Equipment and Calibration History
Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 [27] 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65Ser No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
CR10X 27930 | 2019-08-16 | Lab none | Installed | Bouldin Alfalfa-ReleaseEx Data Logger Near | Installed for CH4 release experiment. This unit was part of the closer (day time) release point. |
PAR-LITE 010161 | 2019-06-18 | Lab none | Calibration | Lab none | New calibration factor of 5.32 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
PAR-LITE 010161 | 2001-07-26 | Factory | Calibration | Lab none | Initial factory calibration 5.27uV/umol/s.m2 |
PAR-LITE 10162 | 2019-06-18 | Lab none | Calibration | Lab none | New calibration factor of 5.76 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
PAR-LITE 10162 | 2001-07-26 | Factory | Calibration | Lab none | Initial factory calibration 5.27uV/umol/s.m2 |
PQS 1 140455 | 2019-08-27 | Lab none | Calibration | Lab none | New calibration factor of 8.30 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2019" for details. |
PAR-LITE 040412 | 2004-05-29 | Factory | Calibration | Lab none | Initial factory calibration 5.58 uV/(umol/m2/s) |
PAR-LITE 030380 | 2003-12-17 | Factory | Calibration | Lab none | Initial factory calibration 5.31 uV/(umol/m2/s) |
PAR-LITE 030380 | 2019-08-16 | Lab none | Calibration | Lab none | New calibration factor of 5.26 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
PAR-LITE 030381 | 2019-06-18 | Lab none | Calibration | Lab none | New calibration factor of 5.43 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
PAR-LITE 030381 | 2003-12-17 | Factory | Calibration | Lab none | Initial factory calibration 5.32 uV/(umol/m2/s) |
BB-Fab BB-TZ_soilco2 | 2019-07-16 | Tonzi | Maintenance | Tonzi | There are six batteries in three pairs for a 24V system – the original four plus two more we added last winter. Two of the originals test a little weak with the load tester so I disconnected them from the bank. |
BB-Fab BB-TZ_soilco2 | 2019-08-22 | Tonzi | Upgrade | Tonzi | Six old weak batteries (four currently connected) were brought back to the lab and four new Duracell deep cycle flooded group 27 batteries were installed in pairs for a 24V system. The new batteries have 09/19 (Sept/2019) stickers. |
LI-7500A 0418 | 2019-08-22 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed on Vaira to replace sn 75H-2668 for calibration. |
LI-7500A 75H-2668 | 2019-08-22 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0418. |
LI-7500A 75H-2180 | 2019-08-22 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed on Tonzi Floor to replace sn 75H-2514 for calibration. |
LI-7500A 75H-2514 | 2019-08-22 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi Floor for calibration. Replaced by sn 75H-2180. |
LI-7500A 0065 | 2019-08-22 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed on Tonzi Tower to replace sn 1594 for calibration. |
LI-7500A 1594 | 2019-08-22 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi Tower for calibration. Replaced by sn 0065. |
LI-7500A 0065 | 2019-08-21 | Lab none | Calibration | Lab none | Last calibration: 2019-04-15 AGC=100 SDM=0 DAC1=- DAC2=- Ambient P=100.3kPa, Tair=27.6C, Tdew=15.0C Zero Air at 1000psi, AMP CO2 408.97 at 800 Tdewspan=20.0 CO2 zero 0.8911 -> 0.8916 Offset: 2.25ppm CO2 span 1.0171 -> 1.0191 Offset: -0.97ppm H2O zero 0.726 -> 0.7278 Offset: 14.2ppm H2O span 1.0118 -> 1.0165 Offset: -0.1ppm Notes: Water zero offset a little large - maybe still acclimating from factory service and new internal chemicals. Signal Strength reset from 101.1 to 100.0 |
LI-7500A 75H-2180 | 2019-08-21 | Lab none | Calibration | Lab none | Last calibration: 2019-04-23
AGC=100.8 SDM=0 DAC1=- DAC2=- Ambient P=100.3kPa, Tair=27.83C, Tdew=15.35C Zero Air at 1000psi, AMP CO2 408.97 at 800 Tdewspan=20.0 CO2 zero 0.9136 -> 0.9137 Offset: 0.45ppm CO2 span 1.0266 -> 1.0266 Offset: -0.02ppm H2O zero 0.8886 -> 0.8884 Offset: -1.3ppm H2O span 1.0236 -> 1.0282 Offset: -0.1ppm Notes: |
LI-7500A 0418 | 2019-08-21 | Lab none | Calibration | Lab none | Last calibration: 2019-04-22 AGC=93.7 SDM=0 DAC1=- DAC2=- Ambient P=100.3kPa, Tair=26.98C, Tdew=14.83C Zero Air at 1000psi, AMP CO2 408.97 at 800 Tdewspan=20.0 CO2 zero 0.8992 -> 0.8990 Offset: -0.55ppm CO2 span 1.0011 -> 0.9976 Offset: 1.71ppm H2O zero 0.8464 -> 0.8466 Offset: -0.8ppm H2O span 0.9866 -> 0.9818 Offset: 0.11ppm Notes: |
NR-LITE 041556 | 2019-08-26 | Lab none | Calibration | Lab none | Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. I calculated a calibration factor of 12.8 uV/(W/m2), but there was a big difference from the calibration factor when the sensor was tested upside down (14.3 uV/(W/m2)), so more testing needed. |
NR-LITE 990353 | 2019-06-18 | Lab none | Calibration | Lab none | New calibration factor of 12.0 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
NR-LITE 990351 | 1999-07-29 | Factory | Calibration | Lab none | Initial factory calibration: 13.9 uV/(W/m2) |
NR-LITE 021029 | 2019-06-18 | Lab none | Calibration | Lab none | New calibration factor of 13.1 uV/(W/m2). Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
NR-LITE 990352 | 2019-08-16 | Lab none | Calibration | Lab none | New calibration factor of 13.7 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
NR-LITE 990351 | 2019-06-18 | Lab none | Calibration | Lab none | New calibration factor of 14.0 uV/(W/m2). Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
NR-LITE 990352 | 1999-07-29 | Factory | Calibration | Lab none | Initial factory calibration 14.2uV/(W/m2) |
NR-LITE 990354 | 1999-07-29 | Factory | Calibration | Lab none | Initial factory calibration 12.9 uV/(W/m2) 1999-07-29 |
GP-001 GP20140026 | 2019-08-19 | West Pond CO2 Probe B | Removed | Lab none | Removed from WP and brought back to the lab for testing out (it was maxing out at 2% even in the air). |
NR-LITE 990354 | 2019-06-18 | Lab none | Calibration | Lab none | Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. I calculated a calibration factor of 12.7 uV/(W/m2), but more testing needs to be done because there was a large calibration difference when testing the sensor upside-down (15.4 uV/(W/m2)). |
LI-7500A 0073 | 2019-08-14 | Lab none | Calibration | Lab none | Last calibration: 2019-04-23 AGC=98.7 SDM= DAC1= DAC2= Ambient P=100.2kPa, Tair=29.6C, Tdew=13.2C Zero Air at 1000psi, AMP CA4396 408.97 at 850 Tdewspan=16.00 CO2 zero 0.9563 -> .9576 Offset: 4.5ppm CO2 span 1.0094 -> 1.0095 Offset: -.13ppm H2O zero 0.7336 -> .7318 Offset: -10.5ppm H2O span 1.0153 -> 1.0247 Offset: -.5ppm Notes: When the SS was 98.7, I reset it to 100. The previous calibration on 2019-04-23, the H2O zero was offset by 12.1 mmol/m3, and this time the offset was -10.5 mmol/m3. |
LI-7500A 0035 | 2019-08-07 | Lab none | Installed | East Pond Gas Analyzer CO2/H2O | Installed at EP to replace sn 0065, which was brought back to the lab for calibration. |
LI-7500A 0065 | 2019-08-07 | East Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from EP and brought back to the lab for calibration. It was replaced by sn 0035. |
U24-002-C 11020258 | 2019-08-07 | East Pond Conductivity Sensor | Removed | Lab none | Removed from EP and brought back to the lab for testing. I haven't been able to connect to it through the HOBOware software for a while. |
LI-7500A 75H-2182 | 2019-08-07 | Lab none | Installed | West Pond Gas Analyzer CO2/H2O | Installed at WP to replace sn 0418, which was brought back to the lab for calibration. |
LI-7500A 0418 | 2019-08-07 | West Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from WP and brought back to the lab for calibration. It was replaced by sn 75H-2182. |
LI-7500A 75H-2150 | 2019-08-07 | Lab none | Installed | Bouldin Alfalfa Gas Analyzer CO2/H2O | Installed at BA to replace sn 75H-2180, which was brought back to the lab for calibration. |
LI-7500A 75H-2180 | 2019-08-07 | Bouldin Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | Removed from BA and brought back to the lab for calibration. It was replaced by sn 75H-2150. |
LI-7500A 75H-2669 | 2019-08-07 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Installed at BC to replace sn 0073, which was brought back to the lab for calibration. |
LI-7500A 0073 | 2019-08-07 | Bouldin Corn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from BC and brought back to the lab for calibration. It was replaced by sn 75H-2669. |
GP-001 GP20150101 | 2019-08-05 | Lab none | Tested | Lab none | This probe is bad - reading 5000 mV consistently. Tested in the lab against 2 new 1% probes in Joe's test chamber (blew into chamber and let it come down to ambient). |
LI-7500A 75H-2182 | 2019-08-01 | Lab none | Calibration | Lab none | Last calibration: 2019-04-15
AGC=87.5 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=23.9C, Tdew=13.6C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.8842 -> .8846 Offset: 1.90ppm CO2 span 1.0156 -> 1.0226 Offset: 3.54ppm H2O zero 0.8459 -> .8456 Offset: -1.6ppm H2O span 1.0189 -> 1.0214 Offset: -.05ppm Notes: When SS was 87.5, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15. |
LI-7500A 75H-2150 | 2019-08-01 | Lab none | Calibration | Lab none | Last calibration: 2019-04-15
AGC=97.7 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=23.4C, Tdew=13.3C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9014 -> .9022 Offset: 3.41ppm CO2 span 1.003 -> 1.0086 Offset: -2.05ppm H2O zero 0.8671 -> .8673 Offset: 1.2ppm H2O span 1.0077 -> 10.114 Offset: -.08ppm Notes: When SS was 97.7, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15. |
LI-7500A 75H-2669 | 2019-08-01 | Lab none | Calibration | Lab none | Last calibration: 2019-04-15
AGC=102.6 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=22.7C, Tdew=14.1C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9186 -> .9178 Offset: -3.12ppm CO2 span 1.0134 -> 1.0125 Offset: .53ppm H2O zero 0.8991 -> .9019 Offset: 17.8ppm H2O span 1.0017 -> 1.015 Offset: -.3ppm Notes: When SS was 102.6, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15. |
LI-7500A 0035 | 2019-08-01 | Lab none | Calibration | Lab none | Last calibration: 2019-04-26
AGC=99.2 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=24.0C, Tdew=13.8C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9371 -> .9374 Offset: 1.23ppm CO2 span 1.0189 -> 1.0207 Offset: -.96ppm H2O zero 0.8989 -> .8984 Offset: -2.9ppm H2O span 1.0186 -> 1.0323 Offset: -.29ppm Notes: When the SS was 99.2, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15. |
LI-7500A 0041 | 2019-07-25 | Lab none | Installed | Sherman Wetland Gas Analyzer CO2/H2O | Installed at SW to replace sn 75H-2150, which was brought back to the lab for calibration. |
LI-7500A 75H-2150 | 2019-07-25 | Sherman Wetland Gas Analyzer CO2/H2O | Removed | Lab none | Removed from SW and brought back to the lab for calibration. It was replaced by sn 0041. |
LI-7500A 75H-2667 | 2019-07-25 | Lab none | Installed | Sherman Barn Gas Analyzer CO2/H2O | Installed at SB to replace sn 75H-2669, which was brought back to the lab for calibration. |
LI-7500A 75H-2669 | 2019-07-25 | Sherman Barn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from SB and brought back to the lab for calibration. it was replaced by sn 75H-2667. |
LI-7500A 0042 | 2019-07-25 | Lab none | Installed | Mayberry Gas Analyzer CO2/H2O | Installed at MB to replace 75H-2182, which was brought back to the lab for calibration. |
LI-7500A 75H-2182 | 2019-07-25 | Mayberry Gas Analyzer CO2/H2O | Removed | Lab none | Removed from MB and brought back to the lab for calibration. It was replaced by sn 0042. |
PAR-LITE 010161 | 2019-07-25 | Lab none | Installed | Mayberry PAR Sensor Reflected Water | Installed at MB (PAR reflected water) to replace sn 040412, which was brought back to the lab for testing. |
PAR-LITE 040412 | 2019-07-25 | Mayberry PAR Sensor Reflected Water | Removed | Lab none | Removed from MB (PAR reflected water) and brought back to the lab for testing (spiky). It was replaced by sn 010161. |
NR-LITE 990351 | 2019-07-25 | Lab none | Installed | Mayberry Net Radiometer Water | Installed at MB over the water to replace sn 990350, which was brought back to the lab for repair. |
NR-LITE 990350 | 2019-07-25 | Mayberry Net Radiometer Water | Removed | Lab none | Removed from MB (Rnet_water) and brought back to the lab for repair (broken cable). It was replaced by sn 990351. |
LI-7500A 75H-2176 | 2019-07-25 | Lab none | Installed | East End Gas Analyzer CO2/H2O | Installed at EE to replace sn 0035, which was brought back to the lab for calibration. |
LI-7500A 0035 | 2019-07-25 | East End Gas Analyzer CO2/H2O | Removed | Lab none | Removed from EE and brought back to the lab for calibration. it was replaced by 75H-2176. |
PAR-LITE 10162 | 2019-07-25 | Lab none | Installed | East End PAR Sensor Incoming | Installed at EE to replace sn 110340, which was removed for spiky behavior. |
PQS 1 110340 | 2019-07-25 | East End PAR Sensor Incoming | Removed | Lab none | Removed from EE because it had been spiky in the last month. It was replaced sn 10162. |
LI-7500A 75H-2667 | 2019-07-24 | Lab none | Calibration | Lab none | Last calibration: 2019-02-11 AGC=104.1 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=26.7C, Tdew=11.7C Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9425 -> .9413 Offset: -4.47ppm CO2 span 1.0143 -> 1.0158 Offset: -.77ppm H2O zero 0.8349 -> .8363 Offset: 8.7ppm H2O span 1.0222 -> 1.0222 Offset: -.02ppm Notes: When the SS was 104.1, I reset it to 100.0 |
LI-7500A 0042 | 2019-07-24 | Lab none | Calibration | Lab none | Last calibration: 2019-01-02
AGC=97.7 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=27.3C, Tdew=14.4C Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9489 -> .9481 Offset: -3.43ppm CO2 span 1.0187 -> 1.0049 Offset: 7.19ppm H2O zero 0.8434 -> .8433 Offset: -1.0ppm H2O span 1.0334 -> 1.0179 Offset: .35ppm Notes: When SS was 97.7, I reset it to 100.0 |
LI-7500A 0041 | 2019-07-24 | Lab none | Calibration | Lab none | Last calibration: 2019-01-02
AGC=97.8 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=27.2C, Tdew=12.7C Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9953 -> .9970 Offset: 5.75ppm CO2 span 1.0194 -> 1.0163 Offset: -1.89ppm H2O zero 0.8248 -> .8201 Offset: -26.5ppm H2O span 1.0312 -> 1.0243 Offset: .16ppm Notes: When SS was 97.8, I reset it to 100.0. Larger H2O zero offset than I expected, but H2O span was ok. Last calibration, the H2O zero offset was a similar magnitude and positive. |
LI-7500A 75H-2176 | 2019-07-24 | Lab none | Calibration | Lab none | Last calibration: 2019-01-02
AGC=102.0 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=26.1C, Tdew=14.5C Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.8676 -> .8672 Offset: -2.14ppm CO2 span 1.0114 -> 1.0061 Offset: 2.7ppm H2O zero 0.8525 -> .8521 Offset: -2.4ppm H2O span 0.8206 -> .9182 Offset: .05ppm Notes: When SS was 102.0, I reset it to 100. |
GP-001 GP20150101 | 2019-07-10 | East End CO2 probe | Removed | Lab none | Removed from EE and brought back to the lab for further testing (it was maxing out, even when out of the water). |
GP-001 GP20150101 | 2019-01-18 | Lab none | Installed | East End CO2 probe | Installed at EE to replace sn GP20140019, which was brought back to the lab for testing. |
LI-7500 0976 | 2009-01-08 | Lab none | Installed | Tonzi-Floor | Installed on the Tonzi Floor to replace sn 0042 for calibration |
LI-7500A 0042 | 2009-01-08 | Tonzi-Floor | Removed | Lab none | Removed from the Tonzi Floor for calibration. Replaced by sn 0976. |
LI-7500A 0418 | 2009-01-08 | Lab Gas Analyzer CO2/H2O | Installed | Tonzi-Floor | Installed on the Tonzi Tower to replace sn 0065 for calibration |
LI-7500A 0065 | 2009-01-08 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab Gas Analyzer CO2/H2O | Removed from the Tonzi Tower for calibration. Replaced by sn 0418. |
LI-7500A 0041 | 2009-01-08 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed on the Vaira tower to replace sn 0073 for calibration |
LI-7500A 0073 | 2009-01-08 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from the Vaira Tower for calibration. Replaced by sn 0041. |
Mark I A001911 | 2019-07-02 | Lab none | Installed | Sherman Wetland Weather Station | Installed at the midpoint of the western railing. Should be remounted on the radiation boom. |
Mark I A000251 | 2019-07-02 | Lab none | Installed | Tonzi | Installed at the top of the tower east of the southeast corner. Should be remounted on the radiation boom. |
URM06 URM06-001 | 2019-06-20 | Bouldin Alfalfa Range Finder | Moved | Bouldin Corn Range Finder | Moved from BA to BC to look at corn height. It was installed on the BC radiation boom. |
miniDOT 1220 | 2019-06-13 | Lab none | Tested | Lab none | Tested but we probably can't use this sensor any more. We put in new AA lithium batteries, and the miniDOT was able to power up, but the LED light continuously blinks 2x, "card error." We tried re-formatting the card, using another 2GB card previously used in a miniDOT (could have been a bad card), and then with a good SD card at EE. In all cases, the LED status light still indicated a "card error." |
miniDOT 1220 | 2019-05-23 | Mayberry DO Sensor E | Removed | Lab none | Brought back to the lab for testing. The o-ring was warped and it was damp inside. There was some corrosion on the board and the batteries were dead. |
PTB110 E1330045 | 2019-05-23 | Mayberry Barrometer | Removed | Lab none | Reading dropped down to 50, so removed and brought back to the lab for inspection. |
miniDOT 1220 | 2018-11-28 | Mayberry DO Sensor E | Maintenance | Mayberry DO Sensor E | Swapped 2x AA lithium batteries |
miniDOT 1221 | 2018-08-23 | Mayberry DO Sensor D | Maintenance | Mayberry DO Sensor D | Swapped 2x AA lithium batteries |
2300-000 2300565 | 2019-06-04 | Lab none | Installed | Mayberry Fluorometer | Installed at MB after a thorough cleaning in the lab. |
2300-000 2300565 | 2019-02-28 | Lab none | Installed | Mayberry Fluorometer | Reinstalled after a thorough cleaning and getting new brushes in the lab. |
2300-000 2300565 | 2019-05-31 | Mayberry Fluorometer | Removed | Lab none | Removed from the tower for a thorough cleaning in the lab. |
SI-131 7385 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near the Floor solar panels looking at the grass to the north. Footprint of about 1.23m x 2.5m diameter |
TS01 TS01-002 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near an oak tree about 25m north of the Floor solar panels. Measuring trunk temperature from about 110cm distance. |
SI-131 7638 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near an oak tree about 25m north of the Floor solar panels. Measuring trunk temperature from about 40cm distance. |
CR10X 47772 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near an oak tree about 25m north of the Floor solar panels to support two IRTs |
TS01 TS01-001 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near the Floor solar panels looking at the grass to the north. Footprint of about 0.7m x 0.35m diameter |
LI-7550 AIU-1866 | 2019-05-31 | Factory | Received | Lab none | Received back from Licor. They replaced the RF chip in the power supply. |
HMP45AC T2130005 | 2019-05-22 | Lab none | Tested | Lab none | I removed the power cord and soldered on an extension cord. Tested in the lab with radiation shields and fans, and compared well with two HMP60s sn L1740398 (Campbell has OK'd this one) and sn G4540023. This sensor is probably ok, but maybe we can test it with a reference AirT/RH sensor from AMP. |
CS451 70011234 | 2018-12-10 | Lab none | Tested | Lab none | Tested in both air and water in the lab and saw an offset of about +0.975 cm. See RT report "Pressure transducer comparison 12-2018." |
CS451 70011229 | 2018-12-10 | Lab none | Tested | Lab none | Tested in both air and water in the lab, but saw diurnal fluctuations of only 0.1 cm, not the 5-6 cm we saw in the field. See RT report "Pressure transducer comparison 12-2018." |
CR1000 47108 | 2019-05-28 | Sherman Barn Data Logger Met | Notes | Sherman Barn Data Logger Met | VX2 may be unreliable. NR01 PRT was drifting constantly until it was re-wired from VX2 to VX3, but that could also have been a wiring issue with another channel. |
AM16/32B 11662 | 2019-05-28 | Sherman Barn Multiplexer Met | Notes | Sherman Barn Multiplexer Met | Channel 28 and 32 are probably bad (see field notes from SB 2019-04-11). |
LI-7550 AIU-1866 | 2019-05-09 | Lab none | Shipped | Factory | Shipped back to Licor for them to inspect. |
LI-7500A 75H-2514 | 2019-05-07 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed on the Floor tower to replace the one that was moved to Tonzi Tower last week. |
LI-7550 AIU-0666 | 2019-05-07 | Lab none | Installed | Tonzi-Floor Interface LI-7500 | Installed on the Floor tower to replace the one that was moved to Tonzi Tower last week. |
LI-7500A 0035 | 2019-05-02 | Lab none | Installed | East End Gas Analyzer CO2/H2O | Installed at EE to replace sn 75H-2667, which was brought back to the lab for calibration. |
LI-7500A 75H-2667 | 2019-05-02 | East End Gas Analyzer CO2/H2O | Removed | Lab none | Removed from EE and brought back to the lab for calibration. it was replaced by sn 0035. |