Equipment and Calibration History

Page: 1 2 3 4 5 6 [7] 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65
Ser No.DateLoc FromTitleLoc ToNotes
LI-7500A 75H-2176 2022-12-05 Lab Calibration Lab Last calibration: 2022-08-23
AGC=101.0 SDM=0 DAC1=- DAC2=-
Ambient P=100.6kPa, Tair=20.5C, Tdew=9.85C
Zero Air at 1700psi, AMP CA04119 524.77 at 1325 Tdewspan=16.00
CO2 zero 0.8678 -> 0.8678 Offset: 0.0ppm
CO2 span 1.0008 -> 1.0144 Offset: -9.57ppm
H2O zero 0.8537 -> 0.8553 Offset: 9.8ppm
H2O span 0.923 -> 0.9281 Offset: -0.12ppm
Notes: Large offset in CO2 span. But this largely undoes some similar previous increases. Calibrated at the same time sn 2180 showed a similar CO2 span offset. This calibration was using 7550 box 0674 instead of the AMP loaner but I don't see how that should effect only CO2 span. Going to go with it.
LI-7500A 0041 2022-11-28 Lab Calibration Lab Last calibration: 2022-11-15
AGC=107 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=22.4C, Tdew=2.2C
Zero Air at 1750psi, AMP CA04119 524.77 at 1300 Tdewspan=16.00
CO2 zero 0.9998 -> .9968 Offset: -9.82ppm
CO2 span 1.0178 -> 1.0199 Offset: -1.37ppm
H2O zero 0.8213 -> .8221 Offset: 4.8ppm
H2O span 1.0308 -> 1.0304 Offset: 0.00ppm
Notes: SS rest from 107 to 100 after calibration. Today's calibration was with our usual box sn AIU-0674. The previous calibration was on a AMP box sn AIU-1039 and I had reset the SS from 93 to 100. Probably some difference in box coefficients that I didn't pay attention to, but the SS should be ok now.
LI-7500A 1594 2022-11-28 Lab Calibration Lab Last calibration: 2022-04-08
AGC=98.3 SDM= DAC1= DAC2=
Ambient P=100.3kPa, Tair=20.9C, Tdew=4.05C
Zero Air at 1800psi, AMP CA04119 524.77 at 1350 Tdewspan=16.00
CO2 zero 0.8643 -> .8643 Offset: -.38ppm
CO2 span 1.0097 -> 1.0168 Offset: -4.93ppm
H2O zero 0.8891 -> .8889 Offset: -1.1ppm
H2O span 1.0106 -> 1.0136 Offset: -.06ppm
Notes: Set H2O span on 2022-11-29 using 16.00C. Ambient T=22.4 and Dew Point = 0.77C.
BB-Fab BB-BA-bncr 2022-11-02 Bouldin Alfalfa-ADEX Battery Bank Bouncer Removed Lab none Removed with decommissioning of advection bouncer tower and brought back to the lab. This battery bank will cease to exist and these new-ish batteries will be moved to Hill Slough to replace old batteries there.
BB-Fab BB-BA-edge 2022-11-02 Bouldin Alfalfa-ADEX Battery Bank Edge Removed Lab none Removed with decommissioning of advection edge tower and brought back to the lab. This battery bank will cease to exist and these new-ish batteries will be moved to Hill Slough to replace old batteries there.
BB-Fab BB-BA-edge 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Battery Bank Edge Two brand-new deep cycle flooded batteries installed to power the advection edge tower
BB-Fab BB-BA-bncr 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Battery Bank Bouncer Two brand-new deep cycle flooded batteries installed to power the advection bouncer tower
BB-Fab BB-HS 2022-07-13 Hill Slough Battery Bank Maintenance Hill Slough Battery Bank We’ve been losing data overnight since mid-June, which is suspicious because the days are long and sunny. 2 batteries tested bad, 2 tested good, and 1 tested good but closer to weak. I removed the bad batteries from the system and wired up the 3 good ones. The good ones all needed water, about 3-4L each. We refilled them with DI water. I cleaned and sprayed all of the terminals.
BB-Fab BB-HS 2022-11-30 Hill Slough Battery Bank Maintenance Hill Slough Battery Bank Removed 3 older, flooded deep cycle batteries. Added 4 new-ish Duracell flooded, deep cycle batteries with 04/22 tags. We purchased these batteries to run the BA advection experiment April-Oct 2022.
HygroVUE10 E3514 2022-11-30 Bouldin Alfalfa-ADEX RH & Temp Main Lower Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
HygroVUE10 E3452 2022-11-30 Bouldin Alfalfa-ADEX RH & Temp Main Upper Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
HMP155 T2930114 2022-11-30 Bouldin Alfalfa-ADEX RH & Temp Bouncer Lower Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
HMP45AC D4420028 2022-11-30 Bouldin Alfalfa-ADEX RH & Temp Edge Upper Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
HMP45AC D4420029 2022-11-30 Bouldin Alfalfa-ADEX RH & Temp Bouncer Upper Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
HMP45AC T1020004 2022-11-30 Bouldin Alfalfa-ADEX RH & Temp Edge Lower Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
WM 1590 152608 2022-11-30 Bouldin Alfalfa-ADEX Anemometer Main Lower Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
WM 1590 152605 2022-11-30 Bouldin Alfalfa-ADEX Anemometer Main Upper Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
WM 1590 152601 2022-11-30 Bouldin Alfalfa-ADEX Anemometer Edge Upper Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
WM 1590 152606 2022-11-30 Bouldin Alfalfa-ADEX Anemometer Bouncer Upper Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
WM 1590 153804 2022-11-30 Bouldin Alfalfa-ADEX Anemometer Bouncer Lower Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
WM 1590 152603 2022-11-30 Bouldin Alfalfa-ADEX Anemometer Edge Lower Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
LI-7500A 0042 2022-11-30 Lab none Installed Hill Slough Gas Analyzer CO2/H2O Installed at Hill Slough to replace sn 75H-2176 for calibration.
LI-7500A 75H-2176 2022-11-30 Hill Slough Gas Analyzer CO2/H2O Removed Lab none Removed from Hill Slough for calibration. Replaced by sn 0042.
Mark II C003251 2022-11-30 Bouldin Corn Weather Station C Removed Lab none Removed and brought back to the lab because we don't need 3 Arable units on one tower. We added a Mark 3 last visit and there is a second Mark II on the tower.
LI-7500A 1594 2022-11-30 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2022-11-30 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 1594.
LI-7700 TG1-0211 2022-11-30 Hill Slough Calibration Hill Slough Calibration Location: Hill Slough
Last calibration: 2022-06-08
RSSI=71%, Optical RH=26%
Ambient P=101.4kPa, Tair=10.3C
Zero Air at 1650psi, AMP CA04396 1.9317 at 1100
CH4 Zero Offset: .005ppm
CH4 Span Offset: .0003ppm
Notes: After calibration we flushed headspace with zero air until optics RH = 8%. We swapped in fresh drierite. The current thermocouple assembly is missing an o-ring (has putty instead in the o-ring groove), so we swapped in a brand-new thermocouple assembly.
LI-7700 TG1-0357 2022-11-30 Bouldin Corn Gas Analyzer CH4 Calibration Bouldin Corn Gas Analyzer CH4 Calibration Location: Bouldin Corn
Last calibration: 2022-06-08
RSSI=80%, Optical RH=23%
Ambient P=101.3kPa, Tair=12.6C
Zero Air at 1650psi, AMP CA04396 1.9317 at 1100
CH4 Zero Offset: .032ppm
CH4 Span Offset: .0083ppm
Notes: Breezy. After calibration we flushed headspace with zero air until optics RH = 8%. We swapped in fresh drierite. Half of the thermocouple tubing had broken off so we put on a new piece of tubing.
WMP 1352 000023 2005-01-27 Lab none Installed Vaira Anemometer Installed at Vaira to replace current sonic - no reason given. Two headings given: 0 and 260 magnetic. Unclear which is new heading and which is old.
WMP 1352 000028 2005-01-27 Vaira Anemometer Removed none none Removed from Vaira to be replaced by another sonic - no reason given. Two headings given: 0 and 260 magnetic. Unclear which is new heading and which is old.
LI-7500A 0065 2005-01-27 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira replacing current IRGA presumably for calibration
Mark 3 B301069 2022-11-18 Lab none Installed Bouldin Corn none Installed production Mark 3 on the west side of the scaffolding. This replaced a prototype Mark 3 that we removed today.
LI-7500A 0073 2022-11-18 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at Bouldin Alfalfa to replace sn 1594, which was removed for calibration.
LI-7500A 1594 2022-11-18 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa for calibration. Replaced by sn 0073.
Mark II C013433 2022-11-18 Bouldin Alfalfa Weather Station B Removed Lab none Removed from BA main tower (south side of scaffolding) since we don't need 3 Arables on one tower. We are installing a Mark 3 today and we already have a Mark II that is connected to the wind set and the Sentek soil moisture profile probe.
Mark 3 B301045 2022-11-18 Lab none Installed Bouldin Alfalfa Weather Station B Installed at BA main tower on the north side of the scaffolding. The solar panel is facing south and the sensors are facing north, above the met/IRT enclosures.
NR-LITE 990350 2022-11-02 Bouldin Alfalfa-ADEX Net Radiometer Edge Removed Lab none Removed from top of the Edge tower at 3.34m alg at the end of the 2022 advection experiment at BA
SQ75 024398E-1-26-04-99974 2022-11-02 Bouldin Alfalfa-ADEX Solar Panel Edge B Removed Lab none Removed from about 10m north of the Edge tower at the end of the 2022 advection experiment
SP75 020854E-6-03-02-06862 2022-11-02 Bouldin Alfalfa-ADEX Solar Panel Edge A Removed Lab none Removed from about 10m north of the Edge tower at the end of the 2022 advection experiment
PS-30 19391162 2022-11-02 Bouldin Alfalfa-ADEX Solar Controller Edge Removed Lab none Removed from the Edge tower at the end of the 2022 advection experiment at BA
ML3 M001972 2022-11-02 Bouldin Alfalfa-ADEX Soil Moisture Probe Bouncer B Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
CS655 17130 2022-11-02 Bouldin Alfalfa-ADEX Soil Moisture Probe Edge A Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
LI-7550 AIU-0674 2022-11-02 Bouldin Alfalfa-ADEX Interface Edge LI-7550 Removed Lab none Removed from the Edge tower at the end of the 2022 advection experiment at BA
HP-127-1.0-1.3-71P PC4 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Edge C Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
HP-127-1.0-1.3-71P PC3 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Edge B Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
HFP01 003073 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Edge A Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
LI-7500A 0042 2022-11-02 Bouldin Alfalfa-ADEX Gas Analyzer Edge CO2/H2O Removed Lab none Removed from top of the Edge tower at 3.96m alg at the end of the 2022 advection experiment at BA
CR23X 2303 2022-11-02 Bouldin Alfalfa-ADEX Data Logger Edge Met Removed Lab none Removed from the Edge tower where it was collecting 30min averages from soil senors at the end of the 2022 advection experiment at BA
NR-LITE 990353 2022-11-02 Bouldin Alfalfa-ADEX Net Radiometer Bouncer Removed Lab none Removed from top of the Bouncer tower at 3.34m alg at the end of the 2022 advection experiment at BA
SQ75 020854E-6-03-02-06832 2022-11-02 Bouldin Alfalfa-ADEX Solar Panel Bouncer B Removed Lab none Removed from about 10m south of the Bouncer tower at the end of the 2022 advection experiment
SQ75 020854E-6-03-02-06825 2022-11-02 Bouldin Alfalfa-ADEX Solar Panel Bouncer A Removed Lab none Removed from about 10m south of the Bouncer tower at the end of the 2022 advection experiment
PS-15 03220377 2022-11-02 Bouldin Alfalfa-ADEX Solar Controller Bouncer Removed Lab none Removed from the Bouncer tower at the end of the 2022 advection experiment at BA
ML3 M003371 2022-11-02 Bouldin Alfalfa-ADEX Soil Moisture Probe Edge B Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
CS655 17131 2022-11-02 Bouldin Alfalfa-ADEX Soil Moisture Probe Bouncer A Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
LI-7550 AIU-2312 2022-11-02 Bouldin Alfalfa-ADEX Interface Bouncer LI-7550 Removed Lab none Removed from the Bouncer tower at the end of the 2022 advection experiment at BA
HP-127-1.0-1.3-71P PC2 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer C Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
HP-127-1.0-1.3-71P PC1 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer B Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
HFP01 003072 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer A Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
LI-7500RS 75H-0128 2022-11-02 Bouldin Alfalfa-ADEX Gas Analyzer Bouncer CO2/H2O Removed Lab none Removed from top of the Bouncer tower at 3.96m alg at the end of the 2022 advection experiment at BA
CR23X 2300 2022-11-02 Bouldin Alfalfa-ADEX Data Logger Bouncer Met Removed Lab none Removed from the Bouncer tower where it was collecting 30min averages from soil senors at the end of the 2022 advection experiment at BA
LI-7500A 0073 2022-11-15 Lab Calibration Lab Last calibration: 2022-04-08
AGC=101.2 SDM= DAC1= DAC2=
Ambient P=101.2kPa, Tair=23.2C, Tdew=1.69C
Zero Air at 1850psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9163 -> .9165 Offset: .68ppm
CO2 span 0.9983 -> 1.0085 Offset: -7.22ppm
H2O zero 0.8246 -> .8296 Offset: 28.9ppm
H2O span 1.0147 -> 1.0258 Offset: -.25ppm
Notes: Used AMP LI-7550 sn AIU-1039
LI-7500A 0042 2022-11-15 Lab Calibration Lab Last calibration: 2022-04-19
AGC=99.1 SDM= DAC1= DAC2=
Ambient P=101.2kPa, Tair=23.0C, Tdew=2.87C
Zero Air at 1850psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9483 -> .9481 Offset: -.73ppm
CO2 span 1.0087 -> 1.0127 Offset: -2.76ppm
H2O zero 0.8444 -> .847 Offset: 16.3ppm
H2O span 1.0191 -> 1.0292 Offset: -.22ppm
Notes: Used AMP LI-7550 sn AIU-1039
LI-7500A 0041 2022-11-15 Lab Calibration Lab Last calibration: 2022-03-30
AGC=93.5 SDM= DAC1= DAC2=
Ambient P=101.2kPa, Tair=22.9C, Tdew=.1C
Zero Air at 1900psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9969 -> .9998 Offset: 9.35ppm
CO2 span 1.0189 -> 1.0178 Offset: .23ppm
H2O zero 0.8214 -> .8213 Offset: -.5ppm
H2O span 1.0337 -> 1.0308 Offset: .07ppm
Notes: Reset SS from 93.5 to 100 before calibrating. Used AMP LI-7550 sn AIU-1039.
EXO fDOM 21K102591 2022-11-03 Gilbert Tract Calibration Gilbert Tract fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.0
UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,0.51 QSU,0.01 QSU,0.00 QSU,16.967 °C
Pt2,316.50 QSU,219.51 QSU,316.49 QSU,0.00 QSU,10.855 °C

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.12 RFU,0.00 RFU,0.00 RFU,15.878 °C
Pt2,105.50 RFU,74.86 RFU,105.50 RFU,0.00 RFU,11.257 °C
EXO fDOM 21K102592 2022-11-03 Gilbert Tract Calibration Gilbert Tract fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.0
UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,1.87 QSU,0.00 QSU,0.00 QSU,16.796 °C
Pt2,316.50 QSU,218.07 QSU,316.49 QSU,0.00 QSU,10.735 °C

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.20 RFU,0.00 RFU,0.00 RFU,15.801 °C
Pt2,105.50 RFU,71.69 RFU,105.50 RFU,0.00 RFU,11.214 °C
EXO fDOM 22C100780 2022-11-03 Gilbert Tract Calibration Gilbert Tract fDOM (QSU and RFU), Last cal: 10/24/2022 8:08:51 AM
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.0
UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-1.28 QSU,0.00 QSU,0.00 QSU,16.696 °C
Pt2,316.50 QSU,315.87 QSU,316.54 QSU,0.00 QSU,10.444 °C

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.11 RFU,0.00 RFU,0.00 RFU,10.610 °C
Pt2,105.50 RFU,92.32 RFU,105.51 RFU,0.00 RFU,11.275 °C
EXO Turbidity 22C100485 2022-11-01 Lab none Calibration Lab none Turbidity (FNU), Last cal: Unknown
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.4
A factory reset was done first before this calibration was done in the lab.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,0.24 FNU,0.00 FNU,0.00 FNU,20.384 °C
Pt2,124.00 FNU,112.11 FNU,123.99 FNU,0.00 FNU,20.389 °C
Pt3,1010.00 FNU,939.77 FNU,1009.99 FNU,0.00 FNU,20.565 °C
EXO pH-ORP 22D102641 2022-09-23 Lab none Calibration Lab none ORP (mV), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,223.0 mV,220.6 mV,223.0 mV,0.0 mV,25.000 °C
EXO pH-ORP 22D102641 2022-09-23 Lab none Calibration Lab none pH, Last cal: 7/12/2022 2:26:48 PM
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.00 pH,6.96 pH,7.00 pH,-4.1 mV,24.500 °C
Pt2,10.01 pH,10.15 pH,10.01 pH,-181.2 mV,24.500 °C
Pt3,4.01 pH,3.83 pH,4.02 pH,169.1 mV,24.500 °C
EXO Turbidity 22E106072 2022-09-23 Lab none Calibration Lab none Turbidity (FNU), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.4
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,-0.19 FNU,0.00 FNU,0.00 FNU,24.000 °C
Pt2,124.00 FNU,124.85 FNU,124.00 FNU,0.00 FNU,24.000 °C
EXO fDOM 22E100649 2022-09-23 Lab none Calibration Lab none fDOM (QSU and RFU), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-1.65 QSU,0.00 QSU,0.00 QSU,25.000 °C
Pt2,308.00 QSU,301.85 QSU,307.99 QSU,0.00 QSU,15.000 °C

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,103.00 RFU,98.95 RFU,103.00 RFU,0.00 RFU,15.000 °C
Pt2,0.00 RFU,-0.68 RFU,0.00 RFU,0.00 RFU,25.000 °C
EXO2 22C103416 2022-09-23 Lab none Calibration Lab none Depth (m), Last cal: 6/27/2022 1:49:10 PM
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,-0.008 m,0.000 m,0.000 m,22.000 °C
EXO pH-ORP 22D102641 2022-07-12 Lab none Calibration Lab none pH, Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
Slope is lower than what it should be for a new sensor,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.00 pH,7.02 pH,7.00 pH,-1.4 mV,20.000 °C
Pt2,10.00 pH,9.89 pH,10.00 pH,-167.8 mV,20.000 °C
Pt3,4.00 pH,4.39 pH,4.02 pH,151.1 mV,20.000 °C
EXO2 22C103416 2022-06-27 Lab none Calibration Lab none Depth (m), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,10.300 m,0.000 m,0.000 m,18.800 °C
PAR-LITE 030382 2022-10-31 Lab none Tested Lab none This sensor was removed because there were spikes in its data on two different mux ports. I pulled apart the splice in the cable and didn't see any problems. From 2022-10-27 to 2022-10-31 I ran an inter-comparison out the lab window with sn 010164. The relationship was tight (r2=0.9999) and the slope was about 1.06. sn 030382 had been inter-compared on the Tonzi tower recently but in the reflected position so the range was not optimal. I think both sensors are functional and should be inter-compared with a good reference in the field to determine appropriate coefficients.
PAR-LITE 010164 2022-10-31 Lab none Repair Lab none One of the wires was broken off right at the base of the sensor. I was able to solder on a lead and spliced a longer cable on and the sensor seemed to respond to light. From 2022-10-27 to 2022-10-31 I ran an inter-comparison out the lab window with sn 030382. The relationship was tight (r2=0.9999) and the slope was about 1.06. sn 030382 had been inter-compared on the Tonzi tower recently but in the reflected position so the range was not optimal. I think both sensors are functional and should be inter-compared with a good reference in the field to determine appropriate coefficients.
CR23X 2143 2022-10-27 Lab none Disposed Disposed none I took this 23x apart and found a surface mount capacitor that had blown. I tried to repair the board trace and replaced the cap with one from a dead CR10x - no way to tell if the components were the same. After putting it back together it was still bricked. We will keep it for parts for a while.
HYDROS21 H21G20001202 2022-10-19 Lab none Installed Gilbert Tract Water Level Installed at GT off of the boardwalk where it is co-located with the eosGP and Campbell conductivity sensor. When the staff gauge read 21cm, the bottom of the sensors was 19cm below the water.
PA-II-SD 84f3ebd5ffb1 2022-10-18 Bouldin Corn Air Quality Sensor Removed Lab none Removed from BC tower and brought back to the lab for further inspection. Data over the summer seemed spiky.
Mark II C013311 2022-10-18 Lab none Installed Bouldin Alfalfa Weather Station C Installed at BA (eastern position) to replace sn C003240, which was removed b/c of a broken T/RH sensor.
Mark II C003240 2022-10-18 Bouldin Alfalfa Weather Station C Removed Lab none Removed from BA (eastern position) because its T/RH sensor is broken. Brought back to the lab to ship back to Arable. It was replaced by sn C013311.
Mark II C013433 2022-10-18 Lab none Installed Bouldin Alfalfa Weather Station B Installed at BA (western position) to replace sn C006164, which was removed b/c of a broken T/RH sensor.
Mark II C006164 2022-10-18 Bouldin Alfalfa Weather Station B Removed Lab none Removed from BA (western position) because its T/RH sensor is broken. Brought back to the lab to ship back to Arable. It was replaced by sn C013433.
OS136-1-MVC NO1107536 2022-10-20 Tonzi none Removed Lab none This sensor has not been sending good data from a while. Seems a woodpecker destroyed its window. Removed and brought in for disposal
SI-131 7385 2022-10-20 Lab none Installed Tonzi Infrared Temperature Ref Panel A Install at the Tonzi cal panel for calibration. But I damaged the data logger so not collecting data until next time.
CR23X 2143 2022-10-20 Lab none Damaged Lab none I tried to install this at Tonzi to collect data from and IRT at the cal panel but fried it with 24V instead of 12V. It flashed "Hello" before something burned on the board.
PAR-LITE 990025 2022-10-20 Lab none Installed Tonzi PAR Sensor Reflected Installed at Tonzi to replace sn 030382 that has been spiky.
PAR-LITE 030382 2022-10-20 Tonzi PAR Sensor Reflected Removed Lab none Removed from Tonzi due to random spikiness. Replaced by sn 990025.
SI-131 7385 2022-10-19 Experiment none Removed Lab none I removed this sensor from the bike kit to calibrate it against the cal panel at Tonzi.
PAR-LITE 990025 2022-10-19 Experiment none Removed Lab none I removed this sensor from the bike kit to use at Tonzi. I also spliced about 3m of extra length on to its cable.
CR23X 2143 2022-10-19 Experiment none Removed Lab none I removed this data logger from the bike kit and prepared it to install at Tonzi to do a calibration on the IRT.
CTD-10 1782303006 2022-10-18 Factory none Disposed Disposed none Apparently the sensor was unrepairable and a new sensor was sent to us a replacement since this sensor was still under warranty. No info on the defect. The new sensor is a new model Hydros21 sn H21G200001202.
CTD-10 1782303006 2022-09-20 Lab none Shipped Factory none After testing in the lab reproduce the errors seen in the field an RMA-008376 was received and this sensors was shipped to METER for repair - it should be under warranty.
SD130BN 0030F4D214A4 2022-10-07 Factory none Returned Lab none Returned from Stardot after inspection determined that a power surge on the ethernet port cooked the innards. Will use the lens on a new body
GMP222 K1310102 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.16 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -16cm
GMP221 K1340006 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.08 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -8cm
GMP222 Z2720015 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.04 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -4cm
GMP222 K1310104 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.02 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -2cm
GMP221 Y2009100 2022-10-05 Tonzi-SoilCO2 CO2 Probe Tree1 Profile -0.16 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Tree1 -16cm
GMT221 G0150036 2022-10-05 Tonzi-SoilCO2 CO2 Probe Tree1 Profile -0.08 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Tree1 -8cm