Equipment and Calibration History

View Equipment |

View history entry 1532 for Licor LI-7500A sn: 75H-2182

Date: 2014-08-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-02-28 AGC=83.4 SDM=1 DAC1=- DAC2=- Ambient P=99.77kPa, Tair=23.79C, Tdew=12.26C Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25 CO2 zero 0.8815 -> 0.8816 Offset: 0.90ppm CO2 span 1.0064 -> 1.0074 Offset: -0.54ppm H2O zero 0.8476 -> 0.8463 Offset: -7.7ppm H2O span 0.9875 -> 1.0039 Offset: -0.33ppm Notes:
Entered by: