Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 [35] 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65
Ser No.DateLoc FromTitleLoc ToNotes
HMP45AC T2130005 2017-12-19 Twitchell Alfalfa Removed Lab none Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP60 G4540025 2017-12-19 Twitchell Alfalfa Removed West Pond RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa and returned to West Pond
HMP60 G4540023 2017-12-19 Twitchell Alfalfa none Removed East End RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa and returned to East End
HMP45AC U3030041 2017-12-19 Twitchell Alfalfa Removed Lab RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP60 G4540021 2017-12-19 Twitchell Alfalfa 0 Removed Lab none Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP45AC W0730036 2017-12-19 Twitchell Alfalfa Removed Lab RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa
U24-002-C 11020258 2017-08-02 East End-Temp3 Conductivity Sensor Moved East End Conductivity Sensor The Hobo conductivity sensor was moved with the miniDOT from the T3 Boardwalk Temporary Tower to the Permanent Tower.
miniDOT 1219 2017-11-10 Lab none Installed East End DO Sensor Installed at East End hanging from the north side of the scaffolding
miniDOT 1219 2017-09-13 East End DO Sensor Moved East End-Temp2 DO Sensor Moved from the Permanent Tower to the T2 Closed Canopy Temporary Tower
U24-002-C 11020258 2017-09-13 East End Conductivity Sensor Moved East End-Temp2 Conductivity Sensor Moved from the Permanent Tower to the T2 Closed Canopy Temporary Tower
miniDOT 1219 2017-08-02 East End-Temp3 DO Sensor Moved East End DO Sensor The miniDOT was moved with the Hobo conductivity sensor from the T3 Boardwalk Temporary Tower to the Permanent Tower.
miniDOT 1219 2017-06-23 Lab none Installed East End-Temp3 DO Sensor Installed at the East End T3 Boardwalk Temporary Tower with the Hobo conductivity sensor.
LI-820 CGA-556 2017-12-14 Tonzi Removed Lab none Removed from the Tower co2 profile for cleaning and calibration. The rest of the profile equipment was turned off
eosFD 20160002 2017-12-14 Tonzi Removed Lab none Removed from the Tonzi soil CO2 profile in the Open (East) location. This FD chamber has been reporting zero flux with nearly identical CO2 readings for both cells. The pump seems to be running though.
eosFD 20160002 2017-10-25 Lab none Installed Tonzi Installed near the Open (East) soil CO2 profile for three month experiment
HMP45A U2120013 2017-12-18 Mayberry RH & Temp Removed Twitchell Alfalfa 0 Removed from Mayberry for 24hour side-by-side comparison at Twitchell Alfalfa
HMP60 L2310736 2017-12-18 Sherman Wetland RH & Temp Removed Twitchell Alfalfa 0 Removed from Sherman Wetland for 24hour side-by-side comparison at Twitchell Alfalfa
HMP60 G4540025 2017-12-18 West Pond RH & Temp Removed Twitchell Alfalfa Removed from West Pond for 24hour side-by-side comparison at Twitchell Alfalfa
LI-7550 AIU-0397 2017-12-18 East End Interface LI-7550 Upgrade East End Interface LI-7550 The embedded software was upgraded from v8.0.0 to v8.5.0
LI-7500A 75H-2669 2017-12-18 Lab none Installed East End Gas Analyzer CO2/H2O Installed on East End to replace sn 0041 for calibration. The AIU software was upgrade from v8.0.0 to v8.5.0
LI-7500A 0041 2017-12-18 East End Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from East End for calibration. Replaced by sn 75H-2669
HMP60 G4540023 2017-12-18 East End RH & Temp Removed Twitchell Alfalfa none Removed from Eat End for a 24hour side-by-side comparison at Twitchell Alfalfa
LI-7500A 75H-2669 2017-12-15 Lab none Calibration Lab none Last calibration: 2017-08-14
AGC=103.3 SDM=0 DAC1=0 DAC2=0
Ambient P=100.7kPa, Tair=22.2C, Tdew=7C
Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.9187 -> .9187 Offset: .39ppm
CO2 span 1.0144 -> 1.0157 Offset: -.73ppm
H2O zero 0.8988 -> .8988 Offset: .2ppm
H2O span 1.0091 -> 1.0033 Offset: .12ppm
Notes:
LI-7500A 75H-2667 2017-12-15 Lab none Calibration Lab none Last calibration: 2017-08-14
AGC=105.9 SDM=0 DAC1=0 DAC2=0
Ambient P=100.7kPa, Tair=21.3C, Tdew=6.46C
Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.9418 -> .9415 Offset: -1.14ppm
CO2 span 1.0162 -> 1.0207 Offset: -3.47ppm
H2O zero 0.8392 -> .8372 Offset: -11.7ppm
H2O span 1.0214 -> 1.0166 Offset: .1ppm
Notes:
LI-7500A 75H-2514 2017-12-15 Lab none Calibration Lab none Last calibration: 2017-08-14
AGC=102.8 SDM=0 DAC1=0 DAC2=0
Ambient P=100.7kPa, Tair=22.1C, Tdew=7.9C
Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.905 -> .9052 Offset: .8ppm
CO2 span 1.0107 -> 1.0118 Offset: -.5ppm
H2O zero 0.8378 -> .8385 Offset: 4.3ppm
H2O span 1.0094 -> 1.0103 Offset: -.01ppm
Notes:
Line Quantum LQ1570 2000-11-04 Lab none Installed Vaira Installed at Vaira with the initial setup of the site.
CR1000 50889 2017-12-11 Lab none Upgrade Lab none Upgraded O.S. from version 27 to version 31.08 (CR1000.Std.31.08.obj). I enabled FTP with the usual username (Tower) and password.
HMP45AC U3030041 2017-12-08 Lab none Repair Lab none The circuit board was corroded and the red wire was broken at the board connection. Joe repaired it. This sensor needs a new o-ring before going outdoors.
HMP60 G4540021 2017-06-23 Lab none Installed East End-Temp3 RH & Temp Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site.
LI-7500A 0418 2017-11-30 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at MB to replace sn 75H-2669.
LI-7500A 75H-2669 2017-11-30 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from MB and brought back to the lab for calibration.
LI-7550 AIU-1285 2017-11-30 Mayberry Interface LI-7550 Upgrade Mayberry Interface LI-7550 Downgraded from embedded software version 8.7.5 to version 8.5.0. This solves the problem of data-header mismatch we had with version 8.7.5.
LI-7700 TG1-0224 2017-11-30 Lab none Notes Lab none Temperature was reading -9C after we installed it at SW, so we brought it back to the lab...The temperature was fine earlier this week when I calibrated it.
LI-7500A 0073 2017-11-30 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at SW to replace sn 75H-2667.
LI-7500A 75H-2667 2017-11-30 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from SW and brought back to the lab for calibration
LI-7550 AIU-1569 2017-11-30 Sherman Wetland Interface LI-7550 Upgrade Sherman Wetland Interface LI-7550 Downgraded from embedded software version 8.7.5 to version 8.5.0. This should solve the issue of data-header mismatch that we found in version 8.7.5.
LI-7500A 1594 2017-11-29 Lab none Installed Tonzi-Floor Installed on the Tonzi Floor to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2017-11-29 Tonzi Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 1594
LI-7500A 75H-2668 2017-11-29 Lab Gas Analyzer CO2/H2O Floor Installed Tonzi-Tower Gas Analyzer CO2/H2O Floor Installed on the Tonzi Tower to replace sn 75H-2182 for calibration.
LI-7500A 75H-2182 2017-11-29 Tonzi Removed Lab none Removed from the Tonzi Tower for calibration. Replaced by sn 75H-2668
LI-820 CGA-556 2017-11-29 Lab none Installed Tonzi Installed in the Tower CO2 profile system after lab cleaning and calibration
LI-7500A 0065 2017-11-29 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 75H-2514 for calibration.
LI-7500A 75H-2514 2017-11-29 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 0065
GMM222 L2010067 2017-11-29 Lab none Installed Vaira CO2 probe East Profile -0.04 Installed in the 4cm location in the East profile at Vaira to replace a probe with erratic values.
GMP222 K1310102 2017-11-29 Vaira CO2 probe East Profile -0.04 Removed Lab none Removed from the 4cm location in the East profile at Vaira. This probe was giving erratic values
LI-7500A 0065 2017-11-28 Lab none Calibration Lab none Last calibration: 2017-05-30
AGC=100.0 SDM=0 DAC1=0 DAC2=0
Ambient P=100.5kPa, Tair=23.0C, Tdew=8.7C
Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.8901 -> .8906 Offset: 2.12ppm
CO2 span 1.013 -> 1.0163 Offset: -1.35ppm
H2O zero 0.7087 -> .7083 Offset: -3.0ppm
H2O span 1.0166 -> 1.0126 Offset: 0.08ppm
Notes: Calibrated on 7550 software version 8.7.5
LI-7500A 1594 2017-11-28 Lab none Calibration Lab none Last calibration: 2017-09-27
AGC=99.9 SDM=0 DAC1=0 DAC2=0
Ambient P=100.5kPa, Tair=23.0C, Tdew=8.7C
Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.8638 -> .8653 Offset: 6.09ppm
CO2 span 1.0168 -> 1.0172 Offset: -.23ppm
H2O zero 0.8878 -> .8868 Offset: -5.6ppm
H2O span 1.0047 -> 1.0000 Offset: 0.1ppm
Notes: Calibrated on 7550 software version 8.7.5
LI-7500A 75H-2668 2017-11-28 Lab none Calibration Lab none Last calibration: 2017-07-25
AGC=99.9 SDM=0 DAC1=0 DAC2=0
Ambient P=100.6kPa, Tair=21.7C, Tdew=8.9C
Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.9395 -> .9392 Offset: -1.56ppm
CO2 span 1.0159 -> 1.0147 Offset: 0.64ppm
H2O zero 0.9047 -> 0.9066 Offset: 11.0ppm
H2O span 1.0107 -> 1.0125 Offset: -0.04ppm
Notes: Calibrated on 7550 software version 8.7.5
LI-7700 TG1-0224 2017-11-27 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-03-22
RSSI=71%, Optical RH=7%
Ambient P=100.9kPa, Tair=22.4C
Zero Air at 350psi, LBL Air 1.9497 at 1750
CH4 Zero Offset: 0.016ppm
CH4 Span Offset: 0.005ppm
Notes:
LI-7500A 0418 2017-11-27 Lab none Calibration Lab none Last calibration: 2017-07-02
AGC=97.1 SDM=0 DAC1=0 DAC2=0
Ambient P=100.9kPa, Tair=22.8C, Tdew=6.8C
Zero Air at 350psi, LBL Air 408.97 at 1750 Tdewspan=16.00
CO2 zero 0.8995 -> .8998 Offset: .87ppm
CO2 span 1.0066 -> 1.006 Offset: .4ppm
H2O zero 0.8436 -> .8437 Offset: .2ppm
H2O span 0.9847 -> .9821 Offset: .05ppm
Notes: Calibrated on 7550 software version 8.5.0
LI-7500A 0073 2017-11-27 Lab none Calibration Lab none Last calibration: 2017-06-22
AGC=100.6 SDM=0 DAC1=0 DAC2=0
Ambient P=101.0kPa, Tair=21.6C, Tdew=6.8C
Zero Air at 350psi, LBL Air 408.97 at 1750 Tdewspan=16.00
CO2 zero 0.9572 -> .9571 Offset: -0.028ppm
CO2 span 1.0084 -> 1.0118 Offset: -2.11ppm
H2O zero 0.7307 -> .7306 Offset: -0.3ppm
H2O span 1.0045 -> 1.0000 Offset: 0.1ppm
Notes: Calibrated on 7550 software version 8.5.0
PTB110 H0160093 2017-11-16 Lab none Tested Lab none Consistently tested as 0V out. Vin was 12V and it was wired correctly. Looking at recent TW data (where this sensor was last) from Mar/Apr 2017, pressure data seems reasonable. Not sure where/why/when it got damaged.
LI-7700 TG1-0357 2017-11-10 Lab none Installed East End Gas Analyzer CH4 Installed at EE to replace sn TG1-0155, which was brought back to the lab for calibration.
LI-7700 TG1-0155 2017-11-10 East End Gas Analyzer CH4 Removed Lab none Removed from EE and brought back to the lab for calibration. It was replaced by sn TG1-0357.
LI-7500A 75H-2176 2017-11-13 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed on Bouldin Corn replacing sn 1594 for calibration.
LI-7500A 1594 2017-11-13 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn because it was calibrated on LI-7550 with embedded software v. 8.7.5 and the LI-7550 at the tower is embedded software v 8.5.0, so we were worried this software mismatch would produce wonky data. Replaced by sn 75H-2176.
SD130BN 0030F4D21491 2017-07-12 Lab none Installed Bouldin Corn Camera This camera was put up on the Bouldin Corn Tower. It was attached to the southwest corner of the scaffolding just below the top railing and connected the to NL115 on the CR1000 to record the images on a CFM card. The CR1000 IP was set to 192.168.2.100 and the FTP user=Tower and pw=biomet.
SD130BN 0030F4D214A4 2016-11-03 Lab none Installed Bouldin Alfalfa Camera This Stardot camera was mounted on the Bouldin Alfalfa tower. It is facing NW off the north side of the tower scaffolding. It is currently mounted near the top of the railing due to a short cable. It will be moved to near the top of the post later.
LI-7500A 75H-2176 2017-11-09 Lab none Calibration Lab none Last calibration: 2017-06-12
AGC=89.6 SDM=0 DAC1=0 DAC2=0
Ambient P=100.6kPa, Tair=22.5C, Tdew=12.2C
Zero Air at 400psi, LBL Air 408.97 at 1800 Tdewspan=16.00
CO2 zero 0.869 -> 0.869 Offset: 0ppm
CO2 span 1.016 -> 1.0225 Offset: -3.37ppm
H2O zero 0.8572 -> .8570 Offset: -1.3ppm
H2O span 1.0022 -> 1.0047 Offset: -0.06ppm
Notes: Calibrated on AIU-0674 (version 8.5.0)
LI-7700 TG1-0224 2017-11-09 Lab none Repair Lab none Replaced thermocouple assembly on 7700. Temperature reading changed from 9999 to 24.5 deg C.
LI-7700 TG1-0357 2017-11-09 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-03-23
RSSI=72.7%, Optical RH=23%
Ambient P=100.6kPa, Tair=21.8C
Zero Air at 400psi, LBL Air 1.9497 at 1800
CH4 Zero Offset: -0.1ppm
CH4 Span Offset: -0.0197ppm
Notes: Optics RH was 23%, so I flushed the head cavity with zero air and replaced the desiccant. We just changed the desiccant last time (March 2017) so I was expecting a RH ~5%. Let's keep an eye on the optics RH.
SD130BN 0030F4D2148D 2017-10-26 Lab none Installed Twitchell Alfalfa Camera Reinstalled at Twitchell Alfalfa. It is missing the plastic slide that clips into the enclosure - to be installed next time.
SD130BN 0030F4D2148D 2017-10-19 Factory Returned Lab none Returned from repair at StarDot
SD130BN 0030F4D2148D 2017-10-16 Factory Repair Factory Repaired by StarDot. Replaced CCD Board.
SD130BN 0030F4D2148D 2017-09-27 Lab none Shipped Factory Sent to StarDot for repair on RMA END-092217AAW1
SD130BN 0030F4D2148D 2017-09-26 Twitchell Alfalfa Camera Removed Lab none Removed from Twitchell Alfalfa because it has been recording bad images with psychedelic colors.
LI-7700 TG1-0223 2017-10-19 Lab none Installed Bouldin Alfalfa Gas Analyzer CH4 Installed at BA (now that the East End temporary towers are done, we have an available LI-7700 to put up here)
eosFD FD20160001 2017-10-27 Lab none Repair Factory Sent back to Eosense for repair.
eosFD FD20160001 2017-10-13 Twitchell Alfalfa Forced Diffusion Removed Lab none Removed from TA because it was giving bad readings (200? ppm +/- 200? ppm).
LI-7700 TG1-0423 2017-10-26 Lab none Installed Mayberry Gas Analyzer CH4 Installed at MB to replace sn TG1-0357
LI-7700 TG1-0357 2017-10-26 Mayberry Gas Analyzer CH4 Removed Lab none Removed from MB and brought back to the lab for calibration
PS-15M 12501584 2017-10-19 East End-Temp1 Solar Controller Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
KD140SX-UFBS 11ZV0M0856 2017-10-19 East End-Temp1 Solar Panel B Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
KD140SX-UFBS 11ZV0M0835 2017-10-19 East End-Temp1 Solar Panel A Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
HMP45AC W0730036 2017-10-19 East End-Temp1 RH & Temp Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
HFP01 10822 2017-10-19 East End-Temp1 Heat Flux Plate B Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
CS450 20010770 2017-10-19 East End-Temp1 Water Level Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
AM25T 2856 2017-10-19 East End-Temp1 Multiplexer Thermocouples Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
CR6 2114 2017-10-19 East End-Temp1 Data Logger Met Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
PAR-LITE 040412 2017-10-19 East End-Temp1 PAR Sensor Reflected Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
NR-LITE 990352 2017-10-19 East End-Temp1 Net Radiometer Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
WM 1590 153803 2017-10-19 East End-Temp1 Anemometer Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
LI-7500A 0065 2017-10-19 East End Gas Analyzer T1 CO2/H2O Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
LI-7700 TG1-0423 2017-10-19 East End-Temp1 Gas Analyzer CH4 Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
LI-7550 AIU-0650 2017-10-19 East End-Temp1 Interface LI-7550 Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
LI-7700 TG1-0421 2017-10-26 Lab none Installed Sherman Wetland Gas Analyzer CH4 Installed at SW to replace sn TG1-0224.
LI-7700 TG1-0224 2017-10-26 Sherman Wetland Gas Analyzer CH4 Removed Lab none Removed from SW tower for calibration and because the temperature sensor was broken (reading 9999). Replaced by sn TG1-0421.
LI-7500A 0042 2001-04-18 Lab none Installed Tonzi-Floor Installed on the Floor tower to replace existing IRGA for calibration.
eosFD 20160003 2017-10-18 Lab none Installed Tonzi Installed near the Tree 1 soil CO2 profile for three month experiment
GP-001 GP20140018 2017-08-16 Lab none Installed West Pond CO2 Probe B Installed at WP to replace a previous bad eosGP
LI-7550 AIU-1569 2017-07-14 Sherman Wetland Interface LI-7550 Upgrade Sherman Wetland Interface LI-7550 Upgraded LI-7550 software from 8.0.0 to 8.7.5.
eosFD 20160006 2017-10-18 Lab none Installed Vaira Forced Diffusion East Installed at Vaira near the East soil CO2 profile. The chamber is on a collar, fastened down with cable ties and power with 12V from a DC-DC power converter.
eosFD 20160005 2017-10-18 Lab none Installed Vaira Forced Diffusion West Installed at Vaira near the West soil CO2 profile. The chamber is on a collar, fastened down with cable ties and power with 12V from a DC-DC power converter.
LI-7500A 1594 2017-10-19 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed on Bouldin Corn to replace sn 75H-2668 for calibration.
LI-7700 TG1-0423 2017-10-23 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-05-31
RSSI=74.6%, Optical RH=3%
Ambient P=101.3kPa, Tair=27.9C
Zero Air at 425psi, AMP Air 1.9497 at 1850
CH4 Zero Offset: -0.055ppm
CH4 Span Offset: 0.0197ppm
Notes:
LI-7500A 75H-2668 2017-10-19 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 1594
LI-7550 AIU-1570 2017-08-23 Bouldin Corn Interface LI-7550 Upgrade Bouldin Corn Interface LI-7550 Upgraded the 7550 firmware from v8.0.0 to v8.7.5. I confirmed the 7550 time, time zone, sonic analog settings, and CO2 and CH4 logging config.
LAI-2050 PCA-1546 2017-10-18 Borrowed Returned Lab none Returned by Sebastian Wolf to the lab
LAI-2000 PCA-1478 2017-10-18 Borrowed Returned Lab none Returned by Sebastian Wolf sewolf@ethz.ch to the lab
LAI-2000 PCA-0846 2017-10-18 Borrowed Returned Lab none Returned by Sebastian Wolf sewolf@ethz.ch to the lab