.5 uV/W/m2
Heater resistance: 109.0 Ohm
Heater area is 0.003885 m2
HFP01 001604 2005-09-29 Factory Calibration Factory Initial Factory calibration 62.3 uV/W/m2
HFP01 001605 2005-09-29 Factory Calibration Factory Initial factory calibration 61.3 uV/W/m2
HFP01 001603 2005-09-29 Factory Calibration Factory Initial factory calibration 60.3 uV/W/m2
LI-7500A 0042 2010-06-03 Sherman Island-Levee Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman Levee tower for lab calibration.
LI-7500A 0418 2010-06-03 Lab none Installed Sherman Island-Levee Gas Analyzer CO2/H2O Installed on the Sherman Levee tower after fresh lab calibration.
LI-7500A 0418 2010-06-02 Lab none Calibration Lab none Last cal 2010-05-18, SMD 7 DAC1 H2O 0-1000 DAC2 CO2 10-25
AGC = 49%, T = 23.11C, P = 100.88kPa, Td = 12.44C
N2 = 0.0 400psi CO2 = 597.4ppm <300psi, Td = 9.7C
CO2 zero = 0.9011-> 0.9015, offset +1.44ppm
CO2 span = 1.0045-> 1.0022, offset +1.5ppm
H2O zero = 0.8524-> 0.8499, offset -0.262mmol/m3
H2O span = 1.0097-> 0.9921, offset +0.31C
FMA RMT-200 08-0114 2010-06-02 Lab none Calibration Lab none Last calibration Apr 6, 2010. Using god bottle 1.8978ppm CH4, <400psi.
Pre cal: 9:09, 21.32C, 142.33Torr, 16.998usec, 1Hz, 1.842+-0.002
Post cal: 1.897+-0.003

Then calibrated 2ppm Praxair tank 1700psi 4.982+-0.002ppm
LI-7500A 0041 2010-05-28 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on Sherman tower after fresh lab calibration.
LI-7500A 1594 2010-05-28 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on Twitchell tower after fresh lab calibration.
LI-7500A 0418 2010-05-28 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none This analyzer was removed from the Twitchell tower for calibration.
LI-7500A 0073 2010-05-28 Sherman Island Gas Analyzer CO2/H2O Removed Lab none This analyzer was removed from the Sherman tower for calibration.
LI-7500A 0035 2010-05-13 Lab none Installed Tonzi-Floor Installed of Tonzi floor tower after fresh calibration in the lab by Siyan.
LI-7500A 1594 2010-05-13 Tonzi-Floor Removed Lab none Swapped off Tonzi floor tower for calibration.
LI-7500A 0065 2010-05-13 Lab none Installed Tonzi-Tower Installed after fresh calibration in the lab.
LI-7500 0976 2010-05-13 Tonzi-Tower Removed Lab none Swapped off Tonzi tower for calibration.
LI-7500A 0041 2010-05-27 Lab none Calibration Lab none Last cal 2010-02-09, SMD 7 DAC1 H2O 0-1000 DAC2 CO2 10-25
AGC = 54%, T = 18.76C, P = 100.02kPa, Td = 15.4C
N2 = 0.0 500psi CO2 = 597.4ppm <300psi
CO2 zero = 0.9881-> 0.9882, offset +0.3ppm
CO2 span = 1.0022-> 0.9984, offset +3.1ppm
H2O zero = 0.7395-> 0.7385, offset -0.095mmol/m3
H2O span = 0.9928-> 0.9804, offset +0.35C
LI-7500A 1594 2010-05-27 Lab none Calibration Lab none Last cal 2010-02-16, SMD 7 DAC1 H2O 0-1000 DAC2 CO2 10-25
AGC = 51%, T = 18.78C, P = 100.3kPa, Td = 15.6C
N2 = 0.0 500psi CO2 = 597.4ppm <300psi
CO2 zero = 0.8645-> 0.8653, offset +3.2ppm
CO2 span = 1.0041 -> 1.0107, offset -5.3ppm
H2O zero = 0.8935 -> 0.8931, offset -0.032mmol/m3
H2O span = 0.9797 -> 0.9943, offset -0.25C
CR10X 25799 2010-03-21 Tonzi Removed Lab none This data logger was swapped out by Siyan from the floor supplemental radiation and temperature system. Final storage data is corrupt.
CR10X 25799 2010-03-16 Lab none Installed Tonzi This data logger was installed with new supplemental radiation and air temperature sensor inside the floor cow fense. Problems getting final storage data dumped - will have it swapped out by next visitor.
AM416 14049 2010-02-10 Tonzi Removed Lab none This mux was removed from Tonzi. It was used for sap flow just west of the floor tower. CR10X sn 38347 was removed with it.
CR10X 38347 2004-03-17 Factory Calibration Factory Factory calibration.
CR10X 38347 2010-02-10 Tonzi Removed Lab none This data logger was removed from Tonzi. It was used for sap flow just west of the floor tower. AM416 sn 14049 was removed with it. Wiring panel is 28250.
CR10X 25799 2010-04-23 Lab none Maintenance Lab none Installed a new internal 3V lithium battery. Duracell DL2430.
HMP45AC W0510026 2001-05-01 Factory Calibration Factory Last factory calibration listed on sticker.
HMP45AC W0510026 2010-03-16 Lab none Installed Tonzi Installed with radiation sensors for Hideki near the tram and floor eddy tower. This sensor is under the oak tree canopy.
PAR-LITE 010163 2010-03-16 Lab none Installed Tonzi Installed at the request of Hideki in an unshaded opening near the east end of the radiation tram. This sensor is pointed downward collecting reflected PAR.
PAR-LITE 010161 2010-03-16 Lab none Installed Tonzi Installed at the request of Hideki in an unshaded opening near the east end of the radiation tram at Tonzi. This sensor is pointing upward collecting incoming PAR.
NR-LITE 990350 2010-03-16 Lab none Installed Tonzi Installed at the request of Hideki under the oak canopy near the radiation tram at Tonzi.
NR-LITE 041556 2010-04-16 Lab none Installed Tonzi-Floor Installed at the request of Hideki in an unshaded opening near the east end of the radiation tram.
FMA RMT-200 05-0010 2010-04-07 Factory Repair Lab none Returned from LGR. LGR bill says pressure controller, filter, and check valve were replaced. Mirrors were cleaned. Running in slow flow mode analyzer is reporting 156.20 Torr and 18.587 usec ring time.
LI-7500A 0042 2010-04-05 Lab none Calibration Lab none Last calibration Dec 8, 2009
AGC=68, Pressure=100.491kPa Temperature=20.42C
CO2 span=597.4ppm, H2O span=17.6C, N2 650psi, CO2 300psi
Precal CO2 zero=0.9615, span=0.9897
Postcal CO2 zero=0.9620, span=0.9862
Precal H2O zero=0.7844, span=0.9773
Postcal H2O zero=0.7840, span=1.0151
CO2 zero off by +1.81ppm, span off by +2.6+-0.2ppm
H2O zero off by -0.05g/m3, span off by -0.77C
CR10X 20140 2010-03-19 Lab none Installed Tonzi Given to Siyan to install in the supplemental solar radiation measurements near the tram and floor system.
HMP45AC U3030032 2007-01-02 Factory Calibration Factory Calibrated at Campbell: Tested and inspected, replaced O-rings, calibrated
HMP45AC U3030032 2010-03-18 Sherman Island Removed Lab none Removed from the soil moisture CR10X at Sherman Island. Values were bad and duplicate measurment are on the tower. Also needed the solar radiation shield for Vaira.
LI-7500A 0042 2010-03-18 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none This LI7500 was removed from Twitchell tower for calibration.
LI-7500A 0035 2010-03-18 Sherman-2ndTower Removed Lab none This LI7500 was removed from the secondary tower at Sherman for calibration.
LI-7500A 0065 2010-03-18 Sherman Island Gas Analyzer CO2/H2O Removed Lab none This was LI7500 was removed from the Sherman tower for calibartion.
LI-7500A 0073 2010-03-18 Lab none Installed Sherman Island Gas Analyzer CO2/H2O This LI7500 was installed at Sherman. No problems.
LI-7500A 0418 2010-03-18 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O This LI7500 was installed at Twitchell Island. Its SDM address was changed to 7 to match CR1000 programs.
LI-7500A 0418 2010-03-18 Lab none Calibration Lab none Calibrated in the lab 100.70kPa, 18.36C, 49% AGC 13.9C Td
N2 tank 0.0 @ 700psi, CO2 tank 597.4ppm @ 200psi
Last cal 11-4-2009
CO2 zero 0.9020 -> 0.9011 offset -3.18ppm
CO2 span 1.0000 -> 1.0042 offset +3.5ppm
H2O zero 0.8486 -> 0.8524 offset +0.410g/m3
H2O span 0.9807 -> 1.0097 offset 0.45C

SDM address 3
LI-7500A 0073 2010-03-18 Lab none Calibration Lab none Calibrated in the lab 100.46kPa, 17.08C, 58% AGC 12.8C Td
N2 tank 0.0 @ 700psi, CO2 tank 597.4ppm @ 200psi
Last cal 11-4-2009
CO2 zero 0.9587 -> 0.9576 offset -6.4ppm
CO2 span 0.9973 -> 0.9960 offset +0.6ppm
H2O zero 0.7325 -> 0.7346 offset +0.238g/m3
H2O span 0.9865 -> 1.0007 offset -0.2C

SDM address 7
LI-7500A 0073 2010-03-17 Lab none Maintenance Lab none Embedded software upgraded from v3.0.0 to v3.0.1
Internal chemicals changed
LI-7500A 0418 2010-03-17 Lab none Maintenance Lab none Embedded software upgraded from v3.0.0 to v3.0.1
Internal chemicals changed
LI-7500A 0041 2009-04-02 Lab none Calibration Lab none AGC = 56%, T = 24.1C, P = 100.0kPa
N2 tank 1200psi, CO2 tank 1400psi 344.95ppm CO2
Last Calibration 1/7/2009

CO2 zero = 1.0024 -> 1.0026
CO2 span = 0.9868 -> 0.9872

H2O zero = 1.1031 -> 1.0377
H2O span = 0.7331 -> 0.7316
FMA RMT-200 05-0010 2009-03-08 Lab none Repair Factory Returned to factory for problem with pressure controller. Factory cleaned dirty pressure controller.
FMA RMT-200 05-0010 2009-08-12 Lab none Calibration Factory Calibrated with 1.898ppm CH4 source - possibly at factory.
FMA RMT-200 05-0010 2009-09-17 Lab none Installed Sherman Island Installed. Calibrated in field with 1.976ppm CH4 secondary standard. Pre-cal LGR read 1.933ppm at 19.58C, 156.38Torr, 18.178usec.
FMA RMT-200 08-0114 2009-09-17 Sherman Island Removed Lab none LGR #2 had low pressure (128Torr) in slow flow and dirty mirrors (10usec ring time). Removed for maintenance. Replaced by LGR #1 sn 05-0010
LI-7500A 1594 2009-12-04 Sherman Island Gas Analyzer CO2/H2O 2ndTower Removed Lab none This analyzer was removed for calibration from the secondary tower for the LGR trailer.
LI-7500A 0035 2009-12-04 Lab none Installed Sherman-2ndTower Freshly calibrated. Installed on secondary tower for LGR trailer.
LI-7500A 0065 2009-12-04 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Freshly calibrated. Installed on permanent tower.
LI-7500A 0042 2009-12-04 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from permanent tower for calibration.
LI-7500A 0042 2009-09-09 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Freshly calibrated, installed.