Equipment and Calibration History
Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 [41] 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65Ser No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
GP-001 GP20150109 | 2017-02-23 | Lab none | Installed | West Pond CO2 Probe B | Installed at West Pond in the same location as the previous one. |
GP-001 GP20150109 | 2016-11-21 | East End CO2 probe | Removed | Lab none | Removed from East End where it was pulling about 6A of power and browning out the entire site. |
GP-001 GP20150100 | 2017-02-13 | West Pond CO2 Probe B | Removed | Lab none | Removed form West Pond due to bad readings. |
HMP60 L1740420 | 2017-02-14 | Lab none | Installed | Twitchell Alfalfa RH & Temp | Installed at Twitchell Alfalfa |
HMP60 G4540021 | 2017-02-14 | Twitchell Alfalfa RH & Temp | Removed | Lab none | Removed from Twitchell Alfalfa b/c RH values were slowly drifting to 0 |
HMP60 L2310711 | 2016-11-02 | Lab none | Installed | Tonzi | Installed on the top of the Tonzi Tower to compare with the existing HMP45 there. |
SD130BN 0030F4D21491 | 2017-01-31 | Sherman Wetland Camera | Removed | Lab none | Removed from Sherman Wetland after being found with the enclosure open and non-operational. |
AM16/32B 3052 | 2017-02-21 | Factory | Disposed | Disposed | Campbell called to say that this AM16/32 was not repairable. I told them to dispose of it. |
AM16/32B 3052 | 2017-02-15 | Lab none | Shipped | Factory | Sent to Campbell on RMA25817 for evaluation and repair. |
WM 1590 074417 | 2017-02-17 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug. The upgrade failed the first time because the Power-on message had been turned off. Without the Power-on message, the upgrade wizard cannot check if the checksums are correct. After I turned on the Power-on message, I used the second upgrade (X074417-s.gcu) file to complete the upgrade. Old SN: Y074417 New SN: X074417 Upgrade file: Y074417-Y124215-m.gcu, X074417-s.gcu Old firmware: 2329-114-01 New firmware: 2329-116-01 All Checksum fields report PASS. Log file: WMUpgradeLog_X074417_170217_121544.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Output rate : P2 2 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ====DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X074417 ----SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-116 ----HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 ADC resolution of attached unit : 14 DAC resolution of attached unit : 14 Number of ADC channels : 4 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ----USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P2 2 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Fixed Field SOS and Sonic Temperature display : SOS Analogue input display : OFF Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound |
SRS-Ni NDVI 882103061 | 2017-02-01 | Lab none | Installed | Twitchell Alfalfa Narrow Band NDVI Reflected | Installed on the Twitchell Alfalfa tower radiation boom to measure outgoing NDVI bands. |
SRS-Ni NDVI 882103036 | 2017-02-01 | Lab none | Installed | Twitchell Alfalfa Narrow Band NDVI Incoming | Installed on the Twitchell Alfalfa tower radiation boom to measure incoming NDVI bands. |
LI-7550 AIU-0674 | 2017-02-13 | Factory | Returned | Lab none | Returned from repair at Licor. |
LI-7500A 75H-2668 | 2017-02-13 | Factory | Returned | Lab none | Returned from Licor after repair on 7550 box and calibration. |
LI-7500A 75H-2668 | 2017-01-30 | Lab none | Calibration | Lab none | Last calibration: 2016-12-23 AGC= SDM= DAC1= DAC2= Ambient P=kPa, Tair=C, Tdew=C at psi, Licor CO2 at 749 Tdewspan=12 CO2 zero 0.9378 -> 0.9392 Offset: ppm CO2 span 1 -> 1.0120 Offset: ppm H2O zero 0.9006 -> 0.9039 Offset: ppm H2O span 0.9663 -> 1.0054 Offset: ppm Notes: Factory calibration after repair of 7550 box and upgrade to embedded software v8.5.0 |
LI-7550 AIU-0674 | 2017-01-10 | Factory | Repair | Factory | Licor says "Inspected the units and found the front panel corroded in the LI-7550 and the LI-7540 in operational condition. Replaced defective parts, updated the embedded software in the LI-7550 to 8.5.0, subjected units to environmental tests and calibrated." They replaced the connector panel assembly, ribbon cables and vent membrane. |
WM 1590 153803 | 2017-02-10 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug. Old SN: Y153803 New SN: X153803 Upgrade file: Y153802-Y153806-m5.gcu Old firmware: 2329-601-03 New firmware: 2329-701-01 All Checksum fields report PASS. Log file: WMUpgradeLog_Y153803_170310_132625.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Output rate : P1 1 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X153803 ---- SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-701 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound Synchronised Polling : Off |
GP-001 GP20150084 | 2016-10-21 | Lab none | Returned | Returned | Returned to Eosense by way of Gregory Maurer at the University of New Mexico. Not precise enough for atmospheric gradient. |
GP-001 GP20160003 | 2016-10-21 | Lab none | Returned | Returned | Returned to Eosense by way of Gregory Maurer at the University of New Mexico. Not precise enough for atmospheric gradient. |
CR10X 35474 | 2016-12-21 | Twitchell Alfalfa Data Logger Radiation | Removed | Lab none | Removed from Twitchell Alfalfa where is was recording data from the radiometers. The radiometers have been moved to the met CR1000. |
SRS-Ni NDVI 882103021 | 2017-02-01 | Lab none | Installed | Bouldin Alfalfa Narrow Band NDVI Reflected | Installed on the Bouldin Alfalfa tower in the reflected/outgoing position to compare against the existing field stop sensor. |
SRS-Ni NDVI 882103061 | 2017-01-30 | Lab none | Maintenance | Lab none | The SDI-12 address was changed from 0 to E. E means reflected hemispherical NDVI. |
SRS-Ni NDVI 882103021 | 2017-01-30 | Lab none | Maintenance | Lab none | The SDI-12 address was changed from 0 to E. E means reflected hemispherical NDVI. |
LI-7500A 1594 | 2017-01-31 | Lab none | Installed | Mayberry Gas Analyzer CO2/H2O | Installed at Mayberry after calibration in the lab, replacing sn 75H-2176. |
LI-7500A 75H-2176 | 2017-01-31 | Mayberry Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Mayberry for calibration in the lab. Replaced by sn 1594 |
WM 1590 120906 | 2017-01-31 | Lab none | Installed | Sherman Wetland Anemometer | Installed at Sherman Wetland tower to replace sonic sn 153803. The old sonic does not have the new firmware fixing the w-bug, this new sonic has the new firmware. |
WM 1590 153803 | 2017-01-31 | Sherman Wetland Anemometer | Removed | Lab none | Removed from Sherman Wetland tower to be replaced with sonic sn 120906 which had a firmware update fixing the w-bug |
LI-7500A 1594 | 2017-01-27 | Lab none | Calibration | Lab none | Last calibration: 2016-08-30 AGC=101.86 SDM= DAC1= DAC2= Ambient P=101.78kPa, Tair=24.47C, Tdew=2.69C Zero Air at 1800psi, God CO2 421.22 at 400 Tdewspan=16.00 CO2 zero 0.8648 -> 0.8654 Offset: 1.16ppm CO2 span 1.0164 -> 1.0194 Offset: -1.63ppm H2O zero 0.8859 -> 0.8868 Offset: 5.5ppm H2O span 1.0113 -> 1.0067 Offset: 0.10ppm Notes: Saved calibration file: sn1594_0614_20170127Lab_cal.L7X |
SRS-Ni NDVI 882103036 | 2017-01-27 | Lab none | Maintenance | Lab none | The SDI-12 address was changed from 0 to C. C means incoming hemispherical NDVI. |
WM 1590 120907 | 2017-01-27 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y120907 New SN: X120907 Upgrade file: Y074417-Y124215-m.gcu Old firmware: 2329-114 New firmware: 2329-116-01 All Checksum fields report PASS. Log file: WMUpgradeLog_Y120907_170127_135608.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Output rate : P1 1 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER ======================= Serial Number : X120907 ---- SW VERSION ------------------------------------------- Anemometer Firmware Version: 2329-116 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound |
WM 1590 153804 | 2017-01-25 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y153804 New SN: X153804 Upgrade file: Y074417-Y124215-m.gcu Old firmware: 2329-601-03 New firmware: 2329-701-01 All Checksum fields report PASS. Log file: .txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Output rate : P1 1 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER ======================== Serial Number : X153804 ---- SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-701 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound Synchronised Polling : Off |
WM 1590 120906 | 2017-01-25 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y120906 New SN: X120906 Upgrade file: Y074417-Y124215-m.gcu Old firmware: 2329-114-01 New firmware: 2329-116-01 All Checksum fields report PASS. Log file: WMUpgradeLog_Y120906_170125_153219.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X120906 ---- SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-116 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound |
NR01 1720 | 2017-01-26 | Lab none | Installed | Mayberry Net Radiometer 4-way | Installed on the Mayberry tower on the boom over the tules. This radiometer still has some marks on the SW domes, which may be cleanable. |
WM 1590 153804 | 2017-01-26 | Lab none | Installed | Mayberry Anemometer | Installed at Mayberry tower. It replaced sn 074417. |
WM 1590 074417 | 2017-01-26 | Mayberry Anemometer | Removed | Lab none | Removed from Mayberry tower so we could take it back to the lab for firmware upgrade. It was replaced by sn 153804. |
AM25T 2424 | 2017-01-26 | Mayberry Multiplexer Thermocouple | Rewired | Mayberry Multiplexer Thermocouple | Moved and re-wired AM25T from an old enclosure to a new enclosure when switching from CR23X and CR1000. |
AM16/32B 3052 | 2017-01-26 | Mayberry Multiplexer Met | Removed | Lab none | Uninstalled when switching from CR23X and CR1000. |
AM16/32A 1430 | 2017-01-26 | Lab none | Installed | Mayberry Multiplexer Met | Installed in new enclosure when switching from CR23X and CR1000. |
CR23X 2143 | 2017-01-26 | Mayberry Data Logger Tule | Removed | Lab none | Uninstalled when switching from CR23X and CR1000. This CR23X was getting unreliable. |
CR1000 38212 | 2017-01-26 | Lab none | Installed | Mayberry Data Logger Tule | Installed in new enclosure and switched all sensors from CR23X system to CR1000 system. |
LI-7700 TG1-0224 | 2016-12-02 | Loaned | Returned | Lab none | Returned from Frank Anderson via FedEx. He ran this instrument for 2-3 weeks at his salt marsh site near Seattle Washington. |
LI-7500A 0073 | 2017-01-17 | Tonzi-Floor | Removed | Lab none | Removed from the Floor tower because of bad CO2 calibration coefficients. Replaced by sn 0418 |
LI-7500A 0418 | 2017-01-17 | Lab none | Installed | Tonzi-Floor | Installed on the Floor tower to replace sn 0073 that had bad CO2 calibration coefficients. |
LI-7500A 0073 | 2017-01-18 | Lab none | Calibration | Lab none | Last calibration: 2016-12-05 AGC=58 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.46kPa, Tair=19.9C, Tdew=5.83C Zero Air at 1750psi, God CO2 421.22 at 400 Tdewspan=16.00 CO2 zero 1.0707 -> 0.9573 Offset: -519ppm CO2 span 1.0029 -> 1.0037 Offset: -0.22ppm H2O zero 0.7391 -> 0.7394 Offset: -1.3ppm H2O span 1.0255 -> 1.0063 Offset: 0.41ppm Notes: Something happened to the CO2 coefficients after the last calibration causing the CO2 channel to be way off. New calibration seems good and matches historic coefficients. |
WM 1590 153805 | 2017-01-17 | Lab none | Installed | Twitchell Rice Anemometer | Installed on Twitchell Rice tower to replace sn 120906. We also replaced the sonic cable. |
WM 1590 120906 | 2017-01-19 | Twitchell Rice Anemometer | Removed | Lab none | Removed from Twitchell Rice tower and returned to lab for firmware upgrade. It was replaced by sn 153805. |
WM 1590 124215 | 2017-01-17 | Lab none | Installed | East End Anemometer | Installed on East End tower to replace sn 153804. |
WM 1590 153804 | 2017-01-17 | East End Anemometer | Removed | Lab none | Returned to lab for firmware upgrade. It was replaced by sn 124215. |
WM 1590 120908 | 2017-01-17 | Lab none | Installed | West Pond Anemometer | Installed on West Pond tower to replace sn 120907. |
WM 1590 120907 | 2017-01-17 | West Pond Anemometer | Removed | Lab none | Removed from West Pond and returned to lab for firmware upgrade. It was replaced by sn 120908. |
LI-7550 AIU-0674 | 2017-01-05 | Lab none | Shipped | Factory | Sent to Licor on RMA 00199505 with 75H-2668. This box had its door left open in the field and got rain inside. It was functional but throwing errors - probably a bad temperature controller. It was sent with the head to Licor for repair. |
LI-7500A 75H-2668 | 2017-01-05 | Lab none | Shipped | Factory | Sent to Licor on RMA 00199505 with AIU-0674. This head was on the box when its door was left open in the field and got rain inside. Subsequent lab calibration looked a little funny so it was sent with the box to Licor for a check. |
LI-7500A 0418 | 2017-01-12 | Lab none | Calibration | Lab none | Last calibration: 2016-08-15 AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.44kPa, Tair=22.75C, Tdew=6.33C Zero Air at 1800psi, God CO2 421.22 at 400 Tdewspan=16.00 CO2 zero 0.9012 -> 0.9009 Offset: ppm CO2 span 1.007 -> 1.0062 Offset: ppm H2O zero 0.8558 -> 0.8570 Offset: 0.17ppm H2O span 0.9911 -> 0.9788 Offset: 0.02ppm Notes: Using a new tank of Zero Air. Forgot to write offsets for CO2. The serial port on bottom of Licor box has missing pin. Connect to laptop using a different cable that has one serial end and one green-guillotine-port end. Connect the green-guillotine-port end inside the box for the "Serial" connection. |
WM 1590 153805 | 2017-01-10 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y153805 New SN: X153805 Upgrade file: Y153802-Y153806-m5.gcu Old firmware: 2329-601-03 New firmware: 2329-701-01 All Checksum fields report PASS. Log file: WMUpgradeLog_Y153805_170110_093737.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. --------------------------------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 Output rate : P1 1 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X153805 ---- SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-701 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound Synchronised Polling : Off |
NULL 0030F4D0A167 | 2011-10-18 | Lab none | Installed | Tonzi | Installed on the Tonzi tower to cover the same area as the existing webcam. The camera is mounted about 20m agl and is pointing about 310deg true and 20deg down from horizontal. Horizontal FOV is about 70deg. |
NULL 0030F4D0A168 | 2011-10-13 | Lab none | Installed | Vaira | Installed on the Vaira radiation tower to cover the same area as the existing webcam. The camera is about 3m agl, pointed at about 290deg true and pointed down 10deg from horizontal. FOV is about 70deg. |
WM 1590 124215 | 2017-01-06 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y124215 New SN: X124215 Upgrade file: Y074417-Y124215-m.gcu, Y012345-s.gcu, X120908 X124215-m.gcu Old firmware: 2329-114 New firmware: 2329-116-01 All Checksum fields report PASS. There are multiple upgrade .gcu files. During the initial upgrade using file "Y074417-Y124215-m.gcu," the sensor was set at 57,600 baud rate. This baud rate caused an error in the upgrade, so the SN reset to Y012345. Gill sent me the second upgrade file, "Y-12345-s.gcu," which reset the sensor to SN X124215. Finally, I used the third upgrade file, "X120908 X124215-m.gcu," to upgrade the firmware. Log file: WMUpgradeLog_X124215_170106_123318.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 Output rate : P1 1 Hz Output rate : P8 20 Hz Output Channel 3 : W output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X124215 ---- SW VERSION -------------------------------------------- Anemometer Firmware Version: 2329-116 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound |
WM 1590 120908 | 2017-01-06 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y120908 New SN: X120908 Upgrade file: Y074417-Y124215-m.gcu, Y012345-s.gcu, X120908 X124215-m.gcu Old firmware: 2329-114 New firmware: 2329-116-01 All Checksum fields report PASS. There are multiple upgrade .gcu files. During the initial upgrade using file "Y074417-Y124215-m.gcu," the sensor was set at 57,600 baud rate. This baud rate caused an error in the upgrade, so the SN reset to Y012345. Gill sent me the second upgrade file, "Y-12345-s.gcu," which reset the sensor to SN X120908. Finally, I used the third upgrade file, "X120908 X124215-m.gcu," to upgrade the firmware. Log file: WMUpgradeLog_X120908_170106_122011.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 Output rate : P1 1 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X120908 ---- SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-116 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound |
WM 1590 153802 | 2017-01-05 | Lab none | Installed | Twitchell Alfalfa Anemometer | Installed at Twitchell Alfalfa to replace sn 153805. Brand new. |
WM 1590 153805 | 2017-01-05 | Twitchell Alfalfa Anemometer | Removed | Lab none | Removed from Twitchell Alfalfa tower for firmware upgrade. Replaced by sn 153802. |
LI-7500A 75H-2514 | 2016-12-29 | Lab none | Installed | Twitchell Alfalfa Gas Analyzer CO2/H2O | Installed at Twitchell Alfalfa with new 7550 box sn AIU-1557 to replace damaged system |
LI-7550 AIU-1557 | 2016-12-29 | Lab none | Installed | Twitchell Alfalfa Interface LI-7550 | Installed at Twitchell Alfalfa after old AIU was damaged in rain. Installed together with 7500 head sn 75H-2514 |
LI-7500A 75H-2180 | 2016-12-29 | Lab none | Installed | Sherman Wetland Gas Analyzer CO2/H2O | Installed at Sherman Wetland after calibration to replace sn 1594 |
LI-7500A 1594 | 2016-12-29 | Sherman Wetland Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Sherman Wetland for calibration. Replaced by sn 75H-2180 |
WM 1590 153802 | 2017-01-03 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y153802 New SN: X153802 Upgrade file: Y153802-Y153806-m5.gcu Old firmware: 2329-601 New firmware: 2329-701-01 All Checksum fields report PASS. Log file: WMUpgradeLog_Y153802_170103_110126.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M2 Polar Output Message Format : M1 UV(W) Resolution settings : Normal resolution Resolution settings : High resolution Output rate : P1 1 Hz Output rate : P8 20 Hz SOS and Sonic Temperature display : Neither SOS and Sonic Temperature display : SOS Analogue output mode : 0v to 5V Analogue output mode : -5V to 5V Output Channel 1 : Polar angle (wrap mode = 360 degree wrap) Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : Polar Magnitude (full range = 5m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 4 : Status (status mode = Good signal high, bad signal low) Output Channel 4 : Speed of Sound THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER ======================== Serial Number : X153802 ---- SW VERSION -------------------------------------------- Anemometer Firmware Version: 2329-701 ---- HW DETAILS OF ANEMOMETER ------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER ---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M2 Polar Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : Normal resolution Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : Neither SOS and Sonic Temperature display : SOS Analogue output mode : 0v to 5V Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : Polar angle (wrap mode = 360 degree wrap) Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : Polar Magnitude (full range = 5m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Status (status mode = Good signal high, bad signal low) Output Channel 4 : Speed of Sound Synchronised Polling : Off |
LI-7500A 75H-2180 | 2016-12-23 | Lab none | Calibration | Lab none | Last calibration: 2016-08-30 AGC=99.5 SDM=6 DAC1=- DAC2=- Ambient P=99.4kPa, Tair=16.12C, Tdew=9.33C Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=14.00 CO2 zero 0.9142 -> 0.9145 Offset: 1.55ppm CO2 span 1.0227 -> 1.0287 Offset: -3.1ppm H2O zero 0.8858 -> 0.8836 Offset: -13.6ppm H2O span 1.0211 -> 1.0124 Offset: 0.17ppm Notes: |
LI-7500A 75H-2668 | 2016-12-23 | Lab none | Calibration | Lab none | Last calibration: 2016-12-06 AGC=102.2 SDM=6 DAC1=- DAC2=- Ambient P=99.57kPa, Tair=14.95C, Tdew=C Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=14.00 CO2 zero 0.9385 -> 0.9378 Offset: -2.65ppm CO2 span 1.0171 -> 1.0000 Offset: 9.43ppm H2O zero 0.9017 -> 0.9006 Offset: -7.5ppm H2O span 1.0841 -> 0.9663 Offset: 2.45ppm Notes: Recal after box failed in the field after getting rain inside because the door was left open. Pressure about 0.5kPa low. Large offsets. Waves with 1 to 2 min period in data at steady state. Sending to licor for check. |
LI-7500A 75H-2514 | 2016-12-23 | Lab none | Calibration | Lab none | Last calibration: 2016-09-12
AGC=103.3 SDM=1 DAC1=- DAC2=- Ambient P=99.98kPa, Tair=15.95C, Tdew=7.22C Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=14.00 CO2 zero 0.9054 -> 0.9054 Offset: -0.1ppm CO2 span 1.0092 -> 1.0135 Offset: -2.47ppm H2O zero 0.8371 -> 0.8356 Offset: -10.3ppm H2O span 1.0239 -> 1.0225 Offset: 0.04ppm Notes: First calibration after upgrading embedded software to v8.0.0. Changed bandwidth to 10Hz for 20Hz sampling. |
LI-7550 AIU-1557 | 2016-12-23 | Lab none | Upgrade | Lab none | I updated the embedded software from v7.5.0 to v8.0.0 and added settings for use at non-CH4 site (TA). |
LI-7500A 75H-2668 | 2016-12-21 | Twitchell Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | This head was deployed with LI7550 box sn AIU-0674 when the box was damaged in the field. Could not determine in the field if the head was working so brought it back to the lab for testing. |
LI-7550 AIU-0674 | 2016-12-21 | Twitchell Alfalfa Interface LI-7550 | Removed | Lab none | This 7550 box was damaged in the field. The door was left open and the circuit board got wet. Brought back to the lab for testing and repairs. LI7500 head sn 75H-2668 was deployed with the unit when it was damaged. |
LI-7500A 75H-2182 | 2016-12-21 | Lab none | Installed | Twitchell Rice Gas Analyzer CO2/H2O | Installed at Twitchell Rice after calibration in the lab to replace 75H-2180. |
LI-7500A 75H-2180 | 2016-12-21 | Twitchell Rice Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Twitchell Rice for calibration in the lab. Replaced by 75H-2182. |
LI-7500A 75H-2669 | 2016-12-21 | Lab none | Installed | East End Gas Analyzer CO2/H2O | Installed at East End after calibration in the lab to replace 75H-2514 |
LI-7500A 75H-2514 | 2016-12-21 | East End Gas Analyzer CO2/H2O | Removed | Lab none | Removed from East End for calibration in the lab. Replaced by 75H-2669. |
LI-7500A 75H-2667 | 2016-12-21 | Lab none | Installed | West Pond Gas Analyzer CO2/H2O | Installed at West Pond after calibration in the lab to replace 75H-2150 |
LI-7500A 75H-2150 | 2016-12-21 | West Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from West Pond for calibration in the lab. Replaced by 75H-2267 |
LI-7500A 75H-2669 | 2016-12-19 | Lab none | Calibration | Lab none | Last calibration: 2016-07-19 AGC=104.5 SDM=6 DAC1=- DAC2=- Ambient P=101.24kPa, Tair=21.21C, Tdew=0.0C Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=16.00 CO2 zero 0.919 -> 0.9184 Offset: -2.14ppm CO2 span 1.0141 -> 1.0119 Offset: 1.13ppm H2O zero 0.9987 -> 0.8989 Offset: 25.1ppm H2O span 1.0143 -> 1.0109 Offset: 0.07ppm Notes: |
LI-7500A 75H-2667 | 2016-12-19 | Lab none | Calibration | Lab none | Last calibration: 2016-08-03 AGC=107.0 SDM=6 DAC1=- DAC2=- Ambient P=101.23kPa, Tair=21.56C, Tdew=-1.32C Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=16.00 CO2 zero 0.9427 -> 0.9421 Offset: -2.08ppm CO2 span 1.0156 -> 1.0156 Offset: -0.02ppm H2O zero 0.8333 -> 0.8378 Offset: 28.1ppm H2O span 1.015 -> 1.0214 Offset: -0.13ppm Notes: |
LI-7500A 75H-2182 | 2016-12-19 | Lab none | Calibration | Lab none | Last calibration: 2016-08-30 AGC=87.2 SDM=6 DAC1=- DAC2=- Ambient P=101.31kPa, Tair=20.29C, Tdew=-0.86C Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=16.00 CO2 zero 0.8831 -> 0.8837 Offset: 2.75ppm CO2 span 1.0142 -> 1.0207 Offset: -3.48ppm H2O zero 0.8464 -> 0.8457 Offset: -4.2ppm H2O span 1.023 -> 1.0186 Offset: 0.09ppm Notes: |
CS655 17135 | 2016-11-22 | Lab none | Installed | Bouldin Alfalfa Soil Moisture Probe B | Installed at the Bouldin Alfalaf Tower to replace the on destroyed by the mower. It was buried horizontally at about 2.5cm depth. The data logger program was not updated for the new SDM address until 2016-12-21 |
HFP01 11334 | 2016-11-22 | Lab none | Installed | Bouldin Alfalfa Heat Flux Plate B | Reinstalled at the Bouldin Alfalfa tower. This SHF plate was installed about 1m west of the tower at about 1cm deep. |
WM 1590 124215 | 2016-12-19 | Lab none | Maintenance | Lab none | This was outputting sonic T instead of SOS. I changed its settings to output SOS. Everything else looked okay. Its firmware is 2329-114 - so far not update tool to fix w bug for this version. |
WMP 1352 000021 | 2016-10-25 | Lab none | Loaned | Loaned | Loaned to Sam in Whendee Silver's lab. |
LI-7500A 0073 | 2016-12-07 | Lab none | Installed | Tonzi-Floor | Installed on the Tonzi Tower to replaced sn 0976 for calibration. |
LI-7500 0976 | 2016-12-07 | Tonzi-Floor | Removed | Lab none | Removed from the Tonzi Floor for calibration. Replaced by sn 0073. |
LI-7500A 0035 | 2016-12-07 | Lab none | Installed | Tonzi-Tower | Installed on the Tonzi Tower to replaced sn 0418 for calibration. |
LI-7500A 0418 | 2016-12-07 | Tonzi-Tower | Removed | Lab none | Removed from the Tonzi Tower for calibration. Replaced by sn 0035. |
LI-7500A 0041 | 2016-12-07 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed on the Vaira Tower to replace sn 0042 for calibration. |
LI-7500A 0042 | 2016-12-07 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from the Vaira Tower for calibration. Replaced by sn 0041. |
LI-7500A 0065 | 2016-12-07 | Lab none | Installed | Bouldin Alfalfa Gas Analyzer CO2/H2O | Installed on the Bouldin Alfalfa tower to replace sn 75H-2667 for calibration. |
LI-7500A 75H-2667 | 2016-12-07 | Bouldin Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Alfalfa for calibration. Replaced by sn 0065 |
LI-7500A 75H-2668 | 2016-12-06 | Lab none | Installed | Twitchell Alfalfa Gas Analyzer CO2/H2O | Installed on the Twitchell Alfalfa tower to replace sn 75H-2182 for calibration |
CS451 14011129 | 2016-11-17 | Lab none | Installed | Vaira Pond | Installed at the main pond at Varia. It was installed in the pond on a wooden stake at 11.75ft of water. |
CR10X 31307 | 2016-11-17 | Lab none | Installed | Vaira Pond | Installed at the main pond at Varia. This will collect water level data. |
LI-7500A 75H-2182 | 2016-12-06 | Twitchell Alfalfa Gas Analyzer CO2/H2O | Calibration | Lab none | Removed from the Twitchell Alfalfa tower for calibration. Replaced by sn 75H-2668. |
LI-7500A 0073 | 2016-12-05 | Lab none | Calibration | Lab none | Last calibration: 2016-05-05 AGC=59 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.65kPa, Tair=20.82C, Tdew=8.53C Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.00 CO2 zero 0.9579 -> 0.9576 Offset: ppm CO2 span 1.0029 -> 1.0707 Offset: -0.30ppm H2O zero 0.7385 -> 0.7391 Offset: 0.09ppm H2O span 1.0159 -> 1.0255 Offset: -0.02ppm Notes: |
LI-7500A 0041 | 2016-12-05 | Lab none | Calibration | Lab none | Last calibration: 2016-05-05 AGC=51 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.66kPa, Tair=20.72C, Tdew=8.12C Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.00 CO2 zero 1.0002 -> 1.0004 Offset: 0.0ppm CO2 span 1.0128 -> 1.0124 Offset: 0.36ppm H2O zero 0.8257 -> 0.8255 Offset: -1.0ppm H2O span 1.0037 -> 1.0000 Offset: -0.01ppm Notes: Had trouble connecting sensor to software at first. Software froze in middle so re-started software to do calibration. |
LI-7500A 0035 | 2016-12-05 | Lab none | Calibration | Lab none | Last calibration: 2016-05-05 AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.73kPa, Tair=20.92C, Tdew=9.8C Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.00 CO2 zero 0.9388 -> 0.9390 Offset: 0.89ppm CO2 span 1.0216 -> 1.0265 Offset: -2.1ppm H2O zero 0.9018 -> 0.9025 Offset: 4.1ppm H2O span 0.9928 -> 1.0029 Offset: -0.27ppm Notes: |
LI-7500A 0065 | 2016-12-06 | Lab none | Calibration | Lab none | Last calibration: 2016-05-13 AGC=74.7 SDM=6 DAC1= DAC2= Ambient P=100.29kPa, Tair=16.04C, Tdew=0.60C Zero Air at 350psi, God CO2 421.22 at 400 Tdewspan=16.02 CO2 zero 0.8898 -> 0.8899 Offset: 0.36ppm CO2 span 1.0149 -> 1.0142 Offset: 0.43ppm H2O zero 0.7086 -> 0.7086 Offset: 0.2ppm H2O span 1.0139 -> 1.0127 Offset: 0.01ppm Notes: H2O span done later in the day. |