Ser No.DateLoc FromTitleLoc ToNotes
eosFD FD20180067 2022-07-14 Lab none Installed Bouldin Corn Forced Diffusion C Installed in the corn at BC. It's in the same location from where it was removed, one row west of the zcam. I connected to the FD chamber to confirm that it was running.
NL115 13056 2022-07-14 Lab none Installed Bouldin Corn Interface NL115 Installed on the BC met CR1000 to replace sn 6761, which was brought back to the lab for testing.
NL115 6761 2022-07-14 Bouldin Corn Interface NL115 Removed Lab none Removed from BC and brought back to the lab for testing. The bottom 1/2-1/3 of phenocam photos is gray. Could be an interface or a card problem. It was replaced by sn 13056.
HFP01 003072 2022-07-14 Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer A Removed Lab none Removed from the BA Bouncer tower because the cable was cut. I will repair it in the lab before we install it again.
WM 1590 153805 2022-07-31 Lab none Tested Lab none Tested the sonic and its cable in the lab with Joe's help over e-mail. Both tested fine. I mated the sonic cable with an Eosense cable and used the conductivity setting on the voltmeter to find the 4 eosense wires that corresponded to outputs 1-4. At rest (while I was standing still next to the sonic), the voltages were around +/- 8mV. I then used a fan to test U, V, W. The voltages went up to around 150mV. Placing the fan on the opposite side of the sonic produced about the same magnitude voltage with the opposite sign. SOS (output 4) was around 1.7V and didn't change much with the fan, as expected. I wired the Eosense cable to a CR1000 with a NL115 and ran it overnight, storing 30min and 20Hz data on an 8gb card. The data the next morning seemed reasonable. SOS had the same pattern as PanelT.

Sonic pin -> Licor color -> Eosense cable
8 (Output 1) -> white -> brown
14 (Output 2) -> green -> white
9 (Output 3) -> grey -> pink
15 (Output 4) -> blue -> black
10 (Analog G) -> red -> grey
16 (Analog G) -> pink -> yellow
17 (Analog G) -> yellow/brown -> blue
WM 1590 153805 2022-08-01 Lab none Installed Mayberry Anemometer Installed on MB after testing in the lab. Sonic heading = 10 deg true N.
LI-7700 TG1-0223 2022-06-08 Lab none Installed Mayberry Gas Analyzer CH4 Installed at MB after attaching a new fuse holder and fuse. This sensor was calibrated in the lab yesterday.
WM 1590 153805 2022-07-26 Mayberry Anemometer Removed Lab none Removed from MB tower for inspection back at the lab. The sonic heading was 7 deg true N before removal.
LI-7700 TG1-0421 2022-07-15 Lab none Loaned Loaned none Installed on the BWW urban tower and updated the GHG columns. I also connected the HMP back to the 7700 auxiliary input. However, I couldn't remember the correct coefficients for the HMP. I tried the same multiplier and offset we use for CRBasic, but that didn't produce reasonable data. I e-mailed Joe for help and he fixed it the following day.

This sensor still needs a wash box and wash power cable. These will have to come from BA after our next visit. The current wash tubing is a spliced one, so I could also replace that with new black tubing.
LI-6400 PSC-0785 2019-06-03 Lab none Loaned Loaned none Loaned to Yvonne Socolar. Other material loaned:
-10 cartridges
-Box of spare parts
-Two 12V Battery
LI-6400 PSC-0785 2022-07-20 Lab none Loaned Loaned none Loaned to Rizwan Rasheed from Ellen Simms' lab for the summer. Other items loaned: 2 9Ah batteries and a 12V power cord. They had originally borrowed Dave Ackerly's 6400 but we couldn't get it to turn on (display remained off).
LI-7700 TG1-0421 2022-06-01 Factory Calibration Factory Calibration Location: Factory
Last calibration: 2021-11-30
RSSI=0%, Optical RH=0%
Ambient P=0kPa, Tair=0C
0 at 0psi, 0 0 at 0
CH4 Zero Offset: 10.1875ppm
CH4 Span Offset: .000103546ppm
Notes: Calibrated at Licor factory after replacing fuse holder and fuse. Not sure of the offset units (but probably not in ppm?)
LI-7700 TG1-0421 2022-07-05 Factory none Returned Lab none Received back in the lab after repair and calibration at Licor. They replaced the fuse and fuse holder and calibrated the instrument.
LI-7700 TG1-0421 2022-04-28 Lab none Shipped Factory none Sent back to Licor for inspection/repair around this date.
eosFD FD20180067 2022-07-11 Lab none Tested Lab none Plugged in and tested in the lab with another eosense cable. The chamber seems to be fine and the clock was still accurate, so I downloaded data from 2022-05-11 to 2022-06-08. I also ran a performance check which looked fine. I did not reset the reference.
Average Flux: -0.2, 0.05 STDEV
Average ATM: 687.6ppm, 29.6C
Average Soil: 677.7ppm, 29.8C
eosFD FD20180067 2022-06-08 Bouldin Corn Forced Diffusion C Removed Lab none Removed from field and brought back to the lab because even though the pump sounds like it's working, we haven't been able to connect to it for the past month
NL115 6761 2022-06-09 Lab none Installed Bouldin Corn Interface NL115 Installed on met CR1000 at BC and connected to Stardot phenocam
eosFD FD20180067 2022-02-17 Lab none Installed Bouldin Corn Forced Diffusion C Installed at BC under the tower
BB-Fab BB-HS 2022-04-20 Hill Slough Battery Bank Maintenance Hill Slough Battery Bank There has been some power losses a couple weeks ago but okay for the last two weeks. All the screw terminals seemed tight. One of the battery terminals had some corrosion - all the terminals should be cleaned and sealed before winter.
BB-Fab BB-HS 2021-03-11 Sherman Wetland-Temp1 Battery Bank Installed Hill Slough Battery Bank We put the battery box from SW on the west side of the platform and installed the five group 31 flooded batteries from the SW temporary tower. The solar controller sn 11400080 and breakers from the SW temporary site were also installed in the battery box. The wiring is not optimal and should be cleaned up. Also because the drill battery died we could not mount the solar controller or bolt the battery box to the platform.
NR01 1454 2011-02-03 Factory none Calibration Factory none Initial factory calibration: Pyranometer Up (mn SR01 sn 1957) 15.3uV/Wm-2, Pyranometer Down (mn SR01 sn 1958) 13.6uV/Wm-2, Pyrgeometer Up (mn IR01 sn 1946) 9.0uV/Wm-2. Pyrgeometer Down (mn IR01 sn 1947) 9.4uV/Wm-2.
LI-7700 TG1-0155 2022-06-09 West Pond Calibration West Pond Calibration Location: West Pond
Last calibration: 2021-12-06
RSSI=90%, Optical RH=3%
Ambient P=101.1kPa, Tair=30.1C
Zero Air at 400psi, AMP CA04119 4.2238 at 1550
CH4 Zero Offset: .016ppm
CH4 Span Offset: .0028ppm
Notes: Hot but relatively still at this site
LI-7700 TG1-0211 2022-06-08 Hill Slough Calibration Hill Slough Calibration Location: Hill Slough
Last calibration: 2021-12-01
RSSI=73%, Optical RH=21%
Ambient P=100.7kPa, Tair=24.0C
Zero Air at 550psi, AMP CA04119 4.2238 at 1525
CH4 Zero Offset: -.002ppm
CH4 Span Offset: .0008ppm
Notes: Very windy day of course. Did not change drierite or flush headspace with zero air. Changed the top/bottom heater to start at 18:00 instead of at 16:00 for summertime.
LI-7700 TG1-0357 2022-06-08 Bouldin Corn Calibration Bouldin Corn Calibration Location: Bouldin Corn
Last calibration: 2021-12-01
RSSI=77%, Optical RH=16%
Ambient P=101kPa, Tair=28C
Zero Air at 600psi, AMP CA04119 4.2238 at 1600
CH4 Zero Offset: -.074ppm
CH4 Span Offset: .0078ppm
Notes: Extra windy day of course. Did not change drierite or flush headspace with zero air.
LI-7700 TG1-0424 2022-06-09 Gilbert Tract Calibration Gilbert Tract Calibration Location: Gilbert Tract
Last calibration: 2021-11-30
RSSI=74%, Optical RH=16%
Ambient P=101kPa, Tair=21.6C
Zero Air at 450psi, AMP CA04119 4.2238 at 1450
CH4 Zero Offset: -.01ppm
CH4 Span Offset: .0698ppm
Notes: Breezy day. Did not change drierite or flush headspace with zero air.
LI-7700 TG1-0423 2022-06-09 East End Calibration East End Calibration Location: East End
Last calibration: 2021-12-01
RSSI=70%, Optical RH=17%
Ambient P=101kPa, Tair=29.5C
Zero Air at 350psi, AMP CA04119 4.2238 at 1550
CH4 Zero Offset: .015ppm
CH4 Span Offset: -.0098ppm
Notes: Hot day, some breeze. Did not change drierite or flush headspace with zero air.
TS01 TS01-002 2022-06-07 Tonzi Removed Lab none Disconnected and packed up on 05/26 but brought back to the lab today - no longer needed
SI-121 1624 2022-06-07 Tonzi Infrared Temperature Ref Panel B Removed Lab none Disconnected and packed up on 05/26 but brought back to the lab today - no longer needed
SI-131 7638 2022-06-07 Tonzi Removed Lab none Disconnected and packed up on 05/26 but brought back to the lab today - no longer needed
SI-131 7385 2022-06-07 Tonzi Infrared Temperature Grass B Removed Lab none Disconnected and packed up on 05/26 but brought back to the lab today - no longer needed
CR10X 47772 2022-06-07 Tonzi Removed Lab none Disconnected and packed up on 05/26 but brought back to the lab today - no longer needed
CR23X 2143 2022-06-07 Tonzi none Removed Lab none Disconnected and packed up on 05/26 but brought back to the lab today - no longer needed
LI-7700 TG1-0223 2022-06-07 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2021-12-03
RSSI=76%, Optical RH=29%
Ambient P=100.0kPa, Tair=21.7C
Zero Air at 625psi, AMP CA04119 4.2238 at 1600
CH4 Zero Offset: -.037ppm
CH4 Span Offset: .0392ppm
Notes: Before this calibration, I replaced fuse and fuse holder. I added some grease to the large o-ring around the base (Licor grease in the small plastic container) before reassembling everything. I followed Joe's advice to use longer helper screws help get the bottom to seat all the way in, and then replace those "helper" screws with the original screws with o-rings. During calibration for the span, I realized at 16:10 that the top heater had been on since 16:00. Probably the zero was calibrated while the heater was on. The span seemed stable from 16:12-16:18 and I don't think turning the heater off really affected it. I spanned the sensor at 16:18. After calibration, I pumped zero air into the headspace until optics RH=10%. I swapped out the old purple desiccant for a bottle of fresh blue desiccant. I added a bit of grease to the desiccant plug as well.
NL115 13056 2022-06-06 Lab none Tested Lab none I tested this CFM interface on the CR1000 in the lab. It didn't seem to have any trouble. Worked with both CF cards that I had trouble with in the field. I could ping the ethernet port. Maybe try updating the CR1000 OS in the field.
CS655 17129 2022-05-27 Bouldin Corn Soil Moisture Probe A Moved Bouldin Corn Soil Moisture Probe A Moved from under the tower in to the field after planting northwest of the tower. The tower was moved about 2m west to the edge of the field
CS655 17133 2022-05-27 Bouldin Corn Soil Moisture Probe C Moved Bouldin Corn Soil Moisture Probe C Moved from under the tower in to the field after planting west of the tower. The tower was moved about 2m west to the edge of the field
CS655 17128 2022-05-27 Bouldin Corn Soil Moisture Probe B Moved Bouldin Corn Soil Moisture Probe B Moved from under the tower in to the field after planting southwest of the tower. The tower was moved about 2m west to the edge of the field
NL115 13056 2022-05-27 Bouldin Corn Interface NL115 Removed Lab none I could not get this card reader to work with either of two CF cards. Removed for testing in the lab.
SoilVUE10 2677 2022-05-26 Lab none Installed Vaira Soil Moisture Probe I Profile Installed 5.3m at heading 220deg from the met box about a meter outside the fence near the gate.
SoilVUE10 2679 2022-05-26 Lab none Installed Vaira Soil Moisture Probe H Profile Installed 7.7m at heading 26deg from the met box near the solar panels and the old Moisture Point probe location
LI-7700 TG1-0223 2022-05-27 Mayberry Gas Analyzer CH4 Removed Lab none Removed from Mayberry because water got inside and rotted out the fuse on the bottom of the can. Same as WP a few weeks ago.
SoilVUE10 2690 2022-05-26 Lab none Installed Vaira Soil Moisture Probe F Profile Installed at 5.4m at heading 345deg from the met box about halfway between the eddy tower and radiation tower
RPI0W RPI0W-002 2021-08-17 Bouldin Corn Removed Lab none Removed from the BC field and brought back to the lab since the canopy has finished developing for this year
RPI0W RPI0W-001 2021-08-17 Bouldin Corn Removed Lab none Removed from the BC field and brought back to the lab since the canopy has finished developing for this year
PS-30 19391162 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Solar Controller Edge Installed on the Edge tower for the advection experiment at BA
PS-15 03220377 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Solar Controller Bouncer Installed on the Bouncer tower for the advection experiment at BA
SQ75 024398E-1-26-04-99974 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Solar Panel Edge B Installed about 10m north of the Edge tower at -10cm for the advection experiment
SP75 020854E-6-03-02-06862 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Solar Panel Edge A Installed about 10m north of the Edge tower at -10cm for the advection experiment
SQ75 020854E-6-03-02-06832 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Solar Panel Bouncer B Installed about 10m south of the Bouncer tower for the advection experiment
SQ75 020854E-6-03-02-06825 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Solar Panel Bouncer A Installed about 10m south of the Bouncer tower for the advection experiment
ML3 M003371 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Soil Moisture Probe Bouncer B Installed about 2m north of the Edge tower at -10cm for the advection experiment
CS655 17130 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Soil Moisture Probe Edge A Installed about 2m north of the Edge tower at -10cm for the advection experiment
ML3 M001972 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Soil Moisture Probe Edge B Installed about 2m south of the Bouncer tower at -10cm for the advection experiment
CS655 17131 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Soil Moisture Probe Bouncer A Installed about 2m south of the Bouncer tower at -10cm for the advection experiment
HP-127-1.0-1.3-71P PC4 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Heat Flux Plate Edge C Installed about 2m north of the Edge tower at -10cm for the advection experiment
HP-127-1.0-1.3-71P PC3 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Heat Flux Plate Edge B Installed about 2m north of the Edge tower at -10cm for the advection experiment
HFP01 003073 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Heat Flux Plate Edge A Installed about 2m north of the Edge tower at -10cm for the advection experiment
HP-127-1.0-1.3-71P PC2 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer C Installed about 2m south of the Bouncer tower at -10cm for the advection experiment
HP-127-1.0-1.3-71P PC1 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer B Installed about 2m south of the Bouncer tower at -10cm for the advection experiment
HFP01 003072 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer A Installed about 2m south of the Bouncer tower at -10cm for the advection experiment
CR23X 2303 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Data Logger Edge Met Installed on the Edge tower to collect 30min averages from soil senors for the advection experiment at BA
CR6 2115 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Data Logger Edge Flux Installed on the Edge tower to collect 20Hz data from the lower sonic and 10sec data from the two RH/Temp sensors for the advection experiment at BA
CR23X 2300 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Data Logger Bouncer Met Installed on the Bouncer tower to collect 30min averages from soil senors for the advection experiment at BA
CR6 2114 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Data Logger Bouncer Flux Installed on the Bouncer tower to collect 20Hz data from the lower sonic and 10sec data from the two RH/Temp sensors for the advection experiment at BA
CR1000 23865 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Data Logger Main Flux Installed on the main tower to collect 20Hz data from the two additional sonics for the advection experiment at BA
LI-7550 AIU-0674 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Interface Edge LI-7550 Installed on the Edge tower for the advection experiment at BA
LI-7550 AIU-2312 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Interface Bouncer LI-7550 Installed on the Bouncer tower for the advection experiment at BA
LI-7500A 0042 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Gas Analyzer Edge CO2/H2O Installed on top of the Edge tower at 3.96m alg for the advection experiment at BA
LI-7500RS 75H-0128 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Gas Analyzer Bouncer CO2/H2O Installed on top of the Bouncer tower at 3.96m alg for the advection experiment at BA
NR-LITE 990350 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Net Radiometer Edge Installed on top of the Edge tower at 334m alg for the advection experiment at BA
NR-LITE 990353 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Net Radiometer Bouncer Installed on top of the Bouncer tower at 3.34m alg for the advection experiment at BA
HMP45AC T1020004 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX RH & Temp Egde Lower Installed on the main post at the Edge tower at 2.42m alg for the advection experiment at BA
HMP45AC D4420028 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX RH & Temp Edge Upper Installed on top of the Edge tower at 3.96m alg for the advection experiment at BA
HMP155 T2930114 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX RH & Temp Bouncer Lower Installed on the main post at the Bouncer tower at 2.42m alg for the advection experiment at BA
HMP45AC D4420029 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX RH & Temp Bouncer Upper Installed on top of the Bouncer tower at 3.96m alg for the advection experiment at BA
HygroVUE10 E3514 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX RH & Temp Main Lower Installed on the existing eddy post at the main tower at 2.42m alg for the advection experiment at BA
HygroVUE10 E3452 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX RH & Temp Main Upper Installed on a new post at the main tower at 6.08m alg for the advection experiment at BA
WM 1590 152603 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Anemometer Egde Lower Installed on a cross arm at the Edge tower at 2.42m alg for the advection experiment at BA
WM 1590 152601 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Anemometer Edge Upper Installed on top of the Edge tower at 3.96m alg for the advection experiment at BA
WM 1590 153804 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Anemometer Bouncer Lower Installed on a cross arm at the Bouncer tower at 2.42m alg for the advection experiment at BA
WM 1590 152606 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Anemometer Bouncer Upper Installed on top of the Bouncer tower at 3.96m alg for the advection experiment at BA
WM 1590 152608 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Anemometer Main Lower Installed on a new cross arm at the main tower at 2.42m alg for the advection experiment at BA
WM 1590 152605 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Anemometer Main Upper Installed on a new post at the main tower at 6.08m alg for the advection experiment at BA
GP-001 GP20160012 2022-05-10 Vaira CO2 probe West Profile -0.16 Removed Lab none Removed with the last of the soil co2 system because data not good and to reduce power load.
GP-001 GP20160117 2022-05-10 Vaira CO2 probe West Profile -0.08 Removed Lab none Removed with the last of the soil co2 system because data not good and to reduce power load.
GP-001 GP20160042 2022-05-10 Vaira CO2 probe West Profile -0.02 Removed Lab none Removed with the last of the soil co2 system because data not good and to reduce power load.
GP-001 GP20170028 2022-05-10 Vaira CO2 probe West Profile -0.04 Removed Lab none Removed with the last of the soil co2 system because data not good and to reduce power load.
CR23X 2299 2022-05-10 Vaira Data Logger Soil CO2 Removed Lab none Removed with the last of the soil co2 system because data not good and to reduce power load.
BB-Fab BB-VR 2022-05-10 Vaira Battery Bank Main Maintenance Vaira Battery Bank Main Fixed broken wire between the two banks of solar panels. Disconnected and load tested all the batteries. Two batteries, second from north in each box, tested bad. These were moved to the north most position and left disconnected. The other six tested good and were reconnected.
LI-7700 TG1-0643 2022-05-09 Borrowed none Removed Returned none Removed from the urban tower on BWW roof because Ameriflux needs it back to do some site visits this summer. The sensor and accessories were returned to Stephen Chan of AMP when he came to our lab to visit the urban tower.
Items returned: 7700, 7700 hat, wash box, brand-new wash power, brand-new wash tubing, wash box mounting hardware
Items I forgot to return: ethernet cable, Licor power cable
PQS 1 212225 2022-05-09 Lab none Returned Returned none Returned to Stephen Chan of AMP when he came to our lab to pick up the methane loaner from the urban tower on BWW roof.
CNR4 120788 2022-05-09 Lab none Returned Returned none Returned to Stephen Chan of AMP when he came to our lab to pick up the methane loaner from the urban tower on BWW roof.
miniDOT 1221 2022-05-02 Lab none Serviced Lab none I connected to this miniDOT using sneakernet and was able to adjust the sampling frequency, but could not sync the clock. Joe thought that maybe the interface is checking a time server that doesn't exist anymore, or maybe the interface cannot be used with Windows 10. He ended up setting the clock by using the old Tonzi laptop (running Windows 7) connected to an external monitor. Campus wi-fi does not allow Windows 7 to connect to the internet b/c of security issues, so he ended up using his phone as a hotspot to connect the old laptop to the internet (after some more hiccups. 2.4 vs. 5.0 GHz wifi?)

The miniDOT clock was behind by 693 seconds (~12min). He synced the clock, set the sampling frequency to 5min, and put in fresh batteries (I think). I put in a fresh desiccant pack.

This miniDOT only accepts 2GB SD cards.
miniDOT 1221 2022-05-06 Lab none Installed Gilbert Tract none Mckenna from USGS installed this at the GT breach site next to the other USGS water sensors in the channel. They plan to service the sensors every 3 months.
EXO fDOM 22C100780 2022-05-06 Lab none Installed Gilbert Tract none Mckenna from USGS installed this at the GT breach site next to the other USGS water sensors in the channel. They plan to service the sensors every 3 months.
Mark II C003231 2022-04-27 Vaira Weather Station B Removed Lab none Removed from Vaira because it has not been working. I connected the wind set to the remaining Mark II unit.
Mark I A000251 2021-08-18 Lab none Disposed Disposed none This unit is dead and Arable doesn't want it back. I took it apart. The lithium battery was all puffy. Birds had pecked off one of the upward sensors. Not much worth salvaging.
Mark II C003771 2021-06-21 Lab none Disposed Disposed none Returned to Loreli from Arable
Mark II C003751 2021-06-21 Lab none Disposed Disposed none Returned to Loreli from Arable
Mark II C006282 2021-06-21 Lab none Disposed Disposed none Returned to Loreli from Arable