Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP221
Serial No:Y2009100
Tag No:
Acquisition Date:0000-00-00
Project:Tonzi
Location on 2022-10-05:Lab,
Notes: 0-3% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6026 2022-10-05 Tonzi-SoilCO2 CO2 Probe Tree1 Profile -0.16 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Tree1 -16cm
3061 2017-09-14 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.16 Installed at Tonzi in the Tree 1 profile at -16cm after lab calibration. This is a new installation with new holes dug.
1879 2015-06-09 Twitchell Alfalfa Removed Lab none Removed from Set B at -2cm with upgrade of soil CO2 system
1527 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in profile B at 2cm to replace sn F4960019 due to zero reading.
1463 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.233 + -902.3; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
1454 2014-04-08 Lab none Calibration Lab none Trueppm = Probeppm*1.047 + -347.2; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
2021 2014-03-18 Vaira CO2 probe East Profile -0.08 Removed Lab none Removed from Vaira in the East profile due to malfunction.
2001 2014-01-29 Lab none Installed Vaira CO2 probe East Profile -0.08 East profile as -8cm with new U-shaped tops on the PVC enclosures.
1432 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*1.031 + -299.9; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
1194 2014-01-24 Vaira CO2 probe East Profile -0.08 Removed Lab none Removed from Vaira east plot at -8cm for lab calibration
1128 2012-03-21 Lab none Installed Vaira CO2 probe East Profile -0.08 Installed at Vaira in the east set at -8cm
1412 2012-02-06 Lab none Calibration Lab none Trueppm = Probeppm*1.036 + -68.6; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan.
1403 2012-02-03 Lab none Calibration Lab none Trueppm = Probeppm*1.003 + -355.6; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.6C, P=101.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan.
1293 2010-04-06 Lab none Installed Sherman Island Installed at Sherman Island in Set B at -8cm.
1358 2010-02-03 Lab none Calibration Lab none Trueppm = Probeppm*1.051 + -207; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=24.1C, P=101.2kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn.
11 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-3% - certificate on file.
Problem: Got probe error 05 and 11. No IR source and memory error.
Action: Replaced probe circuit board and adjusted unit.
 
 