GP-001 GP20150104 2017-09-08 Lab none Replaced Disposed This sensor was given to Jennifer Soong in Margaret Torn's lab with five others in exchange six new ones at a later date. Their new sensors would not arrive in time for deployment.
GP-001 GP20150105 2017-09-08 Lab none Replaced Disposed This sensor was given to Jennifer Soong in Margaret Torn's lab with five others in exchange six new ones at a later date. Their new sensors would not arrive in time for deployment.
GP-001 GP20150103 2017-09-08 Lab none Replaced Disposed This sensor was given to Jennifer Soong in Margaret Torn's lab with five others in exchange six new ones at a later date. Their new sensors would not arrive in time for deployment.
GP-001 GP20150102 2017-09-08 Lab none Replaced Disposed This sensor was given to Jennifer Soong in Margaret Torn's lab with five others in exchange six new ones at a later date. Their new sensors would not arrive in time for deployment.
eosFD FD20160001 2017-09-06 Lab none Installed Twitchell Alfalfa Forced Diffusion Reinstalled with new 90 degree cable after factory refurb.
LI-7500A 0035 2017-09-06 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed on Twitchell Alfalfa to replace sn 1594 for calibration.
LI-7500A 1594 2017-09-06 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Twitchell Alfalfa for calibration. Replaced by sn 0035.
LI-7500A 0035 2017-09-05 Lab none Calibration Lab none Last calibration: 2017-05-08
AGC=99.8 SDM=0 DAC1=0 DAC2=0
Ambient P=100.4kPa, Tair=29.1C, Tdew=18.4C
Zero Air at 1000psi, AMP Air 408.97 at 2100 Tdewspan=16.00
CO2 zero 0.9381 -> .9373 Offset: -3.34ppm
CO2 span 1.023 -> 1.0084 Offset: 7.67ppm
H2O zero 0.8974 -> .8976 Offset: .8ppm
H2O span 1.0261 -> 1.0141 Offset: .25ppm
Notes:
GP-001 GP20140016 2017-08-16 Lab none Installed Sherman Wetland CO2 Probe Installed at SW. I waterproofed the cable-sensor connection with multiple layers of liquid electrical tape and silicone caulk.
GP-001 GP20150101 2017-08-16 Sherman Wetland CO2 Probe Removed Lab none Removed from SW because it did not have a waterproof sensor-cable connection.
LI-7500A 75H-2667 2017-08-16 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at SW to replace sn 75H-2182.
LI-7500A 75H-2182 2017-08-16 Lab none Calibration Lab none Last calibration: 2017-04-20
AGC=87.5 SDM=0 DAC1=0 DAC2=0
Ambient P=100.4kPa, Tair=23.9C, Tdew=12.7C
Zero Air at 1075psi, AMP Air 408.97 at 2000 Tdewspan=16.00
CO2 zero 0.8838 -> .8832 Offset: -2.73ppm
CO2 span 1.0164 -> 1.0123 Offset: 2.13ppm
H2O zero 0.8454 -> .8436 Offset: -11.4ppm
H2O span 1.0139 -> 1.0038 Offset: .2ppm
Notes:
LI-7500A 75H-2180 2017-08-16 Lab none Calibration Lab none Last calibration: 2017-04-20
AGC=100.3 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=23.1C, Tdew=13.9C
Zero Air at 1075psi, AMP Air 408.97 at 2000 Tdewspan=16.00
CO2 zero 0.9144 -> .9129 Offset: -6.82ppm
CO2 span 1.026 -> 1.0145 Offset: 5.95ppm
H2O zero 0.8861 -> .8871 Offset: 6.2ppm
H2O span 1.023 -> 1.0164 Offset: .14ppm
Notes:
LI-7500A 75H-2669 2017-08-16 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at MB to replace sn 75H-2180.
LI-7500A 75H-2180 2017-08-16 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from MB for calibration and replaced with sn 75H-2669
LI-7550 AIU-1285 2017-08-16 Mayberry Interface LI-7550 Upgrade Mayberry Interface LI-7550 The firmware was upgraded from v8.0.0 to v8.7.5
HFP01 001604 2017-08-30 East End-Temp1 Heat Flux Plate A Removed Lab none Removed from Temporary Tower #1 Open Water. Its cable was chewed through.
GMM222 L2030064 2016-01-12 Twitchell Alfalfa Removed Lab none Removed from Twitchell Alfalfa with the FD chamber due to torn membranes.
GMM222 L2030063 2016-01-12 Twitchell Alfalfa Removed Lab none Removed from Twitchell Alfalfa with the FD chamber due to torn membranes.
eosFD FD20160001 2017-07-05 Lab none Shipped Factory Shipped to Eosense for replacement of failed pump. Eosense provided shipping on UPS.
GMP222 Y2030049 2017-08-25 Vaira CO2 probe East Profile -0.08 Removed Lab none Removed from Vaira from East profile -8 cm position for calibration
GMP222 K1310103 2017-08-25 Vaira CO2 probe East Profile -0.04 Removed Lab none Removed from Vaira from East profile -4 cm position for calibration
GMP222 Y2009109 2017-08-25 Vaira CO2 probe East Profile -0.02 Removed Lab none Removed from Vaira from East profile -2 cm position for calibration. Probe labeled 4.
GMM222 L2010068 2017-08-25 Vaira CO2 probe North Profile -0.04 Removed Lab none Removed from Vaira from North profile -4 cm position for calibration
GMT222 G0230062 2017-08-25 Vaira CO2 probe North Profile -0.02 Removed Lab none Removed from Vaira from North profile -2 cm position for calibration
GMP222 K1350101 2017-08-25 Vaira CO2 probe West Profile -0.08 Removed Lab none Removed from Vaira from West profile -8 cm position for calibration
GMP222 K1350102 2017-08-25 Vaira CO2 probe West Profile -0.04 Removed Lab none Removed from Vaira from West profile -4 cm position for calibration
GMP222 K1350103 2017-08-25 Vaira CO2 probe West Profile -0.02 Removed Lab none Removed from Vaira from West profile -2 cm position for calibration
GMP222 Z2720012 2017-08-25 Tonzi Notes Lab none Removed CO2 probe from Tonzi Tree2 profile at -2 cm. It was sn Z2720014, not sn Z2720012. Not sure where sn Z2720012 is.
GMP222 X4840007 2017-08-25 Tonzi Notes Lab none We removed the CO2 probe at the Tree 2 profile at -2 cm position. The probe was sn Z2720014. Not sure where sn X4840007 is.
GMP222 Z2720014 2017-08-25 Tonzi CO2 Probe Tree2 Profile -0.02 Removed Lab none Removed from Tonzi Tree 2 profile at -2 cm position for calibration
GMT221 F4960020 2017-08-25 Tonzi CO2 Probe Tree2 Profile -0.08 Removed Lab none Removed from Tonzi Tree 2 profile at -8 cm position for calibration
GMP222 K1310104 2017-08-25 Tonzi CO2 Probe Tree2 Profile -0.04 Removed Lab none Removed from Tonzi Tree 2 profile at -4 cm position for calibration
GMT221 G0150036 2017-08-25 Tonzi CO2 Probe Tree1 Profile -0.08 Removed Lab none Removed from Tonzi Tree 1 profile at -4 or -8 cm position for calibration. Probe labeled 8.
GMT221 G0150035 2017-08-25 Tonzi CO2 Probe Tree1 Profile -0.04 Removed Lab none Removed from Tonzi Tree 1 profile at -4 or -8 cm position for calibration. Probe labeled 16.
GMM222 L2010069 2017-08-25 Tonzi CO2 Probe Tree1 Profile -0.02 Removed Lab none Removed from Tonzi Tree 1 profile at -2 cm position for calibration
GMP221 K1340006 2017-08-25 Tonzi CO2 Probe Open Profile -0.08 Removed Lab none Removed from Tonzi Open profile at -8 cm position for calibration
GMP222 Z2720015 2017-08-25 Tonzi CO2 Probe Open Profile -0.04 Removed Lab none Removed from Tonzi Open profile at -4 cm position for calibration
GMP222 K1310102 2017-08-25 Tonzi CO2 Probe Open Profile -0.02 Removed Lab none Removed from Tonzi Open profile at -2 cm position for calibration
SRS-Nr NDVI 943103015 2017-08-25 Lab none Installed Tonzi-Tower Installed at Tonzi tower in the outgoing (down) position, pointing toward the grass by the tower.
SRS-Ni NDVI 1066703767 2017-08-25 Lab none Installed Tonzi-Tower Installed at Tonzi tower in the outgoing (down) position.
SRS-Ni NDVI 2046503398 2017-08-25 Lab none Installed Tonzi-Tower Installed at Tonzi tower in the incoming (up) position.
PAR-LITE 010164 2017-08-25 Lab none Installed Tonzi-Tower Installed at Tonzi tower to replace an old PAR sensor sn 010159 that was reading high. It was installed in the incoming (up) position.
PAR-LITE 010159 2017-08-25 Tonzi-Tower Removed Lab none Removed from Tonzi Tower. This sensor has been reading high for a long time. The cable is decayed and the sensor needs to be re-calibrated. This sensor was entered into the database.
PAR-LITE 990017 2017-08-25 none Notes none Removed incoming PAR-Lite from Tonzi tower. It was sn 010159. Not sure where PAR-Lite sn 990017.
CR23X 2300 2017-08-25 Tonzi-Tower Removed Lab none Removed from Tonzi tower met as part of an upgrade to move met sensors from the CR23X to a CR1000.
CMP 11 996981 2017-08-25 Tonzi-Tower Removed Lab none Removed from Tonzi Tower and brought back to the lab. It is a redundant measurement with the 4-way radiometer. Cables are decayed and need to be replaced.
PAR-LITE 060868 2017-08-25 Tonzi-Tower Removed Lab none Removed this sensor as part of a rotating shadow band ensemble from Tonzi Tower because it is a redundant measurement with the BF5.
NR-LITE 990351 2017-08-25 Tonzi-Tower Removed Lab none Removed from Tonzi Tower and brought back to the lab. It is a redundant measurement with the 4-way radiometer. Cables are decayed and need to be replaced.
CR23X 2300 2009-02-02 Lab none Calibration Lab none Calibrated at CSI. Due Date 02 Feb 11.
HMP60 L2310711 2017-08-25 Tonzi Removed Lab none Removed from Tonzi and brought back to the lab. It was installed for a comparison with an HMP45. The sensor is ready to use.
GMM222 L2010070 2017-08-25 Vaira CO2 probe North Profile -0.08 Removed Lab none Removed from Vaira North profile, -8 position and brought back to the lab for calibration
eosFD FD20160001 2017-08-28 Factory Received Lab none Received FD chamber back in the lab. No documentation about what they inspected/fixed.
eosFD FD20160001 2017-06-23 Twitchell Alfalfa Forced Diffusion Removed Lab none Removed from Twitchell Alfalfa and brought back to lab to send back to Eosense
CR1000 47108 2017-08-23 Twitchell Alfalfa Data Logger Met Upgrade Twitchell Alfalfa Data Logger Met The OS was upgraded from probably v24 to v31.08. OS v31.08 is FAT32 and reset the CR1000 to factory settings.
LI-7550 AIU-1557 2017-08-23 Twitchell Alfalfa Interface LI-7550 Upgrade Twitchell Alfalfa Interface LI-7550 The firmware was upgraded from v8.0.0 to v8.7.5
CR23X 2299 2017-08-17 Vaira Data Logger Met Removed Lab none Removed from Vaira where it was collecting met data. It was replaced by a CR1000.
NR-LITE 990353 1999-07-29 Factory Calibration Factory Initial factory calibration: 12.3 uV/W/m2. Recommend recal every two years.
SRS-Ni NDVI 2046503344 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to E indicating a hemispherical incoming orientation and the head was labeled with sn, address, and 'out'.
SRS-Ni NDVI 2046503376 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to E indicating a hemispherical outgoing orientation and the head was labeled with sn, address, and 'out'.
SRS-Ni NDVI 2046503354 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to E indicating a hemispherical incoming orientation and the head was labeled with sn, address, and 'out'.
SRS-Ni NDVI 2046503357 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to E indicating a hemispherical outgoing orientation and the head was labeled with sn, address, and 'out'.
SRS-Ni NDVI 2046503404 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to E indicating a hemispherical outgoing orientation and the head was labeled with sn, address, and 'out'.
SRS-Ni NDVI 1066703780 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to C indicating a hemispherical incoming orientation and the head was labeled with sn, address, and 'in'.
SRS-Ni NDVI 2046503390 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to C indicating a hemispherical incoming orientation and the head was labeled with sn, address, and 'in'.
SRS-Ni NDVI 1066703767 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to E indicating a hemispherical incoming orientation and the head was labeled with sn, address, and 'out'.
SRS-Ni NDVI 2046503398 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to C indicating a hemispherical incoming orientation and the head was labeled with sn, address, and 'in'.
SRS-Ni NDVI 2046503402 2017-08-17 Lab none Installed Vaira Narrow Band Reflected Installed on the Vaira Tower in the outgoing position
SRS-Ni NDVI 2046503368 2017-08-17 Lab none Installed Vaira Narrow Band Incoming Installed on the Vaira Tower in the incoming position.
NR-LITE 990353 2017-08-17 Vaira Net Radiometer Removed Lab none Removed from the Vaira tower with the upgrade on the data logger. This sensor is redundant with the 4-way CNR1 and is very old.
LI-7500A 75H-2514 2017-08-17 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira Tower to replace sn 0035 for calibration
LI-7500A 0035 2017-08-17 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower for calibration. Replaced by sn 75H-2514.
LI-7550 AIU-1870 2017-08-17 Vaira Interface LI-7550 Upgrade Vaira Interface LI-7550 We upgrade the firmware from v8.5.0 to v8.75
LI-7550 AIU-1866 2017-08-17 Tonzi-Tower Upgrade Tonzi-Tower The firmware was upgraded from v8.5.0 to v8.7.5
LI-7500A 75H-2182 2017-08-17 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower to replace sn 0041 for calibration
LI-7500A 0041 2017-08-17 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower for calibration. Replaced by sn 75H-2182
LI-7500A 75H-2180 2017-08-17 Lab none Installed Tonzi-Floor Installed on the Tonzi Floor to replace sn 0042 for calibration
LI-7500A 0042 2017-08-17 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 75H-2180
LI-7550 AIU-1869 2017-08-17 Tonzi-Floor Upgrade Tonzi-Floor The firmware was upgraded from v8.5.0 to v8.7.5
GP-001 GP20140017 2017-08-15 Lab none Tested Lab none Tested in the lab with a CR1000. Scan every second and record every minute. Temperature seems okay (~3000 mV) but CO2 stays high (~5000 mV).
GP-001 GP20150109 2017-08-15 Lab none Tested Lab none Tested in the lab with a CR1000. Scan every second and record every minute. Temperature seems okay (~3000 mV) but CO2 stays low (~25 mV). It responds a little bit when I blow into the sensor, but not much (from 25-100 mV, over a range of 0-5000 mV).
LI-7500A 75H-2669 2017-08-14 Lab none Calibration Lab none Last calibration: 2017-08-07
AGC=103.2 SDM=0 DAC1=0 DAC2=0
Ambient P=100.2kPa, Tair=21.9C, Tdew=12.8C
Zero Air at 1100psi, AMP Air 408.97 at 2000 Tdewspan=16.00
CO2 zero 0.9189 -> .9187 Offset: -1.86ppm
CO2 span 1.0159 -> 1.0144 Offset: -1.51ppm
H2O zero 0.8971 -> .8988 Offset: 6.3ppm
H2O span 1.0282 -> 1.0091 Offset: -0.08ppm
Notes: Previous calibration did not have updated pressure coefficients from 7550 interface, so I re-did the calibration
LI-7500A 75H-2667 2017-08-14 Lab none Calibration Lab none Last calibration: 2017-08-07
AGC=105.8 SDM=0 DAC1=0 DAC2=0
Ambient P=100.2kPa, Tair=21.7C, Tdew=12.8C
Zero Air at 1100psi, AMP Air 408.97 at 2000 Tdewspan=16.00
CO2 zero 0.9419 -> 0.9418 Offset: -0.45ppm
CO2 span 1.0187 -> 1.0162 Offset: 1.31ppm
H2O zero 0.8382 -> 0.8392 Offset: 5.9ppm
H2O span 1.0141 -> 1.0214 Offset: -0.15ppm
Notes: Previous calibration on 2017-08-07 did not use updated pressure coefficients from 7550 interface, so I redid the calibration.
LI-7500A 75H-2514 2017-08-14 Lab none Calibration Lab none Last calibration: 2017-08-03
AGC=102.4 SDM=0 DAC1=0 DAC2=0
Ambient P=100.22kPa, Tair=19.9C, Tdew=13.3C
Zero Air at 1100psi, AMP Air 408.97 at 2000 Tdewspan=16.00
CO2 zero 0.9038 -> .9050 Offset: 4.58ppm
CO2 span 1.018 -> 1.0107 Offset: 3.72ppm
H2O zero 0.8384 -> .8378 Offset: -4.7ppm
H2O span 1.0081 -> 1.0094 Offset: 0.11ppm
Notes:
LI-7500A 75H-2669 2017-08-07 Lab none Calibration Lab none Last calibration: 2017-08-01
AGC=103.5 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=22.9C, Tdew=12.9C
Zero Air at 1150psi, AMP Air 408.97 at 2100 Tdewspan=16.00
CO2 zero 0.9199 -> .9189 Offset: -.7ppm
CO2 span 0.9533 -> 1.0159 Offset: -2.44ppm
H2O zero 0.9118 -> .8971 Offset: -4.7ppm
H2O span 0.8978 -> 1.0282 Offset: -.32ppm
Notes: Previous calibration did not have correct 5th-order-polynomials, which is why the offsets were huge.
LI-7500A 75H-2667 2017-08-07 Lab none Calibration Lab none Last calibration: 2017-08-01
AGC=105.4 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=21.5C, Tdew=13.7C
Zero Air at 1150psi, AMP Air 408.97 at 2100 Tdewspan=16.00
CO2 zero 0.9443 -> .9419 Offset: -1.41ppm
CO2 span 0.9346 -> 1.0187 Offset: -2.3ppm
H2O zero 0.8578 -> .8382 Offset: .9ppm
H2O span 0.9724 -> 1.0141 Offset: 0.16ppm
Notes: Previous calibration did not have correct 5th-order-polynomials, which is why the offsets were huge.
LI-7500A 75H-2514 2017-08-03 Lab none Calibration Lab none Last calibration: 2017-06-20
AGC=101.4 SDM= DAC1= DAC2=
Ambient P=101.3kPa, Tair=24.9C, Tdew=14.3C
Zero Air at 2100psi, AMP Air 408.97 at 2100 Tdewspan=16.00
CO2 zero 0.9041 -> .9038 Offset: -1.36ppm
CO2 span 1.0115 -> 1.018 Offset: -3.31ppm
H2O zero 0.8378 -> .8384 Offset: 4.7ppm
H2O span 1.0019 -> 1.0081 Offset: -.34ppm
Notes:
PS-30M 22678-0999 2014-09-10 Twitchell Rice Removed Lab none Removed from Twitchell Rice. This is the old looking controller and has been slow to start charging batteries. Replace with a new charge controller.
PS-30M 22678-0999 2017-08-07 Lab none Disposed Disposed I decided it time to dispose on this controller. I put it in the dead electronics box.
PS-30M 22678-0999 2017-06-14 Lab none Installed East End-Temp2 Solar Controller Installed at the Temporary Tower 2 Closed Canopy location.
PS-30M 22678-0999 2017-08-04 East End-Temp2 Solar Controller Removed Lab none Removed from the East End Temporary Tower 2 Closed Canopy location. The Solar input terminals looked fried and it was no longer providing power to the tower.
GP-001 GP20140017 2017-08-02 West Pond CO2 Probe B Removed Lab none Removed from West Pond due to railed high CO2 readings and no change when removed from the water. Temperature looked okay
LI-7500A 75H-2514 2017-08-02 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed for calibration and replaced by 75H-2182
LI-7500A 75H-2182 2017-08-02 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Replaced 75H-2514 at Sherman Wetland.
LI-7500A 75H-2669 2017-08-01 Lab none Calibration Lab none Last calibration: 2017-03-21
AGC=100.1 SDM= DAC1= DAC2=
Ambient P=100.2kPa, Tair=24.0C, Tdew=15.8C
Zero Air at 1200psi, LBL Air 408.97 at 2100 Tdewspan=16.00
CO2 zero 0.9191 -> .9199 Offset: 3.65ppm
CO2 span 1.0115 -> .9533 Offset: 33.15ppm
H2O zero 0.8978 -> .9118 Offset: 78.4ppm
H2O span 1.0138 -> .8978 Offset: 2.66ppm
Notes: Manually entered in existing coefficients before the calibration because did not have existing config file (usual laptop was elsewhere). Maybe I had entered the coefficients incorrectly in the database?
LI-7500A 75H-2667 2017-08-01 Lab none Calibration Lab none Last calibration: 2017-03-21
AGC=100 SDM= DAC1= DAC2=
Ambient P=100.1kPa, Tair=23.2C, Tdew=19.4C
Zero Air at 1200psi, LBL Air 408.97 at 2100 Tdewspan=16.00
CO2 zero 0.9423 -> .9443 Offset: 8.88ppm
CO2 span 1.0144 -> .9346 Offset: 59.3ppm
H2O zero 0.838 -> .8578 Offset: 144.2ppm
H2O span 1.0219 -> .9724 Offset: 1.09ppm
Notes: Reset signal strength from 113 to 100. H2O at ambient was about 900 mmol/m3, which seems higher than usual. Manually entered in existing coefficients before the calibration because did not have existing config file (usual laptop was elsewhere). Maybe I had entered the coefficients incorrectly in the database?
CR10X 31308 2017-07-26 Bouldin Corn Data Logger Diffuse PAR Location Bouldin Corn Data Logger Diffuse PAR Check the serial number of the data logger at the site.
LI-7500A 75H-2668 2017-07-26 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Replaced 75H-2182.
LI-7500A 75H-2182 2017-07-26 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Brought down for calibration and replaced by 75H-2668.
GMT221 F4960018 2017-07-12 Twitchell Alfalfa Removed Lab none Removed from Twitchell Alfalfa because it was reading 0. It was in set A, 8 cm (the second depth).