Instrument:Narrow Band
Manufacture:Decagon-METER
Model No:SRS-Pi PRI
Serial No:759503109
Tag No:NULL
Acquisition Date:2014-08-19
Project:DOE-TCP
Location on 2014-09-15:Tonzi Tower, NULL
Notes: 180 degree cosine filters

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
1557 2014-09-11 Lab none Installed Tonzi-Tower I installed this sensor on the top of the Tonzi Tower on a post near the pivot of the radiation boom. This sensor is measuring incoming radiation.
1556 2014-09-09 Lab none Maintenance Lab none I updated the SDI-12 address from the default 0 (zero) to 1 (one) using a CR1000 in SDI-12 transparent mode. I was not able to do this with a 23x.
1555 2014-08-21 Lab none Tested Lab none I tried to install this sensor on the Tonzi Tower. This is a SDI-12 sensor and the 23x was having trouble reading it. This might be due to a default SDI-12 address of 0 (zero). I could not change the address with the 23x terminal in SDI-12 transparent mode, so I brought it back to the lab.
 
 