Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP221
Serial No:K1340006
Tag No:
Acquisition Date:2014-04-03
Project:DOE
Location on 2022-10-05:Lab,
Notes: 0-3% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6029 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.08 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -8cm
3056 2017-09-14 Lab none Installed Tonzi CO2 Probe Open Profile -0.08 Installed at Tonzi in the Open profile at -8cm after lab calibration. This is a new installation with new holes dug.
3011 2017-08-25 Tonzi CO2 Probe Open Profile -0.08 Removed Lab none Removed from Tonzi Open profile at -8 cm position for calibration
2190 2016-02-23 Lab none Installed Tonzi CO2 Probe Open Profile -0.08 Installed in the Tonzi Open profile at -8cm after lab calibration.
1870 2014-09-08 Twitchell Alfalfa Removed Lab none Removed from Alfalfa Set B -8cm because of low readings. Replaced by F4960018
1529 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in profile B at 8cm to replace sn F4960018 due to zero reading.
2093 2014-06-17 Sherman Island Removed Sherman Island Removed from Sherman profile A at -16cm due to low readings. Replaced by sn G0150037
2092 2014-04-10 Lab none Installed Sherman Island Installed in Sherman profile A at -16cm to replace sn G0150035
1453 2014-04-08 Lab none Calibration Lab none Trueppm = Probeppm*1.002 + -37.28; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
 
 

Login