Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 [30] 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65
Ser No.DateLoc FromTitleLoc ToNotes
LI-7500A 75H-2182 2018-10-24 Lab none Calibration Lab none Last calibration: 2018-07-10
AGC=87.3 SDM=0 DAC1=0 DAC2=0
Ambient P=100.55kPa, Tair=26.87C, Tdew=12.61C
Zero Air at 1500psi, AMP Air 408.97 at 1100 Tdewspan=16.00
CO2 zero 0.8841 -> .8844 Offset: 1.11ppm
CO2 span 1.0168 -> 1.0188 Offset: -1.04ppm
H2O zero 0.845 -> .8449 Offset: -.6ppm
H2O span 1.0202 -> 1.0143 Offset: .13ppm
Notes:
LI-7500A 75H-2150 2018-10-24 Lab none Calibration Lab none Last calibration: 2018-07-10
AGC=97.5 SDM=0 DAC1=0 DAC2=0
Ambient P=100.6kPa, Tair=26.7C, Tdew=11.95C
Zero Air at 1500psi, AMP Air 408.97 at 1100 Tdewspan=16.00
CO2 zero 0.9017 -> .9021 Offset: 1.86ppm
CO2 span 1.002 -> 1.0050 Offset: -1.74ppm
H2O zero 0.8662 -> .8660 Offset: -1.9ppm
H2O span 1.0069 -> 1.0069 Offset: 0.00ppm
Notes: No change in H2O span.
LI-7700 TG1-0424 2018-10-24 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2018-04-13
RSSI=82.9%, Optical RH=9%
Ambient P=100.6kPa, Tair=27.8C
Zero Air at 1500psi, AMP CH4 1.9497 at 1150
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: 0.043ppm
Notes: First calibration after replacing power board - no problems found.
LI-7700 TG1-0424 2018-10-22 Lab none Upgrade Lab none I swapped out power board pn 9677-079 per Licor's instructions. According to Licor the original boards Rev 0, could be damaged by exposure to voltage above 24VDC. The new board Rev 0A (V3) should not have that problem.
TE525MM 49187-109 2018-10-18 Sherman Barn Precipitation Calibration Sherman Barn Precipitation Got 94 of 100 tips - adjusted by a total of 3.5 turns CCW (screws were loose, so this total is suspicious; should have been a CW adjustment when all was said and done)

2018-10-18: Got 94 of 100 tips - adjusted by 0.5 turns CW.
2018-12-18: Got 117 of 100 tips - adjusted by 1 turn CCW, but screws feel very loose.
2019-01-03: Got 118 of 100 tips - adjusted by 3 turns CCW. (previous adjustment of 1 turn CCW did not seem to make a difference).
2019-01-24: Got 100 of 100 tips! No adjustments made.
LI-7500A 75H-2669 2018-10-18 Lab none Installed Sherman Barn Gas Analyzer CO2/H2O Installed on Sherman Barn to replace sn 0041 for calibration
LI-7500A 0041 2018-10-18 Sherman Barn Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from Sherman Barn for calibration. Replaced by sn 75H-2669.
GP-001 GP20140016 2018-10-18 Sherman Wetland CO2 Probe Removed Lab none Removed from Sherman Wetland due to railed readings not responding the atmospheric CO2.
TE525MM 65552-615 2018-10-18 Sherman Wetland Precipitation Calibration Sherman Wetland Precipitation Got 94 of 100 tips - adjusted screws a total of 1.5 turns CW

2018-10-18: Got 94 of 100 tips - adjusted screws 0.5 turns CW
2018-10-25: Got 96 of 100 tips - adjusted set screws 1 turn CW.
2018-12-18: 102 tips - no adjustment.
TR-525M 50229-112 2018-10-18 East End Precipitation Calibration East End Precipitation Got 94 of 100 tips - adjusted screws a total of 0.5 turns CW.

2018-10-28: Got 98 of 100 tips and adjusted set screws 0.5 turn CW.
2018-11-01: Got 93 of 100 tips - no adjustment and need to retest
2018-12-18: 98 of 100 tips - no adjustment.
LI-7700 TG1-0424 2018-10-18 Bouldin Alfalfa Gas Analyzer CH4 Removed Lab none Removed from Bouldin Alfalfa for power board upgrade and to swap in for other two 7700s needing upgrade.
LI-7500A 75H-2667 2018-10-18 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed on Bouldin Alfalfa to replace sn 74H-2150 for calibration.
LI-7500A 75H-2150 2018-10-18 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa for calibration. Replaced by sn 75H-2667.
TE525MM 65560-615 2018-10-18 Bouldin Alfalfa Precipitation Calibration Bouldin Alfalfa Precipitation Got 90 of 100 tips. Adjusted screws a total of 2.5 turns CW.

2018-10-18: Got 90 tips and adjusted screws 1 turn CW.
2018-11-01: Got 91 tips and adjusted screws 2 turns CCW accidentally (should have been CW).
2018-12-13: Got 86 tips and adjusted screws 4.5 turns CW.
2018-12-18: Got 107 tips and adjusted screws 1 turn CCW.
LI-7500A 75H-2668 2018-10-18 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed on Bouldin Corn to replace sn 75H-2182 for calibration
LI-7500A 75H-2182 2018-10-18 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 75H-2668.
TE525MM 65548-615 2018-10-18 Bouldin Corn Precipitation Calibration Bouldin Corn Precipitation Got 94 of 100 tips. Adjusted a total of 1 turn CW.

2018-10-18: Got 94 of 100 tips - adjusted screws 0.5 turns CW
2018-10-25: Got 97 tips and adjusted set screws 0.5 turn CW.
2018-11-01: 99 of 100 tips - no adjustment.
LI-7500A 0418 2018-10-17 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on Vaira to replace sn 0065 that had no CO2, H2O or signal strength.
LI-7500A 0065 2018-10-17 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira due to no CO2, H2O or Signal Strength - maybe bad source or detector. Replaced by sn 0418
TR-525I-R1 24875-999 2018-10-17 Vaira Precipitation Calibration Vaira Precipitation Got 78 of 100 tips - adjusted screws 3 turns CW
TE525 48132-611 2018-10-17 Tonzi Calibration Tonzi Got 134 of 100 tips - adjusted screws 4 turns CCW
TE525 48130-611 2018-10-17 Tonzi Calibration Tonzi Got 88 of 100 tips - adjusted screws 2.5 turns CW
TR-525I-R1 24872-999 2018-10-17 Tonzi Calibration Tonzi Got 88 of 100 tips - adjusted screws 2.5 turns CW
LI-7500A 75H-2669 2018-10-10 Lab none Calibration Lab none Last calibration: 2018-10-09
AGC=102 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=26.6C, Tdew=12.9C
Zero Air at 1400psi, AMP Air 408.97 at 1100 Tdewspan=16.00
CO2 zero 0.918 -> .9188 Offset: 3.4ppm
CO2 span 1.0125 -> 1.0106 Offset: .9ppm
H2O zero 0.8989 -> .8992 Offset: 1.6ppm
H2O span 1.0049 -> 1.0034 Offset: .04ppm
Notes: Re-calibrated to check Katrina's calibration yesterday, which looks good.
LI-7500A 75H-2669 2018-10-09 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=103 SDM=0 DAC1=0 DAC2=0
Ambient P=99.88kPa, Tair=27.1C, Tdew=12.9C
Zero Air at 1425psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9182 -> .9180 Offset: -.8ppm
CO2 span 1.0125 -> 1.0125 Offset: 2.2ppm
H2O zero 0.8994 -> .8989 Offset: 52.7ppm
H2O span 1.0062 -> 1.0049 Offset: 16.02ppm
Notes: Calibrated by Katrina. The H2O offset is large, but the change is coefficients is small, so perhaps Katrina wrote down the wrong offset.
LI-7500A 75H-2667 2018-10-10 Lab none Calibration Lab none Last calibration: 2018-10-09
AGC=104 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=16.7C, Tdew=11.5C
Zero Air at 1400psi, AMP Air 408.97 at 1100 Tdewspan=16.00
CO2 zero 0.9413 -> .9412 Offset: -.5ppm
CO2 span 1.0127 -> 1.0145 Offset: -1.57ppm
H2O zero 0.8362 -> .8367 Offset: 3.3ppm
H2O span 1.0328 -> 1.0218 Offset: .22ppm
Notes: Re-calibrated to check Katrina's calibration yesterday, which looks good.
LI-7500A 75H-2667 2018-10-09 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=104 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=27.7C, Tdew=9.5C
Zero Air at 1450psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9414 -> .9413 Offset: -.42ppm
CO2 span 1.0162 -> 1.0127 Offset: -87.77ppm
H2O zero 0.8375 -> .8362 Offset: -7.4ppm
H2O span 1.0237 -> 1.0328 Offset: -.19ppm
Notes: Calibrated by Katrina. The CO2 span offset is large, but the change is coefficients is small, so perhaps Katrina wrote down the wrong offset.
LI-7500A 0418 2018-10-10 Lab none Calibration Lab none Last calibration: 2018-04-13
AGC=93.2 SDM=0 DAC1=0 DAC2=0
Ambient P=99.98kPa, Tair=29.11C, Tdew=15.00C
Zero Air at 1500psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9001 -> .8982 Offset: -7.43ppm
CO2 span 1.0057 -> .9965 Offset: 4.86ppm
H2O zero 0.8426 -> .8454 Offset: 16.8ppm
H2O span 0.9892 -> .9813 Offset: 0.16ppm
Notes: calibrated by Katrina
LI-7500A 75H-2668 2018-10-08 Lab none Calibration Lab none Last calibration: 2018-10-02
AGC=101 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=29.3C, Tdew=5.8C
Zero Air at 1500psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9389 -> .9394 Offset: 2.06ppm
CO2 span 1.0172 -> 1.0164 Offset: .38ppm
H2O zero 0.9072 -> .9057 Offset: -8.2ppm
H2O span 1.0238 -> 1.0157 Offset: .17ppm
Notes: Second calibration since chemical swap. Sensor had been running for 5 days, since 10/3 when it had a huge H2O zero offset when analyzing zero gas. I'm not sure why it was reading weird values on 10/3, but the numbers seem ok today, consistent with a chemical swap. In the future, we should let the sensor run for 1 week after swapping internal chemicals, rather than calibrating after only 24 hours.
LI-7500A 75H-2180 2018-10-04 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed on Sherman Wetland to replace sn 75H-2667 for calibration.
LI-7500A 75H-2667 2018-10-04 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Wetland for calibration. Replaced by sn 75H-2180.
LI-7500A 0035 2018-10-04 Lab none Installed East Pond Gas Analyzer CO2/H2O Installed on the East Pond tower to replace sn 0418 for calibration.
LI-7500A 0418 2018-10-04 East Pond Gas Analyzer CO2/H2O Removed Lab none Removed from the East Pond tower for calibration. Replaced by sn 0035.
Mark I A001911 2018-10-04 Lab none Installed Bouldin Corn Weather Station A Installed on Bouldin Corn to replace sn A000251 which is to be sent back to Arable for inspection of backup temperature sensor. Later we realized we had sent back the incorrect Arable; they had requested sn A000591.
LI-7500A 1594 2018-10-04 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry tower to replace sn 75H-2669 for calibration.
LI-7500A 75H-2669 2018-10-04 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for calibration. Replaced by sn 1594
LI-7500A 75H-2668 2018-10-03 Lab none Tested Lab none Tested with zero air and the most recent configuration file from 2018-10-02, but CO2 was reading 7.2 umol/mol and H2O was reading -50 mmol/m3. I'm going to let it run a few more days before checking again.
LI-7500A 75H-2668 2018-10-02 Lab none Calibration Lab none Last calibration: 2018-09-26
AGC=101 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=26.8C, Tdew=15.9C
Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=20.00
CO2 zero 0.9392 -> .9389 Offset: -1.47ppm
CO2 span 1.0153 -> 1.0172 Offset: -1.09ppm
H2O zero 0.9056 -> .9072 Offset: 9.2ppm
H2O span 1.0166 -> 1.0238 Offset: -.16ppm
Notes: First calibration after swapping out internal chemicals. Kept the sensor on for 3 days after the swap before the calibration.
LI-7500A 1594 2018-10-03 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=98.6 SDM=0 DAC1=0 DAC2=0
Ambient P=100.16kPa, Tair=28.4C, Tdew=15.18C
Zero Air at 1500psi, AMP Air 408.97 at 1200 Tdewspan=20.00
CO2 zero 0.8648 -> .8644 Offset: -1.74ppm
CO2 span 1.0181 -> 1.0150 Offset: 1.68ppm
H2O zero 0.8869 -> .8870 Offset: 0.8ppm
H2O span 1.0106 -> 1.0056 Offset: .11ppm
Notes: Spanned H2O at 20.00 because initial dew point temperature in the room was > 16.00 degC.
LI-7500A 0035 2018-10-02 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=98.9 SDM=0 DAC1=0 DAC2=0
Ambient P=99.95kPa, Tair=26.66C, Tdew=16.36C
Zero Air at 1550psi, AMP Air 408.96 at 1200 Tdewspan=20.00
CO2 zero 0.9374 -> .9375 Offset: .45ppm
CO2 span 1.0296 -> 1.0142 Offset: 1.57ppm
H2O zero 0.8979 -> .8981 Offset: 2.3ppm
H2O span 1.0261 -> 1.0277 Offset: -.04ppm
Notes: Spanned H2O at 20.00 because initial dew point temperature in the room was > 16.00 degC.
LI-7500A 75H-2180 2018-10-02 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=98.6 SDM=0 DAC1=0 DAC2=0
Ambient P=99.86kPa, Tair=27.09C, Tdew=17.03C
Zero Air at 1500psi, AMP Air 408.96 at 1200 Tdewspan=20.00
CO2 zero 0.9141 -> .9137 Offset: -1.7ppm
CO2 span 1.0267 -> 1.023 Offset: 1.86ppm
H2O zero 0.8865 -> .883 Offset: 11.6ppm
H2O span 1.0274 -> .9893 Offset: 4.48ppm
Notes: Calibrated by Katrina. Spanned H2O at 20.00 because initial dew point temperature in the room was > 16.00 degC.
PAR-LITE 010159 2018-09-24 Lab none Repair Lab none Trimmed old cable, soldered on ~10m cable, and secured cable-sensor junction with small L-bracket and zip-ties.
LI-7500A 75H-2668 2018-09-28 Lab none Maintenance Lab none Swapped out internal chemicals (Ascarite II and Mg(ClO4)2) because the last three calibrations have H2O zero offsets of +/-10 deg C, which is larger than expected. I flushed the analyzer housing with zero air for a few minutes before replacing the chemical bottles.
CR6 2115 2018-09-28 Lab none Upgrade Lab none Upgraded OS from CR6.Std.07.02 to CR6.Std.08.00. After upgrading the OS, I used the "Network Services" tab to enable FTP with the username tower and the usual password.
LI-7500A 75H-2514 2018-09-27 Lab none Installed Tonzi-Floor Installed on the Tonzi Tower to replace sn 0035 for calibration.
LI-7500A 0035 2018-09-27 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from Tonzi Tower for calibration. replaced by sn 75H-2514
LI-7500A 0073 2018-09-27 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed on the Tonzi Tower to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2018-09-27 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi Tower for calibration. replaced by sn 0073
LI-7500A 0065 2018-09-27 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira Tower to replace sn 1594 for calibration.
LI-7500A 1594 2018-09-27 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower for calibration. Replaced by sn 0065.
LI-7500A 0065 2018-09-26 Lab none Calibration Lab none Last calibration: 2018-04-04
AGC=99.7 SDM=0 DAC1=0 DAC2=0
Ambient P=100.4kPa, Tair=26.6C, Tdew=13.8C
Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.8905 -> .8903 Offset: -1.09ppm
CO2 span 1.0231 -> 1.0184 Offset: 2.48ppm
H2O zero 0.7082 -> .7086 Offset: 2.7ppm
H2O span 1.0185 -> 1.0155 Offset: .07ppm
Notes:
LI-7500A 0073 2018-09-26 Lab none Calibration Lab none Last calibration: 2018-04-13
AGC=99.7 SDM=0 DAC1=0 DAC2=0
Ambient P=100.5kPa, Tair=25.8C, Tdew=13.3C
Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9575 -> .9569 Offset: -2.02ppm
CO2 span 1.0145 -> 1.0112 Offset: 1.56ppm
H2O zero 0.7295 -> .7314 Offset: 2.0ppm
H2O span 1.0131 -> 1.0183 Offset: -.12ppm
Notes:
LI-7500A 75H-2514 2018-09-26 Lab none Calibration Lab none Last calibration: 2018-04-18
AGC=100.1 SDM=0 DAC1=0 DAC2=0
Ambient P=100.3kPa, Tair=27.3C, Tdew=12.7C
Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.906 -> .9045 Offset: -6.04ppm
CO2 span 1.0106 -> 1.0036 Offset: 3.52ppm
H2O zero 0.8368 -> .8370 Offset: 1.3ppm
H2O span 1.016 -> 1.0165 Offset: -.01ppm
Notes:
LI-7500A 75H-2668 2018-09-26 Lab none Calibration Lab none Last calibration: 2018-04-04
AGC=100.0 SDM=0 DAC1=0 DAC2=0
Ambient P=100.4kPa, Tair=26.5C, Tdew=13.8C
Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9393 -> 0.9392 Offset: -.27ppm
CO2 span 1.0147 -> 1.0153 Offset: -.27ppm
H2O zero 0.9038 -> .9056 Offset: 10.6ppm
H2O span 1.0146 -> 1.0166 Offset: .05ppm
Notes: "Check PLL" error for the first 30 seconds, then status changed to "OK."
LAI-2000 PCA-0846 2018-09-19 Lab none Loaned Loaned Loaned to Sebastian Wolf sewolf@ethz.ch for a few weeks for his work a Sagehen.
LAI-2000 PCA-1478 2018-09-20 Lab none Damaged Lab none The push button on the wand is failing.
GP-001 GP20150109 2018-09-18 Lab none Installed East Pond CO2 Probe Installed at East Pond hanging just below the water surface below the data logger box.
CR10X 47253 2018-05-23 Tonzi Removed Lab none Removed from the Tonzi Tower with the old CO2 profile system. New Vaisala profile system is running
LI-820 CGA-556 2018-05-23 Tonzi Removed Lab none Removed from the Tonzi Tower with the old CO2 profile system. New Vaisala profile system is running
LED-NDVI LED-05 2018-05-23 Tonzi Removed Lab none Removed from the Tonzi Tower because it has not been working for a while and there are Decagon SRS sensors from NDVI and PRI
PQS 1 161073 2017-08-15 Factory Calibration Factory y = 209.39*x; Calibrated by Ameriflux at OSU. 4.78 uV/(umol/m2/s). 1000/4.78 = 209.39 (umol/m2/s)/mV
AM16-32A 1238 2018-07-12 Loaned Disposed Disposed Given to Tibisay Perez in Whendee Silver's lab in exchange for a new AM16/32B sn 27703
LI-190SA Q29881 2007-01-19 Tonzi PAR Sensor Incoming Removed Lab none Removed from Tonzi Tower top.
LI-190SA Q29892 2007-01-19 Tonzi Removed Lab none Removed from Tonzi Tower top.
LI-190SA Q29892 2001-07-10 Factory Calibration Factory y = 248.59*x; Initial factory calibration 6.66 uA/(umol/m2/s). Using a 604Ohm millivolt adapter 1000/(6.66*0.604) = 248.59 (umol/m2/s)/mv
LI-190SA Q29892 2001-09-07 Lab none Installed Tonzi Installed on the Tower top in the outgoing position
LI-190SA Q29881 2001-09-07 Lab none Installed Tonzi PAR Sensor Incoming Installed on the Tower top in the incoming position
LI-190SA Q29881 2001-07-10 Factory Calibration Factory y = 248.97*x; Initial factory calibration 6.65 uA/(umol/m2/s). Using a 604Ohm millivolt adapter 1000/(6.65*0.604) = 248.97 (umol/m2/s)/mv
CNR1 990181 2001-05-20 Lab none Installed Vaira Net Radiometer 4-way Installed on the Vaira tower
NR-LITE 990351 2001-05-20 Lab none Installed Tonzi Installed on the Tower top with the initial set up of the site
PAR-LITE 990017 2001-05-20 Lab none Installed Tonzi Installed on the Tower top in the reflected PAR location with the initial set up of the site
PAR-LITE 990019 2001-05-20 Lab none Installed Tonzi Installed on the Tower Top in the incoming PAR position.
PAR-LITE 990021 2001-05-20 Lab none Installed Tonzi Installed on the Tower top as a calibration sensor
eosFD 20160002 2012-10-17 Lab none Troubleshooting Lab none This chamber was installed at the Tonzi soil co2 profile at the Open location but did not start up. I brought it back in to test. Eosense mentioned a “heat-start-up issue” but thought it had been fixed. It tested fine in the lab.
HMP45AC T2130005 2018-08-23 Lab none Installed Sherman Wetland RH & Temp Installed at SW to replace the other HMP45 with a too short cable. Mounted in the original location. 1% slope in T and 6% slope in RH from TA inter-comparison.
HMP45AC W0730036 2018-08-23 Sherman Wetland RH & Temp Removed Lab none Removed from SW because its cable was to short to allow it to be mounted in the correct place.
HMP45AC T1250012 2018-08-23 Lab none Installed Bouldin Alfalfa RH & Temp Installed on BA to replace an HMP60 with low temperature readings. This sensor was in the TA inter-comparison T slope was 1.0462 and RH slope was 1.0369.
HMP60 L1740398 2018-08-23 Bouldin Alfalfa RH & Temp Removed Lab none Removed from BA due to low temperature readings. Several other HMP60s have similarly failed.
eosFD FD20160001 2018-08-07 Lab none Installed Bouldin Alfalfa Forced Diffusion A Installed under the BA scaffolding over bare soil.
HMP45AC W0730036 2018-08-07 Lab none Installed Sherman Wetland RH & Temp Installed at SW to replace HMP60 sn L2310736. The HMP60 was reading numbers lower than reasonable.
HMP60 L2310736 2018-08-07 Sherman Wetland RH & Temp Removed Lab none Removed from SW because it has been reading too low. It was replaced with HMP45 sn W0730036.
Raven XT 1131631861 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
NR01 1514 2012-01-09 Lab none Installed Tonzi Installed at -8cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
HFP01SC 2500 2012-01-09 Lab none Installed Tonzi Installed at about 1m agl at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
HFP01SC 2495 2012-01-09 Lab none Installed Tonzi Installed at -8cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX350J T21009229B37030 2012-01-09 Lab none Installed Tonzi Installed at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
RF401 35785 2012-01-09 Lab none Installed Tonzi Installed in the Campbell Enclosure at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15731 2012-01-09 Lab none Installed Tonzi Installed at -60cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15729 2012-01-09 Lab none Installed Tonzi Installed at -20cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15802 2012-01-09 Lab none Installed Tonzi Installed at -10cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15852 2012-01-09 Lab none Installed Tonzi Installed at -5cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15730 2012-01-09 Lab none Installed Tonzi Installed at -40cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
SX229 15764 2012-01-09 Lab none Installed Tonzi Installed at -80cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TE 5TE-1400 2012-01-09 Lab none Installed Tonzi Installed at -80cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TE 5TE-1274 2012-01-09 Lab none Installed Tonzi Installed at -60cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TE 5TE-1275 2012-01-09 Lab none Installed Tonzi Installed at -40cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TE 5TE-1393 2012-01-09 Lab none Installed Tonzi Installed at -20cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TE 5TE-1281 2012-01-09 Lab none Installed Tonzi Installed at -5cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TE 5TE-1287 2012-01-09 Lab none Installed Tonzi Installed at -10cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.
STD-5TE 5TE-1278 2012-01-09 Lab none Installed Tonzi Installed at -2cm at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.