Equipment and Calibration History
Page: 1 2 3 4 5 6 7 8 9 [10] 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65Ser No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
SP75 020854E-6-03-02-06905 | 2022-05-04 | Lab none | Repair | Lab none | This panel had been under (salt) water at Yampah. The junction box on the back was all rotted out. I scraped the box off and the four junctions to the two 6V cell strings were still good. I soldered wires to the junction pads and added a screw terminal with two diodes. The panel was 20V and 4.5A in the sun. I cut the structure out of the junction box and reglued it to the back on the panel to protect the wiring and screw terminal. I think this panel is good to go. |
PS-30M 186580399 | 2022-05-04 | Lab none | Disposed | Disposed none | I took this controller apart and discovered that it had been partially submerged in salt water at Yampah. I cleaned it up as best I could and connected it to a battery. It's bricked - not even a puff of smoke. Put it in the e-waste box. |
KC40 04592G0181 | 2021-11-05 | Bouldin Alfalfa-ADEX Solar Panel Bouncer A | Removed | Lab none | Removed from BA bouncer tower at the conclusion of this summer's advection setup. |
KC40 04592G0181 | 2021-07-20 | Lab none | Installed | Bouldin Alfalfa-ADEX Solar Panel Bouncer A | Installed at the Bouncer tower for the advection experiment |
KC40 04592G0181 | 2010-08-19 | Tonzi none | Removed | Lab none | Removed from the Tram charging station so that two 20W panels could be use for a 24V system. |
PS-15 03220377 | 2021-11-05 | Bouldin Alfalfa-ADEX Solar Controller Bouncer | Removed | Lab none | Removed from BA bouncer tower at the conclusion of this summer's advection setup. |
LI-7550 AIU-2312 | 2022-04-24 | Lab none | Upgrade | Lab none | Downgraded firmware from 8.9.0 to 8.8.32. Stephen said I should feel free to downgrade it if I need it to be compatible with my windows interface. |
CR6 2115 | 2022-04-22 | Lab none | Maintenance | Lab none | The internal lithium battery was low <1V. I replaced it with a new Lithium 3.6V AA battery. |
CR6 2114 | 2022-04-22 | Lab none | Maintenance | Lab none | The internal lithium battery was low ~0.9V. I replaced it with a "slightly used" Lithium AA battery that was reading 3.2V on the voltmeter. |
KD140SX-UFBS 11ZV0M0836 | 2022-04-20 | Lab none | Installed | Mayberry none | Installed at MB to replace the 3 Shell solar panels that were removed today to be used for the advection experiment this summer. This panel has been resealed with Qsil and window film. |
SQ75 020854E-6-03-02-06825 | 2022-04-20 | Mayberry | Removed | Lab none | Removed from MB and brought back to the lab. It will be used for the advection experiment this summer. |
SQ75 024398E-1-26-04-99974 | 2022-04-20 | Mayberry | Removed | Lab none | Removed from MB and brought back to the lab. It will be used for the advection experiment this summer. |
SQ75 020854E-6-03-02-06832 | 2022-04-20 | Mayberry | Removed | Lab none | Removed from MB and brought back to the lab. It will be used for the advection experiment this summer. |
LI-7700 TG1-0155 | 2022-04-20 | Lab none | Installed | West Pond Gas Analyzer CH4 | Installed at WP to replace TG1-0421, which was removed last week for power issues. |
LI-7500A 0035 | 2022-04-20 | Lab none | Installed | Hill Slough Gas Analyzer CO2/H2O | Installed at Hill Slough to replace sn 75H-2669 for calibration. |
LI-7500A 75H-2669 | 2022-04-20 | Hill Slough Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Hill Slough for calibration and brought back to the lab. Replaced by sn 0035. |
WM 1590 152605 | 2022-04-20 | Lab none | Adjusted | Lab none | Settings adjusted to output at 20Hz instead of 1Hz and to output SOS instead of sonic temperature |
WM 1590 152606 | 2022-04-20 | Lab none | Adjusted | Lab none | Settings adjusted to output at 20Hz instead of 1Hz and to output SOS instead of sonic temperature |
WM 1590 152608 | 2022-04-20 | Lab none | Adjusted | Lab none | Settings adjusted to output at 20Hz instead of 1Hz and to output SOS instead of sonic temperature |
LI-7500A 75H-2176 | 2022-04-19 | Lab | Calibration | Lab | Last calibration: 2021-12-16 AGC=100.6 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=24.5C, Tdew=10.5C Zero Air at 700psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.867 -> .867 Offset: .13ppm CO2 span 1.0045 -> 1.0081 Offset: -2.65ppm H2O zero 0.8546 -> .8557 Offset: 6.7ppm H2O span 0.8249 -> .9228 Offset: .05ppm Notes: Surprisingly large change in H2OSpan's coefficient despite a small offset. Maybe double check this again before deploying. |
LI-7500A 0042 | 2022-04-19 | Lab | Calibration | Lab | Last calibration: 2021-12-09 AGC=100.4 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=25.8C, Tdew=7.5C Zero Air at 650psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9481 -> .9483 Offset: .85ppm CO2 span 1.0091 -> 1.0087 Offset: .14ppm H2O zero 0.8452 -> .8444 Offset: -4.9ppm H2O span 1.0213 -> 1.0191 Offset: .05ppm Notes: |
LI-7500A 0035 | 2022-04-19 | Lab | Calibration | Lab | Last calibration: 2022-01-11 AGC=99.4 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=25.4C, Tdew=8.2C Zero Air at 650psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9372 -> .9372 Offset: -.09ppm CO2 span 1.0128 -> 1.0167 Offset: -2.67ppm H2O zero 0.8986 -> .8983 Offset: -2.2ppm H2O span 1.0282 -> 1.0310 Offset: -.06ppm Notes: |
CR6 2115 | 2021-11-05 | Bouldin Alfalfa-ADEX Data Logger Upwind | Removed | Lab none | Removed from BA bouncer tower at the conclusion of this summer's advection setup. |
LI-7500A 1594 | 2022-04-13 | Lab none | Installed | Bouldin Alfalfa Heat Flux Plate B | Installed at Bouldin Alfalfa to replace sn 75H-2176 for calibration. |
LI-7500A 75H-2176 | 2022-04-13 | Bouldin Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Alfalfa for calibration. Replaced by sn 1594. |
LI-7700 TG1-0155 | 2022-04-13 | Bouldin Alfalfa Gas Analyzer CH4 | Removed | Lab none | Removed from Bouldin Alfalfa and brought back to the lab because we will use it to replace the faulty one from WP. I removed the power cable as well since the WP power cable could be bad. The wash power, wash tubing, and Ethernet cable are still there. |
LI-7500A 0065 | 2022-04-13 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0042 for calibration. |
LI-7500A 0042 | 2022-04-13 | Bouldin Corn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Corn for calibration and brought back to the lab. Replaced by sn 0065. |
LI-7500A 0073 | 2022-04-13 | Lab none | Installed | West Pond Gas Analyzer CO2/H2O | Installed at West Pond to replace sn 0035 for calibration. |
LI-7500A 0035 | 2022-04-13 | West Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from West Pond for calibration and brought back to the lab. Replaced by sn 0073. |
LI-7700 TG1-0421 | 2022-04-13 | West Pond Gas Analyzer CH4 | Removed | Lab none | Removed from West Pond and brought back to the lab. I couldn't connect to this sensor. The power LED light was blinking intermittently. Ethernet and Status LEDs were off. I removed the power cable as well since there was a crack in the plastic cable collar at the sensor. The wash power, wash tubing, and Ethernet cable are still there. |
SS-6L-12V 21501243 | 2022-04-13 | Lab none | Installed | Gilbert Tract Solar Controller ADCP | Installed for the Gilbert Tract ADCP near the breach. This 6A controller replaced sn 03220377 (15A), which was brought back to the lab so we can use it for the advection set up this summer. |
PS-15 03220377 | 2022-04-13 | Gilbert Tract Solar Controller ADCP | Removed | Lab none | Removed from the ADCP enclosure and brought back to the lab because we want this higher capacity 15A for the advection experiment. It was replaced with a 6A controller sn 21501243. |
LI-7700 TG1-0421 | 2022-04-15 | Lab none | Tested | Lab none | Plugged it in in the lab. The power LED was blinking intermittently and the ethernet and status LEDs were off. The fuse and fuse connector were totally rotten and the inside of the sensor body had condensation and a few tsp of water. We didn't see any obvious corrosion on the other circuit boards and no obvious leaks/cracks in the sensor. We wiped up the water and will let it air dry before sending it back to Licor. |
eosFD FD20180048 | 2022-04-12 | Tonzi-SoilCO2 Forced Diffusion Tree 1 | Removed | Lab none | Removed from Tonzi Tree 1 for testing. Identical CO2 readings (zero flux). Chamber was on its side which would cause this symptom - brought in for testing. |
HMP155 T2930114 | 2022-04-12 | Cal Kit RH & Temp | Removed | Lab none | Removed from the Tonzi Tower at the end of the inter-comparison |
S2-411 1050 | 2022-04-12 | Cal Kit Narrow Band NDVI Reflected | Removed | Lab none | Removed from the Tonzi Tower at the end of the inter-comparison |
S2-411 1049 | 2022-04-12 | Cal Kit Narrow Band NDVI Incoming | Removed | Lab none | Removed from the Tonzi Tower at the end of the inter-comparison |
PQS 1 212225 | 2022-04-12 | Cal Kit PAR Sensor Reflected | Removed | Lab none | Removed from the Tonzi Tower at the end of the inter-comparison |
CNR4 120788 | 2022-04-12 | Cal Kit Net Radiometer 4-way | Removed | Lab none | Removed from the Tonzi Tower at the end of the inter-comparison |
CR1000 23865 | 2022-04-12 | Cal Kit Data Logger | Removed | Lab none | Removed from the Tonzi Tower at the end of the inter-comparison |
PS-30 19391162 | 2022-04-13 | Loaned none | Returned | Returned none | Returned by Patty Oikawa |
TE525 48130-611 | 2021-10-26 | Hill Slough Precipitation | Tested | Hill Slough Precipitation | Got 53/50 tips. I meant to turn the screws 1 turn CCW but I think I may have turned them 2 turns CCW. Needs a retest next time anyways. |
eosFD FD20190028 | 2021-10-26 | Hill Slough Forced Diffusion A | Removed | Lab none | Removed after being underwater when tidal flooding began. The pump was still running. |
PS-30 19391162 | 2022-02-25 | Lab none | Loaned | Loaned none | Loaned to Patty Oikawa for Eden Landing |
LI-7500A 0073 | 2022-04-08 | Lab | Calibration | Lab | Last calibration: 2022-03-30 AGC=98 SDM= DAC1= DAC2= Ambient P=100.8kPa, Tair=28.9C, Tdew=11.3C Zero Air at 775psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9164 -> .9163 Offset: -.24ppm CO2 span 1.0039 -> .9983 Offset: 4.11ppm H2O zero 0.827 -> .8246 Offset: -13.5ppm H2O span 1.0223 -> 1.0147 Offset: .17ppm Notes: During the H2O span, H2O value was trending downward after a while. I added more DI water to the reservoir, and the H2O value stabilized. |
LI-7500A 1594 | 2022-04-08 | Lab | Calibration | Lab | Last calibration: 2021-11-30 AGC=96.7 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=30.8C, Tdew=11.2C Zero Air at 725psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.8645 -> .8643 Offset: -.71ppm CO2 span 1.0128 -> 1.0097 Offset: 2.12ppm H2O zero 0.8877 -> .8891 Offset: 7.7ppm H2O span 1.0045 -> 1.0106 Offset: -.13ppm Notes: |
LI-7500A 0065 | 2022-04-08 | Lab | Calibration | Lab | Last calibration: 2022-01-11 AGC=100.9 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=29.96C, Tdew=12.0C Zero Air at 750psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.8912 -> .8908 Offset: -1.86ppm CO2 span 1.0205 -> 1.0198 Offset: .57ppm H2O zero 0.7314 -> .7308 Offset: -4.2ppm H2O span 1.0259 -> 1.0160 Offset: .22ppm Notes: In the field, this sensor had a SS ~110. In the lab the SS was much more normal ~100. I'm not sure why there is a difference, as the config file should be the same. |
PAR-LITE 040411 | 2022-01-20 | Lab none | Damaged | Lab none | Brought to EE to install as part of the dfPAR set up, but killed the sensor when I accidentally wired its red/black wires into power |
WM 1590 153804 | 2022-04-06 | Lab none | Tested | Lab none | I tested this sensor briefly in the lab against sn 152601. U and V had a much smaller magnitude. I checked the analog settings and they had been set to 5m/s range. I changed U and V from 5m/s to 30m/s. I left W as 5m/s. U, V, W, SOS seem reasonable. I will be moving this intercomparison to my backyard tomorrow to run for a few days outdoors. |
LI-7500A 75H-2180 | 2022-03-31 | Lab none | Installed | Mayberry Gas Analyzer CO2/H2O | Installed at Mayberry to replace sn 1594 for calibration. |
LI-7500A 1594 | 2022-03-31 | Mayberry Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Mayberry for calibration. Replaced by sn 75H-2180. |
LI-7500A 0041 | 2022-03-31 | Lab none | Installed | East End Gas Analyzer CO2/H2O | Installed at East End to replace sn 0065 for calibration. |
LI-7500A 0065 | 2022-03-31 | East End Gas Analyzer CO2/H2O | Removed | Lab none | Removed from East End for calibration. Replaced by sn 0041. |
Mark II C011018 | 2022-03-31 | Lab none | Installed | Bouldin Alfalfa Weather Station PL_S-B | Installed at Bouldin Alfalfa to replace sn C003947. |
Mark II C003947 | 2022-03-31 | Bouldin Alfalfa Weather Station PL_S-B | Removed | Lab none | Removed from Bouldin Alfalfa and replaced by sn C011018. |
Mark II C011081 | 2022-03-31 | Lab none | Installed | Bouldin Corn Weather Station PL_B | Installed at Bouldin Corn to replace sn C004318. |
Mark II C004318 | 2022-03-31 | Bouldin Corn Weather Station PL_B | Removed | Lab none | Removed from Bouldin Corn and replaced by sn C011081. |
ML3 M003370 | 2022-03-31 | Bouldin Corn Soil Moisture Probe Theta | Moved | Bouldin Corn Soil Moisture Probe Theta | Moved from the field back under the tower in anticipation of tilling. This sensor is buried just under the surface. |
CS655 17128 | 2022-03-31 | Bouldin Corn Soil Moisture Probe B | Moved | Bouldin Corn Soil Moisture Probe B | Moved from the field back under the tower in anticipation of tilling. This sensor is to the south at 15cm depth. |
CS655 17133 | 2022-03-31 | Bouldin Corn Soil Moisture Probe C | Moved | Bouldin Corn Soil Moisture Probe C | Moved from the field back under the tower in anticipation of tilling. This sensor is horizontally between the other 2 Campbell sensors at 20cm depth. |
CS655 17129 | 2022-03-31 | Bouldin Corn Soil Moisture Probe A | Moved | Bouldin Corn Soil Moisture Probe A | Moved from the field back under the tower in anticipation of tilling. This sensor is to the north at 10cm depth. |
CS450 20010770 | 2022-03-31 | Hill Slough Water Level | Removed | Lab none | Removed from HS tower because it was replaced with the HYDRO21 CTD sensor. |
NR-LITE 990350 | 2022-03-31 | Hill Slough Net Radiometer | Removed | Lab none | Removed from HS because we are done comparing the NR01 data to it. Planning to use this sensor for the advection setup this summer. |
LI-7500A 75H-2180 | 2022-03-30 | Lab | Calibration | Lab | Last calibration: 2021-11-30 AGC=99.9 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=22.2C, Tdew=10.1C Zero Air at 800psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9136 -> .9141 Offset: 1.91ppm CO2 span 1.0295 -> 1.0373 Offset: 5.34ppm H2O zero 0.8924 -> .8911 Offset: -8.5ppm H2O span 1.0346 -> 1.0345 Offset: .01ppm Notes: |
LI-7500A 0073 | 2022-03-30 | Lab | Calibration | Lab | Last calibration: 2021-11-30 AGC=102 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=21.9C, Tdew=11.1C Zero Air at 800psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9175 -> .9164 Offset: -4.45ppm CO2 span 1 -> 1.0039 Offset: -2.8ppm H2O zero 0.8174 -> .827 Offset: 58.5ppm H2O span 1.0192 -> 1.0223 Offset: -.06ppm Notes: Reset SS from 102 to 100 before the calibration. Huge H2O zero offset even after waiting 45+ min. Need to check this calibration again before deploying in field. |
LI-7500A 0041 | 2022-03-30 | Lab | Calibration | Lab | Last calibration: 2021-12-16 AGC=100.2 SDM= DAC1= DAC2= Ambient P=100.8kPa, Tair=20.9C, Tdew=8.9C Zero Air at 850psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.996 -> .9969 Offset: 2.92ppm CO2 span 1.0172 -> 1.0189 Offset: 1.16ppm H2O zero 0.8227 -> .8214 Offset: -7.4ppm H2O span 1.0335 -> 1.0337 Offset: -.02ppm Notes: H2O span took about 40-50 minutes to stabilize. Dewpoint temp had a range of ~0.2C when setting the H2O span point. |
LI-7500A 75H-2182 | 2022-03-23 | Lab none | Installed | Gilbert Tract Gas Analyzer CO2/H2O | Installed at Gilbert Tract to replace sn 0041 for calibration. |
LI-7500A 0041 | 2022-03-23 | Gilbert Tract Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Gilbert Tract for calibration. Replaced by sn 75H-2182. |
PRB-K Type #8 00000128BDDE8 | 2022-03-24 | Vaira Moisture Point Biomass | Removed | Lab none | Removed from the At biomass fence, one pin broken. Not working for many months |
PRB-K Type #8 00000128B97F3 | 2022-03-24 | Vaira Moisture Point Solar | Removed | Lab none | Removed from near the solar panels, top and bottom pin broken. Not working for many months |
PRB-K Type #8 000007933FDE | 2022-03-24 | Vaira Moisture Point Tower | Removed | Lab none | Removed from near the eddy tower, top and bottom pin broken. Not working for many months |
LI-7500A 75H-2514 | 2022-03-24 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2022-03-24 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 75H-2514. |
LI-7500A 0418 | 2022-03-24 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 75H-2180 for calibration. |
LI-7500A 75H-2180 | 2022-03-24 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0418. |
PQS 1 212225 | 2022-03-24 | Cal Kit PAR Sensor Incoming | Moved | Cal Kit PAR Sensor Reflected | Moved to the reflected position to continue inter-comparison |
LI-7500A 75H-2182 | 2022-03-22 | Lab | Calibration | Lab | Last calibration: 2021-11-30 AGC=99.4 SDM=0 DAC1=- DAC2=- Ambient P=100.8kPa, Tair=29.6C, Tdew=10.7C Zero Air at 900psi, AMP CA04119 524.77 at 1650 Tdewspan=20.0 CO2 zero 0.885 -> 0.8849 Offset: -0.6ppm CO2 span 1.0219 -> 1.0222 Offset: -0.07ppm H2O zero 0.847 -> 0.8483 Offset: 7.9ppm H2O span 1.0264 -> 1.0312 Offset: -0.1ppm Notes: |
LI-7500A 0418 | 2022-03-22 | Lab none | Calibration | Lab none | Last calibration: 2022-03-08
AGC=100.0 SDM=0 DAC1=- DAC2=- Ambient P=100.9kPa, Tair=28.8C, Tdew=10.43C Zero Air at 900psi, AMP CA04119 524.77 at 1650 Tdewspan=20.0 CO2 zero 0.9108 -> 0.9104 Offset: -31.5ppm CO2 span 1.0024 -> 1.0063 Offset: -2.77ppm H2O zero 0.9116 -> 0.9144 Offset: 14.9ppm H2O span 1.0069 -> 1.0127 Offset: -0.13ppm Notes: First lab calibration after Licor replaced chopper motor, light source and detector and factory calibration. All the new configuration coefficients were entered by hand. We always see large offset on first cal - not sure why. |
PS-30M 186580399 | 2022-03-21 | Loaned none | Returned | Lab none | Returned from temporary setup on the Yampah Marsh at Elkhorn Slough. |
SP75 020854E-6-03-02-06905 | 2022-03-21 | Loaned none | Returned | Lab none | Returned from temporary setup on the Yampah Marsh at Elkhorn Slough. |
SP75 020854E-6-03-02-06862 | 2022-03-21 | Loaned none | Returned | Lab none | Returned from temporary setup on the Yampah Marsh at Elkhorn Slough. |
LI-7500A 75H-2514 | 2022-03-14 | Lab | Calibration | Lab | Last calibration: 2022-03-11 AGC=102 SDM= DAC1= DAC2= Ambient P=101.3kPa, Tair=24.9C, Tdew=2.3C Zero Air at 875psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.905 -> .9049 Offset: -.27ppm CO2 span 1.0119 -> 1.0130 Offset: -.72ppm H2O zero 0.8398 -> .84 Offset: 1.8ppm H2O span 1.0137 -> 1.0211 Offset: -.16ppm Notes: Second calibration since chem swap on 3/9. Seems stable now but I'll test it again in a few days to confirm. |
LI-7500A 75H-2514 | 2022-03-17 | Lab | Calibration | Lab | Last calibration: 2022-03-11 AGC=101 SDM= DAC1= DAC2= Ambient P=101.0kPa, Tair=23.6C, Tdew=8.5C Zero Air at 875psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.905 -> .905 Offset: .15ppm CO2 span 1.0119 -> 1.0125 Offset: .25ppm H2O zero 0.8398 -> .8403 Offset: 1.9ppm H2O span 1.0137 -> 1.0196 Offset: .03ppm Notes: Third calibration after chemical swap. The sensor seems stable now. Swapping the chemicals seemed to fix the issue with the large H2O zero offsets. |
PQS 1 212225 | 2022-02-23 | Cal Kit PAR Sensor Incoming | Moved | Cal Kit PAR Sensor Reflected | Moved from the incoming position to the reflected position at VR |
LED-NDVI LED-01 | 2004-07-24 | Lab none | Installed | Vaira LED sensor | Approximate date of installation based on the first data in the database. |
LI-7500A 75H-2514 | 2022-03-11 | Lab | Calibration | Lab | Last calibration: 2022-03-03 AGC=101.8 SDM= DAC1= DAC2= Ambient P=101.6kPa, Tair=25.6C, Tdew=-3.8C Zero Air at 900psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9044 -> .9050 Offset: 2.46ppm CO2 span 1.0101 -> 1.0119 Offset: -1.36ppm H2O zero 0.8433 -> .8398 Offset: -24.2ppm H2O span 1.0199 -> 1.0137 Offset: .14ppm Notes: First calibration after chemical swap. Large negative offset in H2O zero. I'll keep the sensor running over the weekend and will calibrate again on Monday. |
HMP155 T2820183 | 2022-03-10 | Lab none | Installed | West Pond RH & Temp | Installed at West Pond to replace the malfunctioning one removed last week. Broken mounting screw and typo in the data logger program - messy troubleshooting ensued. |
LI-7500A 0418 | 2022-03-08 | Lab | Calibration | Lab | Last calibration: 2021-09-29 AGC= SDM= DAC1= DAC2= Ambient P=kPa, Tair=C, Tdew=C at psi, 750 at Tdewspan=12.00 CO2 zero 0.8995 -> .9108 Offset: ppm CO2 span 1 -> 1.0024 Offset: ppm H2O zero 0.849 -> .9116 Offset: ppm H2O span 0.9783 -> 1.0069 Offset: ppm Notes: Licor factory calibration |
LI-7500A 0418 | 2022-03-11 | Factory none | Received | Lab none | Received back from Licor. The source lamp was defective and we requested that the lamp, chopper motor, and detector all to be replaced. The sensor was recalibrated at the factory. |
HMP155 R0360206 | 2022-03-11 | Factory none | Disposed | Disposed none | Campbell Sci inspected the sensor and found that it was shorted and can't be repaired. Discarded at the factory. |
HMP155 T2930114 | 2022-03-10 | Cal Kit RH & Temp | Moved | Cal Kit RH & Temp | Moved from VR radiometer boom to TZ Tower top |
S2-411 1050 | 2022-03-10 | Cal Kit Narrow Band NDVI Reflected | Moved | Cal Kit Narrow Band NDVI Reflected | Moved from VR radiometer boom to TZ Tower top |
S2-411 1049 | 2022-03-10 | Cal Kit Narrow Band NDVI Incoming | Moved | Cal Kit Narrow Band NDVI Incoming | Moved from VR radiometer boom to TZ Tower top |
PQS 1 212225 | 2022-03-10 | Cal Kit PAR Sensor Reflected | Moved | Cal Kit PAR Sensor Incoming | Moved from VR radiometer boom to TZ Tower top |
CNR4 120788 | 2022-03-10 | Cal Kit Net Radiometer 4-way | Moved | Cal Kit Net Radiometer 4-way | Moved from VR radiometer boom to TZ Tower top |
CR1000 23865 | 2022-03-10 | Cal Kit Data Logger | Moved | Cal Kit Data Logger | Moved from VR radiometer boom to TZ Tower top |
LI-7500A 75H-2667 | 2022-03-10 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 75H-2182 for calibration. |
LI-7500A 75H-2182 | 2022-03-10 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 75H-2667. |
LI-7700 TG1-0643 | 2022-03-07 | Lab none | Loaned | Borrowed none | Installed on the urban flux tower on top of Berkeley Way West building |